RX DC

K163519 · Cefla S.C. · EHD · Jan 13, 2017 · Dental

Device Facts

Record IDK163519
Device NameRX DC
ApplicantCefla S.C.
Product CodeEHD · Dental
Decision DateJan 13, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1800
Device ClassClass 2
AttributesPediatric

Indications for Use

RX DC x-ray unit is designed for use in the dental surgery to make endo-oral x-rays for diagnostic purposes. This equipment can be used to produce traditional x-rays developed using chemicals or, alternatively, it can be used with digital x-ray sensors.

Device Story

RX DC is an extraoral dental X-ray system for intraoral imaging. System comprises a mobile/articulate support arm, control unit, and tube head with X-ray tube. High-frequency, constant-potential generator produces X-rays at 65/70 kV and 8 mA. Device automatically calculates exposure time (0.02s–1.00s) based on selected tooth type and patient size. Used in dental clinics by dental professionals. Output is X-ray radiation for capture on conventional film or digital sensors. Healthcare providers use images for diagnostic assessment of teeth and oral structures. Benefits include optimized exposure parameters for diagnostic quality imaging.

Clinical Evidence

Bench testing only. Safety and EMC testing conducted per IEC 60601-1, 60601-1-6, 62366, 60601-1-3, 60601-2-65, and 60601-1-2. Comparative dosimetric testing measured air kerma across loading factor combinations (patient size/tooth anatomy) against predicate K080076, confirming equivalent X-ray emission performance.

Technological Characteristics

High-frequency, constant-potential X-ray generator; tungsten anode; 65/70 kV tube voltage; 0.4mm/0.7mm focal spot. Articulated support arm (40-90cm). Wired/wireless control. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 62304. Microprocessor-controlled exposure.

Indications for Use

Indicated for dental surgery patients requiring endo-oral x-ray imaging for diagnostic purposes.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 13, 2017 CEFLA S.C. % Maurizio Pantaleoni CEO Isemed Srl Via Argentina Altobelli Bonetti 3A Imola, BO 40026 ITALY Re: K163519 Trade/Device Name: RX DC Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD Dated: November 24, 2016 Received: December 15, 2016 Dear Maurizio Pantaleoni: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K163519 Device Name RXDC Indications for Use (Describe) RX DC x-ray unit is designed for use in the dental surgery to make endo-oral x-rays for diagnostic purposes. This equipment can be used to produce traditional x-rays developed using chemicals or, alternatively, it can be used with digital x-ray sensors. Type of Use (Select one or both, as applicable) IZJ Prescription Use (Part 21 CFR 801 Subpart D) D Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF VolOOl-20 {3}------------------------------------------------ # 510(k) Summary This 510(k) Summary is being submitted as required by 21 CFR 807.92. #### 1. General Information | Submitter : | CEFLA S.C.<br>Via Selice Prov.le 23/a<br>Imola, BO 40026 ITALY<br>Tel. +39 0542 653441<br>Fax +39 0542 653607 | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Consultant/ Contact: | Maurizio Pantaleoni<br>ISEMED srl<br>Via Altobelli Bonetti 3/a<br>Imola, BO 40026 ITALY<br>Mob. +39 348 4435155<br>Tel. +39 0542 683803<br>Fax +39 0542 698456<br>Email: regulatory@isemed.eu | ### Summary Prepared Date: November 24, 2016 ## 2. Names | Device Name: | RX DC | |------------------|-----------------------------------| | Common Name: | Unit, X-Ray, Extraoral with timer | | Regulation Name: | Extraoral Source X-Ray System | | Product Code: | EHD | | Classification: | 21CFR 872.1800; Class II | {4}------------------------------------------------ #### Predicate Devices 3. | Applicant | Device name | 510(k) Number | |-------------|-----------------|---------------| | CEFLA S.C. | RXDC | K080076 | | VIVI S.r.l. | VIVI Ergon-X HF | K120318 | The RX DC, in the new 65/70 kV version, is substantially equivalent to the following device: #### 4. Device Description The subject device RX DC, in the new 65/70 kV version, is an extraoral source dental X-ray system intended for intraoral imaging. The subject device is a device comprises a double mobile and articulate support arm. At the opposite ends of the arm are located, respectively: - . A control unit equipped with wall plate, extension arm and wired/wireless control device; - A tube head with x-ray tube; . X-rays are produced using the high frequency and constant potential generators with a built in round collimator with the high frequency (HF) technology, X-ray emission at 70 kV and 8 mA (maximum power), and x-ray unit automatically calculates the best exposure time (from 0.02 s to 1.00 s) based on the selected tooth and patient size, as well as an adjustable arm allows for easy positioning. The system can be used either with conventional film or a digital imaging system. RX DC, 65/70 kV version, is manufactured by CEFLA S.C. and available in several versions, and it can be sold under different brands and commercial names for commercial needs, as well as with different type of installation, and also different types of maximum anode voltage (65 kV or 70 kV). To identify the different variants of the RX DC product, CEFLA S.C. designed a reference system to identify any variant via the REF code. ## 5. Indications for Use RX DC x-ray unit is designed for use in the dental surgery to make endo-oral x-rays for diagnostic purposes. This equipment can be used to produce traditional x-rays developed using chemicals or, alternatively, it can be used with digital x-ray sensors. ## 6. Comparison of technological characteristics with the predicate device The RX DC, in the new 65/70 kV version, represents a development of the cleared RX DC (K080076), which has a fixed x-ray tube potential 60kV. In addition, concerning the maximum voltage of 70KV, this parameter is identical to that of the other predicate device K120318. {5}------------------------------------------------ ## CEFLA S.C. 510(K) PREMARKET NOTIFICATION All relevant characteristics of the subject device are side-by-side compared with those of the predicate devices in the Comparison table. | | Subject Device | Predicate Devices | | | | |-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | | CEFLA S.C.<br>RX DC | CEFLA S.C.<br>RX DC | VIVI<br>Vivi Ergon- X HF | | | | | | K 080076 | K 120318 | | | | General Features | | | | | | | Classification | II | II | II | | | | Product Code | EHD | EHD | EHD | | | | Intended use | This x-ray unit is designed for use<br>in the dental surgery to make endo-<br>oral x-rays for diagnostic<br>purposes. This equipment can be<br>used to produce traditional x-rays<br>developed using chemicals or,<br>alternatively, it can be used with<br>digital x-ray sensors. | This x-ray unit is designed for use<br>in the dental surgery to make endo-<br>oral x-rays for diagnostic<br>purposes. This equipment can be<br>used to produce traditional x-rays<br>developed using chemicals or,<br>alternatively, it can be used with<br>digital x-ray sensors. | Ergon-X-HF is an extraoral X-<br>ray source system intended to<br>be used for dental<br>radiographic examination and<br>diagnosis of diseases of the<br>teeth, jaws, and oral<br>structures. | | | | Principle of use | X-Ray Tube | X-Ray Tube | X-Ray Tube | | | | Mechanical Features | | | | | | | Installation<br>configuration | Wall-mounted standard version<br>Stand mobile version | Wall-mounted standard version | Wall-mounted standard<br>version<br>Fixed floor Stand<br>Stand mobile version | | | | X-ray emission<br>control | Wired control<br>Wireless Remote control | Remote control | Wired control<br>Wireless remote control | | | | Extension arm | 90 cm (standard)<br>40 cm (optional)<br>60 cm (optional) | 90 cm (standard)<br>40 cm (optional)<br>60 cm (optional) | 30 cm (standard)<br>60 cm (optional)<br>80 cm (optional) | | | | Radiological characteristics | | | | | | | HV generator | High frequency<br>Constant potential | High frequency<br>Constant potential | High Frequency<br>Constant potential | | | | Anode material | Tungsten | Tungsten | Tungsten | | | | Tube Voltage (KV) | 60, 63, 65 kV<br>60,65, 70 kV | 60 kV | 60, 65 or 70 kV | | | | Tube current (mA) | 6, 7 mA @65kV<br>4, 8 mA @ 70 kV | 7mA<br>3.5mA | 7 mA | | | | Exposure time | 0.02s-1s (R 20 steps) | 0.01sec. - 1sec. (in R20 steps) | 0.01 - 2.0 sec | | | | X-ray tube & anode<br>angle | • From 12.5° to 16° | 12.5° | 19° | | | | Focal spot size | • 0.4mm / 0.7mm | 0.4mm | 0.5mm | | | | Leakage radiation | < 0.25 mGy/h | < 0.25 mGy/h | < 0.25 mGy/h | | | {6}------------------------------------------------ | Collimators | | | | |-----------------------------|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | Focus film distance | • Short round (fix): 200 mm (8'')<br>• Long rectangular (removable): 300mm (12'')<br>• Round (removable): 300mm (12'') | • Short round (fix): 200 mm (8'')<br>• Long rectangular (removable): 300mm (12'') | 200 mm | | Diameter of X-ray beam cone | • Short round (fix): Ø 60 mm<br>• Long rectangular (removable): Ø 45x35mm<br>• Round (removable): Ø 55 mm | • Short round (fix): Ø 60 mm<br>• Long rectangular (removable): Ø 45x35mm | Ø 60 mm | | Radiological parameters | | | | | Exposure times control | Microprocessor controlled exposure times<br>preset loading factors<br>or<br>Manual mode | Microprocessor controlled exposure times<br>preset loading factors | Microprocessor controlled exposure times<br>automatic exposure<br>or<br>manual exposure | | Exposure modes | | | | | Selectable parameter | Patient type, anatomical position, | Patient type, anatomical position, | Patient size, anatomical position, film type | | Patient type | Adult-child | Adult -child | | | Tooth type | Molar (upper and lower)<br>Premolars (upper and lower)<br>Incisors/canines (upper and lower)<br>Bite wing | Molar (upper and lower)<br>Premolars (upper and lower)<br>Incisors/canines (upper and lower)<br>Bite wing | Molar (upper and lower)<br>Premolars (upper and lower)<br>Incisors/canines (upper and lower)<br>Bite wing | | Film type | Sensor or photostimulated plate | Sensor or photostimulated plate | Class E - F - D (selectable) | | Standards | | | | | Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 62304 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 62304 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 62304 | According to table above, the subject device has similar technology and features than the predicate devices K080076 and K120318, pointing out only few differences. The differences like the highest energy available is 70 keV or the necessity to manage a wired controller together an others, together with some electronic changes introduced in the subject device, as indicated in the table, have caused the need for software changes in order to allow the firmware porting, but the two mains parts of the software, that consist of the user interface management and the control algorithms that manage the x-ray tube remained substantially the same. Finally this modification has been validated in compliance with the same standard used in the predicate device as indicated in the comparison table Furthermore all these differences have been addressed by dedicated performance tests demonstrating that the technical differences have the equivalent results concerning the performance of the dose released during the radiographic exposition. Therefore, the subject device is able to produce comparable performances in term of emitted dose respect the correspondent dose emitted by RX DC cleared by FDA with K080076. {7}------------------------------------------------ ### 7. Performance Data The following tests were performed for determination of substantial equivalence: 1) Non clinical tests performed on the subject device: A. Safety and EMC tests conducted in compliance with the declared standards: - IEC 60601-1:2005 ● - IEC 60601-1-6:2013 ● - IEC 62366: 2014 ● - IEC 60601-1-3:2013 ● - IEC 60601-2-65:2012 - IEC 60601-1-2:2007 (EMC) ● For all consensus standards here above all requirements have been met. ### B. Comparative dosimetric test: The test purpose is to measure the air kerma emitted by the subject device and RX DC (K080076), for each combination of loading factors accordingly to patient size and anatomical shape of the tooth, in order to verify that the technical differences have the equivalent results concerning the performance of the dose released during the radiographic exposition The test results demonstrated that RX DC, in the new 65/70 kV version, emits the same quantity of x-ray than the predicate device K080076. ### 8. Conclusions In light of evidence discussed above and underlining similar technological features for these three compared devices as well as pointing out the compliance with the same safety-related standards and the similar performance as demonstrated by dedicated performance test, the subject device may be found substantially equivalent to the predicate devices K080076, and K120318. {8}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...