K163376 · Millar, Inc. · DXO · Mar 10, 2017 · Cardiovascular
Device Facts
Record ID
K163376
Device Name
Mikro-Cath Pressure Catheter
Applicant
Millar, Inc.
Product Code
DXO · Cardiovascular
Decision Date
Mar 10, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2870
Device Class
Class 2
Indications for Use
The Mikro-Cath Pressure Catheter is a single-use catheter intended to be used for medical research and diagnostic purposes. The catheter is indicated to measure cardiovascular, intracompartmental, and airway pressures in the human body. The catheter is used as a minimally invasive device under short term limited body contact (<24 hours).
Device Story
Mikro-Cath Pressure Catheter is a single-use, minimally invasive device for measuring cardiovascular, intracompartmental, and airway pressures. A pressure sensor mounted at the distal tip converts sensed pressure or sound into an electrical output signal. The proximal end features a connector for signal transmission. Used in research and diagnostic settings, the device provides real-time pressure data to healthcare providers or researchers to assist in clinical decision-making and physiological monitoring. The device is intended for short-term body contact (<24 hours).
Clinical Evidence
No clinical data provided. No FDA performance standards established. Substantial equivalence is based on identical technological characteristics to the predicate device.
Technological Characteristics
Single-use catheter with a pressure sensor at the distal tip. Electrical output signal proportional to sensed pressure/sound. Proximal connector. Materials, mechanical, and electrical characteristics are identical to K093111.
Indications for Use
Indicated for use in humans for medical research and diagnostic purposes to measure cardiovascular, intracompartmental, and airway pressures. Minimally invasive, short-term use (<24 hours).
Regulatory Classification
Identification
A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2017
Millar, Inc. % Allison Komiyama Principal Consultant Acknowledge Regulatory Strategies 2834 Hawthorn St. San Diego, California 92104
Re: K163376
Trade/Device Name: Mikro-Cath Pressure Catheter, Model 825-0101 Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter Tip Pressure Transducer Regulatory Class: Class II Product Code: DXO Dated: February 6, 2017 Received: February 9, 2017
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Mude Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K163376
Device Name
Mikro-Cath Pressure Catheter, Model 825-0101
#### Indications for Use (Describe)
The Mikro-Cath Pressure Catheter is a single-use catheter intended to be used for medical research and diagnostic purposes. The catheter is indicated to measure cardiovascular, intracompartmental, and airway pressures in the human body. The catheter is used as a minimally invasive device under short term limited body contact (<24 hours).
Type of Use (Select one or both, as applicable)
XPrescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Millar" in a light gray sans-serif font. To the right of the word is a stylized butterfly or flower-like graphic with three petals or wings. The top petal is a gradient of orange and yellow, the middle petal is a gradient of purple and pink, and the bottom petal is a gradient of purple and pink. The overall design is clean and modern.
510(k) Summary
# 510(k) Summary K163376
# DATE PREPARED
February 6, 2017
## MANUFACTURER AND 510(k) OWNER
Millar, Inc. 6001-A Gulf Freeway, Houston, TX 77023, USA Telephone: (832) 667-7000 (713) 714-8498 Fax: Official Contact: Jaqueline Jean-Baptiste, Director of RA/QA
## REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies Telephone: +1 (619) 208-7888 Email: akomiyama@acknowledge-rs.com
## PROPRIETARY NAME OF SUBJECT DEVICE
Mikro-Cath Pressure Catheter, Model 825-0101
COMMON NAME Transducer, Pressure, Catheter Tip
## DEVICE CLASSIFICATION
Catheter tip pressure transducer (21 CFR 870.2870, Product Code DXO, Class II)
#### PREMARKET REVIEW
ODE/DCD/CDDB Cardiovascular Panel
## INDICATIONS FOR USE
The Mikro-Cath Pressure Catheter is a single-use catheter intended to be used for medical research and diagnostic purposes. The catheter is indicated to measure cardiovascular, intracompartmental, and airway pressures in the human body. The catheter is used as a minimally invasive device under short term limited body contact (<24 hours).
## DEVICE DESCRIPTION
The Mikro-Cath Pressure Catheter is a single-use catheter intended to be used for medical research and diagnostic purposes. The catheter is indicated to measure cardiovascular, intracompartmental, and airway pressures in the human body. A pressure sensor is mounted at the distal tip of the catheter. The proximal end terminates in a connector. The sensor produces
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Millar. The logo consists of the word "Millar" in a gray, sans-serif font, positioned to the left of a stylized butterfly or flower-like graphic. The graphic has three petals or wings, colored in shades of orange, yellow, and purple. The overall design is clean and modern.
#### 510(k) Summary
an electrical output signal which varies in direct proportion to the magnitude of sense pressure or sound.
#### PREDICATE DEVICE IDENTIFICATION
The Mikro-Cath Pressure Catheter is substantially equivalent to the following predicates:
| 510(k)<br>Number | Predicate Device Name / Manufacturer | Primary<br>Predicate |
|------------------|-----------------------------------------------------------------------------|----------------------|
| K093111 | Mikro-Cath / Millar Instruments, Inc. | ✓ |
| K061573 | Transpac III Disposable Straight Pressure Transducer (DSPT) / Hospira, Inc. | |
| K830909 | Mikro-Tip Catheter Pressure Transducer / Millar Instruments, Inc. | |
## SUMMARY OF NON-CLINICAL AND CLINICAL TESTING
No FDA performance standards have been established for the Mikro-Cath Pressure Catheter. No additional testing was provided in this submission in order to demonstrate substantial equivalence.
## EQUIVALENCE TO PREDICATE DEVICES
Millar believes that the Mikro-Cath Pressure Catheter is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has identical design and dimensions, and uses similar or identical materials as the device cleared in K093111. The patient contacting materials, mechanical characteristics, and electrical characteristics of the subject device have not changed from the predicate device cleared in K093111. The subject device has the same or similar indications for use as the devices cleared in K093111, K061573, and K830909.
## CONCLUSION
The Mikro-Cath Pressure Catheter is considered substantially equivalent to the predicate devices based on identical technological characteristics. While the subject device has expanded indications (i.e. it may be used to measure intracompartmental and airway pressure), the subject device does not raise new issues of safety or efficacy compared to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.