R2P Metacross RX

K163479 · KANEKA Corporation · LIT · Jan 11, 2017 · Cardiovascular

Device Facts

Record IDK163479
Device NameR2P Metacross RX
ApplicantKANEKA Corporation
Product CodeLIT · Cardiovascular
Decision DateJan 11, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The R2P Metacross RX PTA balloon dilatation catheter is intended to dilate stenoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Device Story

R2P Metacross RX is a rapid exchange PTA balloon dilatation catheter; used for peripheral artery stenosis dilation and AV fistula treatment. Device comprises distal/proximal tubes, balloon, guidewire transition tube, radiopaque markers, hub, and strain relief. Operated by physician in clinical setting; guidewire (max 0.035 inch) inserted through distal lumen; balloon inflated/deflated via inflation device connected to proximal hub. Enables vessel dilation or stent post-dilatation; improves blood flow in peripheral vasculature. Includes flushing needle for heparinized saline preparation.

Clinical Evidence

Bench testing only. Testing included dimensional verification, balloon rated burst pressure, fatigue, compliance, inflation/deflation time, bond strength, flexibility, kink test, torque strength, particulate evaluation, and shelf life. Biocompatibility leveraged from predicate device.

Technological Characteristics

Rapid exchange PTA balloon catheter; 2000 mm working length; 3.0-8.0 mm balloon diameters; 20-200 mm balloon lengths. Compatible with 6 Fr sheath introducers and 0.035 inch guidewires. Materials identical to predicate. Sterilization method not specified.

Indications for Use

Indicated for patients requiring dilation of stenotic iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, or treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae, and for peripheral stent post-dilatation.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. The profiles are all connected to each other, forming a single, unified image. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the image. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 11, 2017 Kaneka Corporation % Mr. Christopher Sloan Principal Consultant Quintiles Consulting 1801 Rockville Pike, Suite 300 Rockville, MD 20852 Re: K163479 Trade/Device Name: R2P Metacross RX PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: December 9, 2016 Received: December 12, 2016 Dear Mr. Sloan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Fernando Aguel Fernando Aguel -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K163479 #### Device Name R2P Metacross RX PTA Balloon Dilatation Catheter Indications for Use (Describe) The R2P Metacross RX PTA balloon dilatation catheter is intended to dilate stenoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY # R2P Metacross RX PTA Balloon Dilatation Catheter #### 510(k) Submitter Kaneka Corporation Medical Device Division 1-12-32, Akasaka, Minato-ku Tokyo, Japan Contact Person: Toshihiko Motomine Telephone: +81-3-5574-8023 Email: Toshihiko.Motomine@kaneka.co.jp #### Official Correspondent Christopher M. Sloan Principal Consultant Quintiles Consulting 1801 Rockville Pike, Suite 300 Rockville, MD 20852 Phone: (301) 272-3114 Email: chris.sloan@quintilesims.com Date Prepared: December 9, 2016 {4}------------------------------------------------ #### Subject Device Name: | Trade Name | R2P Metacross RX PTA Balloon Dilatation<br>Catheter | |----------------------|----------------------------------------------------------------| | Common or usual name | Percutaneous (peripheral) transluminal<br>angioplasty catheter | | Classification name | Percutaneous catheter [21 CFR 870.1250;<br>product code (LIT)] | | Class | II | | Classification Panel | Cardiovascular (74) | #### Predicate Devices: - Metacross RX PTA Balloon Dilatation Catheter [K150865 (Kaneka Corporation)] ● #### Device Description: R2P Metacross RX PTA Balloon Dilatation Catheter (R2P Metacross RX) is a rapid exchange balloon dilation catheter, which consists of a distal tube, proximal tube, balloon, guidewire transition tube, radiopaque markers, hub, and strain relief. A balloon is attached to the distal end of the catheter, and it can be inflated and deflated using the inflation device connected to the hub at the proximal end. R2P Metacross RX has a lumen (guidewire transition tube) at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the balloon and distal tube to the guidewire port for the exit of a guidewire. The maximum compatible diameter of a guidewire used together with R2P Metacross RX in a PTA procedure is 0.035 inches. Additionally, sheath introducers with a minimum diameter of 6 Fr have been deemed to be compatible with R2P Metacross RX. The nominal inflated balloon diameters range from 3.0 mm to 8.0 mm with balloon working lengths of 20 mm to 200 mm. The catheter working length is 2000 mm. The R2P Metacross RX is provided with the following accessory device: - Flushing Needle: provided per one catheter device for the purpose of flushing and filling ● of the guidewire lumen with heparinized saline before use. {5}------------------------------------------------ #### Indications for Use The R2P Metacross RX PTA Balloon Dilatation Catheter is intended to dilate stenosis in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature. ## Comparison of Indications for Use to Predicate Devices The intended use (percutaneous transluminal angioplasty) and indications for use of the R2P Metacross RX are identical to that of the Metacross RX. Therefore, the subject device's indications for use may be considered substantially equivalent to that of the predicate device. #### Comparison of Technological Characteristics to Predicate Devices #### Technological Comparison to Predicate Devices The R2P Metacross RX has the following similarities to the predicate device: - Same intended use . - Same indications for use ● - Same target population - Same operating principle ● - Similar materials - Same fundamental scientific technology - Same sterility assurance level and method of sterilization The principal difference between the devices is that R2P Metacross RX has a lower profile proximal tube (outer dimeter of 1.25 mm) compared to that of the Metacross RX (the outer dimeter of 1.30 mm). In addition, the R2P Metacross RX has a narrower range of balloon diameters (3.0 to 8.0 mm) compared to the Metacross RX (3.0 to 12.0 mm). The range of balloon lengths of R2P Metacross RX and Metacross RX is the same (20 mm to 200 mm). The R2P Metacross RX has only one type of catheter working length (2000 mm) unlike the Metacross RX (900, 1350 and 2000 mm). Other design or dimensional differences include the following: {6}------------------------------------------------ - The R2P Metacross RX has three position markers (900, 1200 and 1500 mm from the . distal tip of the catheter) while the Metacross RX has two (900 and 1000 mm from the distal tip of the catheter). - . The diameter of core wire was narrowed in association with the reduced profile of the proximal tube. ## Performance Testing To demonstrate substantial equivalence of R2P Metacross RX to the predicate devices, the technological characteristics and performance criteria were evaluated using the bench testing recommendations outlined in the FDA Guidance Document "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters" dated September 8, 2010. The following in vitro tests were performed on the subject device: Dimensional Verification, Balloon Preparation, Deployment and Retraction, Balloon Rated Burst Pressure, Balloon Fatigue (Repeat Balloon Inflations), Balloon Compliance, Balloon Inflation and Deflation Time, Catheter Bond Strength, Flexibility and Kink Test, Torque Strength, Particulate Evaluation, Balloon Rated Burst Pressure (in Stent), and Balloon Fatigue (Repeat Balloon Inflations; in Stent), Transportation, and Shelf Life Testing. The results from these tests demonstrate that the technological characteristics and performance of the R2P Metacross RX are substantially equivalent to the predicate device. ## Biocompatibility: The patient-contacting materials of the R2P Metacross RX device are identical to those of the predicate device; therefore, prior biocompatibility testing conducted on the predicate device was leveraged in support of the biological safety of the R2P Metacross RX. ## Conclusions: The R2P Metacross RX met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The R2P Metacross RX PTA Balloon Dilatation Catheter is substantially equivalent to the legally marketed predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%