K163105 · Covidien, LLC · NEY · Dec 6, 2016 · General, Plastic Surgery
Device Facts
Record ID
K163105
Device Name
Emprint Ablation System
Applicant
Covidien, LLC
Product Code
NEY · General, Plastic Surgery
Decision Date
Dec 6, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors. The Emprint Ablation System is not intended for use in cardiac procedures.
Device Story
Emprint Ablation System delivers microwave energy to soft tissue for coagulation/ablation. System comprises 2450 MHz generator and percutaneous antenna. Modified antenna features stiffer cooling jacket material and reinforced proximal joint seal. Used in healthcare facilities by clinicians for percutaneous, laparoscopic, or intraoperative procedures. Generator controls signal frequency, power, and exposure time; antenna design and continuous cooling produce near-spherical ablation zones. Stiffer antenna enhances usability during placement. Output is thermal energy for tissue destruction. Benefits include controlled ablation of non-resectable liver tumors.
Clinical Evidence
Bench testing only. Included mechanical stress testing, product stability testing, biocompatibility testing per ISO 10993-1, and a simulated-use validation study with end users to confirm safety, effectiveness, and usability of the modified antenna.
Technological Characteristics
Microwave ablation system operating at 2450 MHz. Antenna cooling jacket material changed to fiberglass for increased stiffness. Provided sterile (EO). System uses controlled microwave energy delivery for thermal ablation. No software algorithm details provided.
Indications for Use
Indicated for percutaneous, laparoscopic, and intraoperative soft tissue coagulation/ablation, including non-resectable liver tumors. Not for cardiac procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Emprint™ Ablation System with Thermosphere™ Technology (K133821)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and clean, with a focus on the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 6, 2016
Covidien LLC Ms. Patti Arndt Principal Regulatory Affairs Specialist 5920 Longbow Drive Boulder, Colorado 80301
Re: K163105
Trade/Device Name: Emprint Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NEY Dated: November 4, 2016 Received: November 7, 2016
Dear Ms. Arndt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K163105
Device Name
EmprintTM Ablation System
Indications for Use (Describe)
The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors.
The Emprint Ablation System is not intended for use in cardiac procedures.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> <span style="text-decoration: overline;">☑</span> Transportation Use (64-35B-001 Subpart B)</span> <span><span style="padding-left: 5px;">□</span> On-Site Service Use (64-35B-001 Subpart C)</span> |
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
#### l. SUBMITTER
II.
Covidien IIc 5920 Longbow Dr. Boulder, CO 80301
| Contact Person: | Patti Arndt<br>Principal Regulatory Affairs Specialist<br>Phone: 303-581-6668<br>FAX: 303-530-6313 |
|------------------------------|----------------------------------------------------------------------------------------------------|
| Date of Summary Preparation: | November 2, 2016 |
| | DEVICE |
| Device Trade Name: | Emprint™ Ablation System |
| Device Common Name: | Microwave Ablation System |
| Classification Name: | System, Ablation, Microwave and Accessories |
| Regulatory Class: | II |
NEY
#### III. PREDICATE DEVICE
Product Code:
Emprint™ Ablation System with Thermosphere™ Technology, K133821
Modification to System Accessory: Emprint™ Percutaneous Antenna with Thermosphere™ Technology: Catalog Nos.CA15L1 (Short), CA20L1 (Standard), CA30L1 (Long)
The device modification is not the subject of a design-related recall.
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
Covidien has modified the Emprint Ablation System's Percutaneous Antenna (Catalog Numbers: CA15L1, CA20L1 and CA30L1). The modified ablation antenna is identical to the device described in K133821, with two exceptions. The antenna's tissue contacting cooling jacket is manufactured from a new, biocompatible material that significantly increases the antenna's stiffness. Additionally, an adhesive has been added at the antenna shaft's proximal joint to strengthen the seal. By making the antenna stiffer, the cooling jacket material change is expected to enhance the antenna's usability; it does not affect the Emprint Ablation System's essential function, performance or underlying principles of operation.
{5}------------------------------------------------
Like its predicate, the modified antenna (Catalog Numbers CA15L2, CA20L2 CA30L2) is provided sterile (EO) and is intended for use with the Emprint Ablation System in health care facilities.
#### V. INDICATIONS FOR USE
The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic and intraoperative coaqulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors.
The Emprint Ablation System is not intended for use in cardiac procedures.
Indications for the use of the Emprint System and modified Percutaneous Antenna are identical to the predicate device.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The Emprint Ablation System's current percutaneous antenna (predicate device) and the modified ablation antenna share the same underlying technology. Both devices are used to deliver microwave energy to soft tissue in a highly-controlled way. Signal frequency, power and exposure time are controlled by the system's 2450 MHz generator. The design of the radiating portion of the accessory and continuous cooling of the accessory during the ablation procedure determines the shape (near-spherical) and size of the resulting ablations. The modification to the antenna's cooling jacket (the component that forms the antenna's outer structure) changes the stiffness of the antenna's shaft only.
#### VII. PERFORMANCE DATA
The following non-clinical performance data were provided in support of the substantial equivalence determination.
### Bench Verification Testing
Verification testing was conducted to (1) demonstrate the modified antenna's continued robustness to mechanical stresses, (2) confirm that the antenna's printed depth markings were not affected by the underlying cooling jacket material change, (3) quantify the stiffness of the antenna's shaft relative to the 510(k) cleared device and, (4) confirm that the change did not impact the system's essential function - the performance of thermal ablations in soft tissue. Some of the verification tests were conducted using the fully-assembled Emprint Percutaneous Antenna, while other tests required only the cooling jacket sub-assembly.
### Product Stability Testing
Stability testing was conducted to evaluate the effect of the cooling-jacket material change on the antenna's performance over time. The testing was broad in scope and included mechanical stress testing, cytotoxicity testing and performance testing.
{6}------------------------------------------------
### Biocompatibility Testing
Biocompatibility testing of the new fiberglass material and the modified cooling jacket subassembly was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process. Additionally, the modified cooling jacket subassembly underwent a full chemical characterization.
### Validation Testing
Representative end users participated in a simulated-use validation study. The participants completed a series of tasks designed to be representative of those performed with the antenna during ablation procedures. Objective and subjective test results confirmed that (1) essential and safety-critical tasks were performed with the stiffer antenna in a safe and effective manner, (2) the modified device met usability requirements and, (3) there were no new, use-related risks, associated with the change.
#### CONCLUSIONS VIII.
The modified Emprint Percutaneous Antenna is identical to the predicate device, with the exception of an improved seal and a material change that serves to increase the antenna's stiffness. The subject and predicate devices share intended use, technological characteristics, and meet identical performance requirements. The results of comprehensive verification and validation testing raised no new questions of safety or effectiveness and confirmed the usability enhancement. The Emprint Ablation System's Percutaneous Antenna (subject device) and the modified Percutaneous Ablation Antenna (predicate device) are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.