K163377 · Wilson-Cook Medical, Inc. · EOQ · Feb 21, 2018 · Ear, Nose, Throat
Device Facts
Record ID
K163377
Device Name
Captura Disposable Bronchoscope Biopsy Forceps
Applicant
Wilson-Cook Medical, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Feb 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
This device is used in obtaining biopsy samples in the bronchi or lungs. This device is indicated for adult use only.
Device Story
Captura® Bronchoscope Biopsy Forceps are manual surgical instruments used to obtain tissue samples from bronchi or lungs. The device consists of a spool handle, a coated coilspring catheter, a drive cable, and fenestrated forceps cups with a surgical rim edge. The device is passed through a prepositioned endoscope to the target site. The physician operates the spool handle—moving it forward to open the cups and backward to close them—to pinch and excise tissue. The device is intended for use by clinicians in a clinical setting. It facilitates diagnostic biopsy procedures, aiding in the assessment of lung or bronchial tissue.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including shelf life, functional testing, flex and fracture verification, force-to-biopsy testing, and packaging validation (burst and dye leak). Biocompatibility was evaluated per ISO 10993-1, including cytotoxicity, sensitization, irritation, acute toxicity, and material-mediated pyrogenicity.
Technological Characteristics
Manual biopsy forceps consisting of a spool handle, coated coilspring catheter, drive cable, and fenestrated forceps cups with surgical rim edges. Materials were modified from the predicate for improved strength and navigation. The device is a mechanical instrument; no energy source or software is utilized. Biocompatibility testing conducted per ISO 10993-1.
Indications for Use
Indicated for adult patients requiring biopsy samples from the bronchi or lungs.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
February 21, 2018
Wilson-Cook Medical Inc. Ashley Howard Regulatory Affairs Specialist I 4900 Bethania Station Road Winstom-Salem, North Carolina 27105
Re: K163377
Trade/Device Name: Captura® Bronchoscope Biopsy Forceps Regulation Number: 21 CFR 874.5680 Regulation Name: Bronchoscope (Fixed and Rigid) and Accessories Regulatory Class: Class II Product Code: EOO Dated: January 16, 2018 Received: January 22, 2018
Dear Ashley Howard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163377
#### Device Name
Captura® Bronchoscope Biopsy Forceps
Indications for Use (Describe)
This device is used in obtaining biopsy samples in the bronchi or lungs. This device is indicated for adult use only.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 201 Subpart D) </div> |
| <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart C) </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Traditional 510(k) Summary
| Name: | Wilson-Cook Medical, Inc. / Cook Endoscopy |
|----------------------|-------------------------------------------------------------------------------------|
| Address: | 4900 Bethania Station Road<br>Winston-Salem, North Carolina 27105 |
| Phone: | (336) 744-0157 ext. 6174 |
| Contact: | Ashley Howard, Specialist I - Regulatory Affair |
| Date: | February 15, 2018 |
| Trade Name: | Captura® Bronchoscope Biopsy Forceps |
| Common Name: | Bronchoscope Biopsy Forceps |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories<br>21 CFR §874.4680, EOQ, Class II |
| Predicate Device: | Disposable Bronchoscope Biopsy Forceps,<br>K923847, cleared November 18, 1992 |
## Indications for Use
This device is used in obtaining biopsy samples in the bronchi or lungs. This device is indicated for adult use only.
## Device Description
The Captura® Disposable Bronchoscope Biopsy Forceps (subject device) represents modifications made to the Disposable Bronchoscope Biopsy Forceps (predicate device) currently cleared to market via 510K K923847 by Wilson-Cook Medical, Inc. The Captura® Disposable Bronchoscope Biopsy Forceps consist of a spool handle, coated coilspring catheter, drive cable, forceps cups and forceps housing.
The bronchoscope biopsy forceps is used by passing the device through a prepositioned endoscope to the targeted location. The forceps cups are attached on one end of the coilspring sheath with the spool handle attached to the opposite end. The handle is actuated by moving the spool forward to open the forceps cups and backward to close. The Captura® Bronchoscope Biopsy Forceps have a surgical rim edge around the circumference of the cups and the cups are fenestrated. The cups are opened and pressed against the biopsy site. When the cups are closed, they are pulled away from the biopsy site, pinching off the top surface, obtaining a tissue biopsy.
## Substantial Equivalence
Changes were made to the predicate Disposable Bronchoscope Biopsy Forceps cleared to market via K923847. These changes include: a material formulation change for the forceps components, a catheter material change, elimination of a spiked forceps cup option and updated handle design.
These changes to the subject device were made in an effort to improve the strength of the components, improve passage through the endoscope and to facilitate process improvements for components.
{4}------------------------------------------------
The subject device operates in the same manner as the predicate device after these changes. Design verification and design validation testing demonstrated substantial equivalence of the device.
The modified Captura® Disposable Bronchoscope Biopsy Forceps (subject device) is substantially equivalent to the predicate with respect to the intended use, key operating mechanics and technological characteristics.
# Summary of Non-Clinical Testing
The following non-clinical testing was conducted to demonstrate the performance of the subject device and confirmed that the subject device performs as intended.
- Shelf Life Testing .
- o Functional Testing
- Design Validation/Verification Testing .
- o Flex and Fracture Design Verification
- o Force to Biopsy
- Packaging Validation ●
- Burst Testing O
- Dye Leak Testing o
# Summary of Biocompatibility Testing
Biocompatibility testing was performed in accordance with the FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part I: Evaluation and testing within a risk management process." The following tests were performed:
- . Cytotoxicity
- . Sensitization
- Irritation ●
- . Acute toxicity
- Material-mediated Pyrogenicity ●
# Conclusion
We believe that the subject device is substantially equivalent to the predicate device in terms of intended use, key operating principles, materials and technological characteristics. The nonclinical data support the substantial equivalence of the subject device and the performance testing, verification and validation testing demonstrate that the subject device should perform as intended when used as instructed in the instructions for use. The non-clinical data demonstrates that the subject device performs comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.