Parietene DS Composite Mesh

K163212 · Sofradim Production · FTL · Jun 29, 2017 · General, Plastic Surgery

Device Facts

Record IDK163212
Device NameParietene DS Composite Mesh
ApplicantSofradim Production
Product CodeFTL · General, Plastic Surgery
Decision DateJun 29, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Parietene™ DS composite mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists.

Device Story

Parietene DS Composite Mesh is a surgical implant for ventral hernia repair and abdominal fascial deficiency reinforcement. Device features a permanent macroporous polypropylene textile on one side for long-term tissue reinforcement and a fully absorbable synthetic film on the other to minimize visceral tissue attachment. Film degrades via hydrolysis within 105 days. Device includes pre-placed non-absorbable polypropylene/polyethylene sutures and violet markings for orientation. Used in laparoscopic or open surgical procedures by surgeons. Mesh is placed in intraperitoneal sites. Benefits include structural reinforcement of abdominal wall and reduced risk of visceral adhesions.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing and in vivo animal models. Bench tests evaluated pore size, thickness, surface density, bursting strength, breaking strength, elongation, tear strength, and suture pull-out strength. Animal studies evaluated tissue integration and visceral attachment minimization. Biocompatibility confirmed per ISO 10993-1. Usability testing confirmed safety and effectiveness for intended users and environments.

Technological Characteristics

Composite mesh: permanent monofilament polypropylene textile; absorbable synthetic film (glycolide, caprolactone, trimethylene carbonate, lactide copolymer); polycaprolactone binding agent/marking. Pre-placed sutures: polypropylene/polyethylene. Compliant with ISO 10993-1. Mechanical properties assessed per FDA surgical mesh guidance. Sterilized device.

Indications for Use

Indicated for the repair of ventral hernias or other abdominal fascial deficiencies requiring reinforcing material in patients needing abdominal wall soft tissue reinforcement.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half. In the center of the seal is a stylized image of three human profiles facing to the right, with a flowing ribbon-like element connecting them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 29, 2017 Sofradim Production % Ms. Mary Mellows Regulatory Affairs Manager Covidien 60 Middletown Avenue North Haven, Connecticut 06473 Re: K163212 Trade/Device Name: Parietene Ds Composite Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: May 29, 2017 Received: June 5, 2017 Dear Ms. Mellows: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {2}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # David Krause -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES | |-----------------------------------------| | Food and Drug Administration | ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. ## 510(k) Number (if known) K163212 ## Device Name Parietene™ DS Composite Mesh #### Indications for Use (Describe) Parietene™ DS Composite Mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists. It is indicated for the repair of ventral hernias or other abdominal fascial deficiencies that require the addition of a reinforcing material. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ## FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) #### This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (1/14) {4}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness ## Date Prepared: November 14, 2016 # Submitter: Sofradim Production (subsidiary of Covidien IIc) 116, avenue du Formans 01600 Trevoux, France +33 (0)4 74 08 90 00 Telephone: +33 (0) 4 74 08 90 02 Fax: # Contact: Mary Mellows Regulatory Affairs Manager Covidien 60 Middletown Avenue North Haven, CT 06473, USA (203) 492-5284 Telephone: (203) 492-5029 Fax: Email: mary.e.mellows@medtronic.com # Name of device: | | Trade/Proprietary name: | Parietene™ DS Composite Mesh | |--|-------------------------|-----------------------------------------------------------------| | | Common name: | Surgical Mesh | | | Classification name: | Mesh, Surgical, Polymeric<br>• Panel number and product code: 7 | - 9 FTL • Regulation number: 21 CFR 878.3300 ## Predicate Device: | Trade/Proprietary name: | Proceed™ Surgical Mesh | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common name: | Surgical Mesh | | Classification name: | Mesh, Surgical, Polymeric<br>• Panel number and product code: 79 FTL<br>• Regulation number: 21 CFR 878.3300 | | 510(k) Number: | K060713 | | Manufacturer: | Ethicon | | Reason for 510(k) Submission: | To obtain market clearance for the Parietene™ DS Composite Mesh. | | Device Description: | Parietene™ DS Composite Mesh is designed to be placed in an intraperitoneal site by a laparoscopic or open approach. The mesh is composed of a permanent macroporous polypropylene textile on one side and a fully absorbable synthetic film on the other side. The film is adhered to the textile using a binding agent localized on the textile fibers. A violet marking is positioned on the | {5}------------------------------------------------ mesh to help center and orient the mesh. Non-absorbable preplaced sutures are tied to the mesh. -The macroporous textile is knitted from a permanent monofilament polypropylene yarn. - The synthetic film is made out of absorbable synthetic copolymer of glycolide, caprolactone, trimethylene carbonate, and lactide. - The binding agent and the violet marking are made out of absorbable polycaprolactone. The D&C Violet No. 2 dye is used for the marking. - The pre-placed sutures are made out of an isotactic crystalline stereoisomer of polypropylene (a synthetic linear polyolefin) and polyethylene. The Copper Phthalocyanine Blue is used to color the sutures. The permanent macroporous textile is designed to be placed over the abdominal wall to ensure long term reinforcement of soft tissues, while the continuous absorbable film is designed to minimize tissue attachment to the mesh in case of direct contact with the viscera and is essentially degraded within 105 days by hydrolysis. Intended Use: Parietene™ DS composite mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists. Parietene™ DS composite mesh is intended for the reinforcement Indications for use: of abdominal wall soft tissue where a weakness exists. It is indicated for the repair of ventral hernias or other abdominal fascial deficiencies that require the addition of a reinforcing material. Summary comparing The subject Parietene™ DS Composite Mesh is substantially equivalent to the predicate device Proceed™ Surgical Mesh the (K060713) in terms of indications and design for the following characteristics of the subject and predicate technological characteristics: device: - Polypropylene textile performance - Minimizing tissue attachment layer ● - Design: composite mesh with equivalent shapes and sizes ● Materials: Parietene™ DS composite mesh and its constitutive components have been evaluated and found compliant with ISO Standard 10993-1. - Performance data: The following performance data is provided in support of the substantial equivalence determination: In vitro (bench) tests have been performed in accordance with the "Guidance for the Preparation of a Premarket Notification Application of a Surgical Mesh" issued March 2, 1999 to evaluate the performance characteristics of the subject Parietene™ DS Composite Mesh in comparison with the predicate Proceed™ Surgical Mesh (K060713). The following mesh characteristics were assessed: {6}------------------------------------------------ - Pore size - · Thickness - Surface density . - Bursting strength . - . Bursting distension - . Breaking strength - Elongation at break . - · Tear strength - Suture pull-out strength . Knot-pull tensile strength and suture pull-out tests were performed on the pre-placed sutures of the subject Parietene™ DS Composite Mesh to demonstrate conformity to USP monograph for non-absorbable surgical suture and to assess the strength of the textile/suture assembly. Trocar passage testing was conducted to demonstrate that the inteqrity of the mesh is preserved when used by laparoscopy. - . In vivo pre-clinical tests on representative animal models were conducted in comparison with the predicate Proceed™ Surqical Mesh (K060713) to evaluate the tissue integration and to demonstrate the minimizing tissue attachment performance in case of contact with viscera. - . Stability study was conducted and the proposed device shelf life was demonstrated. - . Biocompatibility tests were performed to confirm that Parietene™ DS Composite Mesh and its constitutive components are compliant with ISO Standard 10993-1 for their intended patient contact profile. - . Usability tests were conducted and demonstrate that the subject Parietene™ DS Composite Mesh is safe and effective for the intended users, uses and use environments. This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence. # Conclusion: All testing demonstrates that the subject Parietene™ DS Composite Mesh is substantially equivalent to the predicate device, Proceed™ Surgical Mesh (K060713).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...