K163550 · Habalan Med & Beauty Co.,Ltd · NFO · Dec 20, 2017 · Neurology
Device Facts
Record ID
K163550
Device Name
Pobling MIITY 2
Applicant
Habalan Med & Beauty Co.,Ltd
Product Code
NFO · Neurology
Decision Date
Dec 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pobling MIITY 2 is intended for facial stimulation.
Device Story
Pobling MIITY 2 is a battery-powered, hand-held facial toning device. It delivers micro-current stimulation to healthy facial skin via the main unit head. The device features integrated heating (41°C ±3°C) and cooling (10°C ±3°C) functions, selectable via button. Operated by the user in a home or clinic setting, the device provides non-invasive facial stimulation. The healthcare provider or user observes the device's output performance (voltage/current) to ensure consistent stimulation. The device is intended to provide cosmetic facial toning benefits through electrical stimulation.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993: cytotoxicity, skin sensitization, skin irritation), and thermal safety testing for the heating/cooling functions.
Indicated for over-the-counter cosmetic use for facial stimulation in healthy individuals.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2017
HABALAN Med & Beauty Co., Ltd c/o Peter Chung President Plus Global 300 Atwood Pittsburgh, Pennsylvania 15213
Re: K163550
Trade/Device Name: Pobling MIITY 2 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: December 7, 2017 Received: December 13, 2017
Dear Peter Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vivek J. Pinto -S
for Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163550
Device Name Pobling MIITY 2
Indications for Use (Describe)
The Pobling MITY 2 is intended for facial stimulation and is indicated for over-the counter cosmetic use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for HABALAN, with the words "Different Insight" underneath. The logo is in a bold, sans-serif font. The letters are all capitalized and connected to each other. The words "Different Insight" are in a smaller, sans-serif font.
# 510(k) Summary
Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
### 1. Submitter's Information
- 1) Name : HABALAN Med & Beauty Co., Ltd
- 2) Address : 907, 219, Gasandigital 1-ro, Geumcheon-gu, Seoul, KOREA
- 3) Contact person : Mr. Peter Chung
300 Atwood Street Pittsburgh, PA 15213, USA
Tel : 412-512-8802
- 4) Date prepared : December, 05, 2016
#### 2. Device Information
- 1) Trade/Proprietary Name : Pobling MIITY 2
- 2) Common Name : Facial Toning Device
- 3) Classification Name : Transcutaneous Electrical Nerve Stimulator
- 4) Device Class : Class II
- 5) Product Code : NFO
- 6) Regulation Number : 21 CFR 882.5890
#### 3. Predicate Device
- 1) Carol Cole Company, NuFACE® Plus, K103472
- 2) Nu Skin Enterprises, Inc., NuSkin Facial Spa, K122711
#### 4. Intended Use
The Pobling MIITY 2 is intended for facial stimulation.
#### 5. Indication for Use
The Pobling MIITY 2 is indicated for over-the counter cosmetic use.
#### 6. Device Description
Pobling MIITY 2 is a facial toning device and it is composed of main unit and adaptor. The micro-current is delivered to healthy facial skin through head of main unit to stimulate the face. The device is a battery-powered and hand-held device. There have heating and cooling function by pressing the button additionally.
#### 7 Comparison with Predicate Device
The MITY 2 is identical with predicate devices for intended use and indications for use. It is
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a stylized wordmark for "HABALAN" in a bold, sans-serif font. The letters are interconnected, creating a continuous, flowing design. The color of the text is black, and the background is white, providing a high contrast.
#### Different Insight
used as same technological principle with predicate devices to accomplish its intended use. all are hand-held, battery powered device and the there is no new questions of safety and/or effectiveness.
The determination of substantial equivalence for the MITY2 is based on an assessment of non-clinical performance. This assessment included a comparison of the MIITY2 to that of predicate. The output performance testing included:
- Maximum output voltaget
- Maximum output current
| Features | Subject Device | Predicate Device | | Justification |
|----------------------------------------|----------------------------------------------|--------------------|--------------------------------------------------------------------------------------------------------------------------|---------------|
| | MIITY 2 | NuFACE® Plus | NuSkin Facial Spa | — |
| 510(k) Number | To be assigned | K103472 | K122711 | — |
| Regulation Number | 21 CFR § 882.5890 | | | Same |
| Regulation Name | Transcutaneous Electrical Nerve Stimulator | | | Same |
| Regulatory Class | Class II | | | Same |
| Product Code | NFO | | | Same |
| Intended Use | Stimulate the face | | | Same |
| Indication for Use | Over-the-Counter Cosmetic Use | | | Same |
| Power Source | Battery operated | | | Same |
| Number output<br>channels | 1 | | | Same |
| Automatic shut off | Yes | | | Same |
| Indicator Display | | | | |
| 1) ON / OFF? | Yes | | | Same |
| 2) Low-Battery? | Yes | | | Same |
| On Time (seconds) | Constant | | | Same |
| Off Time (seconds) | None | | | Same |
| Compliance with<br>Voluntary Standards | IEC 60601-1 / IEC 60601-1-2 / IEC 60601-1-11 | | | Same |
| Mode | Two | One | One to Five | Similar |
| Wave form | Direct | Pulsed Monophasic | Direct Current | Same |
| Maximum Output<br>Voltage | @500Ω<br>Mode1: 174mV<br>Mode2: 174mV | @500Ω<br>137mV | @500Ω<br>Mode1: 188mV(214mV)<br>Mode2: 189mV(225mV)<br>Mode3: 185mV(220mV)<br>Mode4: 184mV(212mV)<br>Mode5: 184mV(213mV) | Similar |
| | @2kΩ<br>Mode1: 710mV<br>Mode2: 710mV | @2kΩ<br>769mV | @2kΩ<br>Mode1: 722mV(1527mV)<br>Mode2: 746mV(812mV)<br>Mode3: 735mV(815mV)<br>Mode4: 740mV(802mV)<br>Mode5: 741mV(801mV) | Similar |
| | Subject Device | Predicate Device | | Justification |
| Features | MIITY 2 | NuFACE® Plus | NuSkin Facial Spa | – |
| Maximum Output | @10kΩ | @10kΩ | @10kΩ | Similar |
| Voltage | Model: 3200mV | 3.82V | Model: 2.626V(2.978V) | |
| | Mode2: 3200mV | | Mode2: 3.787V(4.011V) | |
| | | | Mode3: 3.747V(4.014V) | |
| | | | Mode4: 3.503V(3.980V) | |
| | | | Mode5: 3.729V(3.980V) | |
| Maximum Output | @500Ω | @500Ω | @500Ω | Similar |
| Current | Model: 350uA | 274uA | Model: 376uA(427uA) | |
| | Mode2: 350uA | | Mode2: 379uA(449uA) | |
| | | | Mode3: 371uA(440uA) | |
| | | | Mode4: 368uA(424uA) | |
| | | | Mode5: 369uA(427uA) | |
| | @2kΩ | @2kΩ | @2kΩ | Similar |
| | Model: 360uA | 387uA | Model: 361uA(763uA) | |
| | Mode2: 360uA | | Mode2: 373uA(406uA) | |
| | | | Mode3: 367uA(407uA) | |
| | | | Mode4: 370uA(401uA) | |
| | | | Mode5: 370uA(400uA) | |
| | @10kΩ | @10kΩ | @10kΩ | Similar |
| | Model: 320uA | 383uA | Model: 362uA(398uA) | |
| | Mode2: 320uA | | Mode2: 379uA(401uA) | |
| | | | Mode3: 375uA(401uA) | |
| | | | Mode4: 350uA(398uA) | |
| | | | Mode5: 373uA(398uA) | |
| Heating and Cooling | Model: Heating | | | Minor |
| | (Temp. 41℃±3℃) | | | function |
| | Mode2: Cooling | | | |
| | (Temp. 10℃±3℃) | | | |
| Weight | 115g | 9oz | 120g (approx. 4oz) | – |
| | | | Facial Spa with Large | |
| Dimensions | 150 x 500<br>unit : mm | 3" x 5.25" x 1.25" | 143 x 31.2 x 67.5 mm | |
| | | | (5.6" x 1.2" x 2.7") | |
| | | | Facial Spa with Small | |
| | | | 136 x 31.2 x 67.5 mm | |
| | | | (5.4" x 1.2" x 2.7") | |
| Housing material | ABS | Thermo Plastic | ABS, Chrome plate | - |
| and construction | | | ABS, rubber (elastomer) | |
## TABLE : Comparison of MIITY 2 with predicate devices
{6}------------------------------------------------
The heating and cooling are not a major function to be used as accomplished its intended use as stimulator. That are minor for distinguishing the mode. And also conducted the performance data. Therefore, all the difference don't affect the safety and effectiveness which is concluded after all the required testing, So no safety and effectiveness issues relating to the MIITY2
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for HABALAN. The logo is black and consists of the word "HABALAN" in a stylized font. Below the logo, the words "Different Insight" are written in a smaller, sans-serif font.
# 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination 1) Electrical Safety and Electromagnetic compatibility(EMC)
Electrical safety and EMC testing were conducted on MITY2, it comples with the IEC 60601-1:2005 +am1:2002 for safety and the IEC 60601-1-2:2007 standard for EMC.
- 2) Biocompatibility Testing
The biocompatibility evaluation for the contact part with healthy facial skin of MIITY2 was conducted in accordance with the ISO 10993, Cytotoxicity, Skin Sensitization and Skin Irritation testing, it was considered contacting part and duration (Tissue contacting, Duration of less than 24hours)
- 4) Technical characteristic testing
To show the technical characteristic testing about Transcutaneous Electrical Nerve Stimulator, the test was conducted in our testing method.
- 5) Thermal Safety testing
The device have the heating and cooling function with stimulation. To verify the heating and cooling properties, we performed the testing in our testing method.
- 6) Clinical data
Not applicable
# 9. Conclusion
The fundamental technology to be accomplished its intended use in view of technological and operational characteristics between MITY2 and predicate device is same. Non-clinical performance as documented support the safety and effectiveness are similar to the predicate device. In other words, the MIITY2 is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.