K163437 · C.R. Bard, Inc. · PVZ · Jun 23, 2017 · General Hospital
Device Facts
Record ID
K163437
Device Name
hypodermic Pinpoint™ GT Needle
Applicant
C.R. Bard, Inc.
Product Code
PVZ · General Hospital
Decision Date
Jun 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The hypodermic Pinpoint™ GT Needle is intended for the injection into or the withdrawal of body fluids from parts of the body below the surface of the skin. The needle is to be used with syringes for general purpose fluid injection. The needle tip is echogenic and may be used with ultrasound to provide a visual representation of the needle tip throughout the insertion, medication administration and aspiration process.
Device Story
Hypodermic needle for fluid injection/aspiration; features integral passive neodymium iron boron (NdFeB) magnet in hub. Used with ultrasound systems equipped with Pinpoint™ GT Needle Technology (magnet sensing probe and software). Magnet allows ultrasound system to generate virtual image of needle tip on display during insertion, medication administration, and aspiration. Operated by clinicians in clinical settings. Visual output assists needle guidance; improves accuracy of needle placement; reduces potential for multiple insertion attempts. Does not impact basic needle function.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 7864, ISO 9626, ISO 594-1/2, and USP <788>. Biocompatibility testing performed per ISO 10993-1, -5, -10, -11, and -12. Testing confirmed lumen patency, tensile force, tip inspection, surface finish, dimensions, stiffness, corrosion resistance, leakage, particulate matter, echogenicity, and magnetic axis orientation.
Technological Characteristics
Hypodermic needle with echogenic tip and passive NdFeB magnet (nickel-coated) in hub. Hub: Makrolon® 2458 polycarbonate. Shaft: silicone-coated alloy. Dimensions: 18G-24G, 1.5"-3.0". Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: passive magnetic field detection via external ultrasound system.
Indications for Use
Indicated for patients requiring injection of medication or withdrawal of body fluids from below the skin surface using syringes.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or a stream.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2017
C. R. Bard, Inc. Jamie Howell Associate Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, Utah 84116
Re: K163437
Trade/Device Name: hypodermic Pinpoint™ GT Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: PVZ Dated: June 19, 2017 Received: June 20, 2017
Dear Jamie Howell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
## Lori A. Wiggins -S6
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K163437
Device Name
hypodermic Pinpoint™ GT Needle
#### Indications for Use (Describe)
The hypodermic Pinpoint™ GT Needle is intended for the injection into or the withdrawal of body fluids from parts of the body below the surface of the skin. The needle is to be used with syringes for general purpose fluid injection. The needle tip is echogenic and may be used with ultrasound to provide a visual representation of the needle tip throughout the insertion, medication administration and aspiration process.
| Type of Use <i>(Select one or both, as applicable)</i> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> <div style="display: inline-block; vertical-align: middle;"> <svg height="16px" width="16px"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M2,14 L14,2" stroke="black" stroke-width="2"></path> </svg> </div> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="padding-right: 5px;"> <div style="display: inline-block; vertical-align: middle;"> <svg height="16px" width="16px"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> </svg> </div> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BARD ACCESS SYSTEMS. The word "BARD" is in large, bold, sans-serif font. Below the word "BARD" is the phrase "ACCESS SYSTEMS" in a smaller, sans-serif font.
# K163437
510(k) Summary
21 CFR 807.92(a)
| | Submitter Name:<br>Address: | Bard Access Systems, Inc.<br>605 North 5600 West<br>Salt Lake City, UT 84116 |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General<br>Provisions | Contact Person:<br>Telephone Number:<br>Fax Number:<br>Date of Preparation: | Jamie (Howell) Duke<br>Associate Regulatory Affairs Specialist<br>(801) 522-5465<br>(801) 522-4969<br>23 June 2017 |
| Subject<br>Device | Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Manufacturer:<br>Classification Panel: | hypodermic Pinpoint™ GT Needle<br>Hypodermic Needle<br>21 CFR §880.5570<br>Needle, Hypodermic, Single Lumen<br>II<br>PVZ<br>Bard Access Systems, Inc.<br>General Hospital |
| Primary<br>Predicate<br>Device | Premarket Notification:<br>Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Manufacturer:<br>Classification Panel: | K021475; date of clearance, July 19, 2002<br>BD Single Lumen Needle<br>Hypodermic Needle<br>21 CFR §880.5570<br>Needle, Hypodermic, Single Lumen<br>II<br>FMI<br>Becton Dickinson & Company<br>General Hospital |
| Reference<br>Device | Premarket Notification:<br>Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Manufacturer:<br>Classification Panel: | K142445, date of clearance, April 13, 2015<br>Pinpoint™ GT Safety Introducer Needle<br>Safety Introducer Needle<br>21 CFR §870.1340<br>Catheter Introducer<br>II<br>DYB<br>Bard Access Systems, Inc.<br>Cardiovascular |
{4}------------------------------------------------
| Device<br>Description | Bard Access Systems, Inc's hypodermic Pinpoint™ GT Needles are<br>designed to inject medication or aspirate fluid. The hypodermic Pinpoint™<br>GT Needles contain a magnet which emits a passive magnetic field that can<br>be detected by ultrasound systems equipped with Pinpoint™ GT Needle<br>Technology. The hypodermic Pinpoint™ GT Needles, when used with an<br>ultrasound system equipped with Pinpoint™ GT Needle Technology, creates<br>a virtual image of the needle on the ultrasound display, providing clinicians<br>with a visual representation of the needle throughout the insertion,<br>medication administration and aspiration process.<br>There are four (4) hypodermic Pinpoint™ GT Needle configurations included<br>in this submission, as shown in the table below. | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| | hypodermic Pinpoint™<br>GT Needles | Needle configurations |
| | | 18G x 2.75", 19° |
| | | 21G x 3", 30° |
| | | 22G x 2", 30° |
| | | 24G x 1.5", 30° |
| Intended Use | The hypodermic Pinpoint™ GT Needles are intended for the injection of<br>medication into or the withdrawal of body fluids from parts of the body below<br>the surface of the skin. The needle is to be used with syringes for general<br>purpose fluid injection/aspiration. | |
| Indications For<br>Use | The hypodermic Pinpoint™ GT Needle is intended for the injection of<br>medication into or the withdrawal of body fluids from parts of the body below<br>the surface of the skin. The needle is to be used with syringes for general<br>purpose fluid injection/aspiration. The needle tip is echogenic and may be<br>used with ultrasound to provide a visual representation of the needle tip<br>throughout the insertion, medication administration and aspiration process. | |
| Technological<br>Characteristics | The technological characteristics of the subject hypodermic Pinpoint™ GT<br>Needles are substantially equivalent with respect to the basic design and<br>function to those of the cited predicate device. The differences between the<br>subject and predicate device (e.g. the addition of the passive magnet) are not<br>critical to the intended use of the device and do not raise any new or different<br>questions regarding equivalence.<br><br>The subject and predicate devices share equivalent intended use and patient<br>population. The subject device differs from its primary predicate in terms of in<br>types of materials used, dimensions (gauges, lengths), needle tip<br>configurations and the addition of a passive magnet on the subject device.<br>The difference in materials, dimensions (gauges, lengths), and needle tip<br>configurations do not alter the indications for use or intended use of the<br>needles, nor do they raise different questions of equivalence between the<br>subject device and the primary predicate device. A reference device is<br>introduced to support the material and dimensional features of the subject<br>device. Nonclinical testing further demonstrates the equivalence of the<br>subject device to its predicates. | |
{5}------------------------------------------------
| | Attribute | Subject:<br>hypodermic<br>Pinpoint™ GT<br>Needles | Primary Predicate:<br>BD Single Lumen<br>Needle | Reference:<br>Pinpoint™ GT<br>Safety Introducer<br>Needle |
|-----------------------------------------------|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Owner | Bard Access Systems,<br>Inc. | Becton Dickinson &<br>Company | Bard Access Systems,<br>Inc. |
| | Classification | 21 CFR §880.5570<br>Hypodermic Needle | FMI<br>21 CFR §880.5570<br>Hypodermic Needle | DYB<br>21 CFR §870.1340<br>Catheter Introducer |
| | 510k Status | Subject of this<br>Premarket Notification | K021475 - date of<br>clearance July 19, 2002 | K142445 – date of<br>clearance April 13,<br>2015 |
| | Commercial<br>Name | hypodermic Pinpoint™<br>GT Needles | BD Single Lumen<br>Needle | Pinpoint™ GT Safety<br>Introducer Needle |
| Technological<br>Characteristics<br>Continued | Indications for<br>Use | The hypodermic<br>Pinpoint™ GT Needle<br>is intended for the<br>injection of medication<br>into or the withdrawal<br>of body fluids from<br>parts of the body<br>below the surface of<br>the skin. The needle is<br>to be used with<br>syringes for general<br>purpose fluid<br>injection/aspiration.<br>The needle tip is<br>echogenic and may be<br>used with ultrasound<br>to provide a visual<br>representation of the<br>needle tip throughout<br>the insertion,<br>medication<br>administration and<br>aspiration process. | These needles are<br>intended for general<br>purpose fluid<br>injection/aspiration,<br>infusion, venipuncture<br>to obtain blood<br>collection and insulin<br>injection. | The Pinpoint™ GT<br>Safety Introducer<br>Needle is intended for<br>patients requiring<br>percutaneous access<br>to place a guidewire<br>for subsequent<br>placement of<br>catheters or other<br>medical procedures<br>requiring introducer<br>needle access.<br>The Pinpoint™ GT<br>Safety Introducer<br>Needle may be used<br>in any appropriate<br>patient population. |
| | Intended Use | The hypodermic<br>Pinpoint™ GT<br>Needles are intended<br>for the injection of<br>medication into or the<br>withdrawal of body<br>fluids from parts of the<br>body below the<br>surface of the skin.<br>The needle is to be<br>used with syringes for<br>general purpose fluid<br>injection/aspiration. | These needles are<br>intended for general<br>purpose fluid<br>injection/aspiration,<br>infusion, venipuncture<br>to obtain blood<br>collection and insulin<br>injection. | The Pinpoint™ GT<br>Safety Introducer<br>Needle is designed for<br>percutaneous<br>vascular access or<br>procedures requiring<br>the placement of a<br>guidewire. |
| | Scientific<br>Technology<br>Description | The primary intent of<br>the subject device,<br>hypodermic Pinpoint™<br>GT Needles, are to<br>inject medication or<br>aspirate fluid from<br>parts of the body<br>below the surface of<br>the skin. | Not publically available. | The primary intent of<br>the reference device,<br>Pinpoint™ GT Safety<br>Introducer Needle, is<br>to assist with<br>percutaneous<br>vasculature access or<br>procedures requiring<br>the placement of a<br>guidewire. |
| | Subject and Primary Predicate Device Comparison Table | | | |
| | Attribute | Subject:<br>hypodermic Pinpoint™<br>GT Needles | Primary Predicate:<br>BD Single Lumen<br>Needle | Reference:<br>Pinpoint™ GT<br>Safety Introducer<br>Needle |
| Technological<br>Characteristics<br>Continued | Scientific<br>Technology<br>Description<br>Continued | Additionally, the<br>hypodermic Pinpoint™<br>GT Needles contain,<br>integral, within the<br>needle hub a passive<br>magnet. The needle's<br>incorporated passive<br>magnet can be detected<br>by an ultrasound system<br>equipped with<br>Pinpoint™ GT Needle<br>Technology (not the<br>subject of this<br>submission). The<br>Pinpoint™ GT Needle<br>Technology consists of<br>a magnet sensing probe<br>and software loaded on<br>ultrasound equipment,<br>creating the Pinpoint™<br>GT System; this system<br>is capable of displaying<br>a visual representation<br>of the needle on an<br>ultrasound image. The<br>detection of the needle's<br>passive magnet by the<br>Pinpoint™ GT Needle<br>Technology is an<br>optional feature and<br>tool/device offered to<br>clinicians for visual<br>representation of a<br>needle. The presence of<br>the passive magnet<br>does not impact the<br>ability of the device to<br>perform as a<br>hypodermic needle. | Not publically<br>available. | Additionally, the<br>Pinpoint™ GT Safety<br>Introducer Needle<br>contains, integral,<br>within the needle<br>hub an active safety<br>mechanism and a<br>passive magnet. The<br>needle's<br>incorporated passive<br>magnet can be<br>detected by an<br>ultrasound system<br>equipped with<br>Pinpoint™ GT<br>Technology. The<br>Pinpoint™ GT<br>Technology (not the<br>subject of this<br>submission)<br>consists of a magnet<br>sensing probe and<br>software loaded on<br>ultrasound<br>equipment, creating<br>the Pinpoint™ GT<br>System; this system<br>is capable of<br>displaying a visual<br>representation of the<br>needle on an<br>ultrasound image.<br>The detection of the<br>needle's passive<br>magnet by the<br>Pinpoint™ GT<br>technology is an<br>optional feature and<br>tool/device offered to<br>clinicians for visual<br>representation of a<br>needle throughout<br>the insertion<br>process; the<br>presence of the<br>passive magnet<br>does not impact the<br>ability of the device<br>to perform as an<br>introducer needle. |
| | Subject and Primary Predicate Device Comparison Table | | | |
| | Attribute | Subject:<br>hypodermic<br>Pinpoint™ GT<br>Needles | Primary Predicate:<br>BD Single Lumen<br>Needle | Reference:<br>Pinpoint™ GT<br>Safety Introducer<br>Needle |
| | Needle<br>Components | Needle Shaft:<br>Silicone coated<br>Needle tip:<br>Echogenic<br>A-Bevel 19° or Short<br>30°<br>Hub:<br>Open ended luer<br>locking hub<br>Bevel Indicator<br>Passive Magnet<br>Safety Mechanism:<br>None | Needle Shaft:<br>Silicone coated<br>Needle tip:<br>Regular, Short,<br>Intradermal Bevel<br>Hub:<br>Open ended luer locking<br>hub<br>Safety Mechanism:<br>None | Needle Shaft:<br>Silicone coated<br>Needle tip:<br>Echogenic<br>A-Bevel 12°<br>Hub:<br>Open ended luer<br>locking hub<br>Bevel Indicator<br>Passive Magnet<br>Safety Mechanism:<br>Incorporated active<br>safety mechanism |
| Technological<br>Characteristics<br>Continued | Needle Materials | Luer Hub:<br>Makrolon®2458<br>Polycarbonate<br>Clear<br>Needle Shaft:<br>Silicone Coated alloy<br>Magnet:<br>Neodymium Iron<br>Boron (NdFeB)<br>Nickel (Ni) Coated<br>Needle Protective<br>Cover (packaging<br>retainer):<br>High Density<br>Polyethylene (HDPE)<br>No colorant<br>Safety Mechanism:<br>None | Luer Hub:<br>Polypropylene<br>Needle shaft:<br>Stainless Steel<br>Silicone Coated<br>Magnet:<br>None<br>Needle Protective Cover<br>(packaging retainer):<br>Protective Plastic sheath<br>Details not publically<br>available<br>Safety Mechanism:<br>None | Luer Hub:<br>Makrolon®2458<br>Polycarbonate<br>Clear<br>Needle Shaft:<br>Silicone Coated alloy<br>Magnet:<br>Neodymium Iron<br>Boron (NdFeB)<br>Nickel (Ni) Coated<br>Needle Protective<br>Cover (packaging<br>retainer):<br>Clear Amorphous<br>Polyethylene<br>Terephtalate (APET)<br>Safety Mechanism:<br>Translucent Green<br>Polycarbonate<br>Silicone O-ring<br>301 Stainless Steel<br>Cap polycarbonate |
| | Needle<br>Dimensions | There are four (4)<br>total subject device<br>configurations<br>consisting of the<br>following gauge sizes<br>and lengths:<br>Diameter:<br>18, 21, 22, 24 gauge<br>Length:<br>1.50", 2.00", 2.75",<br>3.00" | Diameter:<br>16 to 30 gauge<br>Length:<br>3/8" to 2" | Diameter:<br>21 gauge<br>Length:<br>2.75" |
| | Subject and Primary Predicate Device Comparison Table | | | |
| Technological<br>Characteristics<br>Continued | Attribute | Subject:<br>hypodermic<br>Pinpoint™ GT<br>Needles | Primary Predicate:<br>BD Single Lumen<br>Needle | Reference:<br>Pinpoint™ GT<br>Safety Introducer<br>Needle |
| | Needle Labeling | Gauge Size:<br>No gauge size<br>coloring*<br>*Note: Needle hubs for<br>the different needle<br>gauge sizes are not<br>colored to ensure<br>visualization of blood<br>flashback. Colors used in<br>the subject device<br>labeling are based on<br>needle length and are<br>consistent with the<br>needle color coding used<br>with the Pinpoint™ GT<br>Needle Technology. | Gauge Size:<br>Translucent, color-coded<br>hubs:<br>Gray (27-gauge), Tan<br>(26), Blue (25),<br>Turquoise (23), Black<br>(22), Green (21), Yellow<br>(20), Brown (19), Pink<br>(18), and Purple (16) | Gauge Size:<br>Safety mechanism is<br>a translucent green<br>color |
| | | Bevel-Up Indicator:<br>V in ribs of the hub<br>(extra ribs on the hub<br>form a V-shape on<br>the hub that indicates<br>bevel-up). | Bevel-Up Indicator:<br>None | Bevel-Up Indicator:<br>Raised embossed<br>arrow on the hub |
| | Sterility | Provided Sterile<br>SAL 10-6<br>Ethylene Oxide | Provided Sterile<br>SAL 10-6<br>Gamma irradiation or<br>Ethylene Oxide | Provided Sterile<br>SAL 10-6<br>Ethylene Oxide |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Verification and validation tests were designed and performed in accordance with Design Controls as per 21 CFR §820.30. The following table identifies the performance tests conducted per standards in conjunction with in-house protocols to determine appropriate methods for evaluating the performance of the device.
| Performance Tests | Test Description / Standard Utilized | |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Needle Lumen Patency | | |
| Needle-Hub Tensile Force | • ISO 7864:1993, Sterile hypodermic needles<br>for single use | |
| Effective Needle Length | | |
| Needle Tip Inspection | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.