The Yooeehouse breast pump is intended to be used by lactating women to express and collect milk from their breast. It can be used as a single pump and as a double pump. The unit is intended for indoor use only and is intended for single users. The Yooeehouse pumpset should be used in combination with Yooeehouse breast pump and is intended to be used by lactating women to express and collect milk from their breast. The pumpset can be used both as a single pumpset and as a double pumpset.
Device Story
Powered breast pump for lactating women; home use. Device consists of pump unit, non-sterile pumpset (silicone, polypropylene, ABS, polyphenylene sulfone), and USB power adapter. Microcomputer-controlled negative pressure module creates vacuum to extract milk. User selects mode (massage or expression), velocity, and intensity. Solenoid valve provides variable vacuum cycling (24-110 cycles/min). Massage mode: 30 mmHg; expression mode: 60-220 mmHg. Models S08/S10 feature LED indicators; L08/L10 feature LCD. Device provides single or double pumping. Healthcare provider does not operate; patient self-use. Output is expressed milk collected in containers. Benefits include assistance with milk expression for breastfeeding.
Clinical Evidence
Bench testing only. Testing included backflow protection, vacuum profile (min/max levels, cycle rates), and durability. Electrical safety (AAMI/ANSI ES60601-1, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10) were verified. Software validation performed per FDA guidance.
Technological Characteristics
Powered breast pump; micro air pump mechanism. Materials: silicone, polypropylene, ABS, polyphenylene sulfone. Power: USB adapter. Vacuum: 60-220 mmHg (expression), 30 mmHg (massage). Cycling: 24-110 cycles/min. Interface: LED or LCD. Backflow protection via valve. Microcomputer-controlled.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for single-user, indoor use; supports single or double pumping.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Ardo Carum and Calypso Powered Breast Pumps (K141742)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 2, 2017
Nantong Poly Technology Co., Ltd. % Jonathan Hu Technical Manager Medwheat (Shanghai) Medical Technology Co.,Ltd. Yangpu District Liaoyuan East Road Shuangyang First Suite No.33 Room 303 Shanghai, 200093 China
Re: K163183
Trade/Device Name: Yooeehouse Breast Pump (Models: S08, S10, L08, L10) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: August 15, 2017 Received: August 22, 2017
Dear Jonathan Hu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -S" in a large, sans-serif font. The text is black and appears to be the main subject of the image. The background is plain white, which makes the text stand out clearly. There is a faint watermark in the background.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163183
Device Name
Yooeehouse Breast Pump (Models: S08, S10, L08, L10)
Indications for Use (Describe)
The Yooeehouse breast pump is intended to be used by lactating women to express and collect milk from their breast. It can be used as a single pump and as a double pump. The unit is intended for indoor use only and is intended for single users.
The Yooeehouse pumpset should be used in combination with Yooeehouse breast pump and is intended to be used by lactating women to express and collect milk from their breast. The pumpset can be used both as a single pumpset and as a double pumpset.
Type of Use (Select one or both, as applicable)
| | Investor/Insider Use (See SEC Rule 144 & Section 16) |
|--|------------------------------------------------------|
| | Issuer Use: Compensation and/or SEC Rule 10b5-1 Plan |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K163183 Page 1 of 4
# 510(k) Summary
### Submitter's Information
Name of Sponsor:
Contact Name: Telephone No .: Fax No .: Email Address:
### Date Prepared
September 28, 2017
### Correspondent's Information
Company Name: Address:
Correspondent Name: Telephone No .: Email Address:
## Medwheat ( Shanghai) Medical Technology Co., Ltd. Yangpu District Liaoyuan East Road Shuangyang, First Suite No, 33, Room 303, Shanghai, China Jonathan Hu 86-021-65181421 Jonathan.hu@medwheat.com
### Device Name and Classification
Trade Name: Model Names: Common Name: Classification Number: Classification Name: Product Code: Regulatory Class:
Yooeehouse Breast Pump S08. S10. L08. L10 Powered Breast Pump 21 CFR 884.5160 Powered Breast Pump HGX (Pump, Breast, Powered) II
### Identification of Predicate Device(s)
K141742 - Ardo Carum and Calypso Powered Breast Pumps
The predicate devices have not been subject to a design-related recall.
### Description of the Device
The subject device is a breast pump that is intended to be a single patient reusable device. The device is used in a home environment and is capable of single and dual pumping. The submission includes four different breast pumps: L08, L10, S08, and S10.
The breast pump uses a USB adapter. The breast pump consists of a pump, a non-sterile pumpset (composed of silicone, polypropylene, acrylonitrile butadiene styrene, and polyphenylene sulfone), and a power adapter. The breast pump has a negative pressure
Nantong Poly Technology Co., Ltd. No.332, Xinxing Road, Development Zone, Nantong, Jiangsu, China Rongmin Lin 86-0513 81528229 86-0513 81528229 lin_11000@126.com
{5}------------------------------------------------
module that utilizes a microcomputer-controlled system to set and adjust the mode (massage or expression), velocity and intensity of stimulation and suction. The vacuum pump component of the device creates the negative pressure used to extract milk from the breast. The vacuum pump is connected to a solenoid valve that provides variable vacuum cycling rates ranging from 24-110 cycles/minutes. In the massage mode, there is single vacuum pressure of 30 mmHg, while the expression mode has vacuum pressures ranging from 60 to 220 mmHg depending on the expression setting selected. A light-emitting diode indicator (Models S08 and S10) or liquid-crystal display (models L08 and L10) indicate the status of operation.
### Indication for Use
The Yooeehouse breast pump is intended to be used by lactating women to express and collect milk from their breast. It can be used as a single pump and as a double pump. The unit is intended for indoor use only and is intended for single users.
The Yooeehouse pumpset should be used in combination with Yooeehouse powered breast pumps and is intended to be used by lactating women to express and collect milk from their breast. The pumpset can be used both as a single pumpset and as a double pumpset.
| | K 163183 | K141742 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Yooeehouse breast pump is<br>intended to be used by lactating women<br>to express and collect milk from their<br>breast. It can be used as a single pump<br>and as a double pump. The unit is<br>intended for indoor use only and is<br>intended for single users.<br><br>The Yooeehouse pumpset should be<br>used in combination with Yooeehouse<br>powered breast pumps and is intended<br>to be used by lactating women to<br>express and collect milk from their<br>breast. The pumpset can be used both as<br>a single pumpset and as a double<br>pumpset. | The ARDO Carum powered breast<br>pump is intended to be used by<br>lactating women to express and collect<br>milk from their breast. It can be used as<br>a single pump and as a double pump.<br>The unit is intended for indoor use only<br>and is intended for multiple users.<br><br>The ARDO Calypso breast pump is<br>intended to be used by lactating women<br>to express and collect milk from their<br>breast. It can be used as a single pump<br>and as a double pump. The unit is<br>intended for indoor use only and is<br>intended for single users.<br><br>The ARDO Pumpset should be used in<br>combination with ARDO breast pumps<br>and is intended to be used by lactating<br>women to express and collect milk<br>from their breast. The Pumpset can be<br>used both as a single pumpset and as a<br>double pumpset |
Comparison of Indication for Use with Predicate Device
The subject device's indication for use statement is not identical to the predicate device. The predicate device has separate indication for use statements for each version included in the predicate submission; however, both the subject and predicate device have the same intended use to express and collect milk from lactating women for breast feeding.
{6}------------------------------------------------
| Device & Predicate Device(s): | K163183 | | | K141742 | | | |
|----------------------------------------|---------------------|---------------------|---------------------|---------------------|------------------------------------------------------------|--|--|
| General Device Characteristics | | | | | | | |
| Models | S08 | S10 | L08 | L10 | | | |
| Code | HGX | HGX | HGX | HGX | | | |
| Single User | Yes | Yes | Yes | Yes | Carum : multiuser<br>Calypso: single user | | |
| Single/double pump | Single or<br>double | Single or<br>double | Single or<br>double | Single or<br>double | Single or double | | |
| Pump Type | Micro air<br>pump | Micro air<br>pump | Micro air<br>pump | Micro air<br>pump | Micro air pump | | |
| Media separation (backflow protection) | Valve | Valve | Valve | Valve | Valve | | |
| Expression pattern | 2- phase | 2- phase | 2- phase | 2- phase | 2- phase | | |
| Specifications | | | | | | | |
| Power supply | Micro USB | Micro USB | Micro<br>USB | Micro USB | Carum: AC or rechargeable<br>Calypso: Rechargeable battery | | |
| Suction levels (massage/stimulation) | 1 | 1 | 1 | 1 | 7 (Carum) | | |
| Suction levels (expression) | 9 | 9 | 9 | 9 | 8 | | |
| Suction Strength (expression) | 60- 220<br>mmHg | 60- 220<br>mmHg | 60- 220<br>mmHg | 60- 220<br>mmHg | Carum: 22-250 mmHg<br>Calypso: 37-250 mmHg | | |
| Suction levels (Expression) | 9 | 9 | 9 | 9 | 8 | | |
| Cycles per minutes (expression) | 24-65 | 24-65 | 24-65 | 24-65 | 30-60 | | |
| Visual indicator | LED | LED | LCD | LCD | LCD | | |
### Comparison of Technological Characteristics with Predicate Device
The subject device and the predicate device do not have the same technological characteristics. The maximum vacuum pressure, suction levels, cycle speed, and power supply are different between the subject device and the predicate device. These differences do not raise different questions of safety and effectiveness, and can be addressed through performance testing.
### Non-Clinical Performance Test Summary
The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination:
- Bench Testing
- Backflow Protection Testing: Performance testing was conducted to assess backflow . protection of the subject device (i.e., prevention of milk ingress into the pump tubing or pump).
- . Vacuum Profile Testing: Performance testing was conducted at minimum and maximum vacuum settings to determine the minimum and maximum vacuum levels of the pump as compared to its specifications. The specifications were met for vacuum level, cycle rate, and backflow protection. Testing was conducted under single and double pumping conditions with appropriate power sources.
- . Durability testing: performance testing was conducted to assess use-life of the device.
- Electrical Safety
- AAMI / ANSI ES60601-1:2005/ (R) 2012 and A1:2012, C1:2009/(R) 2012 and .
{7}------------------------------------------------
K163183 Page 4 of 4
- A2:2010/(R) 2012
- . IEC 60601-1-11:2015
- Electromagnetic Compatibility
- IEC 60601-1-2:2014 ●
- . Biocompatibility
- Sensitization ISO 10993-10:2010 •
- . Irritation - ISO 10993-10:2010
- . Cytotoxicity - ISO10993-5:2009
- Software Validation ●
- FDA Guidance: "Guidance for the Content of Premarket Submissions for Software • Contained in Medical Devices" issued May 11, 2005.
### Conclusion
Based on the comparison and analysis above, the Yooeehouse Breast Pump is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.