Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL)

K163412 · Girgis Scope, LLC · CCW · Jun 22, 2017 · Anesthesiology

Device Facts

Record IDK163412
Device NameGirgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL)
ApplicantGirgis Scope, LLC
Product CodeCCW · Anesthesiology
Decision DateJun 22, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5540
Device ClassClass 1

Indications for Use

The Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) is indicated to improve direct visualization of larynx and assist in both non-complicated and difficult intubation in situation of unpredictable airway condition.

Device Story

Device is a rigid, double-bladed video laryngoscope used by clinicians (anesthesiologists, ER physicians, EMS) to assist in tracheal intubation. It features a stationary handle with fiber-optic illumination and a movable blade equipped with a camera. The device provides both direct visualization (via traditional blade design) and indirect visualization (via the camera accessory) of the laryngeal anatomy. The movable blade, controlled by a levering system, articulates to create a pathway for endotracheal tube insertion. The camera output is displayed on a DVR monitor to assist in difficult intubation scenarios. The device is reusable, disassemblable for cleaning, and sterilized via pre-vacuum steam. It is intended to improve visibility of the vocal cords and epiglottis while reducing potential for teeth damage during intubation procedures.

Clinical Evidence

No clinical studies were conducted. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 60335), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993 series, ASTM F 756-08, ASTM F 619-03), and a simulated use/usability study involving anesthesiologists, ER physicians, and EMS personnel.

Technological Characteristics

Rigid laryngoscope with double-blade design; fiber-optic illumination; 3.7V Lithium-Ion rechargeable battery; lever-actuated flexible tip (70-degree range); ratchet spring holder with ball joint socket; camera accessory for indirect visualization. Materials tested per USP 24/NF 19 Class VI and ISO 10993. Sterilization: Pre-vacuum steam (132°C for 4 minutes).

Indications for Use

Indicated for patients requiring tracheal intubation, including those with non-complicated or difficult/unpredictable airway conditions, to improve direct visualization of the larynx.

Regulatory Classification

Identification

A rigid laryngoscope is a device used to examine and visualize a patient's upper airway and aid placement of a tracheal tube.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around it. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three curved lines. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 22, 2017 Girgis Scope LLC % E. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton. Marvland 21114 Re: K163412 Trade/Device Name: Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) Regulation Number: 21 CFR 868.5540 Regulation Name: Rigid Laryngoscope Regulatory Class: Class I Product Code: CCW Dated: May 23, 2017 Received: Mav 26, 2017 Dear E. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Mark S. Fellman -S for Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163412 #### Device Name Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) Indications for Use (Describe) The Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) is indicated to improve direct visualization of larynx and assist in both non-complicated and difficult intubation of unpredictable airway condition. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary | Sponsor | | |---------------------------|-----------------------------------------------------------------| | Company Name: | Girgis Scope LLC | | Company Address: | 2 Stonegate Drive<br>Monroe Township, NJ 08831 | | Telephone: | 1-732-710-9724 | | Fax: | 1-732-605-1631 | | Contact Person: | Madgy Girgis, M.D. | | Summary Preparation Date: | June 22, 2017 | | Trade Name: | Girgis Scope Video 200 Doubled Bladed Laryngoscope (GSV200.DBL) | | Common/Usual Name: | Video Laryngoscope | | Classification Name: | Laryngoscope Rigid Accessory | | Regulation Code: | 21 CFR 868.5540 | | Product Code: | CCW | | Device Class: | Class I | | Panel: | Division of Anesthesiology | Predicate Device | Manufacturer | Brand Name | Classification | |----------------------------------------|------------------------------------|----------------| | Propper Manufacturing<br>Company, Inc. | Propper Flip-Tip™™<br>Laryngoscope | K915804 | #### Device Description GSV200.DBL is designed to assist in tracheal intubation procedure by direct visualization of laryngeal anatomy in closed position and use movable blade with camera as accessory from different angle in open position. The physical shape of GSV200.DBL is a modification of traditional Macintosh blade similar to Propper flip tip laryngoscope with added movable blade with camera as accessory to provide visualization to insert endotracheal tube. Girgis Scope Video DBL is designed to use two power sources, one in the stationary handle {4}------------------------------------------------ for fiber optic illumination of the airway and the second in the DVR monitor to illuminate the Larynx for the video camera. Battery chargers, cables to computer and cable to connect to TV screen are accessories included in the kit. The GSV200.DBL is provided non-sterile and is reusable and can be disassembled to provide cleaning and disinfection. # Indications for Use Statement The Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) is indicated to improve direct visualization of larynx and assist in both non-complicated and difficult intubation in situation of unpredictable airway condition. | Parameters | Girgis Scope LLC<br>(K163412) | Propper Manufacturing<br>Company, Inc.<br>(K915804) | Comments | |---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The Girgis Scope Video<br>200 Double Bladed<br>Laryngoscope<br>(GSV200.DBL) is<br>indicated to improve<br>direct visualization of<br>larynx and assist in<br>both non-complicated<br>and difficult intubation<br>in situation of<br>unpredictable airway<br>condition. | Propper Flip Tip™<br>laryngoscope is<br>indicated to improve<br>direct visualization of<br>larynx and assist in<br>both non-complicated<br>and difficult intubation<br>in situation of<br>unpredictable airway<br>condition. | Identical | | Type of Device | GVS200.DBL is a<br>modification of the<br>traditional MacIntosh<br>laryngoscope.<br>In addition to proper<br>design movable blade<br>and ratchet spring<br>holder are attached. | Propper Flip-Tip™<br>Laryngoscope is a<br>modification of the<br>traditional MacIntosh<br>laryngoscope. | Similar - Propper offers<br>a flip tip at the distal<br>end of the<br>laryngoscope blade and<br>the GSV200.DBL offers<br>a second blade. | | Design | Anatomically designed<br>to improve visibility of<br>the vocal cords and<br>reduce teeth damage | Anatomically designed<br>to improve visibility of<br>the vocal cords and<br>reduce teeth damage | Identical | | Laryngoscope Flexi-Tip<br>and Lever Control | Flexi-tip is adjustable to<br>70 degrees using<br>levering movable<br>handle to provide a<br>precise control to<br>expose the epiglottis. | Flexi-tip is adjustable to<br>70 degrees using<br>levering movable<br>handle to provide a<br>precise control to<br>expose the epiglottis. | Identical | # Predicate Product Comparison Table {5}------------------------------------------------ | Fiber Optic Intubating<br>Laryngoscope | Stationary handle for<br>fiber optic illumination<br>of the mouth | Stationary handle for<br>fiber optic illumination<br>of the mouth | Identical | |----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------| | Power Supply and Light<br>Source | Stationary<br>laryngoscope handle<br>containing a Lithium<br>Ion rechargeable<br>battery (3.7V).<br>GSV200.DBL utilizes a<br>rechargeable battery to<br>power light bulb and<br>through their 5mm<br>fibro-optic bundle to<br>provide bright white<br>illumination. | Stationary<br>laryngoscope handle<br>containing battery.<br>Propper utilizes AA and<br>C Batteries to power<br>light bulb and through<br>their 5mm fibro-optic<br>bundle to provide<br>bright while<br>illumination. | Predicate uses C<br>batteries and the<br>GSV200.DBL utilizes a<br>rechargeable battery | | Sterilization | Pre-vacuum sterilizer<br>132 degrees for 4<br>minutes | Pre-vacuum sterilizer<br>132 degrees for 4<br>minutes | Identical | | Laryngoscope Blades | Reusable | Reusable | Identical | #### Discussion of Technological Characteristics - 1. Propper Flip-Tip™ Laryngoscope blade (K915804) and the GBV200.DBL Laryngoscope both have a flexible tip that is adjustable to 70 degrees using a levering system which includes a movable handle, tip and spring mechanism and both offer fiber optic illumination. - 2. Both devices are anatomically designed to improve visibility of the epiglottis and reduce teeth damage. Both devices provide a laryngoscope stationary handle containing batteries, Propper utilizes C Batteries and GBV200.DBL a rechargeable battery, to power light bulb through their fibro-optic bundle which provides bright white illumination. - Both systems have a stationary handle and in addition to the Propper design the 4. GSV200.DBL uses a ratchet spring holder with a ball joint socket that locks the movable blade in closed position while introducing the two blades in patient mouth and unlocks to open position in step by step progression. - In addition to the Propper laryngoscope design, GSV200.DBL uses the movable blade, which 5. is attached to the movable handle and articulates with stationary blade so movement in the movable handle is transmitted to the movable blade to: - A. To create a pathway to introduce the endotracheal tube with ease - B. The movable blade carries camera arrangement as accessory for indirect visualization of laryngeal anatomy in unpredictable difficult intubation. The camera angle will be looking upward and forward to show the most anterior position of larynx in difficult intubation - 5. Both offer a reusable Flexi-Tip style laryngoscope blade, both recommend pre-vacuum sterilization at 132°C for 4 minutes. {6}------------------------------------------------ ### Non-Clinical Performance Data - IEC 60335-1 Corrigendum 1 Household and similar electrical appliances Safety Part ● 1: General Requirements - . IEC 60335-2-29 Safety of household and similar electrical appliances - Safety - Part 2-29 Particular requirements for battery charges - IEC 60950-1 Information technology equipment – Safety – Part 1: General requirements - IEC 60601-1 + CORR 1 (2006) + CORR 2 (2007) + A1 (2012) Medical electrical equipment . Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: Collateral standard: Electromagnetic compatibility - Requirements and tests - . USP 24, NF 19 Biological Test for Plastics, Class VI 70 degrees C - . AAMI/ANSI ST79 Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities - . ISO 10933-5 Biological Evaluation of Medical Devices – Part 5: Tests for in vitro Cytotoxicity L929 MEM Elution Test - ISO 10993-10 Biological Evaluation of Medical Devices – Part 10: Test for Irritation and Skin Sensitization Kligman Maximization Test - ISO 10993-10 Biological Evaluation of Medical Devices – Part 10: Test for Irritation and Skin Sensitization Intracutaneous Injection Test - ISO 10993-11 Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity - ASTM F 756-08 Standard Practice for Assessment of Hemolytic Properties of Materials - ASTN F 619-03 Standard Practice for Extraction of Medical Plastics - A Simulated Use/Usability study was conducted using Anesthesiologists, Emergency Room Physicians, and Emergency Medical Service (EMS). - . Software Validation per Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices # Clinical Study No clinical studies were conducted. # Conclusion The Girgis Scope Video 200 Double Bladed Laryngoscope model (GSV200.DBL) has the same indications for use, flexible tip design, fiber optic bundle for lighting, laryngoscope blade design, laryngoscope handle design, same methods of sterilization, and similar nonclinical testing and are battery operated as the Propper Flip-Tip Laryngoscope (K915804). The DVR accessory improves visualization of the epiglottis and the usability study demonstrated ease-of-use of the system. Based on the similarities of the two laryngoscope systems no different questions of safety and effectiveness have been raised and we conclude that our laryngoscope system is substantially equivalent to the predicate laryngoscope system.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...