The Poseidon Ultrasound System is indicated for use by qualified and trained healthcare professionals to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediatric patients for the following clinical applications: Peripheral Vessel (including carotid and arterial studies), Procedural Guidance, Small Organs (including thyroid), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional) and Musculoskeletal (superficial).
Device Story
Poseidon Ultrasound System is a general-purpose diagnostic imaging system; consists of a single transducer with broad imaging capabilities; connects via USB to a standard handheld commercial off-the-shelf (COTS) mobile device. Device uses capacitive micromachined ultrasonic transducer (CMUT) array technology to capture ultrasound signals; transforms inputs into B-mode, M-mode, and Color Flow Doppler imagery. Operated by qualified healthcare professionals in clinical settings. User interface on mobile device provides touchscreen menus, buttons, and navigation icons for system control, image viewing, and measurement. Output assists clinicians in visualizing anatomical structures and fluids, supporting diagnostic decision-making and procedural guidance. Benefits include portable, broad-capability diagnostic imaging.
Clinical Evidence
Bench testing only. Verification and validation activities were performed to demonstrate compliance with predetermined performance specifications and safety standards, including IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, and NEMA UD-2.
Technological Characteristics
System utilizes a capacitive micromachined ultrasonic transducer (CMUT) array. Supports B-mode, M-mode, and Color Flow Doppler imaging. Connectivity via USB to a COTS mobile device. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-37 (ultrasound safety), ISO 10993 (biocompatibility), and NEMA UD-2 (acoustic output).
Indications for Use
Indicated for diagnostic ultrasound imaging and measurement of anatomical structures and fluids in adult and pediatric patients across peripheral vessel, procedural guidance, small organs, cardiac, abdominal, urology, fetal/obstetric, gynecological, and musculoskeletal applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Philips Healthcare, Lumify Ultrasound System (K162549)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 6, 2017
Butterfly Network, Inc. % Mr. Brian Sawin Sr. Regulatory Affairs Manager 530 Old Whitfield Street GUILFORD CT 06437
Re: K163510
Trade/Device Name: Poseidon Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 9, 2017 Received: August 10, 2017
Dear Mr. Sawin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara
Robert A. Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163510
Device Name Poseidon Ultrasound System
#### Indications for Use (Describe)
The Poseidon Ultrasound System is indicated for use by qualified and trained healthcare professionals to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediativ for the following clinical applications: Peripheral Vessel (including carotid and arterial studies), Procedural Guidance, Small Organs (including thyroid), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional) and Musculoskeletal (superficial).
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> <span>Prescription Use (Part 21 CER 801 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------|
| <div> <span>☐</span> <span>Over-The-Counter Use (21 CER 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Diagnostic Ultrasound Indications For Use Table
System: Poseidon Ultrasound System
#### Transducer: Poseidon Ultrasound System transducer
#### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|--------------------------------------------|-------------------|---|-----|-----|--------------------|-----------------------|---------------------|
| General (Track 1<br>Only) | Specific (Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal/Obstetric | N | N | | | N | B mode + M<br>mode | |
| | Abdominal | N | N | | | N | B mode + M<br>mode | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | N | | | N | B mode + M<br>mode | |
| | Small Organ (including<br>thyroid) | N | N | | | N | B mode + M<br>mode | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | N | N | | | N | B mode + M<br>mode | |
| | Intravascular | | | | | | | |
| | Other (Musculo-skeletal -<br>Conventional) | N | N | | | N | B mode + M<br>mode | |
| | Other (Gynecological) | N | N | | N | B mode + M<br>mode | | |
| | Other (Urology) | N | N | | N | B mode + M<br>mode | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | N | N | | N | B mode + M<br>mode | | |
| | Cardiac Pediatric | N | N | | N | B mode + M<br>mode | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral Vessel | Peripheral vessel | N | N | | N | B mode + M<br>mode | | |
| | Other (Carotid, arterial<br>studies) | N | N | | N | B mode + M<br>mode | | |
| | Other (Procedural Guidance) | N | N | | N | B mode + M<br>mode | | |
{4}------------------------------------------------
N = new indication; P = previously cleared by FDA; E = added under this appendix
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Butterfly Network. The logo consists of a blue butterfly icon on the left and the words "Butterfly Network" in black on the right. The word "Butterfly" is in a larger font than the word "Network", and there is a trademark symbol next to the word "Butterfly".
## 510(k) Summary of Safety and Effectiveness
### Submitter Information
#### Submitter Name and Address
Butterfly Network, Inc. 530 Old Whitfield St. Guilford, CT 06437 USA (tel.) 203.204.6600 (fax) 203.458.2514 www.butterflynetwork.com
#### Contact Person
Brian Sawin Sr. Requlatory Affairs Manager 203.204.6600 bsawin@butterflynetinc.com
#### Date Prepared
August 9, 2017
### Subject Device - Proprietary/Trade Name
Poseidon Ultrasound System
## Subject Device - Common Name
Ultrasound Imaging System
#### Classification Name
| | Regulation Number | Product Code |
|------------------------------------------|-------------------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
### Classification
Class II
#### Predicate Device:
K162549 - Philips Healthcare, Lumify Ultrasound System (Clearance Date:
{6}------------------------------------------------
## 10/03/2016)
## Device Summary:
The Poseidon Ultrasound System is a general-purpose diagnostic imaging system for use by qualified and trained healthcare professionals enabling visualization and measurement of anatomical structures and fluid on adult and pediatric patients. The system consists of a single transducer with broad imaging capabilities connected via a USB cable to a standard handheld commercial off the shelf (COTS) mobile device. In addition to M-mode and B-mode imaging the instrument also supports Color Flow Doppler imaging.
The user interface includes touchscreen menus, buttons, controls, indicators and navigation icons that allow the operator to control the system and to view and measure ultrasound imagery.
## Indications for Use:
The Poseidon Ultrasound System is indicated for use by qualified and trained healthcare professionals to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediatric patients for the following clinical applications: Peripheral Vessel (including carotid and arterial studies), Procedural Guidance, Small Organs (including thyroid), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional) and Musculoskeletal (superficial).
## Summary of Technological Characteristics
There are no technological characteristics, features or indications for use in this submission that are not previously evaluated and cleared in the predicate device, with the exception of utilizing a capacitive micromachined ultrasonic transducer (CMUT) array rather than a traditional crystalline piezoelectric array. This technology meets the same intended use and performs the same actions as piezoelectric material.
## Summary of Safety and Performance
Verification and validation activities were designed and performed to demonstrate that the Poseidon Ultrasound System meets predetermined performance specifications. The following standards were used to determine appropriate methods for evaluating the performance of the device:
IEC 60601-1: Medical Electrical Equipment – Part 1: General Requirements for Safety.
IEC 60601-1-2: Medical Electrical Equipment - Part 1-2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
IEC 60601-2-37: Medical Electrical Equipment – Part 2-37: Particular
{7}------------------------------------------------
Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment.
ISO 10993:2009 Biological Evaluation of Medical Devices. Part 1
NEMA UD-2: Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment.
## Summary of Substantial Equivalence:
Based on the indications for use, technological characteristics, and safety and performance testing, the subject device meets the requirements that are considered adequate for its intended use and is substantially equivalent in design, principles of operation and indications for use to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.