SpeedStitch Suture Passer; SpeedStitch Suture Cartridges, White Suture, Blue Cobraid, Black Cobraid, Mixed Pack; SpeedStitch Needle Loader and Unloader
K163142 · ArthroCare Corporation · GAT · Jan 6, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163142
Device Name
SpeedStitch Suture Passer; SpeedStitch Suture Cartridges, White Suture, Blue Cobraid, Black Cobraid, Mixed Pack; SpeedStitch Needle Loader and Unloader
Applicant
ArthroCare Corporation
Product Code
GAT · General, Plastic Surgery
Decision Date
Jan 6, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SpeedStitch System is indicated for use in the placement of suture(s) through soft tissue in arthroscopic and limited access procedures.
Device Story
The SpeedStitch System is a manual, handheld orthopedic/arthroscopic instrument designed for mechanical delivery of sutures through soft tissue. The system comprises a handle, a needle cassette, and a suture cartridge containing MagnumWire suture. In operation, a clinician uses the device to grasp soft tissue, insert a needle through the tissue, and capture the suture from the cartridge, pulling it through the tissue to complete a simple stitch. The device is used in clinical settings for arthroscopic or limited access procedures. It provides a mechanical means for suture placement, aiding in tissue repair. The system is provided non-sterile (handle) and sterile (needle cassette/suture cartridge) and is intended for use by trained healthcare professionals.
Clinical Evidence
No clinical or animal data included. Bench testing only. In vitro testing verified tissue grasping, needle insertion, and suture capture/delivery performance. Design verification and simulated use testing confirmed that the device meets design inputs and performance specifications.
Technological Characteristics
Manual handheld suture passer. Materials: 300/301/304/316 stainless steel, 17-4/17-7/455 stainless steel, polycarbonate, glass-filled nylon, UV adhesive. Components: handle, needle cassette, suture cartridge. Sterilization: ethylene oxide for needle cassette and suture cartridge. Mechanical operation; no energy source or software.
Indications for Use
Indicated for placement of suture(s) through soft tissue during arthroscopic and limited access procedures.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with three intertwined snakes and a pair of wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2017
ArthroCare Corporation Ms. Laura Kasperowicz Principle RA Specialist 15285 Alton Parkway Suite 200 Irvine, California 92618
Re: K163142
Trade/Device Name: Speedstitch Suture Passer; Speedstitch Suture Cartridges, White Suture, Blue Cobraid, Black Cobraid, Mixed Pack; Speedstitch Needle Loader And Unloader Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT, NBH, HXK Dated: November 7, 2016 Received: November 9, 2016
Dear Ms. Kasperowicz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
## David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
Device Name
SpeedStitch Suture Passer System
Indications for Use (Describe)
The SpeedStitch System is indicated for use in the placement of suture(s) through soft tissue in arthroscopic and limited access procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
November 7, 2016
## 510(k) Summary
ArthroCare® Corporation
SpeedStitch® System
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### General Information
| Submitter Name: | ArthroCare Corporation, a Smith & Nephew Company |
|-----------------|----------------------------------------------------------------------------------------------------------|
| Address: | 15285 Alton Parkway, Suite 200<br>Irvine, CA. 92618 |
| Contact Person: | Laura Kasperowicz<br>Principle Regulatory Affairs Specialist<br>Phone: 949-585-2406<br>Fax: 949-585-2401 |
| Date Prepared: | November 7, 2016 |
#### Device System Names/Components
The SpeedStitch System is a Class II device consisting of the following components:
| Proprietary: | SpeedStitch Suture Passer |
|--------------|--------------------------------------------------------|
| | Common: Suture Passer |
| | Classification: Class I |
| | Product Code: NBH |
| | CFR Section: 21 CFR 888.1100 |
| Proprietary: | SpeedStitch Needle Cassette |
| | Common: Holder, Needle, Orthopedic |
| | Classification: Class I |
| | Product Code: HXK |
| | CFR Section: 21 CFR 888.4540 |
| Proprietary: | Suture Cartridge containing MagnumWire® Suture |
| | Common: Suture, Nonabsorbable, Synthetic, Polyethylene |
| | Classification: Class II |
| | Product Code: GAT |
| | CFR Section: 21 CFR 878.5000 |
#### Predicate Device
The SpeedStitch System is substantially equivalent to the following:
- SpeedStitch System (K042031, October 19, 2004)
- I M-Connector Ultra System (K121306, July 12, 2012)
{4}------------------------------------------------
### Description
The SpeedStitch System is an orthopedic/arthroscopic instrument for mechanically delivering sutures through soft tissue based on the basic design concepts of handheld, manual suture passers.
The handle is provided non-sterile. The needle cassette and suture cartridge are sterilized by ethylene oxide. The finished products are packaged in a Tyvek® pouch inserted into a chipboard carton.
## Intended Use/Indications For Use
The SpeedStitch System is indicated for use in the placement of suture(s) through soft tissue in arthroscopic and limited access procedures.
### Non-Clinical Data
Bench testing was performed on both the proposed and predicate devices. This in vitro testing involved grasping of a tissue model with the suture passer, insertion of needles through the tissue model and capturing suture from suture cartridges, which was successfully pulled through the tissue model using the suture passer. Design Verification testing was performed in order to verify the design outputs meet the design inputs. Simulated use testing was performed on the device and accessory devices (suture cartridges) which will be promoted for use with the SpeedStitch system in order to affirm safety and effectiveness.
The test results demonstrate that the proposed SpeedStitch meets its design, performance, and safety specifications. Based on the test results, the proposed device performs as intended and mechanical properties are substantially equivalent to the predicate device when used in accordance with labeling.
#### Clinical Data
No clinical or animal data are included in this submission.
#### Summary
All testing demonstrates that the proposed SpeedStitch performs as intended and has acceptable performance when used in accordance with the labeling.
As the intended use, operating principle, materials and technological characteristics are comparable to the predicate device, the proposed SpeedStitch System is substantially equivalent. The minor differences between the proposed SpeedStitch System and predicate device do not raise any new questions of safety or effectiveness.
{5}------------------------------------------------
# ArthroCare® Corporation Premarket Notification
SPEEDSTITCH® SYSTEM
November 7, 2016
| Comparison of Device Characteristics | | | |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Parameter | M-Connector Ultra<br>Predicate Device | SpeedStitch<br>Predicate Device | SpeedStitch<br>Proposed Device |
| 510(k) | K121306 | K042031 | TBD |
| Panel | General & Plastic Surgery,<br>Orthopedic | Same | Same |
| Product Code(s) | GAT & NBH | GAT | GAT/NBH/HXK |
| Classification Section | 878.5000<br>888.1100 | 878.5000 | 878.5000<br>888.1100<br>888.4540 |
| Classification Name | > Suture, Nonabsorbable,<br>Synthetic, Polyethylene<br>> Arthroscope & Accessories<br>(Suture Passer) | > Suture, Nonabsorbable,<br>Synthetic, Polyethylene | > Suture, Nonabsorbable,<br>Synthetic, Polyethylene<br>> Arthroscope & Accessories<br>(Suture Passer)<br>> Holder, Needle, Orthopedic |
| Device Classification | Class II (Sutures) &<br>Class I (Handle, Connector) | Class II | Class II (Sutures),<br>Class I (Handle),<br>Class I (Needle Holder) |
| Intended Use | For use in the placement of<br>suture(s) through soft tissue in<br>arthroscopic and limited access<br>procedures. | For use in the placement of<br>suture(s) through soft tissue in<br>endoscopic and limited access<br>procedures. | For use in the placement of<br>suture(s) through soft tissue in<br>arthroscopic and limited access<br>procedures. |
| Design Technology | Suture Passer with 2 needles | Suture Passer with 1 needle | Suture Passer with 1 needle |
| Suture Delivery<br>Method | Suture preloaded in a cartridge<br>for arthroscopic or limited<br>access delivery | Same | Same |
| Suture Specifications | USP #2 Nonabsorbable<br>Synthetic Suture,<br>Polyethylene | Same | Same |
| Suture Placement | Interrupted Mattress Stitch | Simple Stitch | Simple Stitch |
| Main Components | ■ Handle<br>■ Connector<br>■ Suture Cartridge with<br>suture | ■ Handle with Connector<br>■ Suture Cartridge with<br>suture<br>■ Needle Loader & Needle<br>Extractor | ■ Handle with Connector<br>■ Suture Cartridge with<br>suture<br>■ Needle Loader & Needle<br>Extractor |
| Main Materials | 300/301/304/316 Stainless Steel<br>17-4/17-7/455 Stainless Steel<br>Polycarbonate<br>Glass Filled Nylon<br>UV Adhesive | Same | Same |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.