The COOLIEF* Cooled Radiofrequency Probe is to be used in conjunction with a radiofrequency generator to create lesions in nervous tissue. This device is also indicated for creating radiofrequency lesions of the genicular nerves for the management of moderate to severe knee pain of more than 6 months with conservative therapy, including medication, in patients with radiologically-confirmed osteoarthritis (grade 2-4) and a positive response (≥50% reduction in pain) to a diagnostic genicular nerve block.
Device Story
COOLIEF* Cooled RF Probe is a sterile, single-use device for creating lesions in nervous tissue. Input: RF energy delivered via probe tip; cooling provided by circulating sterile water. Transformation: Internal water circulation cools electrode tip during ablation, enabling larger, more homogeneous RF heating area in target tissue. Output: RF lesions in nervous tissue. Used in clinical settings by physicians; device connects to Halyard RF Generator. Healthcare providers use output to perform genicular neurotomy; procedure blocks pain signal transmission. Benefits: Provides sustained pain reduction and functional improvement for patients with chronic knee osteoarthritis.
Clinical Evidence
Prospective, multicenter, randomized comparative study (n=151). Primary endpoint: ≥50% knee pain reduction (NRS) at 6 months. Results: 67.2% success in COOLIEF group vs 15.7% in corticosteroid injection group. Mean NRS drop: 4.9 points (COOLIEF) vs 1.3 points (control). Functional improvement: 39.7% (COOLIEF) vs 3% (control). Adverse events were similar between groups; no significant safety concerns identified.
Indicated for patients with moderate to severe knee pain (>6 months) refractory to conservative therapy, including medication, with radiologically-confirmed osteoarthritis (grade 2-4) and a positive response (≥50% reduction in pain) to a diagnostic genicular nerve block.
Regulatory Classification
Identification
A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a family, with three profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2017
Halyard Health, Inc. Monica King Associate Director, Regulatory Affairs 5405 Windward Parkway Alpharetta, Georgia 30004
Re: K163461
Trade/Device Name: Coolief* Cooled RF Probe Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency Lesion Probe Regulatory Class: Class II Product Code: GXI Dated: March 13, 2017 Received: March 14, 2017
Dear Ms. King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163461
Device Name COOLIEF* Cooled RF Probe
#### Indications for Use (Describe)
The COOLIEF* Cooled Radiofrequency Probe is to be used in conjunction with a radiofrequency generator to create lesions in nervous tissue. This device is also indicated for creating radiofrequency lessons of the management of moderate to severe knee pain of more than 6 months with conservative therapy, including medication, in patients with radiologically-confirmed osteoarthritis (grade 2-4) and a possitive response (≥50% reduction in pain) to a diagnostic genicular nerve block.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span style="letter-spacing: normal;"> <span style="line-height: normal;"> <span style="orphans: auto;"> <span style="text-align: start;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: auto;"> <span style="word-spacing: 0px;"> <span style="-webkit-text-stroke-width: 0px;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary as Required by 21 CFR §807.92(c)
As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR §807.92(a), the 510(k)-summary provided below is of sufficient detail to provide an understanding of the basis for a determination of substantial equivalence.
# 510(k) Summary
#### 1. Contact Details
| Date Summary Prepared | 04/05/2017 |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Applicant Name,<br>Address,<br>Website | Halyard Health, Inc.<br>5405 Windward Parkway<br>Alpharetta, GA 30004<br>www.halyardhealth.com |
| Applicant Contact Person | Monica King<br>Associate Director, Regulatory Affairs<br>Phone: (678) 477-4165<br>FAX: (678) 254-0347<br>Email: monica.king@hyh.com |
## 2. Device Information
| Trade Name | COOLIEF* Cooled RF Probe |
|---------------------|------------------------------|
| Common Name | Radiofrequency Lesion Probe |
| Models | CRP-, CRK-, MCK- |
| Classification | II |
| Classification Name | Probe, Radiofrequency Lesion |
| Regulation Number | 21 CFR §882.4725 |
| Product Code | GXI |
| Review Panel | 84 Neurology |
{4}------------------------------------------------
#### 3. Legally Marketed Predicate Device:
| Trade Name | Pain Management Cooled Probe |
|---------------|------------------------------|
| 510(k) Number | K053082 |
| Product Code | GXI |
| Manufacturer | Halyard Health |
#### 4. Description of Device:
The COOLIEF* Cooled Radiofrequency (RF) Probe is a sterile, single-use device that delivers RF energy within the area of the active probe tip, while the probe tip is cooled by sterile water that circulates within the probe. Cooling the probe tip creates a larger. more homogenous RF heating area that results in a larger RF lesion in the target tissue. COOLIEF* Cooled RF Probe is used in conjunction with the Halyard RF Generator to create RF lesions in nervous tissue. The shaft of the probe is insulated with a polyimide sheath, and the distal tip consists of a medical grade stainless steel electrode. Sterile water circulates through a cavity in the electrode to cool the electrode tip during the cooled RF ablation procedure. The COOLIEF* Cooled Radiofrequency (RF) Probe is sterilized by ethylene oxide.
## Proposed Indication for Use:
The COOLIEF* Cooled Radiofrequency Probe is to be used in conjunction with a radiofrequency generator to create lesions in nervous tissue. This device is also indicated for creating radiofrequency lesions of the genicular nerves for the management of moderate to severe knee pain of more than 6 months with conservative therapy, including medication, in patients with radiologicallyconfirmed osteoarthritis (grade 2-4) and a positive response (≥50% reduction in pain) to a diagnostic genicular nerve block.
## 5. Substantial Equivalence Comparison
The following table compares the subject COOLIEF* Cooled RF Probe to the predicate Pain Management Cooled Probe (K053082) to support substantial equivalence.
{5}------------------------------------------------
| Characteristic | COOLIEF*<br>Cooled RF<br>Probe<br>(K163461) | Pain<br>Management<br>Cooled Probe<br>(K053082) | Comments |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The COOLIEF* Cooled<br>Radiofrequency Probe is to<br>be used in conjunction<br>with a radiofrequency<br>generator to create lesions<br>in nervous tissue. This<br>device is also indicated for<br>creating radiofrequency<br>lesions of the genicular<br>nerves for the management<br>of moderate to severe knee<br>pain of more than 6<br>months with conservative<br>therapy, including<br>medication, in patients<br>with radiologically-<br>confirmed osteoarthritis<br>(grade 2-4) and a positive<br>response (≥50% reduction<br>in pain) to a diagnostic<br>genicular nerve block. | Used in conjunction<br>with a radiofrequency<br>generator to create<br>radiofrequency lesions<br>in nervous tissue. | Clinical data to support the<br>proposed indication is<br>submitted in the Protocol A<br>Prospective, Multi-center,<br>Randomized, Clinical Trial<br>Evaluating the Safety and<br>Effectiveness of Using<br>Coolief™ Cooled<br>Radiofrequency Probe to<br>Create Lesions of the<br>Genicular Nerves and<br>Comparing Corticosteroid<br>Injection in the Management<br>of Knee Pain |
| Probe Shaft<br>Length | Overall Useable Length<br>150mm, 100mm, 75mm,<br>50mm<br>Shaft Outer Diameter<br>18Ga | Overall Useable Length<br>150mm, 100mm, 75mm,<br>50mm<br>Shaft Outer Diameter<br>18Ga | Equivalent |
| Distal Tip<br>Length | 6mm probe electrode<br>with various active<br>lengths when combined<br>with a mating introducer<br>of 2, 4, and 5.5mm | 6mm probe electrode<br>with various active<br>lengths when combined<br>with a mating introducer<br>of 2, 4, and 5.5mm | Equivalent |
| Lesion Size | 10-12mm,<br>Spherical | 10-12mm,<br>Spherical | Equivalent |
| Temperature<br>Measurement<br>Accuracy | ± 3° C | ± 3° C | Equivalent |
| Temperature<br>Measurement<br>Device | Thermocouple | Thermocouple | Equivalent |
| Characteristic | COOLIEF*<br>Cooled RF<br>Probe<br>(K163461) | Pain Management<br>Cooled Probe<br>(K053082) | Comments |
| Temperature<br>Range | 38° - 95° C | 38° - 95° C | Equivalent |
| Temperature<br>Increment | 1°C | 1°C | Equivalent |
| Single Use | Yes | Yes | Equivalent |
| Compatible RF<br>System | Halyard<br>COOLIEF*<br>System Only | Halyard<br>COOLIEF*<br>System Only | Equivalent |
| Disposable | Yes | Yes | Equivalent |
| Biocompatibility | Conforms to<br>ISO10993 | Conforms to<br>ISO10993 | Equivalent |
| Sensitization<br>ISO 10993-10 | Conforms to<br>ISO10993-10 | Conforms to<br>ISO10993-<br>10 | Equivalent |
| Irritation<br>ISO 10993-20 | Conforms to<br>ISO10993-20 | Conforms to<br>ISO10993-<br>20 | Equivalent |
| Cytotoxicity<br>ISO 10993-5 | Conforms to<br>ISO10993-5 | Conforms<br>to<br>ISO10993-5 | Equivalent |
| Characteristic | COOLIEF*<br>Cooled RF<br>Probe<br>(K163461) | Pain Management<br>Cooled Probe<br>(K053082) | Comments |
| Systemic<br>toxicity<br>ISO 10993-11 | Conforms to<br>ISO10993-11 | Conforms to<br>ISO10993-11 | Equivalent |
| Sterility | Sterilized by EO<br>SAL = 10-6 | Sterilized by EO<br>SAL = 10-6 | Equivalent |
| Packaging | Device contained in a<br>single use Tyvek sealed<br>tray | Device contained in a<br>single use Tyvek sealed<br>tray | Equivalent |
{6}------------------------------------------------
{7}------------------------------------------------
The difference in the Indications for Use statements do not raise new questions of safety and effectiveness. Cooled radiofrequency (RF) is a well-established method for delivering lesions into nervous tissue to accomplish neurotomy procedures. The use of the Cooled RF probe to perform the genicular neurotomy procedure is like the other minimally invasive cooled radiofrequency ablation procedures in that lesions are created in targeted sensory nerves to block the transmission of pain signals. The RF lesions created in the genicular neurotomy procedure are the same size and shape as in other RF procedures. The clinical data collected to support the proposed indication demonstrate that the COOLIEF* Cooled RF Probe does not present safety or effectiveness issues related to the proposed indication for use. Data collected at the primary endpoint supports the conclusion that the COOLIEF* Cooled RF Probe used for genicular nerve ablation is superior to corticosteroid injection in osteoarthritic subjects for managing knee pain.
#### Reference Device
A reference device is presented below regarding the biocompatibility data set for this device. Biocompatibility testing data is derived from testing conducted on the representative Halyard* TransDiscal* RF probe (K031951): the TransDiscal* RF probe and the Cooled Radiofrequency Probe are composed of the same raw materials, manufactured using similar processes within the same facility, sterilized using the same Ethylene Oxide cycles and chambers, and packaged using the same packaging materials.
{8}------------------------------------------------
| Comparison to Reference Device | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device: COOLIEF* Cooled RF Probe | Reference Device: TransDiscal* RF probe |
| K163481 | K031951 |
| The COOLIEF* Cooled<br>Radiofrequency Probe is to be used in<br>conjunction with a radiofrequency<br>generator to create lesions in nervous<br>tissue. This device is also indicated<br>for creating radiofrequency lesions of<br>the genicular nerves for the<br>management of moderate to severe<br>knee pain of more than 6 months with<br>conservative therapy, including<br>medication, in patients with<br>radiologically-confirmed<br>osteoarthritis (grade 2-4) and a<br>positive response (≥50% reduction in<br>pain) to a diagnostic genicular nerve<br>block. | The Transdiscal system, in combination with<br>the Baylis Pain Management Generator-TD<br>(PMG-TD), is indicated for the coagulation and<br>decompression of disc material to treat<br>symptomatic patients with contained herniated<br>discs. |
| Sterility: Sterilized by | Sterility: Sterilized |
| EO SAL = 10-6 | by EO SAL = 10-6 |
| Packaging: Device contained in a single use<br>Tyvek sealed tray | Packaging: Device contained in a single use<br>Tyvek sealed tray |
# 6. Non-Clinical Testing
The table below describes the test type, standard reference, acceptance criteria, and result summary.
| Test Type | Standard | Test Name | Criteria | Result |
|-----------------------------------------------|--------------|----------------------------------------------|----------------------------------------------------------------------|-------------------------------------------|
| Cytotoxicity | ISO 10993-5 | In Vitro Cytotoxicity,<br>Direct and Extract | Qualitative Grade = 0<br>Quantitative = cell death<br>< 30% | Pass: No<br>Cytotoxic effect |
| Sensitization | ISO 10993-10 | In Vivo, Animal GPMT | Challenge Phase = Less<br>than Grade 1 and less<br>than the controls | Pass: No signs<br>of sensitization |
| Irritation or<br>Intracutaneous<br>reactivity | ISO 10993-10 | In Vivo, Animal<br>Irritation (Rabbits) | Test sample score < 1.0<br>for Erythema and Edema<br>grading | Pass: No signs<br>of Irritation |
| Systemic Toxicity<br>(acute) | ISO 10993-11 | In Vivo, Animal<br>Toxicity (Mice) | No animal death | Pass: No signs<br>of systemic<br>toxicity |
Biocompatibility Testing
{9}------------------------------------------------
| Test | Test Description | Results |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Electrical safety | IEC 60601-1:2005<br>Medical electrical equipment<br>Part 1: General requirements for basic safety and essential performance | Passed |
| High Frequency surgical equipment | IEC 60601-2-2: 2009 (Fifth Ed.)<br>Medical electrical equipment<br>Part 2-2: Requirements for basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories | Passed |
| Electromagnetic Compatibility (EMC) | IEC 60601-1-2:2007 / AC 2010<br>Medical electrical equipment<br>Part 1-2: General requirements for safety –<br>Collateral standard: Electromagnetic compatibility | Passed |
#### Summary of IEC 60601 Compliance
#### 7. Clinical Testing
Halyard conducted a prospective, multicenter, randomized comparative human study, to confirm the safety and effectiveness of COOLIEF* Cooled RF Probe for creating lesions of the genicular nerves for pain management of the knee. The COOLIEF* Cooled RF Probe was compared to corticosteroid injection.
The primary effectiveness endpoint of the clinical study was the proportion of subjects whose knee pain is reduced by ≥ 50% based on the Numeric Rating Scale (NRS) at the 6month study time point. At the 6-month study time point, based on Intention to Treat (ITT) ≥50% pain relief over baseline was experienced by 67.2% of the COOLIEF* Cooled RF study group vs. 15.7% of the comparison group (i.e., corticosteroid injection), using the Numeric Rating Scale (Pain rating scale 1 to 10).
Significant and sustained pain reduction was observed in the study group: subjects had a 4.9-point mean drop in NRS from a baseline mean of 7.3 to a mean of 2.5 at 6 months, while the comparison group had a 1.3-point mean drop in NRS from a baseline of 7.2 to a mean of 5.9 at 6 months. Significant functional improvement occurred in the study group: 39.7% reported "Satisfactory Joint Function" vs. 3% in comparison group. Global Perceived Effect Knee condition was reported as "improved" in 91.4% of the study group vs. 23.9% in the comparison group.
#### Adverse Events
The proportion of study subjects that had adverse events (AEs) in each cohort was: CRFA, 45% (34/76); IAS, 40% (30/75). The number of AEs reported in each study group was similar (CRFA = 61 events, IAS = 65 events). Most AEs during the study were non-serious, mild or moderate in severity, and were determined to be not related to study treatment. The AEs with Possible, Probable, or Definite relationship to procedure are:
{10}------------------------------------------------
- CRFA Group (14 events in 13 subjects): post-procedure pain (9 events), ecchymosis (1), pruritic skin lesion (1), swelling and redness - infection (1), mild tenderness to touch (1), increased knee pain - severe (1)
- IAS Group (2 events in 2 subjects): white discoloration at injection site (1), ● fluctuating blood sugar levels (1)
- Post-procedural "fall" incidence: ■
- CRFA Group (2 events in 1 subject) ●
- . IAS Group (4 events in 4 subjects)
- I Serious AEs:
- . CRFA Group (4 events in 2 subjects): 1) pyelonephritis, 2) exacerbation of asthma, 3) severe acute asthma, and 4) acute respiratory failure
- IAS Group (8 events in 7 subjects): 1) opioid overdose, 2) heart attack (two ● subjects), 3) death, 4) nausea and vomiting, 5) worsening of hiatal hernia, 6) gastric volvulus, and 7) abdominal pain secondary to small bowel obstruction.
## Medication Use
The study demonstrated a reduction in non-opioid pain medication that was consistent with the clinically relevant improvements demonstrated in the primary endpoint. Table 76 below describes the pain medication usage for subjects taking non-morphine medication at the baseline through 6 months' visits.
{11}------------------------------------------------
| | Baseline | | 1 Month | | 3 Month | | 6 Month | |
|-------------------------------------------------------------------------------------------|----------------------|--------|---------------------|---------|---------------------|---------|----------------------|---------|
| | CRFA | IAS | CRFA | IAS | CRFA | IAS | CRFA | IAS |
| Non-Morphine Pain Medication<br>Usage - Total Daily Dose (mg) | | | | | | | | |
| N | 33 | 35 | 33 | 34 | 32 | 34 | 29 | 35 |
| Mean | 899.5 | 497.4 | 899.5 | 537.9 | 865.2 | 561.2 | 834.8 | 621.4 |
| SD | 625.1 | 437.3 | 625.1 | 484.4 | 636.4 | 461.4 | 682.0 | 497.5 |
| Median | 700.0 | 470.0 | 700.0 | 472.5 | 675.0 | 498.0 | 650.0 | 545.0 |
| Minimum | 150.0 | 0.0 | 150.0 | 0.0 | 150.0 | 0.0 | 0.0 | 0.0 |
| Maximum | 3000.0 | 2000.0 | 3000.0 | 2000.0 | 3000.0 | 2000.0 | 3000.0 | 2000.0 |
| Difference between means<br>(CRFA-IAS) and 95% Cl | 402.1 (138.7, 685.5) | | 361.7 (89.3, 634.0) | | 304.0 (31.8, 576.1) | | 213.5 (-81.7, 508.6) | |
| P-value (difference between<br>groups) | 0.0012** | | 0.0038** | | 0.0250** | | 0.2143** | |
| Change from Baseline in Non-<br>Morphine Pain Medication<br>Usage - Total Daily Dose (mg) | | | | | | | | |
| N | - | - | 33 | 34 | 32 | 34 | 29 | 35 |
| Mean | - | - | 0.0 | 44.9 | -15.6 | 60.9 | -34.5 | 123.9 |
| SD | - | - | 0.0 | 226.3 | 88.4 | 277.7 | 128.9 | 375.4 |
| Median | - | - | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| Minimum | - | - | 0.0 | 0.0 | -500.0 | 0.0 | -500.0 | -440.0 |
| Maximum | - | - | 0.0 | 1307.5 | 0.0 | 1600.0 | 0.0 | 1600.0 |
| Difference between means<br>(CRFA-IAS) and 95% Cl | - | | -44.9 (-124, 34.0) | | -76.5 (-178, 24.8) | | -158 (-295, -21.7) | |
| P-value (difference between<br>groups) | - | | 0.1668** | | 0.0506** | | 0.0229** | |
| P-value (change from<br>Baseline) | - | | - | 0.2552§ | 0.3251§ | 0.2100§ | 0.1609§ | 0.0591§ |
Table 76. Pain Medication Usage for Subjects Taking Non-Morphine Pain Medication at Any Visit Through 6 Months - TDD
Table disclays total daily dose of normally for the subjects taking nor-morphine pan medication at any follow-up visit. Subjects with only the Baseine visit are excluded.
T-test for two independent means, Wilcoxon rank sum test for two Independent samples, 3paired t-test
Program: HYHOS outout TDD Non-Morphine At Any Visit OM SBS Data Source: hyh03 painmedsub} Date Run: 01MAR2017 - 15:30
An analysis of the primary endpoint for the patients who were opioid dependent at baseline and were evaluated for the primary endpoint demonstrated that there was no relationship between outcome and opioid status as described in Table 82 below. Of the 43 Cooled Radiofrequency Ablation (CRFA) successes, only 10 patients were on opioids at Baseline (23.3%) as compared to 46.7% of the CRFA group who failed the primary endpoint (7/15). Overall, opioid status did not influence the outcomes (p= 0.4073).
{12}------------------------------------------------
| | Cooled Radiofrequency<br>Ablation (CRFA) | Intraarticular<br>Steroid (IAS) | Overall |
|-----------------------------------------------|------------------------------------------|---------------------------------|-----------------|
| Subjects that are Primary<br>Endpoint success | 10/43 (23.3) | 5/11 (45.5) | 15/54<br>(27.8) |
| Subjects that are Primary<br>Endpoint failure | 7/15 (46.7) | 18/57 (31.6) | 25/72 (34.7) |
| P-value | 0.1073† | 0.4889† | 0.4073†† |
## Table 82: Primary Endpoint Success or Failure for Subjects Taking Opioids at Baseline n/N%
+Fisher exact test for proportions, ++Chi-square test for proportions
The results of the pre-planned statistical analysis of the primary endpoint supports the conclusion that the COOLIEF* Cooled Radiofrequency Probe used for genicular nerve ablation is superior to corticosteroid injection in osteoarthritic subjects for managing knee pain.
# 8. Conclusion
The non-clinical data demonstrate that the COOLIEF* Cooled RF probe devices perform equivalently to the predicate device that is currently marketed. The clinical data demonstrate that the COOLIEF* Cooled RF probe does not present safety or effectiveness issues related to the proposed indication for use. Data collected at the primary endpoint supports the conclusion that the COOLIEF* Cooled Radiofrequency Probe used for genicular nerve ablation is superior to corticosteroid injection in osteoarthritic subjects for managing knee pain. The change to the indications for use does not raise different questions of safety or effectiveness.
End of 510(k) Summary
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.