← Product Code [KWY](/productcode/KWY) · K163081

# 40-42mm BiPolar Head and 22mm Femoral Head (K163081)

_Signature Orthopaedics Pty, Ltd. · KWY · Jan 29, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K163081

## Device Facts

- **Applicant:** Signature Orthopaedics Pty, Ltd.
- **Product Code:** [KWY](/productcode/KWY.md)
- **Decision Date:** Jan 29, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3390
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemilhip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as: - Acute femoral head or neck fracture - Fracture dislocation of the hip - Avascular necrosis of the femoral head - Non-union of femoral neck fractures - Certain high subcapital and femoral neck fractures in the elderly - Degenerative arthritis involving only the femoral head

## Device Story

Device consists of 40-42mm BiPolar Head and 22mm Femoral Head for hip hemiarthroplasty; used when natural acetabulum remains intact. BiPolar Head features stainless steel outer shell articulating against natural acetabulum and UHMWPE insert articulating against 22mm cobalt-chrome femoral head. Femoral head connects to existing Signature Orthopaedics femoral stems. Device implanted by orthopedic surgeons in clinical/OR settings. Provides stable articulation for patients with femoral head/neck pathologies; restores hip function; avoids total hip replacement when acetabulum is healthy.

## Clinical Evidence

Bench testing only. Evaluations included bacterial endotoxin levels (LAL method), range of motion analysis, component connection strength, fretting corrosion testing, and BiPolar head disclamping resistance testing to verify performance for in-vivo use.

## Technological Characteristics

BiPolar Head: Stainless steel outer shell (ISO 5832-9) and UHMWPE insert (ASTM F648). Femoral Head: Cobalt-chrome. Dimensions: 40-42mm BiPolar head diameter; 22mm femoral head diameter. Mechanical articulation. Non-active, implantable orthopedic prosthesis.

## Regulatory Identification

A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.

## Predicate Devices

- Signature Orthopaedics BiPolar Head ([K133370](/device/K133370.md))
- Signature Orthopaedics Femoral Head ([K121297](/device/K121297.md))
- DePuy Self-Centering Hip ([K812672](/device/K812672.md))
- Medacta Bipolar Heads ([K091967](/device/K091967.md))

## Reference Devices

- Signature Orthopaedics UniPolar Head ([K143184](/device/K143184.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is composed of three intertwined lines that form the shape of a bird or a wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 29, 2018

Signature Orthopaedics Pty Ltd. Declan Brazil Managing Director 7 Sirius Road Lane Cove, NSW 2066 AUSTRALIA

Re: K163081 Trade/Device Name: 40-42mm BiPolar Head and 22mm Femoral Head Regulation Number: 21 CFR 888.3390 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWY Dated: January 19, 2018 Received: January 22, 2018

Dear Declan Brazil:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Katherine D. Kavlock -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K163081

Device Name

40-42mm BiPolar Head and 22mm Femoral Head

#### Indications for Use (Describe)

Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemilhip arthroplasty only, where the natural acetabulum does not require replacement. The Evolve UniPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:

- · Acute femoral head or neck fracture
- · Fracture dislocation of the hip
- Avascular necrosis of the femoral head
- · Non-union of femoral neck fractures
- · Certain high subcapital and femoral neck fractures in the elderly
- · Degenerative arthritis involving only the femoral head

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div>  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 2 510(K) SUMMARY

| Manufacturer:      | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia                                                                                                                                                                                                                       |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | 40-42mm BiPolar Head and 22mm Femoral Head                                                                                                                                                                                                                                                                |
| Common Name:       | Hip Prosthesis                                                                                                                                                                                                                                                                                            |
| Contact:           | Dr. Declan Brazil<br>Managing Director of Signature Orthopaedics                                                                                                                                                                                                                                          |
| Prepared By:       | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br>Phone: +61 (2) 9428 5181<br>Fax: +61 (2) 8456 6065                                                                                                                                                                 |
| Date Prepared:     | January 19th, 2018                                                                                                                                                                                                                                                                                        |
| Classification:    | Class II per 21 CFR 888.3390: Hip joint femoral (hemi-hip)<br>metal/polymer cemented or uncemented prosthesis (KWY)                                                                                                                                                                                       |
| Predicate Devices: | Primary Predicate:<br>• Signature Orthopaedics BiPolar Head (K133370)<br>• Signature Orthopaedics Femoral Head (K121297)<br>Secondary Predicate:<br>• DePuy Self-Centering Hip (K812672)<br>• Medacta Bipolar Heads (K091967)<br>Reference Predicate:<br>• Signature Orthopaedics UniPolar Head (K143184) |

### Device Description:

The BiPolar Head consists of a stainless steel outer shell (per ISO5832-9) and a UHMWPE insert (per ASTM F648). The outer shell is highly polished to articulate against the patient's natural acetabulum. The insert articulates against a Signature Orthopaedics 22 or 28mm cobalt-chrome femoral head (K121297). The femoral head connects to a femoral stem from Signature Orthopaedics' range to complete the hip hemiarthroplasty.

### Indications for Use:

Signature Orthopaedics' Evolve UniPolar Head and BiPolar Head are intended for hemihip arthroplasty only, where the natural acetabulum does not require replacement. The 

{4}------------------------------------------------

Evolve UniPolar Head and BiPolar Head are indicated for bone fractures or pathologies involving only the femoral head/neck and/or proximal femur, such as:

- . Acute femoral head or neck fracture
- Fracture dislocation of the hip
- . Avascular necrosis of the femoral head
- . Non-union of femoral neck fractures
- Certain high subcapital and femoral neck fractures in the elderly
- . Degenerative arthritis involving only the femoral head

### Performance Testing:

Bacterial endotoxin levels were evaluated on the finished form of the subject device using the LAL method. Non-clinical testing and engineering evaluations were conducted to verify that the

performance of the 40-42mm BiPolar Heads are adequate for anticipated in-vivo use. Non-clinical testing included:

- Range of motion analysis
- Component connection strength and fretting corrosion testing
- . BiPolar head disclamping resistance testing

## Substantial Equivalence:

The 40-42mm BiPolar Heads have the same intended use. indications for use, materials and design as the Signature Orthopaedics BiPolar Heads (per K133370). The 40-42mm BiPolar Heads differ from the Signature Orthopaedics BiPolar Heads (per K133370) only in the size range. The 40-42mm BiPolar Heads expand Signature Orthopaedics BiPolar Heads' size range to match the 40-42mm diameter DePuy Self Centering Hip (per K812672) and Medacta Bipolar Head (per K091967) sizes. The 22mm Femoral Head is identical to the Signature Orthopaedics Femoral Heads (per K121297) excluding the outer diameter, which has been reduced to 22mm for compatibility with the 40-42mm BiPolar Heads. Non-clinical testing results support the substantial equivalence claim.

---

**Source:** [https://fda.innolitics.com/device/K163081](https://fda.innolitics.com/device/K163081)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
