CeraClean

K163171 · Bisco, Inc. · PME · Apr 14, 2017 · Dental

Device Facts

Record IDK163171
Device NameCeraClean
ApplicantBisco, Inc.
Product CodePME · Dental
Decision DateApr 14, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3260
Device ClassClass 2

Intended Use

Ceraclean is an extraoral cleaner of pre-treated ceramic, zirconia and metal restoration surfaces which have been contaminated during intraoral try-in.

Device Story

CeraClean is an extraoral cleaning gel used by dental professionals to remove saliva contamination from prosthetic restorations (ceramic, zirconia, metal) after intraoral try-in. The device is applied to the restoration surface via syringe tip or brush, then rinsed off with water prior to final cementation or priming. Unlike abrasive cleaners, CeraClean is a non-abrasive, water-based gel containing potassium hydroxide. It functions by chemically cleaning the substrate; the gel is fully water-soluble and leaves a neutral pH surface after rinsing. By effectively removing contaminants, the device ensures optimal bonding conditions for the final placement of the restoration, potentially improving clinical outcomes and restoration longevity.

Clinical Evidence

Bench testing only. Evaluated shear bond strength using modified ISO 29022:2013, water solubility, and pH levels. SEM EDX analysis confirmed complete removal of the cleaning agent from the substrate. Biocompatibility testing performed per ISO 7405:2008 and ISO 10993-1:2009, including cytotoxicity testing, demonstrated compliance with requirements.

Technological Characteristics

Water-based gel; active ingredient: potassium hydroxide; viscosity modifier: xanthum gum; pigmented (blue). Non-abrasive; water-soluble. Applied via syringe or brush. Neutral pH after rinsing. Biocompatibility tested per ISO 7405:2008 and ISO 10993-1:2009.

Indications for Use

Indicated for cleaning pre-treated ceramic, zirconia, and metal dental restoration surfaces contaminated during intraoral try-in.

Regulatory Classification

Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 14, 2017 Bisco, Inc. Ryan Hobson Regulatory Affairs Product Registration Coordinator 1100 West Irving Park Road Schaumburg, Illinois 60193 Re: K163171 Trade/Device Name: CeraClean Regulation Number: 21 CFR 872.3260 Regulation Name: External Cleaning Solution Regulatory Class: Class II Product Code: PME Dated: March 23, 2017 Received: March 20, 2017 Dear Ryan Hobson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163171 Device Name CeraClean Indications for Use (Describe) Ceraclean is an extraoral cleaner of pre-treated ceramic, zirconia and metal restoration surfaces which have been contaminated during intraoral try-in. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------| | Research Use (Per 21 CFR 801.3 and/or R) <input checked="checked" type="checkbox"/> | | For In-Vitro Use Only (21 CFR 801.5 and/or G) <input type="checkbox"/> | > Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for BISCO. The logo consists of five connected, three-dimensional hexagonal shapes. The shapes alternate in color between red and blue. Below the shapes is the word "BISCO" in a gray, sans-serif font. The letters are bold and slightly rounded. ### 510(k) SUMMARY | Applicant: | Bisco, Inc.<br>1100 W. Irving Park Road<br>Schaumburg IL, 60193 | |-----------------|--------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Ryan Hobson<br>Bisco, Inc.<br>1100 W. Irving Park Road<br>Schaumburg IL, 60193<br>Tel: 847-534-6143<br>Fax: 847-534-6143 | | Date Prepared: | 13 April 2017 | | Trade Name: | CeraClean | | Common Name: | Restoration Cleaner | | Product Code: | PME | Predicate Device: Classification/Name: Resolve 2.0, K152322 by Denali Corporation ### CeraClean Indication for Use: CeraClean is an extraoral cleaner of pre-treated ceramic, zirconia and metal restoration surfaces which have been contaminated during intraoral try-in. External Cleaning Solution Class II per 21 CFR 872.3260 #### Description of Applicant Device: CeraClean is used outside the mouth, as an extraoral cleaning gel suitable for non-abrasive cleaning of the bonding surfaces of prosthetic restorations after intraoral try-in. CeraClean is applied through a single-use tip or application to the prosthetic restorations and rinsed off prior to the application of a primer or cement for final placement of the restoration. > BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A. 800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the BISCO company logo. The logo consists of five connected hexagons, colored in red and blue. Below the hexagons, the word "BISCO" is written in bold, gray letters, with the bottom portion of the letters in black. # 510(k) SUMMARY (continued) # Comparison of Technological Characteristics: ### Indications for Use | Resolve 2.0<br>K152322 (predicate device) | CeraClean | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Resolve 2.0 is recommended for use on dental<br>restorations, external to the mouth, prior to<br>insertion into the mouth. After applying<br>Resolve 2.0 to cleanse the restoration, the<br>surface of the restoration should be washed<br>thoroughly with water, and then air dried<br>before placement in the mouth. | CeraClean is an extraoral cleaner of pre-treated<br>ceramic, zirconia and metal restoration<br>surfaces which have been contaminated during<br>intraoral try-in. | CeraClean's Indications for Use Statement does not include the cleaning information included in the predicate's statement and list pre-treated ceramic, zirconia and metal restorations instead of just general restorations. These differences are not critical to the intended use of the subject device as equivalent cleaning instructions are included within the instructions for use. The increased specificity list the most common materials used in dental prosthetic restorations and fall within the general indication and no not alter the device risk. The subject device is just as safe and effective as compared to the predicate device. #### Chemical Composition | Chemical Composition | Resolve 2.0<br>K152322 (predicate device) | CeraClean | |----------------------|-------------------------------------------|---------------------| | Composition | Acrylate resin based gel | Water based gel | | Viscosity modifier | Silicate | Xanthum Gum | | Active ingredient | Chlorohexidine | Potassium Hydroxide | | pH after rinsing | Neutral | Neutral | | Water Solubility | Water Soluble | Water Soluble | | Pigmented | Yes - Pink | Yes - Blue | 800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com K163171 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the BISCO company logo. The logo consists of five connected hexagons, alternating in color between red and blue. Below the hexagons is the word "BISCO" in a bold, sans-serif font, with the letters in gray and black. ### 510(k) SUMMARY (continued) Comparison of Technological Characteristics (continued): | Physical / Mechanical<br>Property | Resolve 2.0<br>K152322 (predicate device) | CeraClean | |--------------------------------------------------|-------------------------------------------|------------------------------| | Shear Bond Strength<br>(modified ISO 29022:2013) | Removes saliva contamination | Removes saliva contamination | | Method of application | Syringe tip w/brush | Syringe tip or brush | | Method of cleaning | Abrasive | Non-abrasive | | Method of Removal | Water spray & air dried | Water spray & air dried | | Delivery configuration | Syringe | Syringe or Bottle | | Removal | Water Soluble | Water Soluble | The major differences between Resolve 2.0 and CeraClean are that Resolve 2.0 is a pigmented acrylate resin based gel that uses silicate as a viscosity modifier, includes Chlorohexidine, and is abrasive. CeraClean however, is a pigmented water based gel that uses xanthum gum as a viscosity modifier and does not require abrasive scrubbing to clean the substrate of saliva contamination in preparation for final restoration placement. The lack of acrylate resins in CeraClean does not raise new questions of safety or effectiveness as both devices are water soluble, used outside the mouth, and rinsed off the restorative device prior to placement in the mouth. CeraClean does not include the antimicrobial agent Chlorohexidine. Both devices are pigmented, water soluble, and cleaned substrates are neutral after rinsing. > BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A. 800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for BISCO. The logo consists of five geometric shapes arranged horizontally, with the first and last shapes being red, the second and fourth shapes being blue, and the middle shape being a lighter shade of blue. Below the shapes is the word "BISCO" in a bold, sans-serif font, with the "BI" in gray and the "SCO" in black. ### 510(k) SUMMARY (continued) ### Performance Data: Shear bond strength was tested based upon ISO 29022:2013 Dentistry - Adhesion - Notchededge shear bond strength test. Additionally, water solubility and pH (before and after rinse) were tested. A comparison of the physical/mechanical properties of CeraClean to the predicate device is provided in the table below. | Physical / Mechanical<br>Property | CeraClean | |--------------------------------------------------|---------------------------------------------------| | Shear Bond Strength<br>(modified ISO 29022:2013) | Substantially equivalent to the predicate device. | | Water Solubility | Substantially equivalent to the predicate device. | | pH (before rinse / after rinse) | Predicate: pH = 6/7, CeraClean pH = 14/7 | | SEM EDX | Demonstrates device is removed from substrate. | # Biocompatibility: An evaluation of biocompatibility was conducted using ISO 7405:2008 and ISO 10993-1:2009. It is concluded from the biocompatibility evaluation and the results of the cytotoxicity testing that CeraClean met the requirements of the test. #### Conclusion: It is concluded that the information supplied in this submission has demonstrated that CeraClean is substantially equivalent in design, composition, performance, and intended use to the legally marketed predicate device. > BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A. 800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com