Ceraclean is an extraoral cleaner of pre-treated ceramic, zirconia and metal restoration surfaces which have been contaminated during intraoral try-in.
Device Story
CeraClean is an extraoral cleaning gel used by dental professionals to remove saliva contamination from prosthetic restorations (ceramic, zirconia, metal) after intraoral try-in. The device is applied to the restoration surface via syringe tip or brush, then rinsed off with water prior to final cementation or priming. Unlike abrasive cleaners, CeraClean is a non-abrasive, water-based gel containing potassium hydroxide. It functions by chemically cleaning the substrate; the gel is fully water-soluble and leaves a neutral pH surface after rinsing. By effectively removing contaminants, the device ensures optimal bonding conditions for the final placement of the restoration, potentially improving clinical outcomes and restoration longevity.
Clinical Evidence
Bench testing only. Evaluated shear bond strength using modified ISO 29022:2013, water solubility, and pH levels. SEM EDX analysis confirmed complete removal of the cleaning agent from the substrate. Biocompatibility testing performed per ISO 7405:2008 and ISO 10993-1:2009, including cytotoxicity testing, demonstrated compliance with requirements.
Technological Characteristics
Water-based gel; active ingredient: potassium hydroxide; viscosity modifier: xanthum gum; pigmented (blue). Non-abrasive; water-soluble. Applied via syringe or brush. Neutral pH after rinsing. Biocompatibility tested per ISO 7405:2008 and ISO 10993-1:2009.
Indications for Use
Indicated for cleaning pre-treated ceramic, zirconia, and metal dental restoration surfaces contaminated during intraoral try-in.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2017
Bisco, Inc. Ryan Hobson Regulatory Affairs Product Registration Coordinator 1100 West Irving Park Road Schaumburg, Illinois 60193
Re: K163171
Trade/Device Name: CeraClean Regulation Number: 21 CFR 872.3260 Regulation Name: External Cleaning Solution Regulatory Class: Class II Product Code: PME Dated: March 23, 2017 Received: March 20, 2017
Dear Ryan Hobson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163171
Device Name CeraClean
Indications for Use (Describe)
Ceraclean is an extraoral cleaner of pre-treated ceramic, zirconia and metal restoration surfaces which have been contaminated during intraoral try-in.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------|
| Research Use (Per 21 CFR 801.3 and/or R) <input checked="checked" type="checkbox"/> |
| For In-Vitro Use Only (21 CFR 801.5 and/or G) <input type="checkbox"/> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BISCO. The logo consists of five connected, three-dimensional hexagonal shapes. The shapes alternate in color between red and blue. Below the shapes is the word "BISCO" in a gray, sans-serif font. The letters are bold and slightly rounded.
### 510(k) SUMMARY
| Applicant: | Bisco, Inc.<br>1100 W. Irving Park Road<br>Schaumburg IL, 60193 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ryan Hobson<br>Bisco, Inc.<br>1100 W. Irving Park Road<br>Schaumburg IL, 60193<br>Tel: 847-534-6143<br>Fax: 847-534-6143 |
| Date Prepared: | 13 April 2017 |
| Trade Name: | CeraClean |
| Common Name: | Restoration Cleaner |
| Product Code: | PME |
Predicate Device:
Classification/Name:
Resolve 2.0, K152322 by Denali Corporation
### CeraClean Indication for Use:
CeraClean is an extraoral cleaner of pre-treated ceramic, zirconia and metal restoration surfaces which have been contaminated during intraoral try-in.
External Cleaning Solution
Class II per 21 CFR 872.3260
#### Description of Applicant Device:
CeraClean is used outside the mouth, as an extraoral cleaning gel suitable for non-abrasive cleaning of the bonding surfaces of prosthetic restorations after intraoral try-in. CeraClean is applied through a single-use tip or application to the prosthetic restorations and rinsed off prior to the application of a primer or cement for final placement of the restoration.
> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the BISCO company logo. The logo consists of five connected hexagons, colored in red and blue. Below the hexagons, the word "BISCO" is written in bold, gray letters, with the bottom portion of the letters in black.
# 510(k) SUMMARY (continued)
# Comparison of Technological Characteristics:
### Indications for Use
| Resolve 2.0<br>K152322 (predicate device) | CeraClean |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Resolve 2.0 is recommended for use on dental<br>restorations, external to the mouth, prior to<br>insertion into the mouth. After applying<br>Resolve 2.0 to cleanse the restoration, the<br>surface of the restoration should be washed<br>thoroughly with water, and then air dried<br>before placement in the mouth. | CeraClean is an extraoral cleaner of pre-treated<br>ceramic, zirconia and metal restoration<br>surfaces which have been contaminated during<br>intraoral try-in. |
CeraClean's Indications for Use Statement does not include the cleaning information included in the predicate's statement and list pre-treated ceramic, zirconia and metal restorations instead of just general restorations. These differences are not critical to the intended use of the subject device as equivalent cleaning instructions are included within the instructions for use. The increased specificity list the most common materials used in dental prosthetic restorations and fall within the general indication and no not alter the device risk. The subject device is just as safe and effective as compared to the predicate device.
#### Chemical Composition
| Chemical Composition | Resolve 2.0<br>K152322 (predicate device) | CeraClean |
|----------------------|-------------------------------------------|---------------------|
| Composition | Acrylate resin based gel | Water based gel |
| Viscosity modifier | Silicate | Xanthum Gum |
| Active ingredient | Chlorohexidine | Potassium Hydroxide |
| pH after rinsing | Neutral | Neutral |
| Water Solubility | Water Soluble | Water Soluble |
| Pigmented | Yes - Pink | Yes - Blue |
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
K163171
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the BISCO company logo. The logo consists of five connected hexagons, alternating in color between red and blue. Below the hexagons is the word "BISCO" in a bold, sans-serif font, with the letters in gray and black.
### 510(k) SUMMARY (continued)
Comparison of Technological Characteristics (continued):
| Physical / Mechanical<br>Property | Resolve 2.0<br>K152322 (predicate device) | CeraClean |
|--------------------------------------------------|-------------------------------------------|------------------------------|
| Shear Bond Strength<br>(modified ISO 29022:2013) | Removes saliva contamination | Removes saliva contamination |
| Method of application | Syringe tip w/brush | Syringe tip or brush |
| Method of cleaning | Abrasive | Non-abrasive |
| Method of Removal | Water spray & air dried | Water spray & air dried |
| Delivery configuration | Syringe | Syringe or Bottle |
| Removal | Water Soluble | Water Soluble |
The major differences between Resolve 2.0 and CeraClean are that Resolve 2.0 is a pigmented acrylate resin based gel that uses silicate as a viscosity modifier, includes Chlorohexidine, and is abrasive. CeraClean however, is a pigmented water based gel that uses xanthum gum as a viscosity modifier and does not require abrasive scrubbing to clean the substrate of saliva contamination in preparation for final restoration placement. The lack of acrylate resins in CeraClean does not raise new questions of safety or effectiveness as both devices are water soluble, used outside the mouth, and rinsed off the restorative device prior to placement in the mouth. CeraClean does not include the antimicrobial agent Chlorohexidine. Both devices are pigmented, water soluble, and cleaned substrates are neutral after rinsing.
> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for BISCO. The logo consists of five geometric shapes arranged horizontally, with the first and last shapes being red, the second and fourth shapes being blue, and the middle shape being a lighter shade of blue. Below the shapes is the word "BISCO" in a bold, sans-serif font, with the "BI" in gray and the "SCO" in black.
### 510(k) SUMMARY (continued)
### Performance Data:
Shear bond strength was tested based upon ISO 29022:2013 Dentistry - Adhesion - Notchededge shear bond strength test. Additionally, water solubility and pH (before and after rinse) were tested. A comparison of the physical/mechanical properties of CeraClean to the predicate device is provided in the table below.
| Physical / Mechanical<br>Property | CeraClean |
|--------------------------------------------------|---------------------------------------------------|
| Shear Bond Strength<br>(modified ISO 29022:2013) | Substantially equivalent to the predicate device. |
| Water Solubility | Substantially equivalent to the predicate device. |
| pH (before rinse / after rinse) | Predicate: pH = 6/7, CeraClean pH = 14/7 |
| SEM EDX | Demonstrates device is removed from substrate. |
# Biocompatibility:
An evaluation of biocompatibility was conducted using ISO 7405:2008 and ISO 10993-1:2009. It is concluded from the biocompatibility evaluation and the results of the cytotoxicity testing that CeraClean met the requirements of the test.
#### Conclusion:
It is concluded that the information supplied in this submission has demonstrated that CeraClean is substantially equivalent in design, composition, performance, and intended use to the legally marketed predicate device.
> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.