Atellica CH Phencyclidine (Pcp)

K163220 · Siemens Healthcare Diagnostics, Inc. · LCM · Apr 6, 2017 · TX

Device Facts

Record IDK163220
Device NameAtellica CH Phencyclidine (Pcp)
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeLCM · TX
Decision DateApr 6, 2017
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K163220 · Apr 6, 2017Atellica CH Phencyclidine (Pcp)Siemens Healthcare Diagnostics, Inc.Anonymous, discarded clinical urine samplesMethod comparison study comparing the Atellica CH Phencyclidine (Pcp) assay performance against a GC/MS reference method using clinical urine samples.Method comparison; Clinical urine samples; Diagnostic accuracy

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Method comparison studyClinical urine samples; Sample Size: Not explicitly stated (total count derived from contingency table: 112 samples)GC/MS (Gas Chromatography/Mass Spectrometry)Qualitative and semi-quantitative agreement with reference method

Indications for Use

The Atellica™ CH Phencyclidine (Pcp) assay is for in vitro diagnostic use in the qualitative or semiquantitative analyses of phencyclidine in human urine using the Atellica CH Analyzer, using a cutoff of 25 ng/mL. The Pcp assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/ Mass Spectrometty (GC-MS) or Liquid Chromatography/ Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Device Story

Atellica CH Phencyclidine (Pcp) assay is a homogenous enzyme immunoassay; utilizes competition between drug in urine specimen and drug labeled with glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. G6PDH activity inversely proportional to drug concentration; active enzyme converts NAD+ to NADH, measured spectrophotometrically at 340/410 nm. Used on Atellica CH Analyzer in clinical laboratories. Provides preliminary qualitative or semiquantitative results; requires confirmatory testing via GC/MS or LC-MS/MS. Assists clinicians in identifying phencyclidine use or overdose; results require professional judgment.

Clinical Evidence

Bench testing only. Precision determined per CLSI EP05-A3 (N=80 per concentration). Method comparison against GC/MS reference method showed 98% positive agreement and 95% negative agreement. Interference testing per CLSI EP07-A2 confirmed assay robustness against endogenous compounds, pH, and specific gravity; boric acid identified as causing false negatives.

Technological Characteristics

Homogeneous enzyme immunoassay; liquid reagents (polyclonal sheep anti-PCP antibodies, PCP-labeled G6PDH). Spectrophotometric detection at 340/410 nm. Calibrated for 60-day interval. Onboard reagent stability 30 days. Referenced CLSI guidelines: EP05-A3, EP07-A2, EP17-A2, EP25-A. Standalone assay for Atellica CH Analyzer.

Indications for Use

Indicated for qualitative or semiquantitative detection of phencyclidine in human urine at a 25 ng/mL cutoff. Intended for use as a preliminary analytical test; requires confirmation by GC/MS or LC-MS/MS. For professional use in clinical laboratories.

Predicate Devices

Submission Summary (Full Text)

{1} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k163220 B. Purpose for Submission: New device C. Measurand: Phencyclidine (PCP) D. Type of Test: Enzyme immunoassay E. Applicant: Siemens Healthcare Diagnostics Inc. F. Proprietary and Established Names: Atellica CH Phencyclidine (PCP) G. Regulatory Information: 1. Regulation section: Enzyme Immunoassay, Phencyclidine 2. Classification: Unclassified, 510(k) required 3. Product code: LCM 4. Panel: Toxicology (91) {2} H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: The Atellica CH Phencyclidine (PCP) assay is for in vitro diagnostic use in the qualitative or semiquantitative analyses of phencyclidine in human urine using the Atellica CH Analyzer, using a cutoff of 25 ng/mL. The PCP assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography/ Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used. 3. Special conditions for use statement(s): Prescription use only. For In Vitro Diagnostic Use only. 4. Special instrument requirements: Atellica CH Analyzer I. Device Description: The Atellica CH Phencyclidine reagents are liquid, ready to use. There are two reagents packaged in two separate reagent packs. These include a reagent with antibodies to phencyclidine (polyclonal sheep) and glucose-6-phosphate. The second reagent is phencyclidine labeled with bacterial glucose-6-phosphate dehydrogenase. The Atellica CH Analyzer has been previously cleared as part of the Trinidad CH System under k151767. The assay uses a previously cleared calibrator (k993755). {3} 3 J. Substantial Equivalence Information: 1. Predicate device name(s): Siemens Urine phencyclidine (PCP) screen flex reagent cartridge 2. Predicate 510(k) number(s): k000462 3. Comparison with predicate: | Similarities/Differences | | | | --- | --- | --- | | Item | k163220 Atellica CH Phencyclidine (PCP) Candidate Device | k000462 Urine Phencyclidine (PCP) Screen Flex Reagent Cartridge Predicate Device | | Intended Use: | Qualitative or semiquantitative analysis of phencyclidine (PCP) in human urine using the Atellica CH analyzer. | Qualitative or semiquantitative analysis of phencyclidine (PCP) in human urine using the Dimension clinical chemistry system. | | Methodology: | Enzyme Immunoassay | Same | | Type of Test: | Qualitative or semiquantitative | Same | | Specimen Type: | Human urine | Same | | Cutoff: | 25 ng/mL | Same | | Intended Users: | Prescription use only | Same | | Calibration Frequency: | 60 days | 30 days | K. Standard/Guidance Document Referenced (if applicable): The following guidelines from the Clinical and Laboratory Standards Institute (CLSI) were referenced: EP05-A3. Evaluation of Precision Performance of Quantitative Measurement Methods, Approved Guideline; Third edition EP07-A2. Interference Testing of Clinical Chemistry; Approved Guidelines EP17-A2. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, Approved Guideline, Second edition {4} EP25-A. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline L. Test Principle: The Atellica CH PCP assay is a homogenous enzyme immunoassay based on competition between drug present in the specimen and drug labeled-glucose-6-phosphate dehydrogenase (PCP-G6PDH) for antibodies raised to PCP. PCP-G6PDH activity decreases upon binding to the anti-PCP antibodies and free PCP in the specimen competitively prevents this binding, so that PCP-G6PDH enzyme activity is proportional to drug concentration in the specimen. Active PCP-G6PDH enzyme converts nicotinamide adenine dinucleotide (NAD+) to NADH in the presence of glucose-6-phosphate, resulting in an absorbance change that is measured spectrophotometrically at 340/410 nm. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision study samples were prepared from negative urine samples spiked with nine different concentrations: +100%, +75%, +50%, +25%, cut-off, -25%, -50%, -75% and -100% of the drug cutoff concentration (25 ng/mL) for PCP. Precision was tested using two replicates, two times a day for at least 20 days for a total of at least 80 replicates. The results in the qualitative mode and semi-quantitative mode are identical. The results are summarized below. | Urine Pool (ng/mL) | % of Cutoff | # of Results | Repeatability and Within-Laboratory Results | | --- | --- | --- | --- | | 0 | -100 | 80 | 80 Negative | | 6.25 | -75 | 80 | 80 Negative | | 12.5 | -50 | 80 | 80 Negative | | 18.75 | -25 | 80 | 80 Negative | | 25 | Cut-off | 80 | 60 Positive / 20 Negative | | 31.25 | 25 | 80 | 80 Positive | | 37.5 | 50 | 80 | 80 Positive | | 43.75 | 75 | 80 | 80 Positive | | 50 | 100 | 80 | 80 Positive | b. Linearity/assay reportable range: Urine samples were spiked with PCP concentrations ranging from 4.0 - 80.0 ng/mL. Seven replicates were processed in the same analytical run. For each known {5} concentration, drug recovery was calculated using the mean concentration of the replicates. The data support the semi-quantitative reportable range of 5 - 75 ng/mL. The data are summarized below. | Sample ID | Spiked PCP (ng/mL) | Mean PCP (ng/mL) | % Recovery | | --- | --- | --- | --- | | 1 | 4.0 | 4.3 | 107.5 | | 2 | 5.0 | 5.6 | 112.0 | | 3 | 10.0 | 10.1 | 101.0 | | 4 | 15.0 | 15.0 | 100.0 | | 5 | 20.0 | 19.7 | 98.5 | | 6 | 25.0 | 25.1 | 100.4 | | 7 | 30.0 | 30.0 | 100.0 | | 8 | 40.0 | 41.0 | 102.5 | | 9 | 60.0 | 60.9 | 101.5 | | 10 | 80.0 | 83.7 | 104.6 | c. Traceability, Stability, Expected values (controls, calibrators, or methods): The assay uses previously cleared calibrators (k993755). Reagent Stability: A real time stability study to support a claim of 12 months shelf life is ongoing. Unopened reagents are stable until the expiration date on the product when stored at 2 - 8 °C. Reagents are stable onboard the system for 30 days. Calibration Interval: All protocols and acceptance criteria for Lot Calibration and Pack Calibration Intervals were reviewed and found to be acceptable. The study results support a calibration interval of 60 days and a pack calibration interval of 19 days. d. Detection limit: Not applicable. e. Analytical specificity: Cross-reactivity was evaluated by spiking the structurally similar compounds shown below into drug free urine. All samples were tested in replicates of N = 6. The results are summarized in the table below. {6} 6 | Compound | % Cross-Reactivity | | --- | --- | | 1-(4-Hydroxypiperidino)phenylcyclohexane | 5.97 | | 1-(1-Phenylcyclohexyl)pyrrolidine | 38.33 | | 1-[1-(2-Thienyl)-cyclohexyl]piperidine | 58.11 | | trans-4-phenyl-4-Piperidinocyclohexanol | 74.38 | | Chlorpromazine | 0.02 | | Clomipramine | 0.02 | | Cyclobenzaprine | 0.03 | | Dextromethorphan | 0.02 | | Diphenhydramine | 0.01 | | Doxepin | 0.01 | | Imipramine | 0.01 | | Methoxetamine | 0.03 | | 4-Methoxyphencyclidine | 8.43 | | Thioridazine | 0.04 | | Venlafaxine | 0.01 | Endogenous interferents were evaluated by spiking urine aliquots, with PCP concentrations at $+/-25\%$ of the cutoff, with endogenous interferents at the below indicated concentrations. No positive or negative interference was detected at the indicated concentrations. | Compound | Concentration Tested | | --- | --- | | Acetone | 1.0 g/dL | | Ascorbic Acid | 0.75 g/dL | | Conjugated bilirubin | 0.25 mg/dL | | Creatinine | 0.5 g/dL | | Ethanol | 1.0 g/dL | | Gamma Globulin | 0.5 g/dL | | Galactose | 0.01 g/dL | | Glucose | 2.0 g/dL | | Hemoglobin | 115 mg/dL | | Human Serum Albumin | 0.5 g/dL | {7} 7 | Oxalic Acid | 0.1 g/dL | | --- | --- | | Riboflavin | 7.5 mg/dL | | Sodium Azide | 1% (w/v) | | Sodium Chloride | 1.5 g/dL | | Sodium Fluoride | 1% (w/v) | | Urea | 6.0 g/dL | Structurally unrelated interferents were evaluated by spiking urine aliquots, with PCP with concentrations at +/-25% of the cutoff, with structurally unrelated interferents at the below indicated concentrations. No positive or negative interference was detected at the indicated concentrations. | Compound | Concentration Tested (ng/mL) | | --- | --- | | Acetaminophen | 500,000 | | I-α-Acetylmethadol (LAAM) | 25,000 | | N-Acetyl Procainamide (NAPA) | 100,000 | | Acetylsalicylic Acid | 500,000 | | Amitriptyline | 8,750 | | S-(+)-Amphetamine | 100,000 | | Benzoylecgonine | 100,000 | | Buprenorphine | 100,000 | | Caffeine | 500,000 | | Cannabinol | 100,000 | | Carbamazepine | 100,000 | | Chlordiazepoxide | 100,000 | | Cimetidine | 100,000 | | Clonidine | 100,000 | | Codeine | 25,000 | | Cotinine | 100,000 | | Desipramine | 75,000 | | Dextrorphan | 781 | | Diazepam | 100,000 | | Digoxin | 100,000 | | 2-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 12,500 | | EMDP | 100,000 | | 1R,2S-Ephedrine | 100,000 | | 1S,2R-Ephedrine | 100,000 | | Fluoxetine | 75,000 | {8} | Compound | Concentration Tested (ng/mL) | | --- | --- | | Flurazepam | 50,000 | | Glutethimide | 100,000 | | Haloperidol | 100,000 | | Heroin | 25,000 | | Hydrocodone | 25,000 | | Ibuprofen | 500,000 | | Ketamine | 75,000 | | Ketorolac Tromethamine | 100,000 | | Lidocaine | 100,000 | | Lorazepam | 100,000 | | Lormetazepam | 100,000 | | LSD | 100,000 | | MDMA | 100,000 | | Meperidine | 1,563 | | Methadone | 50,000 | | S(+) - Methamphetamine | 100,000 | | Methaqualone | 100,000 | | Morphine | 75,000 | | Naproxen | 100,000 | | Nordiazepam | 100,000 | | Nortriptyline | 75,000 | | Oxazepam | 100,000 | | Oxycodone | 100,000 | | Phenobarbital | 100,000 | | Phenylephrine | 100,000 | | Phenytoin | 100,000 | | Promethazine | 3,125 | | Propoxyphene | 100,000 | | Propranolol | 100,000 | | Protriptyline | 75,000 | | R,R - Pseudoephedrine | 100,000 | | S,S - Pseudoephedrine | 100,000 | | Ranitidine | 100,000 | | Ritalinic Acid | 100,000 | | Salicylic Acid | 100,000 | | Scopolamine | 100,000 | | Secobarbital | 100,000 | | Tapentadol | 50,000 | | 11-nor-Δ9-THC-9-COOH | 100,000 | | Tramadol | 50,000 | | Trazodone | 100,000 | | Tyramine | 100,000 | 8 {9} | Compound | Concentration Tested (ng/mL) | | --- | --- | | Verapamil | 60,000 | | Zidovudine (AZT) | 100,000 | | Zolpidem | 100,000 | Boric acid 1% (w/v) results in a false negative result. The labeling states that boric acid not be used as a preservative for urine samples. Effect of specific gravity: To evaluate the effect of specific gravity, the specific gravity of drug-free urine samples was adjusted with water or addition of creatinine to obtain the following values: 1.000, 1.002, 1.005, 1.010, 1.015, 1.020, 1.025, and 1.030. Specimens were spiked with PCP at ± 25% of the cutoff values. Six replicates of each specific gravity value and PCP concentration were performed. No positive or negative interference was detected with changes in specific gravity. Effect of pH: To evaluate the effects of pH, drug-free urine was adjusted to 3.0 to 11.0 (+/- 0.2) in increments of 1 pH unit using 0.1N HCL or 0.1N NaOH. Urine pools were spiked with PCP at 25% below and 25% above the cutoff concentrations. Six replicates of each pH value and PCP concentration were analyzed. The pH ranges tested did not affect the results from the device. f. Assay cut-off: Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision section, M.1.a., above. 2. Comparison studies: a. Method comparison with predicate device: A total of 106, unaltered phencyclidine samples, and 6 altered samples (to achieve below cutoff levels), were analyzed using the Atellica CH PCP Assay and the reference method GC/MS. Results were obtained as positive or negative relative to the 25 ng/mL assay cutoff (qualitative mode) or in analyte units (semi-quantitative mode) on the Atellica CH. One replicate was processed for each sample. Twenty-eight samples were within +/- 50% of the cutoff by GC/MS. The results are summarized below by GC/MS and concordance agreement by Atellica CH Phencyclidine: {10} | | GC/MS Results | | | | | | --- | --- | --- | --- | --- | --- | | Atellica PCP | Neg (< 13 ng/mL) | Neg Within 50% below the cutoff (13 - 24 ng/mL) | Pos Within 50% above the cutoff (25 - 38 ng/mL) | Pos (>38 ng/mL) | % Agreement | | Qualitative | | | | | | | Atellica Pos | 0 | 1 | 17 | 36 | 98% | | Atellica Neg | 48 | 7 | 3 | 0 | 95% | | Semi-Quantitative | | | | | | | Atellica Pos | 0 | 1 | 17 | 36 | 98% | | Atellica Neg | 48 | 7 | 3 | 0 | 95% | Discordant Samples between Atellica CH Phencyclidine and GC/MS: | Sample ID | Atellica Semi-quant. Value | GC/MS Value (ng/mL) | Atellica Pos/Neg | GC/MS Pos/Neg | | --- | --- | --- | --- | --- | | 53 | 25 | 23.7 | Pos | Neg | | 57 | 23 | 25.3 | Neg | Pos | | 59 | 23 | 28.2 | Neg | Pos | | 61 | 22 | 30.0 | Neg | Pos | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): None {11} 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 11
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...