← Product Code [DQA](/productcode/DQA) · K163135

# Finger Pulse Oximeter (K163135)

_Shenzhen Fitfaith Technology Co.,Ltd · DQA · Sep 19, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K163135

## Device Facts

- **Applicant:** Shenzhen Fitfaith Technology Co.,Ltd
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Sep 19, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Fingertip Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).

## Device Story

Fingertip pulse oximeter; portable, battery-powered device. Inputs: red (660nm) and infrared (905nm) light beams transmitted through fingertip. Principle: Photoelectric Oxyhemoglobin Inspection Technology; Capacity Pulse Scanning and Recording; Lambert-Beer Law applied to differential light absorption of oxyhemoglobin and reductive hemoglobin. Output: SpO2 and pulse rate displayed on device screen. Used in homes and clinical environments (surgery, anesthesia, ICU) by patients or clinicians for spot-checking. Provides real-time physiological data to assist in monitoring patient oxygenation status; aids in identifying potential hypoxemia or pulse rate abnormalities.

## Clinical Evidence

Controlled desaturation study conducted on 12 healthy, non-smoking adults (ages 21-32). 294 data sets collected (288 valid). Subject device SpO2 compared against CO-oximeter (control) across 70-100% range. Results met ISO 80601-2-61 criteria for accuracy and precision. Bland-Altman analysis showed majority of data within 95% limits of agreement. No adverse effects reported.

## Technological Characteristics

Materials: ABS enclosure, silica gel clip. Sensing: Photoelectric sensor (660nm red, 905nm IR). Power: 2x AAA 1.5V alkaline batteries. Form factor: Portable fingertip clip. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5/10, ISO 80601-2-61.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- JPD-500A Fingertip Pulse Oximeter ([K140582](/device/K140582.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 19, 2017

Shenzhen Fitfaith Technology Co.,ltd % Migo. Yang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1122#, Internation Mayor Communication Center, Baishizhong Rd 55#, Nanshan District, Shenzhen, 518000 CN

Re: K163135

Trade/Device Name: Fingertip Pulse Oximeter (Model: A300, A310, M110, M120, M130, M150, M160, M170, M230) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: July 15, 2017 Received: August 7, 2017

Dear Migo. Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Tara A. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K163135

#### Device Name

Fingertip Pulse Oximeter (Model: A300, A310, M100, M110,M120,M130,M150,M170,M230)

Indications for Use (Describe)

The Fingertip Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery, anesthesia, intensive care, etc).

| Type of Use (Select one or both, as applicable)                                                                               |                                                                                                                              |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10pt;">☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </div> | <div> <span> <span style="font-size: 10pt;">☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This summary of 510(k) information is submitted as required by requirements of SMDA and 21 CFR §807.92.

## 1 Administrative Information

| Submission Date                      | Aug 18,2016                                                                                                                                                                                                                                                                                                             |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information          | Submitter's Name: Shenzhen Fitfaith Technology Co.,Ltd.<br>Address: Area B, Floor 9, Building D1, Tangwei Industrial Park<br>Donglong Road, Guangming New District, ShenZhen,<br>Guangdong, China.<br>Contact person: Yuan.Junfeng<br>TEL: +86-0755-29544253-803<br>FAX: +86-0755-29544253-801<br>E-Mail: 160666@qq.com |
| Submission<br>Correspondent          | Contact person: Miss Migo. Yang<br>E-Mail: migo@cefda.com<br>Shenzhen Joyantech Consulting Co., Ltd.<br>1122#, International Mayor Communication Center,<br>Baishizhong Road 55#, Nanshan District, Shenzhen,<br>Guangdong, China.                                                                                      |
| Image: Logo                          | Contact person: Mr. Field.Fu<br>E-Mail: cefda13485@163.com<br>Shenzhen Joyantech Consulting Co., Ltd.<br>1122#, International Mayor Communication Center,<br>Baishizhong Road 55#, Nanshan District, Shenzhen,<br>Guangdong, China                                                                                      |
| Establishment<br>registration number | NA                                                                                                                                                                                                                                                                                                                      |

## 2 Device Information

| Type of 510(k)<br>submission: | Traditional                                                                               |
|-------------------------------|-------------------------------------------------------------------------------------------|
| Trade Name:                   | Fingertip Pulse Oximeter (Model: A300, A310, M100,<br>M110,M120,M130,M150,M160,M170,M230) |
| Classification<br>name:       | Oximeter                                                                                  |
| Review Panel:                 | Anesthesiology                                                                            |
| Product Code:                 | DQA                                                                                       |
| Device Class:                 | II                                                                                        |
| Regulation                    | 870.2700                                                                                  |
| Number:                       | K163135                                                                                   |

{4}------------------------------------------------

#### 3 Predicate Device Information

Shenzhen Jumper Medical Equipment Co., Ltd. Sponsor: JPD-500A Fingertip Pulse Oximeter Device: K140582 510(K) Number:

#### 4 Device Description

The Fingertip Pulse Oximeteris intended for spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of adult and pediatric patients in the home and hospital.

The fingertip pulse oximeter features a small size, low power consumption, a convenient operation, and portability.It is only necessary for a patient to put one of his/her fingers into the fingertip clips for measurement.

Principle of the fingertip pulse oximeter as follows:

A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in red and infrared zones.

Operation principle of the instrument:

Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. Relevant data is shown on the Oximeter's display through electronic circuits and a microprocessor.

#### 5 Intended Use

The subject device is intended for measuring the functional oxygen saturation and pulse rate (PR) through a patient's finger. It is applicable for spot-checking SpO2 and pulse rate (PR) of adult and pediatric patients in homes and clinics.

#### 6 Indications for Use

The Fingertip Pulse Oximeter is non-invasive device intended for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The portable fingertip device is indicated for adult and pediatric patients in home and hospital environments (including clinical use in internist/ surgery. anesthesia, intensive care, etc).

{5}------------------------------------------------

#### 7 Contraindications

- High-frequency electrosurgical
- Placement of a sensor on an extremity with a blood pressure cuff arterial catheter, or intravascular line
- The patient has hypotension severe vasoconstriction severe anemia or hypothermia.
- The patient is in cardiac arrest or is in shock.
- Fingernail polish or false fingernails may cause inaccurate SpO2 readings.

{6}------------------------------------------------

## 8 SE Comparison

| Characteristics                       | Subject device                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate device<br>(k140582)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Remark                                   |                  |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------|
| Device name                           | Fingertip Pulse Oximeter<br>A300,A310,M100,M110,M<br>120,M130,M150,M160,M1<br>70,M230                                                                                                                                                                                                                                                                                                                          | Fingertip Pulse Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | /                                        |                  |
| Model                                 | Note:<br>A300 and M110 are totally<br>identical model except for<br>model name based on<br>customer's requirement;<br>A310 and M100 are totally<br>identical model except for<br>model name based on<br>customer's requirement;                                                                                                                                                                                | JPD-500A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | /                                        |                  |
| Manufacturer                          | Shenzhen Fitfaith<br>Technology Co.,Ltd.                                                                                                                                                                                                                                                                                                                                                                       | Shenzhen Jumper<br>Medical Equipment<br>Co.,Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | /                                        |                  |
| Intended patient<br>population        | Adult & Pediatric                                                                                                                                                                                                                                                                                                                                                                                              | Adult & Pediatric                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | same                                     |                  |
| Intended application<br>site          | Fingertip                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | same                                     |                  |
| Intended<br>Environments              | home & hospital                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | same                                     |                  |
| Prescription & OTC                    | Prescription                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | same                                     |                  |
| Intended use                          | The subject device is intended for measuring the<br>functional oxygen saturation and pulse rate through<br>patient's finger. It is applicable for spot-checking SpO2<br>and pulse rate of adult and pediatric patients in homes<br>and clinics.                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | same                                     |                  |
| Indications for Use                   | The Fingertip Pulse<br>Oximeter is non-invasive<br>device intended for spot-<br>checking of functional<br>oxygen saturation of<br>arterial hemoglobin<br>(SpO2) and pulse rate<br>(PR). The portable<br>fingertip device is<br>indicated for adult and<br>pediatric patients in home<br>and hospital environments<br>(including clinical use in<br>internist/ surgery,<br>anesthesia, intensive<br>care, etc). | The JPD-500A Fingertip<br>Pulse Oximeter is non-<br>invasive device intended<br>for spot-checking of<br>functional oxygen<br>saturation of arterial<br>hemoglobin (Spo2) and<br>pulse rate. The portable<br>fingertip device is<br>indicated for adult and<br>pediatric patients in home<br>and hospital<br>environments (including<br>clinical use in internist/<br>surgery, anesthesia,<br>intensive care, etc).                                                                                                                                                                                                                                                                                                      | Similar(Note01)                          |                  |
| Characteristics                       |                                                                                                                                                                                                                                                                                                                                                                                                                | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Predicate device<br>(k140582)            | Remark           |
| Working Principle                     |                                                                                                                                                                                                                                                                                                                                                                                                                | A mathematical formula is established making use of<br>Lambert Beer Law according to Spectrum Absorption<br>Characteristics of Reductive hemoglobin (RHb) and<br>Oxyhemoglobin (HbO2) in red and near-infrared<br>zones. Operation principle of the instrument:<br>Photoelectric Oxyhemoglobin Inspection Technology is<br>adopted in accordance with Capacity Pulse<br>Scanning and Recording Technology, so that two<br>beams of different wavelength of lights can be focused<br>onto a human nail tip through a clamping finger-type<br>sensor. A measured signal obtained by a<br>photosensitive element, will be shown on the<br>Oximeter's display through process in electronic<br>circuits and microprocessor. |                                          | same             |
| Light<br>Specification<br>on          | wavelength                                                                                                                                                                                                                                                                                                                                                                                                     | 660nm±3nm for red light,<br>905nm±5nm for infrared light (IR)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                          | same             |
|                                       | maximum<br>optical<br>power                                                                                                                                                                                                                                                                                                                                                                                    | 1.5 mW for red light (660nm)<br>1.2 mW for IR (905nm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                          | same             |
| Contact Material                      |                                                                                                                                                                                                                                                                                                                                                                                                                | ABS for enclosure, silica gel for clip                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                          | same             |
| Internal Power supply                 |                                                                                                                                                                                                                                                                                                                                                                                                                | 2*AAA 1.5v alkaline battery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                          | same             |
| Working Current                       |                                                                                                                                                                                                                                                                                                                                                                                                                | Less than 40mA(Normal)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 35mA                                     | Similar(Note02)  |
| Resolution                            | SpO2                                                                                                                                                                                                                                                                                                                                                                                                           | 1%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                          | same             |
|                                       | Pulse<br>rate                                                                                                                                                                                                                                                                                                                                                                                                  | 1 bpm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                          | same             |
| Measurement<br>Range                  | SpO2                                                                                                                                                                                                                                                                                                                                                                                                           | 0%~100%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                          | same             |
|                                       | Pulse<br>rate                                                                                                                                                                                                                                                                                                                                                                                                  | 25~250 bpm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                          | same             |
| Measurement<br>Accuracy               | SpO2                                                                                                                                                                                                                                                                                                                                                                                                           | ±2%: (70%~100%)<br>Unspecified : (0%~69%)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | ±2%: (70%~100%)<br>no definition(0%~69%) | same             |
|                                       | Pulse<br>rate                                                                                                                                                                                                                                                                                                                                                                                                  | ±2 bpm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | ±2 bpm                                   | same             |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                          |                  |
| Operating<br>Environment              | Temperature                                                                                                                                                                                                                                                                                                                                                                                                    | 5°C~40°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                          | Similar (Note03) |
|                                       | Humidity                                                                                                                                                                                                                                                                                                                                                                                                       | 15%~85% non-<br>condensing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 15%~80% non-<br>condensing               |                  |
| Storage &<br>Transport<br>Environment | Temperature                                                                                                                                                                                                                                                                                                                                                                                                    | -20°C~+55°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | -10°C~+50°C                              | Similar(Note04)  |
|                                       | Humidity                                                                                                                                                                                                                                                                                                                                                                                                       | 10%~95%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 10%~93%                                  |                  |

### Table 1. Substantial Equivalence Comparison

{7}------------------------------------------------

Note01:

The trade name is different, but SE determination does not depend on trade name.

Note02:

The working currents of subject device and predicate device are similar. The subject device met the requirements of IEC 60601-1.

{8}------------------------------------------------

Note03:

The humidity under operating environment of subject device and predicate device are similar. The subject device met the requirements of ISO 80601-2-61. Note04:

The Storage & Transport environment of subject device and predicate device are similar. The subject device met the requirements of ISO 80601-2-61.

The subject device is the same as the predicate device in intended use, intended patient population, intended application site, working principle, contact material, internal power supply, display range, resolution. Only their appearance, indication for use (Note01), working current (Note02), operation environment (Note03), storage transport environment (Note04), are a little bit different. However, the minor differences do not raise different questions of safety and effectiveness.

#### 9 Brief discussions of tests

#### 9.1 Brief discussions of the nonclinical tests

The subject device conforms to the following standards:

- ゃ IEC 60601-1:2005+A1:2012: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General ゃ Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests;
- や IEC 60601-1-11: Requirements for Medical Electrical (ME) Equipment and ME Systems used in the Home Healthcare Environment;
- ゃ ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for In Vitro cytotoxicity;
- � ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization;
- ゃ ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment;
- ゃ Cleaning and low level disinfection validation.

#### 9.2 Brief discussions of clinical tests

Clinical testing according to ISO 14155:2011 and ISO 80601-2-61: 2011 has also been performed on the device.

Clinical hypoxia accuracy testing (controlled desaturation study) was conducted during induced hypoxia studies on healthy, non-smoking, light-to-dark-skinned subjects in an independent research laboratory. The measured arterial

{9}------------------------------------------------

hemoglobin saturation value (SpO2) of the subject device was compared with arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter (control device). The accuracy of the subject device is in comparison with the control device over the SpO2 range of 70~100%. Data was calculated and analyzed using the mean bias (B), root-mean-square (Ams), PRECISION (standard deviation of the residuals (Sres) for all subjects, per ISO 80601-2-61, the result showed that the error is far less than the scope specified in the ISO 80601-2-61;

Besides, the Agreement between methods of measurement with multiple observations for both all subjects pooled and individual test subjects were analyzed with the Bland and Altman statistics, the analysis demonstrated that the vast majority of data is within ±95% limit of agreement, the data points beyond or below this scope were regarded as outliers. The outliers only occurred occasionally and after being analyzed, it was determined that the outliers do not raise performance concerns regarding the accuracy and precision of the device.

During the clinical study, 12 subjects were enrolled, who are healthy, nonsmoking, competent adults, between twenty-one and thirty-two (21-32) years of age, and they were provided EC (Ethics Committee)-approved informed consent as documented on an informed consent form. No case was lost in this trial. The trial completed 12 cases, on which 294 data sets were collected, of which 288 were valid. 6 data sets which less than 70% were excluded. The result met the criteria specified in the ISO 80601-2-61; In addition, there were no reported adverse effects during these investigations.

#### 10 Other information (such as required by FDA guidance)

No other information.

#### 11 Conclusions

Based on the clinical and non-clinical testing performed, the results demonstrate that the subject device fingertip pulse oximeter is substantially equivalent to the predicate device JPD-500A Fingertip Pulse Oximeter manufactured by Shenzhen Jumper Medical Equipment Co., Ltd.

---

**Source:** [https://fda.innolitics.com/device/K163135](https://fda.innolitics.com/device/K163135)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
