MALDI Biotyper CA (MBT-CA) System, MBT smart CA System
Applicant
Bruker Daltonik GmbH
Product Code
PEX · Microbiology
Decision Date
Jul 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3361
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K163536 · Jul 26, 2017
MALDI Biotyper CA (MBT-CA) System, MBT smart CA System
Bruker Daltonik GmbH
Clinical isolates (fresh and stored) from clinical study sites
Clinical isolates were used to perform a method comparison study to validate the identification performance of the MBT-CA System against 16S rRNA/ITS sequencing reference methods.
Method Comparison Study; Method comparison study using clinical isolates
2,091 fresh and stored clinical isolates (bacteria and yeast); Sample Size: 2,091; Number of Sites: 4 US clinical test sites and in-house laboratory
16S rRNA or ITS sequencing and protein gene sequencing
Identification accuracy (log score >= 2.0)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Microorganism identification
Biostatistical algorithm comparing peak lists to reference spectra library
—
98.45% high resolution species identification accuracy
—
—
Method comparison study: 2,091 fresh and stored organisms tested at four US clinical test sites and in-house laboratory
—
Indications for Use
The MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens. The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast infections.
Device Story
The MBT-CA System uses MALDI-TOF mass spectrometry to identify microorganisms from cultured human specimens. Input: isolated microbial colonies from various agar media. Process: colonies are transferred to a target plate, overlaid with matrix (α-Cyano-4-hydroxycinnamic acid), and ionized via laser (nitrogen or smartbeam). The system measures the time-of-flight of ionized ribosomal proteins to generate a mass spectrum. Software compares this spectrum against a reference library using a biostatistical algorithm to produce a log(score) (0.00–3.00). Output: identification result with confidence level. Used in clinical laboratories by trained microbiologists. Results are interpreted alongside Gram stain, morphology, and clinical context to aid in diagnosing bacterial and yeast infections.
Clinical Evidence
Clinical performance evaluated at four US sites using fresh and stored isolates. Study compared MBT-CA results against reference sequencing (16S rRNA/ITS). Overall performance for bacteria (n=2011) showed 98.7% correct genus/species identification. Yeast performance (n=80) showed 98.8% correct genus/species identification. Reproducibility study (n=180) demonstrated 99-100% agreement across sites, operators, and instruments.
Technological Characteristics
Mass spectrometer: microflex LT/SH or microflex LT/SH smart (nitrogen or smartbeam laser). Energy source: 337nm nitrogen laser. Matrix: alpha-Cyano-4-hydroxycinnamic acid (HCCA). Target plates: US IVD 48 Spot Target (steel) or MBT Biotarget 96 US IVD (disposable). Connectivity: Standalone system with barcode reader support. Software: MBT-CA System (client/server architecture).
Indications for Use
Indicated for identification of cultured bacterial and yeast microorganisms from human specimens to aid in diagnosis of infections. For use by trained laboratory personnel.
Regulatory Classification
Identification
A mass spectrometer system for clinical use for the identification of microorganisms is a qualitative in vitro diagnostic device intended for the identification of microorganisms cultured from human specimens. The device is comprised of an ionization source, a mass analyzer, and a spectral database. The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.
(2) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols.
(3) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(4) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(5) Premarket notification submissions must include details on the appropriate end user device training program that will be offered while marketing the device.
{1}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
A. 510(k)
K163536
B. Purpose for Submission:
To obtain a substantial equivalent determination for the MALDI Biotyper CA System.
C. Measurand:
See Intended Use.
D. Type of Test:
The MALDI Biotyper CA System is a qualitative in vitro diagnostic device intended for the identification of Gram-negative bacterial colonies cultured from human specimens. The device is comprised of an ionization source, a mass analyzer and a spectral database. The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of Gram-negative bacterial infections.
E. Applicant:
Bruker Daltonik GmbH
F. Proprietary and Established Names:
Trade Name: MALDI Biotyper CA (MBT-CA) System, MBT smart CA System
Common Names: MBT-CA, System, mass spectrometry
G. Regulatory Information:
1. Regulation section: 21 CFR 866.3361 Instrumentation for clinical multiplex test systems
2. Classification: Class II (special controls)
3. Product code: PEX
4. Panel: Microbiology (83)
H. Intended Use:
1. Intended use(s):
The MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens.
The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast infections.
The following organisms are claimed:
1
{2}
| Bacteria: | |
| --- | --- |
| Abiotrophia defectiva | Achromobacter xylosoxidans |
| Acinetobacter baumannii / nosocomialis group | Acinetobacter calcoaceticus |
| Acinetobacter haemolyticus | Acinetobacter johnsonii |
| Acinetobacter junii | Acinetobacter lwoffii |
| Acinetobacter pittii | Acinetobacter radioresistens |
| Acinetobacter ursingii | Actinomyces europaeus |
| Actinomyces funkei | Actinomyces graevenitzii |
| Actinomyces hyovaginalis | Actinomyces meyeri |
| Actinomyces neuii | Actinomyces odontolyticus |
| Actinomyces oris | Actinomyces radingae |
| Actinomyces turicensis | Actinomyces urogenitalis |
| Actinotignum schaalii group | Aerococcus sanguinicola |
| Aerococcus urinae | Aerococcus viridans |
| Aeromonas salmonicida | Aeromonas sp[7] |
| Aggregatibacter actinomycetemcomitans | Aggregatibacter aphrophilus |
| Aggregatibacter segnis | Alcaligenes faecalis |
| Alloiococcus otitis | Alloscardovia omnicolens |
| Anaerococcus murdochii | Anaerococcus vaginalis |
| Arthrobacter cumminsii | Bacteroides caccae |
| Bacteroides fragilis | Bacteroides nordii |
| Bacteroides ovatus group | Bacteroides pyogenes |
| Bacteroides salyersiae | Bacteroides stercoris group |
| Bacteroides thetaiotaomicron group | Bacteroides uniformis |
| Bacteroides vulgatus group | Bifidobacterium breve |
| Bordetella group [3] | Bordetella hinzii |
| Brevibacterium casei | Brevundimonas diminuta group |
| Burkholderia cepacia complex [13] | Burkholderia gladioli |
| Burkholderia multivorans | Campylobacter coli |
| Campylobacter jejuni | Campylobacter ureolyticus |
| Capnocytophaga ochracea | Capnocytophaga sputigena |
| Chryseobacterium gleum | Chryseobacterium indologenes |
| Citrobacter amalonaticus complex | Citrobacter freundii complex |
| Citrobacter koseri | Clostridium beijerinckii |
| Clostridium bifermentans | Clostridium butyricum |
| Clostridium clostridioforme group | Clostridium difficile |
| Clostridium innocuum | Clostridium paraputrificum |
2
{3}
| Bacteria: | |
| --- | --- |
| Clostridium perfringens | Clostridium ramosum |
| Clostridium septicum | Clostridium sordellii |
| Clostridium sporogenes / Clostridium botulinum (group I) | Clostridium tertium |
| Corynebacterium accolens | Corynebacterium afermentans group |
| Corynebacterium amycolatum | Corynebacterium aurimucosum group |
| Corynebacterium bovis | Corynebacterium coyleae |
| Corynebacterium diphtheriae | Corynebacterium freneyi |
| Corynebacterium glucuronolyticum | Corynebacterium glutamicum |
| Corynebacterium jeikeium | Corynebacterium kroppenstedtii |
| Corynebacterium macginleyi | Corynebacterium minutissimum |
| Corynebacterium mucifaciens / ureicelerivorans group | Corynebacterium propinquum |
| Corynebacterium pseudodiphtheriticum | Corynebacterium pseudotuberculosis |
| Corynebacterium resistens | Corynebacterium riegelii |
| Corynebacterium striatum group | Corynebacterium tuberculostearicum |
| Corynebacterium ulcerans | Corynebacterium urealyticum |
| Corynebacterium xerosis | Cronobacter sakazakii group |
| Cupriavidus pauculus group | Delftia acidovorans group |
| Dermabacter hominis | Dermacoccus nishinomiyaensis |
| Edwardsiella tarda | Eikenella corrodens |
| Elizabethkingia meningoseptica group | Enterobacter aerogenes |
| Enterobacter amnigenus | Enterobacter cloacae complex |
| Enterococcus avium | Enterococcus casseliflavus |
| Enterococcus durans | Enterococcus faecalis |
| Enterococcus faecium | Enterococcus gallinarum |
| Enterococcus hirae | Enterococcus mundtii |
| Enterococcus raffinosus | Escherichia coli |
| Escherichia hermannii | Escherichia vulneris |
| Ewingella americana | Facklamia hominis |
| Finegoldia magna | Fluoribacter bozemanae |
| Fusobacterium canifelinum | Fusobacterium necrophorum |
| Fusobacterium nucleatum | Gardnerella vaginalis |
| Gemella haemolysans | Gemella morbillorum |
| Gemella sanguinis | Granulicatella adiacens |
| Haemophilus haemolyticus | Haemophilus influenzae |
3
{4}
| Bacteria: | |
| --- | --- |
| Haemophilus parahaemolyticus group | Haemophilus parainfluenzae |
| Hafnia alvei | Helcococcus kunzii |
| Kingella denitrificans | Kingella kingae |
| Klebsiella oxytoca / Raoultella ornithinolytica | Klebsiella pneumoniae |
| Klebsiella variicola | Kocuria kristinae |
| Kytococcus sedentarius | Lactobacillus gasseri |
| Lactobacillus jensenii | Lactobacillus rhamnosus |
| Lactococcus garvieae | Lactococcus lactis |
| Leclercia adecarboxylata | Legionella longbeachae |
| Legionella pneumophila | Leuconostoc citreum |
| Leuconostoc mesenteroides | Leuconostoc pseudomesenteroides |
| Listeria monocytogenes | Macrococcus caseolyticus |
| Mannheimia haemolytica group | Micrococcus luteus |
| Micrococcus lylae | Mobiluncus curtisii |
| Moraxella sg Branhamella catarrhalis* | Moraxella sg Moraxella nonliquefaciens* |
| Moraxella sg Moraxella osloensis* | Morganella morganii |
| Myroides odoratimimus | Myroides odoratus |
| Neisseria bacilliformis | Neisseria cinerea |
| Neisseria elongata | Neisseria flavescens / subflava group |
| Neisseria gonorrhoeae | Neisseria lactamica |
| Neisseria meningitidis | Neisseria sicca group |
| Neisseria weaveri | Nocardia brasiliensis |
| Nocardia cyriacigeorgica | Nocardia farcinica group |
| Nocardia nova | Nocardia otitidiscaviarum |
| Ochrobactrum anthropi | Oligella ureolytica |
| Oligella urethralis | Pantoea agglomerans |
| Parabacteroides distasonis | Parabacteroides goldsteinii |
| Parabacteroides johnsonii / merdae group | Parvimonas micra |
| Pasteurella multocida | Pediococcus acidilactici |
| Pediococcus pentosaceus | Peptoniphilus harei group |
| Peptostreptococcus anaerobius | Plesiomonas shigelloides |
| Pluralibacter gergoviae | Porphyromonas gingivalis |
| Porphyromonas somerae | Prevotella bivia |
| Prevotella buccae | Prevotella denticola |
| Prevotella intermedia | Prevotella melaninogenica |
4
{5}
| Bacteria: | |
| --- | --- |
| Propionibacterium acnes | Proteus mirabilis |
| Proteus vulgaris group | Providencia rettgeri |
| Providencia stuartii | Pseudomonas aeruginosa |
| Pseudomonas fluorescens group | Pseudomonas oryzihabitans |
| Pseudomonas putida group | Pseudomonas stutzeri |
| Ralstonia pickettii | Rhizobium radiobacter |
| Rothia aeria | Rothia dentocariosa |
| Rothia mucilaginosa | Salmonella sp** |
| Serratia fonticola | Serratia liquefaciens |
| Serratia marcescens | Serratia odorifera |
| Serratia plymuthica | Serratia rubidaea |
| Sphingobacterium multivorum | Sphingobacterium spiritivorum |
| Sphingomonas paucimobilis group | Staphylococcus aureus |
| Staphylococcus auricularis | Staphylococcus capitis |
| Staphylococcus caprae | Staphylococcus carnosus |
| Staphylococcus cohnii | Staphylococcus delphini |
| Staphylococcus epidermidis | Staphylococcus equorum |
| Staphylococcus felis | Staphylococcus haemolyticus |
| Staphylococcus hominis | Staphylococcus intermedius |
| Staphylococcus lentus | Staphylococcus lugdunensis |
| Staphylococcus pasteuri | Staphylococcus pettenkoferi |
| Staphylococcus pseudintermedius | Staphylococcus saccharolyticus |
| Staphylococcus saprophyticus | Staphylococcus schleiferi |
| Staphylococcus sciuri | Staphylococcus simulans |
| Staphylococcus vitulinus | Staphylococcus warneri |
| Staphylococcus xylosus | Stenotrophomonas maltophilia |
| Streptococcus agalactiae | Streptococcus anginosus |
| Streptococcus canis | Streptococcus constellatus |
| Streptococcus dysgalactiae | Streptococcus equi |
| Streptococcus gallolyticus | Streptococcus gordonii |
| Streptococcus intermedius | Streptococcus lutetiensis |
| Streptococcus mitis / oralis group | Streptococcus mutans |
| Streptococcus parasanguinis | Streptococcus pneumoniae |
| Streptococcus pyogenes | Streptococcus salivarius / vestibularis group |
| Streptococcus sanguinis | Streptococcus sobrinus |
5
{6}
| Bacteria: | |
| --- | --- |
| Streptococcus thermophilus | Sutterella wadsworthensis |
| Trueperella bernardiae | Turicella otitidis |
| Vagococcus fluvialis | Veillonella parvula group |
| Vibrio parahaemolyticus | Vibrio vulnificus |
| Weeksella virosa | Yersinia enterocolitica |
| Yersinia frederiksenii | Yersinia intermedia |
| Yersinia kristensenii | Yersinia pseudotuberculosis |
| * = subgenus | |
| sp** = species | |
| Yeasts: | |
| --- | --- |
| Candida albicans | Candida boidinii |
| Candida dubliniensis | Candida duobushaemulonii |
| Candida famata | Candida glabrata |
| Candida guilliermondii | Candida haemulonis |
| Candida inconspicua | Candida intermedia |
| Candida kefyr | Candida krusei |
| Candida lambica | Candida lipolytica |
| Candida lusitaniae | Candida metapsilosis |
| Candida norvegensis | Candida orthopsilosis |
| Candida parapsilosis | Candida pararugosa |
| Candida pelliculosa | Candida tropicalis |
| Candida valida | Candida zeylanoides |
| Cryptococcus gattii | Cryptococcus neoformans var grubii* |
| Cryptococcus neoformans var neoformans* | Cyberlindnera jadinii |
| Geotrichum candidum | Geotrichum capitatum |
| Kloeckera apiculata | Malassezia furfur |
| Malassezia pachydermatis | Pichia ohmeri |
| Rhodotorula mucilaginosa | Saccharomyces cerevisiae |
| Trichosporon asahii | Trichosporon inkin |
| Trichosporon mucoides group | |
| * = variety | |
2. Indication(s) for use: Same as intended use.
3. Special conditions for use statement(s):
The MALDI Biotyper CA System is for prescription use only in accordance with 21 CFR
6
{7}
801.109.
# 4. Special instrument requirements:
Mass Spectrometer: microflex LT/SH mass spectrometer
Target Plates: US IVD 48 Spot Target
Reagents:
• US IVD Bacterial Test Standard (BTS)
• US IVD HCCA portioned ( \( \alpha \) -Cyano-4-hydroxycinnamic acid)
Database: MALDI Biotyper for Clinical Applications (MBT-CA)
Software:
• MBT-CA System Software Package:
MBT-CA System client software displaying the user interface
MBT-CA System Server
MBT-CA System DB Server
- flexControl Software Package:
GTPS firmware
flexControl acquisition software
# I. Device Description:
The MBT-CA System consists of the microflex LT/SH mass spectrometer, reference library, kit reagents (US IVD HCCA, US IVD Bacterial Test Standard), US IVD 48 Spot Target or MBT Biotarget 96 US IVD plate, and software. The MALDI Biotyper CA System with closed safety covers is a Class 1 Laser product. With the safety cover opened it becomes a Class 4 Laser product. The laser is a \(337\mathrm{nm}\) fixed focus, nitrogen laser. The MALDI Biotyper CA System is also referred to as the MBT-CA System.
The reference library, MALDI Biotyper for Clinical Applications, includes type strains, clinical strains and culture collection strains. The MALDI Biotyper CA System reference library was established by analyzing the type strain from each claimed species combined with 5 to 10 additional strains from the same species provided by clinical laboratories or different commercial strain collections for a total of 528 strains. Library mass spectra used for matching contain up to 70 peaks. The MALDI Biotyper for Clinical Applications is also referred to as the MBT-CA.
US IVD HCCA portioned ( \( \alpha \) -Cyano-4-hydroxycinnamic acid) is a solution that is used when processing test organisms for identification on the MALDI Biotyper CA System. US IVD HCCA is reconstituted in accordance with instructions provided using recommended solvent. 1.0 \( \mu \) L of the matrix is added to the spot with the sample and allowed to dry.
US IVD Bacterial Test Standard (BTS) is an in-vitro-diagnostic product used for quality control and validation of the microflex LT/SH mass spectrometers. US IVD BTS contains a manufactured extract of Escherichia coli DH5 alpha that demonstrates a characteristic peptide and protein profile mass spectrum, when tested on the MALDI Biotyper CA System. US IVD BTS is spiked with two additional proteins that extend the upper boundary of the mass range of the US IVD BTS. The overall mass range covered by US IVD BTS is 3.6 to 17 kDa.
US IVD 48 Spot Target plates are reusable steel plates which have been developed for the
7
{8}
preparation and identification of test organisms using the MALDI Biotyper CA System. The target allows for the identification of 48 test organisms. The target has five cross-joint positions which should be used for US IVD BTS control. Target plate cleaning is performed after each run.
MALDI Biotyper CA System client software displays a user-interface which guides the user through the MALDI Biotyper CA System workflow. The MALDI Biotyper CA System client also interfaces to the flexControl software for automated acquisition of mass spectra on the microflex LT/SH instrument.
The MALDI Biotyper CA System server communicates with the MALDI Biotyper CA System client and the MBT-DB server. It performs preprocessing on acquired spectra, and matches peaks lists against the Main Spectrum (reference pattern, (MSP)) for matching and calculates the score value (log (score)).
The MBT-DB server stores all information for the MALDI Biotyper CA System. The MBT-DB maintains spectra data (creation information and mass/intensity lists), project data (results of defined and executed runs), method data (parameter lists for spectra preprocessing and identification), user management data, reference patterns and other peak lists plus additional maintenance data.
GTPS firmware communicates with the flexControl PC software, controls and monitors the vacuum, moves the sample carrier and performs the docking of the target plate, controls and monitors high voltages in the ion source, generates trigger signals, and monitors instrument status.
The flexControl acquisition software communicates with the MALDI Biotyper CA System client, loads automatic run jobs, communicates with the GTPS firmware, communicates with the laser in the microflex LT/SH instrument, sets the acquisition parameters in the digitizer and reads the acquired data from the digitizer, performs automated data acquisition, evaluates acquired spectra, adjusts the laser power during automatic data acquisition, performs a recalibration of the time-of-flight to mass transformation, stored acquired spectra on disk and performs source cleaning. The flexControl software does not display a user interface.
The optional Honeywell (Hyperion 1300g) Barcode Reader USB cable is connected to the MALDI Biotyper CA System computer. The barcode reader scans the unique ten-digit target ID which appears in the Target ID box on the target plate. After the target ID has been entered, the a new Run page opens and the ten-digit target ID appears as the Plate Id and is appended to the Run name. Sample identifications are entered into the computer corresponding to the target plate position for that run.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
MALDI Biotyper CA System
2. Predicate 510(k) number(s):
K142677
3. Comparison with predicate:
8
{9}
| SIMILARITIES | | |
| --- | --- | --- |
| Characteristic | NEW DEVICE (K163536) MALDI Biotyper CA System | PREDICATE DEVICE (K142677) MALDI Biotyper CA System |
| Product Code | PEX | PEX |
| Intended use | The MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens. The MALDI Biotyper CA System is a qualitative *in vitro* diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast infections. | The MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens. The MALDI Biotyper CA System is a qualitative *in vitro* diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast infections. |
9
{10}
| SIMILARITIES | | |
| --- | --- | --- |
| Characteristic | NEW DEVICE (K163536) MALDI Biotyper CA System | PREDICATE DEVICE (K142677) MALDI Biotyper CA System |
| Sample type | Isolated colony from any patient sample source. Acceptable media: Columbia blood agar with 5% sheep blood (Gram-negative bacteria) Trypticase soy agar with 5% sheep blood (Gram-negative bacteria, Gram-positive bacteria, yeasts) Chocolate agar (Gram-negative bacteria, Gram-positive bacteria) MacConkey Agar (Gram-negative bacteria) Columbia CNA agar with 5% sheep blood (Gram-positive bacteria) Brucella Agar with 5% horse blood (Gram-negative anaerobic bacteria, Gram-positive anaerobic bacteria) CDC anaerobe Agar with 5% sheep blood (Gram-negative anaerobic bacteria, Gram-positive anaerobic bacteria) CDC anaerobe 5% sheep blood Agar with phenylethyl alcohol (Gram-negative anaerobic bacteria, Gram-positive anaerobic bacteria) CDC anaerobe laked sheep blood Agar with kanamycin and vancomycin (Gram-negative anaerobic bacilli) Bacteroides bile esculin Agar with amikacin (Bacteroides species) Clostridium difficile Agar with 7% sheep blood (Clostridium difficile) Sabouraud-Dextrose Agar (Yeasts) Brain Heart Infusion Agar (Yeasts) Campylobacter Agar with 5 Antimicrobics and 10% Sheep Blood (Campylobacter species) Bordet Gengou Agar with 15% sheep blood (Bordetella species) | Isolated colony from any patient sample source. Acceptable media: Columbia blood agar with 5% sheep blood (Gram-negative bacteria) Trypticase soy agar with 5% sheep blood (Gram-negative bacteria, Gram-positive bacteria, yeasts) Chocolate agar (Gram-negative bacteria, Gram-positive bacteria) MacConkey Agar (Gram-negative bacteria) Columbia CNA agar with 5% sheep blood (Gram-positive bacteria) Brucella Agar with 5% horse blood (Gram-negative anaerobic bacteria, Gram-positive anaerobic bacteria) CDC anaerobe Agar with 5% sheep blood (Gram-negative anaerobic bacteria, Gram-positive anaerobic bacteria) CDC anaerobe 5% sheep blood Agar with phenylethyl alcohol (Gram-negative anaerobic bacteria, Gram-positive anaerobic bacteria) CDC anaerobe laked sheep blood Agar with kanamycin and vancomycin (Gram-negative anaerobic bacilli) Bacteroides bile esculin Agar with amikacin (Bacteroides species) Clostridium difficile Agar with 7% sheep blood (Clostridium difficile) Sabouraud-Dextrose Agar (Yeasts) Brain Heart Infusion Agar (Yeasts) Campylobacter Agar with 5 Antimicrobics and 10% Sheep Blood (Campylobacter species) Bordet Gengou Agar with 15% sheep blood (Bordetella species) |
| Type of Test | Automated Mass Spectrometry System | Automated Mass Spectrometry System |
| Matrix | α-Cyano-4-hydroxycinnamic acid | α-Cyano-4-hydroxycinnamic acid |
10
{11}
| SIMILARITIES | | |
| --- | --- | --- |
| Characteristic | NEW DEVICE (K163536) MALDI Biotyper CA System | PREDICATE DEVICE (K142677) MALDI Biotyper CA System |
| Method of Testing | Bacteria & Yeast: Direct testing If after initial analysis the log(score) is reported at <2.00, organisms may be processed using the Extraction (Ext) procedure or extended Direct Transfer (eDT, 70% aqueous formic acid) procedure. If eDT procedure still yields log (score) <2.00, organisms may be processed via Ext procedure. | Bacteria & Yeast: Direct testing If after initial analysis the log(score) is reported at <2.00, organisms may be processed using the Extraction (Ext) procedure or extended Direct Transfer (eDT, 70% aqueous formic acid) procedure. If eDT procedure still yields log (score) <2.00, organisms may be processed via Ext procedure. |
| Result Reporting | Organism identification is reported with high confidence if the log(score) is ≥2.00 An organism identification is reported with low confidence if the log (score) is between 1.70 and <2.00. | Organism identification is reported with high confidence if the log(score) is ≥2.00 An organism identification is reported with low confidence if the log (score) is between 1.70 and <2.00. |
| Matching Algorithm | Calculates matches by comparing a new spectrum against each single reference entry of a reference database. | Calculates matches by comparing a new spectrum against each single reference entry of a reference database. |
| Mass range | 2,000 - 20,000 m/z | 2,000 - 20,000 m/z |
| Calibration | Bruker US IVD Bacterial Test Standard | Bruker US IVD Bacterial Test Standard |
| Database | MALDI Biotyper Reference Library for Clinical Applications (MBT-CA)-update | MALDI Biotyper Reference Library for Clinical Applications (MBT-CA) |
| DIFFERENCES | | |
| --- | --- | --- |
| Characteristic | NEW DEVICE (K163536) MALDI Biotyper CA System | PREDICATE DEVICE (K142677) MALDI Biotyper CA System |
| System Update | System claim additional organisms but no additional changes | N/A |
| Media | Additional Media validated: - Buffered Charcoal Yeast Extract Agar (Legionella species) - Buffered Charcoal Yeast Extract Selective Agar with polymyxin, anisomycin and vancomycin (Nocardia species) - Modified Thayer-Martin Agar (Neisseria species) | See Similarities |
| MALDI-TOF MS instruments | Bruker microflex LT/SH (benchtop) Bruker microflex LT/SH smart (benchtop) | Bruker microflex LT/SH (benchtop) |
| MALDI Target Plates | US IVD 48 Spot Target (48 positions reusable steel targets) MBT Biotarget 96 US IVD (96 positions disposable targets) | US IVD 48 Spot Target (48 positions reusable steel targets) |
11
{12}
The differences do not affect substantial equivalence of the predicate and the new device. Both systems are mass spectrometers using matrix-assisted laser desorption/ionization-time to flight (MALDI-TOF) for the identification of microorganisms cultured from human specimens. The differences noted above do not impact the intended use and do not raise new questions related to the safety and effectiveness of the test (new) device.
# K. Standard/Guidance Document Referenced (if applicable):
| | Standards No. | Recognition No.(FDA) | Standards Title | Date |
| --- | --- | --- | --- | --- |
| 1 | CLSI MM-18A | 7-192 | Criteria for Identification of Bacteria and Fungi by DNA Target Sequencing; Approved Guideline, 1^{st} Edition | 4/28/2008 |
| 2 | CLSI EP09-A2-IR | 7-92 | Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Second Edition (Interim Revision), | 7/30/2010 |
# L. Test Principle:
Biochemical methods are currently the most commonly used approach for the identification of microorganisms. Organisms are tested against a range of reagents and organism identification is based on a microorganism's reaction to these reagents.
The MBT-CA System uses a different methodology for organism identification based on unique protein patterns of the microorganisms obtained from mass spectrometry. The test organism's spectrum (a pattern of mass peaks) is compared with a reference spectra library (database). Using biostatistical analysis, a probability ranking of the organism identification is generated. The probability ranking is represented as a log (score) between 0.00 and 3.00. Organism identification is reported with high confidence if the log (score) is ≥2.00. An organism identification is reported with low confidence if the log (score) is between 1.70 and <2.00.
Organisms to be identified with the MBT-CA System should be isolated for purity on appropriate isolation media.
Direct Transfer procedure (DT): An individual colony from a subculture plate is transferred to a selected position on a US IVD 48 Spot Target plate or a MBT Biotarget 96 US IVD Target plate (targets) and overlaid with US IVD HCCA portioned (matrix). The standard solvent (50% acetonitrile / 47.5% H2O / 2.5% trifluoroacetic acid) in the matrix solution extracts proteins (mainly ribosomal proteins, which are present in high concentration) from the microorganisms. When dried matrix crystallizes, the inoculated target is ready to be analyzed on the MBT-CA System. If after initial analysis the log (score) is reported as <2.00, organisms can be processed using the extended Direct Transfer (eDT) procedure or the Extraction procedure and analysis repeated. If eDT is employed and log (score) is reported as <2.00, reanalysis via the Extraction procedure may be used.
extended Direct Transfer procedure (eDT): If DT analysis yields a (log(score) <2.00) result, an individual colony from a subculture plate may be transferred to a selected position on a target and overlaid with 70% aqueous formic acid solution. The target is air-dried and then matrix is overlaid. When dried matrix crystallizes, the inoculated target is ready to be analyzed on the MBT-CA System. If a high confidence result is not achieved (log (score) is reported at <2.00), organisms can be processed using the Extraction procedure and analysis repeated.
12
{13}
Extraction procedure (Ext): If after initial analysis and eDT procedure the log (score) is reported at <2.00, organisms are processed using the Extraction procedure and analysis repeated. For this purpose, isolated colonies from the subculture plate are extracted in accordance with instructions of the user manual (Ext sample preparation procedure). An aliquot of extracted material is transferred to a selected position on a target, air-dried and then overlaid with matrix material. When dried matrix crystallizes, the inoculated target is ready to be analyzed on the MBT-CA System.
MALDI-TOF Analysis: Samples are analyzed using MALDI (matrix-assisted laser desorption/ionization) TOF (time of- flight) mass spectrometry. The matrix transfers protons onto the extracted proteins and absorbs UV light. A laser in the MALDI- TOF mass spectrometer irradiates the matrix sample composite, causing evaporation and release of positively charged intact proteins and peptides ("soft" ionization technique). These ions are electrostatically accelerated over a short distance and arrive in the flight tube at a mass-dependent speed. As different proteins/peptides have different masses, ions arrive at the detector at different times (time of flight). The system measures the time (in the nanosecond range) between pulsed acceleration and the corresponding detector signal, the speed is converted into an exact molecular mass. The mass-to-charge ratio of an ion is proportional to the square of its drift time.
Highly abundant microbial proteins (mainly ribosomal proteins) result in a mass spectrum with characteristic mass and intensity distribution. It is species-specific for many bacteria and is interpreted as a molecular fingerprint to identify the test organism.
Data acquisition is controlled with MBT-CA System software. The spectrum of the unknown organism is first transformed into a peak list. Using a biostatistical algorithm, this peak list is compared to the reference peak lists of organisms in the reference library (database) and a log (score) is generated. A higher log (score) indicates a higher degree of similarity to the organism in the reference library. Organism identification is reported with high confidence if the log (score) is ≥2.00. An organism identification is reported with low confidence if the log (score) is between 1.70 and <2.00.
The log (score) ranges, defined in the MBT-CA System, are indicative of the probability of organism identification. Results should be reviewed by a trained microbiologist and final organism identification should be based on all relevant information available. This information includes but is not limited to: Gram staining, colony morphology, growth characteristics, sample matrix, etc.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
# Reproducibility
The reproducibility study for Gram-negative and Gram-positive aerobic bacteria, Gram-negative and Gram-positive microaerophilic bacteria, Gram-negative and Gram-positive anaerobic bacteria and yeasts was carried out to confirm day-to-day reproducibility and precision of the MALDI Biotyper CA System at different clinical study sites. The study was conducted for five days with two runs and two operators each day per clinical site.
The sources of variability tested were:
- Two operators/each clinical study site
- Three clinical study sites
13
{14}
- At least two target plates/each clinical study site
- Four microflex LT/SH instruments
Ten well-characterized organisms were chosen for this study and tested in duplicate via Direct Transfer procedure in accordance with product instructions. When the DT log (score) was <2.00, per product instructions, the test organism was tested following the extended Direct Transfer and Extraction procedure. The results are shown in the table below.
**Table: Reproducibility Study Summary:**
| Blinded Test Organism | Reproducibility Panel | Site A: MBT-CA ID (DT+eDT+Ext) | Site B: MBT-CA ID (DT+eDT+Ext) | Site C: MBT-CAID (DT+eDT+Ext) |
| --- | --- | --- | --- | --- |
| *Dermabacter hominis* | REPRO-01 | 20/20 (100%) | 20/20 (100%) | 20/20 (100%) |
| *Listeria monocytogenes* | REPRO-02* | N/A | N/A | N/A |
| *Nocardia farcinica* group | REPRO-03 | 19/20 (95%) | 19/20 (95%) | 19/20 (95%) |
| *Legionella pneumophila* | REPRO-04 | 20/20 (100%) | 20/20 (100%) | 20/20 (100%) |
| *Clostridium tertium* | REPRO-05 | 20/20 (100%) | 20/20 (100%) | 20/20 (100%) |
| *Facklamia hominis* | REPRO-06 | 20/20 (100%) | 20/20 (100%) | 20/20 (100%) |
| *Bacteroides caccae* | REPRO-07 | 20/20 (100%) | 20/20 (100%) | 20/20 (100%) |
| *Trueperella bernardiae* | REPRO-08 | 20/20 (100%) | 20/20 (100%) | 20/20 (100%) |
| *Neisseria meningitidis* | REPRO-09 | 20/20 (100%) | 20/20 (100%) | 20/20 (100%) |
| *Rhodotorula mucilaginosa* | REPRO-10 | 20/20 (100%) | 20/20 (100%) | 20/20 (100%) |
| **TOTAL** | | **179/180 (99%)** | **179/180 (99%)** | **179/180 (99%)** |
* REPRO-02: Organism skipped as MBT-CA ID applying DT procedure was not confirmed via eDT and Ext procedure.
The data showed that 99% of all test organisms were correctly identified with a log (score) ≥2.00 at each test site after final extraction testing confirming reproducibility of the MALDI Biotyper CA System.
The reproducibility of the MALDI Biotyper smart CA System was evaluated at one laboratory site (in-house). Reproducibility was assessed with ten well-characterized organisms that were chosen for this study and tested in duplicate via Direct Transfer procedure in accordance with MALDI Biotyper CA System instruction for use. When the DT log (score) was <2.00, per product instructions, the test organism was tested following the extended Direct Transfer and Extraction procedure. The study was conducted for five days with two runs and two operators each day per clinical site. Results are summarized below:
**Table: Reproducibility Study Summary:**
| Blinded Test Organism | Reproducibility Panel | Site C: MBT-smart CA ID (DT+eDT+Ext) |
| --- | --- | --- |
| *Dermabacter hominis* | REPRO-01 | 20/20 (100%) |
| *Listeria monocytogenes* | REPRO-02* | N/A |
| *Nocardia farcinica* group | REPRO-03 | 20/20 (100%) |
14
{15}
| *Legionella pneumophila* | REPRO-04 | 20/20 (100%) |
| --- | --- | --- |
| *Clostridium tertium* | REPRO-05 | 20/20 (100%) |
| *Facklamia hominis* | REPRO-06 | 20/20 (100%) |
| *Bacteroides caccae* | REPRO-07 | 20/20 (100%) |
| *Trueperella bernardiae* | REPRO-08 | 20/20 (100%) |
| *Neisseria meningitidis* | REPRO-09 | 20/20 (100%) |
| *Rhodotorula mucilaginosa* | REPRO-10 | 20/20 (100%) |
| **TOTAL** | | **180/180 (100%)** |
* REPRO-02: Organism skipped as MBT-CA ID applying DT procedure was not confirmed via eDT and Ext procedure.
The data show that 100% of all test organisms were correctly identified with a log (score) >2.00 at each test site after final extraction testing confirming reproducibility of the MALDI Biotyper smart CA System.
b. Linearity/assay reportable range:
Not applicable, qualitative assay.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrator:
US IVD BTS is used for mass spectrum calibration and optimization as well as a performance control for the identification of microorganisms with the MALDI Biotyper CA System. US IVD BTS contains a manufactured extract of Escherichia coli DH5 alpha that demonstrates a characteristic peptide and protein profile mass spectrum, when tested on the MALDI Biotyper CA System. US IVD BTS is spiked with two additional proteins that extend the upper boundary of the mass range of the US IVD BTS. The overall mass range covered by US IVD BTS is 3.6 to 17 kDa. Two US IVD BTS control positions on a US IVD 48 Spot Target are selected and inoculated with US IVD BTS solution. The US IVD BTS solution is allowed to dry at room temperature and then overlaid with reconstituted US IVD HCCA portioned solution. If US IVD BTS does not meet all required performance specifications, the test run will be invalid. If US IVD BTS is not inoculated onto a target prior to processing, the test run will be invalid.
Controls:
Klebsiella pneumoniae, Haemophilus influenzae, Proteus vulgaris, Pseudomonas aeruginosa, and Esherichia coli, were used as controls. Of the 193 Quality Control runs conducted during the course of the method comparison study, there were five instances where a Quality Control organism failed to yield an expected result. As a result, all isolates included in that plate run were repeated using a fresh QC organism transfer. The overall plate repeat rate was 2.6% (5/193).
d. Assay cut-off:
The assay cut-off was established and reported in 510(k) K142677.
e. Detection limit:
The Limit of Detection/Dynamic Range study for Gram-negative bacteria was previously performed and reported in 510(k) K130831 (pp. 46-50). The Limit of Detection/Dynamic Range study for Gram-positive bacteria and yeasts was performed and reported in 510(k) K142677.
15
{16}
# *f. Analytical specificity:*
The Interference & Specificity Study was previously established and reported in 510(k) K130831.
# *g. Sample stability studies:*
# Sample Stability after Matrix Overlay:
The sample stability study on target plates for Gram-negative bacteria was previously validated and reported in 510(k) K130831. The sample stability on target plates for Gram-positive bacteria and yeasts was validated and reported in 510(k) K142677.
# US IVD Bacterial Test Standard (BTS)
BTS Stability was established and described in 510(k) K130831.
# HCCA portioned (Matrix) Stability
HCCA portioned (Matrix) Stability was established and described in 510(k) K130831.
# Target plates stability:
Target plate stability was established and described in 510(k) K142677.
# *h. Carry-Over and Cross Contamination:*
The carry-over, cross-contamination and target cleaning study was previously performed and reported in 510(k) K130831.
# *i. Influence of Agar Media*
The validation of sample preparation of test organism to demonstrate that culture media inoculated onto US IVD 48 Spot targets with or without an organism present does not interfere with system performance was previously performed and reported in 510(k) K130831.
# *j. Organism Stability*
# Media and Colony Stability
In accordance with device instructions for use, primary or secondary isolation plates of recommended media [Trypticase Soy Agar with 5% sheep blood (TSA), Buffered Charcoal Yeast Extract Agar (BCYE), Buffered Charcoal Yeast Extract Selective Agar with polymyxin, anisomycin and vancomycin (BCYE/PAV) and Modified Thayer-Martin Agar (MTM))] may be held for up to 12 hours at room temperature prior to testing on the MALDI Biotyper CA System.
In the previous submission (K142677) TSA medium was only validated for Gram-positive /Gram-negative aerobic bacteria and yeast. This submission was intended to establish the suitability of this medium for Gram-positive /-negative microaerophilic and anaerobic bacteria.
Testing was conducted using one Gram-negative bacterium, one microaerophilic
Gram-negative bacterium, one Gram-positive bacterium, one microaerophilic Gram-positive bacterium, two Legionella species, three Nocardia species and three Neisseria species, at varying incubation time points in replicates of eight. Additionally, thirteen clinical isolates derived from Neisseria species were tested in duplicate at varying incubation time points. Results are summarized below:
16
{17}
Table: TSA Summary: Anaerobic Gram-negative/Gram-positive Bacteria (applying MBT-CA System workflow)
| Testing Condition | ≥2.00 ID (DT) | false ID (DT) | ≥2.00 ID (DT+eDT) | false ID (DT+eDT) | ≥2.00 ID (DT+eDT+Ex) | false ID (DT+eDT+Ex) |
| --- | --- | --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 18h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 24h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 24h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 48h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 48h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
Table: TSA Summary: Anaerobic Gram-negative/Gram-positive Bacteria (applying eDT and Ext procedure)
| Testing Condition | >2.00 ID (eDT) | false ID (eDT) | >2.00 ID (Ext) | false ID (Ext) |
| --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 18h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 24h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 24h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 48h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 48h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
Table: TSA Summary: Microaerophilic Gram-negative/Gram-positive Bacteria (applying MBT-CA System workflow)
| Testing Condition | ≥2.00 ID (DT) | false ID (DT) | ≥2.00 ID (DT+eDT) | false ID (DT+eDT) | ≥2.00 ID (DT+eDT+Ext) | false ID (DT+eDT+Ext) |
| --- | --- | --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 18h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 24h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 24h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 48h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 48h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
Table: TSA Summary: Microaerophilic Gram-negative/Gram-positive Bacteria (applying eDT and Ext procedure)
| Testing Condition | >2.00 ID (eDT) | false ID (eDT) | >2.00 ID (Ext) | false ID (Ext) |
| --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 18h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 24h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 24h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
17
{18}
| 48h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| --- | --- | --- | --- | --- |
| 48h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
Table: BCYE Summary: *Legionella* organisms (applying *MBT-CA System workflow*)
| Testing Condition | ≥2.00 ID (DT) | false ID (DT) | ≥2.00 ID (DT+eDT) | false ID (DT+eDT) | ≥2.00 ID (DT+eDT+Ext) | false ID (DT+eDT+Ext) |
| --- | --- | --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 18h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 24h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 24h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 48h/37°C, 0h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
| 48h/37°C, 12h/25°C | 16/16 | 0/16 | NA | NA | NA | NA |
Table: BCYE Summary: *Legionella* organisms (applying *eDT* and *Ext* procedure)
| Testing Condition | >2.00 ID (eDT) | false ID (eDT) | >2.00 ID (Ext) | false ID (Ext) |
| --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 18h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 24h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 24h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 48h/37°C, 0h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
| 48h/37°C, 12h/25°C | 4/4 | 0/4 | 4/4 | 0/4 |
Table: BCYE/PAV Summary: *Nocardia* organisms (applying *MBT-CA System workflow*)
| Testing Condition | ≥2.00 ID (DT) | false ID (DT) | ≥2.00 ID (DT+eDT) | false ID (DT+eDT) | ≥2.00 ID (DT+eDT+Ex) | false ID (DT+eDT+Ext) |
| --- | --- | --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 24/24 | 0/24 | NA | NA | NA | NA |
| 18h/37°C, 12h/25°C | 24/24 | 0/24 | NA | NA | NA | NA |
| 24h/37°C, 0h/25°C | 24/24 | 0/24 | NA | NA | NA | NA |
| 24h/37°C, 12h/25°C | 24/24 | 0/24 | NA | NA | NA | NA |
| 48h/37°C, 0h/25°C | 19/24 | 0/24 | 23/24 | 0/24 | 24/24 | 0/24 |
| 48h/37°C, 12h/25°C | 20/24 | 0/24 | 24/24 | 0/24 | NA | NA |
18
{19}
Table: BCYE/PAV Summary: Nocardia organisms (applying eDT and Ext procedure)
| Testing Condition | >2.00 ID (eDT) | false ID (eDT) | >2.00 ID (Ext) | false ID (Ext) |
| --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 6/6 | 0/6 | 6/6 | 0/6 |
| 18h/37°C, 12h/25°C | 6/6 | 0/6 | 6/6 | 0/6 |
| 24h/37°C, 0h/25°C | 6/6 | 0/6 | 6/6 | 0/6 |
| 24h/37°C, 12h/25°C | 5/6 | 0/6 | 6/6 | 0/6 |
| 48h/37°C, 0h/25°C | 6/6 | 0/6 | 6/6 | 0/6 |
| 48h/37°C, 12h/25°C | 5/6 | 0/6 | 6/6 | 0/6 |
Table: MTM Summary: Neisseria organisms (applying MBT-CA System workflow)
| Testing Condition | ≥2.00 ID (DT) | false ID (DT) | ≥2.00 ID (DT+eDT) | false ID (DT+eDT) | ≥2.00 ID (DT+eDT+Ext) | false ID (DT+eDT+Ext) |
| --- | --- | --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 24/24 | 0/24 | NA | NA | NA | NA |
| 18h/37°C, 12h/25°C | 24/24 | 0/24 | NA | NA | NA | NA |
| 24h/37°C, 0h/25°C | 24/24 | 0/24 | NA | NA | NA | NA |
| 24h/37°C, 12h/25°C | 24/24 | 0/24 | NA | NA | NA | NA |
| 48h/37°C, 0h/25°C | 17/24 | 0/24 | 19/24 | 0/24 | 23/24 | 0/24 |
| 48h/37°C, 12h/25°C | 20/24 | 0/24 | 20/24 | 0/24 | 24/24 | 0/24 |
Table: MTM Summary: Neisseria organisms (applying eDT and Ext procedure)
| Testing Condition | >2.00 ID (eDT) | false ID (eDT) | >2.00 ID (Ext) | false ID (Ext) |
| --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 6/6 | 0/6 | 6/6 | 0/6 |
| 18h/37°C, 12h/25°C | 6/6 | 0/6 | 6/6 | 0/6 |
| 24h/37°C, 0h/25°C | 6/6 | 0/6 | 6/6 | 0/6 |
| 24h/37°C, 12h/25°C | 4/6 | 0/6 | 6/6 | 0/6 |
| 48h/37°C, 0h/25°C | 6/6 | 0/6 | 5/6 | 0/6 |
| 48h/37°C, 12h/25°C | 4/6 | 0/6 | 5/6 | 0/6 |
The study results confirm that the following culture media can be used on the MALDI Biotyper CA System (18 – 48 hours incubation at 37 (± 2) °C):
Anaerobic Gram-negative/-positive bacteria and microaerophilic Gram-negative/Gram-positive bacteria:
- Trypticase Soy Agar with 5% sheep blood (TSA)
Legionella organisms:
- Buffered Charcoal Yeast Extract Agar (BCYE)
Nocardia organisms:
- Buffered Charcoal Yeast Extract Selective Agar with polymyxin, anisomycin and vancomycin (BCYE/PAV)
19
{20}
The cultivation of Neisseria using Modified Thayer-Martin Agar (MTM) should not be longer than 24 hours.
## Organism Stability prior to MALDI Biotyper CA System Analysis
The organism stability study for Gram-negative bacteria prior to MALDI Biotyper CA analysis was previously validated and reported in 510(k) K130831. Organism stability for Gram-positive bacteria and yeasts was previously established and reported in 510(k) K142677.
Other supportive Instrument Performance Characteristics
### Mixed Culture:
The validation of mixed cultures derived from a target organism and varying amounts of non-target organisms was previously performed and reported in 510(k) K130831.
### Viability Study
Viability studies with gram negative rods mixed with matrix on the target plate were not performed. The user is advised to consider all samples, microbial cultures and inoculated products as infectious. Aseptic techniques and usual precautions for handling organisms should be observed throughout the MALDI Biotyper CA System workflow according to "CLSI M29-A, Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline - Current revision". For additional handling precautions, refer to "Biosafety in Microbiological and Biomedical Laboratories - CDC/NIH -Latest edition".
### Resolution of the Acinetobacter baumannii complex:
A study was performed to evaluate the resolution of the Acinetobacter baumannii complex to the species level after final extraction procedure. Sixty-six (66) clinical isolates of Acinetobacter baumannii complex (members: A. baumannii, A. calcoaceticus, A. nosocomialis, A. pittii) collected during initial Clinical Method Comparison protocol (see K130831) and identified by 16S rRNA and/or protein gene sequencing were re-tested using Direct Transfer, extended Direct Transfer and Extraction procedure in parallel. The results are summarized below:
Table: MBT-CA ID of clinical isolates from Acinteobacter baumannii complex
| Test sequence | High confidence ID (DT) | false ID (DT) | high confidence ID (eDT) | false ID (eDT) | high confidence ID (Ext) | false ID (Ext) |
| --- | --- | --- | --- | --- | --- | --- |
| 1^{st} Day | 65/66 | 0/66 | 66/66 | 0/66 | 66/66 | 0/66 |
| 2^{nd} Day | 66/66 | 0/66 | 66/66 | 0/66 | 66/66 | 0/66 |
The study results demonstrated that the Acinetobacter baumannii complex cannot be completely resolved applying MBT-CA System workflow. The reference library/software was updated to report complex Acinetobacter baumannii and Acinetobacter nosocomialis under the Acinetobacter baumannii/nosocomialis group. After formation of an Acinetobacter baumannii/nosocomialis group unambiguous MBT-CA identification with high confidence (log(score) ≥2.00) of Acinetobacter baumannii/nosocomialis group, Acinetobacter calcoaceticus and Acinetobacter pittii is possible if the final extraction procedure has been applied. A matching hint will be included in the package insert which contains the following alert: The displayed species should be considered a member of the Acinetobacter baumannii complex. For organisms identified by the MBT-CA System as
20
{21}
Acinetobacter calcoaceticus, Acinetobacter pittii or Acinetobacter baumannii/nosocomialis group the full Extraction procedure (Ext) is mandatory for secure species differentiation.
Cultivation of yeast organisms at 37 °C:
A study was performed showing the general applicability of the MBT-CA System workflow for identification of yeasts cultivated at 37 (± 2)°C. Testing was conducted using three yeast species at varying incubation time points in replicates of eight. After initial testing, isolates were further tested at room temperature after twelve hours post-incubation. The results are summarized below:
Evaluation of cultivation time for yeast organisms on TSA media at 37 °C (applying MBT-CA System workflow)
| Testing Condition | ≥2.00 ID (DT) | false ID (DT) | ≥2.00 ID (DT+eDT) | false ID (DT+eDT) | ≥2.00 ID (DT+eDT+Ext) | false ID (DT+eDT+Ext) |
| --- | --- | --- | --- | --- | --- | --- |
| 18h/37°C, 0h/25°C | 4/24 | 0/24 | 7/24 | 0/24 | 24/24 | 0/24 |
| 18h/37°C, 12h/25°C | 6/24 | 0/24 | 24/24 | 0/24 | NA | NA |
| 24h/37°C, 0h/25°C | 10/24 | 0/24 | 14/24 | 0/24 | 24/24 | 0/24 |
| 24h/37°C, 12h/25°C | 12/24 | 0/24 | 23/24 | 0/24 | 24/24 | 0/24 |
| 48h/37°C, 0h/25°C | 9/24 | 0/24 | 13/24 | 0/24 | 24/24 | 0/24 |
| 48h/37°C, 12h/25°C | 9/24 | 0/24 | 20/24 | 0/24 | 24/24 | 0/24 |
The study results confirmed the acceptability of 37 (± 2)°C cultivation of yeasts and sample colony stability of up to 12 hours.
Biological / Technical Equivalency Study - MBT-CA smart System:
A study was performed to demonstrate equivalence of the MBT-CA output when using MALDI-TOF mass spectrometers equipped with smartbeam laser technology and nitrogen laser technology.
A panel of thirty four species which are part of the MBT-CA library were measured, with the 34 species representing Gram negative, Gram positive, and yeast. All three sample preparation techniques (DT, eDT, Ext) were used in parallel (eDT and Ext were always additionally performed independent from the DT result). Each sample preparation technique (DT, eDT, Ext) was spotted eight times onto the MALDI target. The measurement was performed on two nitrogen laser instruments and three smart laser instruments.
Overall 4080 spectra were collected in this study. The performance of all spectra for each single species, each single sample prep (DT, eDT and Ext), each MBT-CA result (no ID “red”, low confidence ID “yellow”, high confidence ID “green”) for each instrument type is summarized in the table below.
Table: Biological / Technical Equivalency Study Results; Overall performance of 4080 measured samples. “High confidence ID” (green), “low confidence ID” (yellow) and “no ID” (red) are shown.
21
{22}
| | Nitrogen (544) | | | Smart (816) | | |
| --- | --- | --- | --- | --- | --- | --- |
| DT | 187 | 72 | 285 | 300 | 117 | 399 |
| eDT | 37 | 61 | 446 | 63 | 120 | 633 |
| Ext | 2 | 3 | 539 | 0 | 13 | 803 |
| Σ | 226 | 136 | 1270 | 363 | 250 | 1835 |
| --- | --- | --- | --- | --- | --- | --- |
| | Nitrogen (544) | | | Smart (816) | | |
| --- | --- | --- | --- | --- | --- | --- |
| DT | 34.4% | 13.2% | 52.4% | 36.8% | 14.3% | 48.9% |
| eDT | 6.8% | 11.2% | 82.0% | 7.7% | 14.7% | 77.6% |
| Ext | 0.4% | 0.6% | 99.1% | 0.0% | 1.6% | 98.4% |
| Σ | 13.8% | 8.3% | 77.8% | 14.8% | 10.2% | 75.0% |
| --- | --- | --- | --- | --- | --- | --- |
The study results demonstrated equivalence between the Nitrogen Laser System and the Smartbeam Laser System.
Equivalence Study MBT Biotarget 96 US IVD (Nitrogen Laser):
Studies were performed to verify and validate the use of the MBT Biotarget 96 US IVD in conjunction with the MBT-CA System (Nitrogen Laser). The study goal was to show the equivalency between the cleared US IVD MSP 48 Target Polished Steel and the new MBT Biotarget 96 US IVD. The functionality and performance of the MBT-CA System using the MBT Biotarget 96 US IVD under varying conditions was shown. Test runs were performed always in parallel using US IVD MSP 48 Target Polished Steel plates. The following studies were performed:
Repeatability and Precision:
The repeatability and precision study of the MBT Biotarget 96 US IVD was evaluated using three different testing-operators, three unique Biotarget 96 lots, three unique HCCA lots, three unique BTS lots and two unique MBT-CA Systems. The tests were performed on six working-days with two assay runs per day (microbial samples prepared as triplicate). The data showed \(100\%\) of test organisms were correctly MBT-CA identified at high confidence level \([\log (\text{score}) \geq 2.0]\) applying the MBT-CA System workflow (combination of DT, eDT and Ext procedure). The rate of MBT-CA System false identification was \(0\%\) for all test samples after final sample preparation procedure.
The study results demonstrate that both target types performed similarly.
Tolerance Range (Dynamic Range):
The Limit of Detection study was designed to establish the estimated dynamic range of sample size of Gram-negative and Gram-positive bacteria and yeast enabling automated mass spectra acquisition and species-identification using the Biotarget 96. Ten different species were analyzed representing Gram positive, Gram negative bacteria and yeast. All sample preparation techniques (DT, eDT and Ext) were conducted. Up to five dilutions steps were performed for each scenario. Cell density and cell concentration were estimated by measuring the optical density of the suspension at a wavelength of \(600\mathrm{nm}\).
The rate of MBT-CA System false identifications at low confidence level \( [\log(\text{score}) \geq 1.70] \) was 0%. The estimated dynamic range (limit of detection) of bacteria and yeast onto the MBT Biotarget 96 US IVD / US IVD MSP 48 Target Polished Steel required for MBT-CA System identification did not show significant differences.
Study results demonstrated that the estimated dynamic range for the direct (or cell equivalents using extraction) technique is \( 5 \times 10^{5} \) cells/uL.
Sample Stability prior to Matrix application:
This study was conducted to assess sample stability of microbial material (bacteria and yeasts isolates) on Biotarget 96 prior to addition of HCCA solution. Five species were
22
{23}
analyzed on two instruments. Measurement was applied directly after matrix application and after 15, 30, 60 and 120 minutes and was carried out at ambient temperature (23 ± 2) °C to simulate working conditions. All sample preparation techniques (DT, eDT and Ext) were conducted. Each sample was spotted and measured eight times. For comparison, evaluation of sample stability was performed in parallel using Steel Targets. The rate of MBT-CA System false identifications was 0% for all tests.
The study results demonstrate that both target types performed similarly.
# Sample Stability post Matrix application:
This study was conducted to access the stability of the transferred microbial samples after addition of HCCA solution on a Biotarget 96 / Steel Target the MBT-CA workflow (DT, eDT, Ext) was tested in parallel. Two temperature conditions were analyzed in this study (21°C and 25°C). All measurements were conducted using five different species on two MALDI instruments. All sample preparation techniques (DT, eDT and Ext) were performed and each sample was spotted eight times on both target types. The ready prepared sample stability was analyzed immediately after matrix application (standard condition) and after 4h, 8h and 24h of aging.
No MBT-CA false identifications occurred. The study results demonstrate that both target types performed similarly.
# Validation Study - Validation of 50 Representative Claimed Species:
This study was conducted to validate the Biotarget 96 performance using 50 different organisms 22 Gram-negative aerobic/anaerobic bacteria, 22 Gram-positive aerobic/anaerobic bacteria and six which are content of the FDA cleared MBT-CA System software and reference library. Single sample preparations for each of the test organisms were carried out using MBT-CA System preparation procedures (DT, eDT and Ext) in parallel. For comparison purposes, this validation study was performed in parallel using six steel Targets. In this study 50 claimed species representing Gram negative aerobic / anaerobic, Gram positive aerobic / anaerobic bacteria and yeast were analyzed. All sample preparation techniques (DT, eDT and Ext) were conducted. The rate of MBT-CA false identifications was 0% for this test. The overall identification rate of test samples at high confidence level (log (score) ≥2.0) was 100%.
The study results demonstrate that both target types performed similarly.
# Technical Study - Mass Accuracy/Target Edge Effects:
This study was designed for validation of Biotarget 96 in terms of "mass accuracy" using BTS, MBTCA System software and reference library and at least two microflex LT/SH mass spectrometer systems. Targets were prepared with dissolved BTS, dried and then overlaid with HCCA solution.
Two parameters were evaluated during this study: Target Flatness (the mass accuracy was used as an indicator for target flatness) and edge effects (comparison of mass accuracy from edge positions vs inner target positions). For evaluation, one prominent reference protein (e.g. 6255.4 m/z) within the mass range of the BTS was analyzed concerning mass reproducibility. Additionally, the log (scores) of each BTS spectrum were calculated as an indicator of identification performance.
# Equivalence Study MBT Biotarget 96 US IVD (Smartbeam Laser):
The study design of this study was identical to the previous section (Equivalence Study MBT Biotarget 96 US IVD (Nitrogen Laser)).
23
{24}
The following six studies were performed:
- Repeatability / Precision
- Tolerance range (dynamic range) - limit of detection (LOD)
- Sample stability prior to matrix application
- Sample stability post matrix application
- Mass Accuracy / Target Edge Effects
- Identification of 50 FDA cleared organisms according to MBT-CA System workflow.
Generally it can be stated that the technical performance between the MBT Biotarget 96 US IVD compared to the US IVD MSP 48 Target Polished Steel was comparable.
# *Nocardia Study*
This study was designed to verify and validate the measurement and identification of different strains and different species of the genus *Nocardia* by using the MBT-CA System as well as the MBT smart CA System in parallel. In addition the study was performed using the US IVD MSP 48 Target Polished Steel and the MBT Biotarget 96 US IVD in parallel. The results of this study can be evaluated as analytical study for identification performance of the MBT-CA System for the genus *Nocardia* and additionally as demonstration of equivalence between the “MBT-CA and MBT smart CA Systems” as well as equivalence of the “MBT Biotarget 96 US IVD and US IVD MSP 48 Target Polished Steel”.
Analytical Studies:
- Measurement of 30 strains covering 6 *Nocardia* species.
- DT, eDT and Ext was used for sample preparation.
- Each sample preparation technique was spotted 8 times on the targets.
- Each spot was measured twice.
- MALDI targets “US IVD MSP 48 Target Polished Steel” and “MBT Biotarget 96 US IVD” was used.
- Measurement on MBT smart CA System and MBT-CA System.
- *Nocardia* were measured after 24h, 48h and after 5 days of cultivation.
The identification performance of *Nocardia* using the standard identification MBT-CA workflow is shown in the diagram below. The general log(score) distribution was similar and the equivalence for all examined components (MBT-CA System, MBT smart CA System, MBT Biotarget 96 US IVD, and US IVD MSP 48 Target Polished Steel) was shown.
24
{25}
Table: An example all 5760 log (scores) after 48h cultivation

From 5760 spectra derived log(scores) after 48h cultivation overlaid. Each “log(score) lane” represents 480 single log(scores).
Data showed that the MBT smart System and the MBT System, as well as the MBT Biotarget 96 US IVD and US IVD MSP 48 Target Polished Steel had equivalent performance
Comparison studies:
Method comparison with predicate device:
Not applicable. Refer to the Clinical Studies section of this document.
Matrix comparison:
Not applicable
Clinical studies:
Clinical Sensitivity:
Challenge Panel
To demonstrate intra-laboratory performance, a challenge panel of 46 organisms (12 Gram-positive aerobic bacteria, eight Gram-negative aerobic bacteria, four Gram-positive microaerophilic bacteria, five Gram-negative microaerophilic bacteria, 10 Gram-positive anaerobic bacteria, four Gram-negative anaerobic bacteria, three yeasts) was tested at three study sites. All of the 46 organisms included in the panel were selected from stored organisms tested during the clinical study. A reference laboratory prepared the panel. Organism identifications were blinded to test sites. Each site tested the challenge panel member via Direct Transfer, extended Direct Transfer and Extraction procedure in parallel. Results are summarized below:
25
{26}
**Table: Challenge Panel Study Summary**
| Test procedure | Site A* | Site B** | Site C*** |
| --- | --- | --- | --- |
| | MALDI Biotyper CA System | | |
| DT method | 43/45 (96%) | 38/44 (86%) | 36/41 (88%) |
| eDT method | 44/45 (98%) | 36/44 (82%) | 38/41 (93%) |
| Ext method | 43/45 (96%) | 36/44 (82%) | 39/41 (95%) |
| **MBT-CA workflow** | **45/45 (100%)** | **40/44 (91%)** | **40/41 (98%)** |
* One (1) sample was not identified due to isolate failure to grow.
** Two (2) samples were not identified due to isolate failure to grow.
*** Five (5) samples were not identified due to isolate failure to grow.
# **Method Comparison:**
To demonstrate performance of the MALDI Biotyper CA (MBT-CA) System and the MALDI Biotyper smart CA (MBT-smart CA), a method comparison study was performed at four US clinical test sites and in-house laboratory. Fresh and stored organisms were tested on the MALDI Biotyper CA System in accordance to manufacturer's instructions for use. All organisms included in the study were sub-cultured for purity. Testing on the MBT-CA System was done from a freshly isolated colony.
All organisms included in the study and tested by US study sites were also sub-cultured onto an agar slant or appropriate media for isolation and shipped to the study reference laboratory. The reference laboratory stored all organisms included in the study and when requested, sent organisms to the sequencing reference laboratory for 16S rRNA or ITS sequencing and protein gene sequencing when requested. In-house laboratory and some testing sites processed their own isolates to the sequencing reference laboratory. The results are summarized below:
26
{27}
Table: Overall Performance
| Overall Performance | | | | |
| --- | --- | --- | --- | --- |
| MBT-CA RESULT | REFERENCE ALGORITHM | | | Total |
| | high resolution species | low resolution species / genus | Negative | |
| Organism ID ≥ 2.0 (High Confidence) | 1906 | 131 | 18^{1)} + 2) | 2055 |
| Organism ID (≥1.7; <2.0) (Low Confidence) | 23 | 4 | 5^{3)} + 4) | 32 |
| - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | 3^{5)} | 1^{5)} | n/a | 4 |
| Total | 1932 | 136 | 23 | 2091 |
| Positive | | Negative |
| --- | --- | --- |
| high confidence ID | high & low confidence ID | |
| 98.5% | 99.8% | n/a |
Discordant results:
1) MBT-CA Organism ID >2.0; Correct Genus ID - Incorrect Species ID (17 isolates); Reference Method reported a different organism.
2) MBT-CA Organism ID >2.0; Incorrect Genus ID (4 isolates); Reference Method reported a different organism.
3) MBT-CA Organism ID (≥1.7; <2.0); Correct Genus ID - Incorrect Species ID (3 isolates); Reference Method reported a different organism.
4) MBT-CA Organism ID (≥1.7; <2.0); Incorrect Genus ID (1 isolate); Reference Method reported a different organism.
5) MBT-CA Organism ID <1.7; no ID (4 isolates); Reference Method reported a species from the included in the MBT-CA Reference Library.
| Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA <1.7 | |
| RESOLUTION REFERENCE ALGORITHM | | | | | | | | | | | | | |
| high | low | high | low | high | low | high | low | high | low | high | low | high | low |
| 1906 | 131 | 23 | 4 | 5 | 9 | 2 | 2 | 0 | 4 | 0 | 1 | 3 | 1 |
| 91.2% | 6.3% | 1.1% | 0.2% | 0.2% | 0.4% | 0.1% | 0.1% | 0.0% | 0.2% | 0.0% | 0.0% | 0.1% | 0.0% |
| 2037 | | 27 | | 14 | | 4 | | 4 | | 1 | | 4 | |
| 97.4% | | 1.29% | | 0.67% | | 0.19% | | 0.19% | | 0.05% | | 0.19% | |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | Incorrect: Genus ID | no ID |
| --- | --- | --- | --- |
| 2064 | 18 | 5 | 4 |
| 98.7% | 0.86% | 0.24% | 0.19% |
27
{28}
Table: Overall Performance – Bacteria:
| Performance BACTERIA | | | | |
| --- | --- | --- | --- | --- |
| MBT-CA RESULT | REFERENCE ALGORITHM | | | |
| | high resolution species | low resolution species / genus | Negative | Total |
| Organism ID ≥ 2.0 (High Confidence) | 1827 | 130 | 21 | 1978 |
| Organism ID (≥1.7; <2.0) (Low Confidence) | 21 | 5 | 4 | 30 |
| - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | 2 | 1 | n/a | 3 |
| Total | 1850 | 136 | 25 | 2011 |
| Positive | | Negative |
| --- | --- | --- |
| high resolution | high & low resolution | |
| 98.54% | 99.85% | n/a |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA <1.7 | |
| RESOLUTION REFERENCE ALGORITHM | | | | | | | | | | | | | |
| high | low | high | low | high | low | high | low | high | low | high | low | high | low |
| 1829 | 131 | 21 | 4 | 5 | 9 | 2 | 2 | 0 | 4 | 0 | 1 | 2 | 1 |
| 90.9% | 6.5% | 1.0% | 0.2% | 0.2% | 0.4% | 0.1% | 0.1% | 0.0% | 0.2% | 0.0% | 0.0% | 0.1% | 0.0% |
| 1960 | | 25 | | 14 | | 4 | | 4 | | 1 | | 3 | |
| 97.5% | | 1.24% | | 0.70% | | 0.20% | | 0.20% | | 0.05% | | 0.15% | |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | Incorrect: Genus ID | no ID |
| --- | --- | --- | --- |
| 1985 | 18 | 5 | 3 |
| 98.7% | 0.90% | 0.25% | 0.15% |
28
{29}
Table: Overall Performance – Gram Negative Bacteria:
| Performance Gram negative bacteria | | | | |
| --- | --- | --- | --- | --- |
| MBT-CA RESULT | REFERENCE ALGORITHM | | | Total |
| | high resolution species | low resolution species / genus | Negative | |
| Organism ID ≥ 2.0 (High Confidence) | 728 | 26 | 7 | 761 |
| Organism ID (≥1.7; <2.0) (Low Confidence) | 0 | 0 | 4 | 4 |
| - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | 0 | 0 | n/a | 0 |
| Total | 728 | 26 | 11 | 765 |
| Positive | | Negative |
| --- | --- | --- |
| high confidence ID | high & low confidence ID | |
| 100.0% | 100.0% | n/a |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA <1.7 | |
| RESOLUTION REFERENCE ALGORITHM | | | | | | | | | | | | | |
| high | low | high | low | high | low | high | low | high | low | high | low | high | low |
| 728 | 26 | 0 | 0 | 3 | 0 | 2 | 1 | 0 | 4 | 0 | 1 | 0 | 0 |
| 95.2% | 3.4% | 0.0% | 0.0% | 0.4% | 0.0% | 0.3% | 0.1% | 0.0% | 0.5% | 0.0% | 0.1% | 0.0% | 0.0% |
| 754 | | 0 | | 3 | | 3 | | 4 | | 1 | | 0 | |
| 98.6% | | 0.00% | | 0.39% | | 0.39% | | 0.52% | | 0.13% | | 0.00% | |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| 754 | | | | 6 | | | | 5 | | | | 0 | |
| 98.6% | | | | 0.78% | | | | 0.65% | | | | 0.00% | |
29
{30}
Table: Overall Performance – Gram Positive Bacteria:
| Performance Gram positive bacteria | | | | |
| --- | --- | --- | --- | --- |
| MBT-CA RESULT | REFERENCE ALGORITHM | | | Total |
| | high resolution species | low resolution species / genus | Negative | |
| Organism ID ≥ 2.0 (High Confidence) | 1101 | 105 | 11 | 1217 |
| Organism ID (≥1.7; <2.0) (Low Confidence) | 21 | 4 | 1 | 26 |
| - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | 2 | 1 | n/a | 3 |
| Total | 1124 | 110 | 12 | 1246 |
| Positive | | Negative |
| --- | --- | --- |
| high confidence ID | high & low confidence ID | |
| 97.7% | 99.8% | n/a |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA <1.7 | |
| RESOLUTION REFERENCE ALGORITHM | | | | | | | | | | | | | |
| high | low | high | low | high | low | high | low | high | low | high | low | high | low |
| 1101 | 105 | 21 | 4 | 2 | 9 | 0 | 1 | 0 | 0 | 0 | 0 | 2 | 1 |
| 88.4% | 8.4% | 1.7% | 0.3% | 0.2% | 0.7% | 0.0% | 0.1% | 0.0% | 0.0% | 0.0% | 0.0% | 0.2% | 0.1% |
| 1206 | | 25 | | 11 | | 1 | | 0 | | 0 | | 3 | |
| 96.8% | | 2.01% | | 0.88% | | 0.08% | | 0.00% | | 0.00% | | 0.24% | |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | Incorrect: Genus ID | no ID |
| --- | --- | --- | --- |
| 1231 | 12 | 0 | 3 |
| 98.8% | 0.96% | 0.00% | 0.24% |
30
{31}
Table: Overall Performance - Yeast:
| Performance YEAST | | | | |
| --- | --- | --- | --- | --- |
| MBT-CA RESULT | REFERENCE ALGORITHM | | | Total |
| | high resolution species | low resolution species / genus | Negative | |
| Organism ID ≥ 2.0 (High Confidence) | 77 | 0 | 0 | 77 |
| Organism ID (≥1.7; <2.0) (Low Confidence) | 2 | 0 | 0 | 2 |
| - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | 1 | 0 | n/a | 1 |
| Total | 80 | 0 | 0 | 80 |
| Positive | | Negative |
| --- | --- | --- |
| high confidence ID | high & low confidence ID | |
| 96.3% | 98.8% | n/a |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA <1.7 | |
| RESOLUTION REFERENCE ALGORITHM | | | | | | | | | | | | | |
| high | low | high | low | high | low | high | low | high | low | high | low | high | low |
| 77 | 0 | 2 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 |
| 96.3% | 0.0% | 2.5% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 1.3% | 0.0% |
| 77 | | 2 | | 0 | | 0 | | 0 | | 0 | | 1 | |
| 96.3% | | 2.50% | | 0.00% | | 0.00% | | 0.00% | | 0.00% | | 1.25% | |
| Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| 79 | | | | 0 | | | | 0 | | | | 1 | |
| 98.8% | | | | 0.00% | | | | 0.00% | | | | 1.25% | |
31
{32}
### Performance - single species:
Interpretation of Bruker Performance Report “Single Species”:
The performance table of the single species Arthrobacter cumminsii was selected exemplarily to illustrate the meanings of the single sectors within the table:
| Arthrobacter cumminsii | | | | |
| --- | --- | --- | --- | --- |
| MBT-CA RESULT | REFERENCE ALGORITHM | | | Total |
| | high resolution species | low resolution species / genus | Negative | |
| Organism ID ≥ 2.0 (High Confidence) | 4 | 8\( ^{a} \)(1) | 0 | 12 |
| Organism ID (≥1.7; <2.0) (Low Confidence) | 1 | 1 | 0 | 2 |
| - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | 0 | 0 | | 0 |
| Total | 5 | 9 | 0 | 14 |
| # One isolate was identified with high confidence by eDT procedure only. Matching hint is included in the labeling. | | | | |
| MBT-CA RESULT | REFERENCE ALGORITHM | | | Total |
| --- | --- | --- | --- | --- |
| | high resolution species | low resolution species / genus | Negative | |
| Organism ID ≥ 2.7 (High Confidence) | I | II | E | |
| Organism ID (≥1.7; <2.0) (Low Confidence) | III | IV | F | |
| - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | G | H | | |
| Total | | | | |
| comment field (J) | | | | |
The MBT-CA results as well as the results of the reference algorithm (gold standard) were subdivided in three parts. The MBT-CA results can be interpreted as “high confidence species ID”, “low confidence species ID” and (combined) “incorrect ID” and “no ID”. The reference algorithm ID was subdivided in “high resolution species” if the results was unambiguous and “low resolution species / genus” if the result was not unambiguous. Additionally the result of the reference algorithm could be different from the MBT-CA result (“negative”).
| I | The MBT-CA reported “high confidence species ID” and the result of the reference algorithm was unambiguous (both reported A. cumminsii). |
| --- | --- |
| II | The MBT-CA reported “high confidence species ID” and the result of the reference algorithm was not unambiguous (the MBT-CA reported A. cumminsii and the reference algorithm reported A. cumminsii and A. albus equivalently). |
| III | The MBT-CA reported “low confidence species ID” and the result of the reference algorithm was unambiguous (both reported A. cumminsii). |
| IV | The MBT-CA reported “low confidence species ID” and the result of the reference algorithm was not unambiguous (the MBT-CA reported A. cumminsii and the reference algorithm reported A. cumminsii and A. albus equivalently). |
| E | The MBT-CA reported “high confidence species ID” and the result of the reference algorithm was different (the MBT-CA would report A. cumminsii but the reference algorithm would report a different species, false positive for A. cumminsii). |
| F | The MBT-CA reported “low confidence species ID” and the result of the reference algorithm was different (the MBT-CA would report A. cumminsii but the reference algorithm would report a different species, false positive for A. cumminsii). |
| G | The MBT-CA reported “incorrect ID” or “no ID” and the result of the reference algorithm was unambiguous (the MBT-CA would report a different species or no ID but the reference algorithm would report A. cumminsii, false negative for A. cumminsii). |
| H | The MBT-CA reported “incorrect ID” or “no ID” and the result of the reference algorithm was not unambiguous (the MBT-CA would report a different species or no ID but the reference algorithm would report A. cumminsii, false negative for A. cumminsii). |
| J | Room for further information. |
32
{33}
In addition to the performance table further tables with deeper differentiation of the incorrect results were created for demonstrating the MBT-CA performance. The incorrect MBT-CA results were divided into “correct genus, incorrect species”, “incorrect genus” and “no ID”.
The transition, accumulation and counting of data are shown below:
| Single Species Performance Table | | | | | Single Species Performance Table | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| MBT-CA RESULT | REFERENCE ALGORITHM | | | Total | MBT-CA RESULT | REFERENCE ALGORITHM | | | Total |
| | high resolution species | low resolution species / genus | Negative | | | high resolution species | low resolution species / genus | Negative | |
| Organism ID ≥ 2.0 (High Confidence) | I | II | E | | Organism ID ≥ 2.0 (High Confidence) | I | II | A + B + E1 + E2 | |
| Organism ID (≥1.7; <2.0) (Low Confidence) | III | IV | F | | Organism ID (≥1.7; <2.0) (Low Confidence) | III | IV | C + D + F1 + F2 | |
| - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | G | H | | | - INCORRECT MBT-CA ID (≥1.7) - NO ID (<1.7) | G | H | | |
| Total | | | | | Total | | | | |
| species | Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA <1.7 | |
| | RESOLUTION REFERENCE ALGORITHM | | | | | | | | | | | | | |
| | high | low | high | low | high | low | high | low | high | low | high | low | high | low |
| | I | II | III | IV | A | B | C | D | E1 | E2 | F1 | F2 | G | H |
| Arthrobacter cumminsii | 4 | 8 | 1 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| | 28.6% | 57.1% | 7.1% | 7.1% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% |
| | 12 | | 2 | | 0 | | 0 | | 0 | | 0 | | 0 | |
| | 85.7% | | 14.3% | | 0.0% | | 0.0% | | 0.0% | | 0.0% | | 0.0% | |
| | 14 | | | | | | | | | | | | | |
| | 100.0% | | | | | | | | | | | | | |
| species | Correct: Genus ID Correct: Species ID or Group / Complex ID | | | | Correct: Genus ID Incorrect: Species ID or Group / Complex ID | | | | Incorrect: Genus ID | | | | no ID | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA ≥2.0 | | MBT-CA ≥1.7 to <2.0 | | MBT-CA <1.7 | |
| | RESOLUTION REFERENCE ALGORITHM | | | | | | | | | | | | | |
| | high | low | high | low | high | low | high | low | high | low | high | low | high | low |
| Arthrobacter cumminsii | 4 | 8 | 1 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| | 28.6% | 57.1% | 7.1% | 7.1% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% | 0.0% |
| | 12 | | 2 | | 0 | | 0 | | 0 | | 0 | | 0 | |
| | 85.7% | | 14.3% | | 0.0% | | 0.0% | | 0.0% | | 0.0% | | 0.0% | |
| | 14 | | | | | | | | | | | | | |
| | 100.0% | | | | | | | | | | | | | |
All incorrect or “no ID” results of the performance tables from the single species (sections E and F in the above left table) were separately considered in…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.