POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs
K163146 · Hebei Hongsen Plastics Technology Co, Ltd. · LZC · Feb 13, 2017 · General Hospital
Device Facts
Record ID
K163146
Device Name
POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs
Applicant
Hebei Hongsen Plastics Technology Co, Ltd.
Product Code
LZC · General Hospital
Decision Date
Feb 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Device Story
Disposable nitrile patient examination glove; worn by clinicians to provide barrier protection against contamination between patient and examiner. Device tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings; non-sterile. Provides physical barrier; protects healthcare workers from exposure to hazardous chemotherapy agents. Benefits include reduced risk of cross-contamination and chemical exposure.
Clinical Evidence
Bench testing only. No clinical data. Compliance verified via ASTM D6319-10 (physical properties), ASTM D5151-06 (freedom from holes), ASTM D6124-06 (residual powder), ASTM D6978-05 (chemotherapy drug permeation), and ISO 10993-10:2010 (irritation and sensitization).
Technological Characteristics
Material: Nitrile. Form factor: Ambidextrous, powder-free, blue, non-sterile. Standards: ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ASTM D6978-05, ISO 2859-1, ISO 10993-10. Dimensions: Length 230mm, varying widths/thicknesses per size.
Indications for Use
Indicated for use as a disposable medical examination glove worn on the examiner's hands to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-Sterile, Tested for use with Chemotherapy Drugs (K141982)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular text element and a stylized graphic. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The graphic element features a stylized depiction of three human profiles facing to the right, layered on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 13, 2017
Hebei Hongsen Plastics Technology Co, Ltd % Ray Wang Official Correspondent Beijing Believe Tech. Service Co., Ltd. 5-1206, Build 332, Dafangju, No.25 Banbidian Rd.,liyuan, Tongzhou District Beijing, 101121 CN
Re: K163146 Trade/Device Name: POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: January 3. 2017 Received: January 9, 2017
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D.
Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K163146
### Device Name
POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs
#### Indications for Use (Describe)
The POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed device was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Chemotherapy Drug Permeation
| The following chemicals have been tested with proposed device. | | |
|----------------------------------------------------------------|-------------------------|----------------------------------------|
| Chemotherapy Drug | Concentration | Breakthrough Detection Time in Minutes |
| Fluorouracil | 50.0 mg/ml (50,000 ppm) | > 240 |
| Etoposide (Toposar) | 20.0 mg/ml (20,000 ppm) | > 240 |
| Cyclophosphamid (Cytoxan) | 20.0 mg/ml (20,000 ppm) | > 240 |
| *Carmustine (BCNU) | 3.3 mg/ml (3,300 ppm) | 45.0 |
| *Thiotepa | 10.0 mg/ml (10,000 ppm) | 30.0 |
| Paclitaxel (Taxol) | 6.0 mg/ml (6,000 ppm) | > 240 |
| Doxorubicin Hydrochloride | 2.0 mg/ml (2,000 ppm) | > 240 |
| Dacarbazine (DTIC) | 10.0 mg/ml (10,000 ppm) | > 240 |
| Cisplatin | 1.0 mg/ml (1,000 ppm) | > 240 |
| Carboplatin | 10.0 mg/ml (10,000 ppm) | > 240 |
| Docetaxel | 10.0 mg/ml (10,000 ppm) | > 240 |
| Ifosfamide | 50.0 mg/ml (50,000 ppm) | > 240 |
| Irinotecan | 20.0 mg/ml (20,000 ppm) | > 240 |
| Mechlorethamine HCL | 1.0 mg/ml (1,000 ppm) | > 240 |
| Methotrexate | 25.0 mg/ml (25,000 ppm) | > 240 |
| Mitomycin C | 0.5 mg/ml (500 ppm) | > 240 |
| Mitoxantrone | 2.0 mg/ml (2,000 ppm) | > 240 |
| Vincristine Sulfate | 1.0 mg/ml (1,000 ppm) | > 240 |
* Please note that the following drugs have low permeation times: Carmustine (BCNU): 45 minutes and Thiotepa: 30 minutes
Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K163146
- 1. Date of Preparation: 02/08/2017
- 2. Sponsor Identification
# HEBEI HONGSEN PLASTICS TECHNOLOGY CO., LTD EASTERN INDUSTRIAL ZONE, NANGONG CITY, HEBEI PROVINCE, CHINA
Contact Person: Mr. ShaoZhang Nan Tel: +86-0319-7295820 Fax: +86-0319-7295801 Email: nanshaozhang@163.com
- 3. Designated Submission Correspondent
Mr. Ray Wang
## Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd., LiYuan Town, TongZhou District, Beijing, 101121, China
Tel: +86-21-59120817 Fax: +86-21-59120817 Email: Ray.Wang@believe-med.com
{5}------------------------------------------------
#### 4. Proposed Device Identification
Trade Name: POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs Device Name: NITRILE Patient Examination Gloves (Powder Free) Common Name: Patient Examination Gloves
### Regulatory Information
Classification: I Product Code: LZA, LZC Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital
#### Indication for Use:
The POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Chemotherapy Drug Permeation
The following chemicals have been tested with proposed device.
| Chemotherapy Drug | Concentration | Breakthrough Detection Time in Minutes |
|---------------------------|-------------------------|----------------------------------------|
| Fluorouracil | 50.0 mg/ml (50,000 ppm) | > 240 |
| Etoposide (Toposar) | 20.0 mg/ml (20,000 ppm) | > 240 |
| Cyclophosphamid (Cytoxan) | 20.0 mg/ml (20,000 ppm) | > 240 |
| *Carmustine (BCNU) | 3.3 mg/ml (3,300 ppm) | 45.0 |
| *Thiotepa | 10.0 mg/ml (10,000 ppm) | 30.0 |
| Paclitaxel (Taxol) | 6.0 mg/ml (6,000 ppm) | > 240 |
| Doxorubicin Hydrochloride | 2.0 mg/ml (2,000 ppm) | > 240 |
| Dacarbazine (DTIC) | 10.0 mg/ml (10,000 ppm) | > 240 |
| Cisplatin | 1.0 mg/ml (1,000 ppm) | > 240 |
| Carboplatin | 10.0 mg/ml (10,000 ppm) | > 240 |
| Docetaxel | 10.0 mg/ml (10,000 ppm) | > 240 |
| Ifosfamide | 50.0 mg/ml (50,000 ppm) | > 240 |
| Irinotecan | 20.0 mg/ml (20,000 ppm) | > 240 |
| Mechlorethamine HCL | 1.0 mg/ml (1,000 ppm) | > 240 |
| Methotrexate | 25.0 mg/ml (25,000 ppm) | > 240 |
| Mitomycin C | 0.5 mg/ml (500 ppm) | > 240 |
| Mitoxantrone | 2.0 mg/ml (2,000 ppm) | > 240 |
| Vincristine Sulfate | 1.0 mg/ml (1,000 ppm) | > 240 |
*Please note that the following drugs have low permeation times:
Carmustine (BCNU): 45 minutes and Thiotepa: 30 minutes
{6}------------------------------------------------
- 5. Predicate Device Identification
510(k) Number: K141982
Product Name: Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-Sterile, Tested for use with Chemotherapy Drugs
Manufacturer: WRP Asia Pacific Sdn Bhd.
- Device Description 6.
The proposed device, POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device is a Powder Free Nitrile Patient Examination Glove that is available in multiple sizes
The proposed device is provided non-sterile. The proposed device is made of Nitrile. The proposed device acts as a barrier.
The proposed device was tested according to the following standards: ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, and ASTM D6978-05. These standards are identified in the following section "Non-clinical test conclusion.
- 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all specifications and the proposed device is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.
ASTM D5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves.
ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs
ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
{7}------------------------------------------------
| Item | Proposed Device<br>POWDER FREE Blue Nitrile<br>GLOVES, Tested for Use with<br>Chemotherapy Drugs<br>(K163146) | Predicate Device<br>Dermagrip Powder Free Blue Nitrile<br>Patient Examination Gloves Tested for<br>Use with Chemotherapy Drugs<br>(K141982) | Remark | | | | Size | | | | | |
|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------|-------------------|--|------|-----|-----|-----|-----|-----------|
| Product Code | LZA, LZC | LZA, LZC | Same | Proposed Device | Designation | | XS | S | M | L | XL | Tolerance |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same | POWDER FREE Blue Nitrile GLOVES, Tested for Use<br>with Chemotherapy Drugs<br>(K163146) | Length, mm | | 230 | 230 | 230 | 230 | 230 | min |
| Class | I | I | Same | | Width, mm | | 70 | 80 | 95 | 110 | 120 | ±10 |
| Indication for use | The POWDER FREE Blue Nitrile<br>GLOVES, Tested for Use with<br>Chemotherapy Drugs is a disposable<br>device intended for medical purposes<br>that is worn on the examiner's hands<br>to prevent contamination between<br>patient and examiner. | A patient examination glove is a<br>disposable device intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between patient and<br>examiner. | Same | | Thickness, mm: | | | | | | | |
| Powdered or<br>Powered free | Powdered free | Powdered free | Same | | Finger | | 0.10 | | | | | ±0.03 |
| Design Feature | ambidextrous | ambidextrous | Same | | Palm | | 0.08 | | | | | ±0.03 |
| Labeling<br>Information | Single-use indication, powder free,<br>device name, glove size and quantity,<br>Nitrile Examination Gloves,<br>Non-Sterile | Single-use indication, powder free,<br>device name, glove size and quantity,<br>Nitrile Examination Gloves,<br>Non-Sterile | Same | | Cuff | | 0.06 | | | | | ±0.03 |
| Chemotherapy Drug<br>Permeation Claim | Fluorouracil, Etoposide (Toposar),<br>Cyclophosphamid (Cytoxan),<br>Carmustine (BCNU), Thiotepa,<br>Paclitaxel (Taxol), Doxorubicin<br>Hydrochloride, Dacarbazine (DTIC),<br>Cisplatin, Carboplatin, Docetaxel,<br>Ifosfamide, Irinotecan,<br>Mechlorethamine HCL,<br>Methotrexate, Mitomycin C,<br>Mitoxantrone, Vincristine Sulfate | Fluorouracil, Etoposide (Toposar),<br>Cyclophosphamid (Cytoxan),<br>Carmustine (BCNU), Thiotepa,<br>Paclitaxel (Taxol), Doxorubicin<br>Hydrochloride, Dacarbazine (DTIC),<br>Cisplatin, Ifosfamide, Mitoxantrone,<br>Vincristine Sulfate | Analysis 1 | Predicate Device | Size: Min. 240 mm | | | | | | | |
| Dermagrip Powder Free Blue Nitrile Patient<br>Examination Gloves Tested for Use with Chemotherapy<br>Drugs<br>(K141982) | Thickness: Finger (0.07-0.10); Palm(0.07-0.09); Cuff (0.06-0.08) | | | | | | | | | | | |
| Remark | Analysis 2 | | | | | | | | | | | |
#### 8. Substantially Equivalent (SE) Comparison Conclusion
Table 1 General Comparison
Analysis 1:
The proposed and predicate devices both have a tested for use with chemotherapy drugs claim. However, different chemotherapy drugs have been tested for the proposed device and the results meet the specifications of ASTM D6978
{8}------------------------------------------------
### Table 2 Device Dimensions Comparison
Analysis 2:
The proposed device has different size specification as compared to the predicate device, but the proposed device meets the specifications of ASTM D6319.
{9}------------------------------------------------
| Item | | | Proposed Device<br>POWDER FREE Blue Nitrile GLOVES,<br>Tested for Use with Chemotherapy Drugs<br>(K163146) | Predicate Device<br>Dermagrip Powder Free<br>Blue Nitrile Patient<br>Examination Gloves Tested<br>for Use with Chemotherapy<br>Drugs<br>(K141982) | Remark |
|------------------------|-----------------|------------------------|------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Colorant | | | Blue | Blue | Same |
| Physical<br>properties | Before<br>Aging | Tensile<br>Strength | 15 Mpa, min | Meet the Requirements of<br>ASTM D 6319 | Analysis 3 |
| | | Ultimate<br>Elongation | 500% min | Meet the Requirements of<br>ASTM D 6319 | |
| | After<br>Aging | Tensile<br>Strength | 14 MPa, min | Meet the Requirements of<br>ASTM D 6319 | |
| | | Ultimate<br>Elongation | 400% min | Meet the Requirements of<br>ASTM D 6319 | |
| | | | Comply with ASTM D6319 | Comply with ASTM D6319<br>14 MPa. Min/500% min.<br>before aging;<br>14 MPa. Min./400% min.<br>After aging | Same |
| Freedom from Holes | | | Be free from holes when tested in<br>accordance with ASTM D5151<br>AQL 1.5 | Be free from holes when<br>tested in accordance with<br>ASTM D5151<br>under AQL 2.5/Inspection<br>Level G-I | Same |
| Powder Content | | | Max. 0.32 mg per glove | Meet the requirements of<br>ASTM D6319<br>Less than 2mg per glove | Same |
Table 3 Performance Comparison
Analysis 3:
The proposed device has a different Ultimate Elongation after aging as compared to the predicate device, but the proposed device meets the specifications of ASTM D6319.
{10}------------------------------------------------
| Item | Proposed Device<br>POWDER FREE Blue<br>Nitrile GLOVES, Tested for<br>Use with Chemotherapy<br>Drugs<br>(K163146) | Predicate Device<br>Dermagrip Powder Free Blue Nitrile<br>Patient Examination Gloves Tested for Use<br>with Chemotherapy Drugs<br>(K141982) | Remark | |
|--------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|------|
| Material | Nitrile | Nitrile | Same | |
| Biocompatibility | Irritation | Under the conditions of the<br>study, not an irritant | Comply with ISO 10993-10 | Same |
| | Sensitization | Under the conditions of the<br>study, not a sensitizer | | |
| Label and Labeling | Meet FDA's<br>Recommendations | Meet FDA's Recommendations | Same | |
### Table 4 Safety Comparison
#### 9. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device, POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs, is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Dermagrip Powder Free Blue Nitrile Patient Examination Gloves Tested for Use with Chemotherapy Drugs.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.