K163067 · Cymedica Orthopedics, Inc. · IPF · Dec 1, 2016 · Physical Medicine
Device Facts
Record ID
K163067
Device Name
CyMedica e-vive™ System; CY-1000
Applicant
Cymedica Orthopedics, Inc.
Product Code
IPF · Physical Medicine
Decision Date
Dec 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The CyMedica e-vive System is a multifunctional electrotherapy device with two treatment modes that allow for neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS).
Device Story
Multifunctional electrotherapy device (NMES/TENS) for knee rehabilitation; includes integrated goniometer for range-of-motion measurement. System comprises conductive knee garment, controller, and mobile app. Controller delivers electrical stimulation via electrodes to motor points (NMES) or nerve pathways (TENS). Goniometer uses dual accelerometers (one in controller, one in sensor pod) to track knee flexion/extension. Mobile app provides virtual control panel, manages stimulation programs, and displays real-time range-of-motion data via Bluetooth Low Energy (BLE 4.1). Used in clinical or home settings under physician prescription. Patients operate device by selecting programs and adjusting intensity. Closed-loop feedback minimizes energy delivery. Benefits include muscle strengthening, pain relief, and objective tracking of joint recovery.
Clinical Evidence
Bench testing only. No clinical data. Safety and performance demonstrated via electrical safety (IEC 60601-1, 60601-2-10, 60601-1-11), EMC (IEC 60601-1-2), biocompatibility (ISO 10993), and usability (IEC 62366) testing. Wireless coexistence verified for BLE module. Software validated per FDA guidance and IEC 62304.
Technological Characteristics
Materials: Molded PC/ABS (Bayblend FR3010). Sensing: Accelerometer-based goniometer. Energy: 3.7V Li-Po rechargeable battery. Connectivity: BLE 4.1 wireless to mobile app. Software: Microprocessor-controlled stimulation with closed-loop feedback. Form factor: Wearable conductive knee garment with integrated controller and sensor pod.
Indications for Use
Indicated for patients requiring NMES for relaxation of muscle spasms, retardation/prevention of disuse atrophy, increasing local blood circulation, muscle re-education, post-surgical calf muscle stimulation to prevent venous thrombosis, and maintaining/increasing range of motion. Indicated for TENS for symptomatic relief/management of chronic intractable pain and adjunctive treatment for post-surgical and post-trauma acute pain.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are depicted in a single, continuous line. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2016
CyMedica Orthopedics, Inc. % Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, Minnesota 55313
Re: K163067
Trade/Device Name: CyMedica e-vive™ System; CY-1000 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ, KQX Dated: October 31, 2016 Received: November 2, 2016
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Michael J. Hoffmann -A" in a large, sans-serif font. The text is black and appears to be centered on a white background. The name is written in a clear and legible manner.
- for Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163067
#### Device Name
CyMedica e-vive System; CY-1000
Indications for Use (Describe)
The CyMedica e-vive System is a multifunctional electrotherapy device with two treatment modes that allow for neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS).
Indications for Use:
- As an NMES device, indications are for the following conditions:
- Relaxation of muscle spasms
- Retardation or prevention of disuse atrophy
- Increasing local blood circulation
- Re-educating muscles
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or increasing range of motion
As a TENS device, indications are for the following conditions:
- Symptomatic relief and management of chronic intractable pain
- Adjunctive treatment for post-surgical and post-trauma acute pain
Type of Use (*Select one or both, as applicable*)
| <div> <span style="font-size:10px"> <b> <span style="text-decoration:overline">X</span> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K163067 510(k) Summary
# CyMedica Orthopedics, Inc.
CyMedica e-vive™ System
## 1- SUBMITTER
| Manufacturer Name: | CyMedica Orthopedics, Inc.<br>19120 N. Pima Rd. Suite 135<br>Scottsdale, AZ 85255<br>Telephone (480) 664-1282<br>FAX (866) 296-2772 |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Kereshmeh Shahriari<br>19120 N. Pima Rd. Suite 135<br>Scottsdale, AZ 85255<br>kereshmeh@cymedicaortho.com<br>Telephone (480) 664-1282<br>FAX (866) 296-2772 |
| 510(k) Summary Preparation<br>Date | November 30, 2016 |
| 2- DEVICE | |
| Trade/Proprietary Name: | CyMedica e-vive™ System; CY-1000 |
| Common Name: | Muscle stimulator & Electrogoniometer |
| Classification Names: | Powered muscle stimulator (21 CFR 890.5850)<br>Transcutaneous Electrical Nerve Stimulator for Pain Relief (21<br>CFR 882.5890)<br>Goniometer, AC-powered (21 CFR 888.1500) |
| Product Codes: | Powered Muscle Stimulator or Neuromuscular Electrical<br>Stimulator (NMES); IPF<br>Transcutaneous Nerve Stimulator (TENS) for pain relief;<br>GZJ<br>Goniometer; KQX |
| Device Class: | Powered Muscle Stimulator & TENS: Class II<br>Goniometer: Class I, 510(k) Exempt |
{4}------------------------------------------------
CyMedica e-vive™ System K163067
NMES and TENS devices are reviewed by the Division of Neurological and Physical Medicine Devices (DNPMD).
## 3- PREDICATE DEVICE
Name & 510(k) Number:
K150413 QB1 NMES & TENS System CyMedica Orthopedics, Inc.
## 4- DEVICE DESCRIPTION
The CyMedica e-vive™ System is a multifunctional electrotherapy device with two stimulation channels and two treatment modes that allows for neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS). The principles of electrotherapy emulate the process observed during a voluntary muscle contraction. The e-vive™ system delivers stimulation based on the principles of NMES and TENS. NMES pulses stimulate motor points of target muscles, causing a muscle contraction. This can help re-educate and strengthen muscles following an injury or surgery. TENS blocks the pain signal sent from the affected area on nerve pathways.
The e-vive device is equipped with a goniometer (Class I, 510(k) Exempt), which is intended to measure and record joint range of motion.
e-vive™ electrical stimulation technology is based on the electrical stimulation technology developed for CyMedica QB1™ system. K150413. The e-vive™ electrotherapy programs are delivered to the patient directly by a User Interface controller integrated in a conductive garment. The stimulation programs are managed wirelessly using a mobile application developed for use on smart phones or tablets that interact with the User Interface controller.
The e-vive™ system are prescription devices in the USA and are intended to be used following the directions of a healthcare provider. The device may be used in a healthcare facility setting or by a patient or lay operator in a home environment.
In NMES mode, the e-vive™ system provides two therapeutic treatment programs: Post-Operative and Strength. Its simplified programming makes the device convenient for home use; after placing the electrodes and selecting the program as prescribed by a healthcare professional, the patient only needs to increase the intensity to a comfortable level to begin therapy. The evive™ NMES Post-Operative and Strength programs utilize an electrical stimulus that, when properly applied, activates specific muscles or muscle groups to help treat disuse muscle atrophy and to reeducate muscles. This is achieved via a closed loop feedback system that minimizes energy delivery to the targeted treatment areas.
In NMES mode, the e-vive™ system consists of a knee conductive garment with an incorporated Controller, docking receptacle, range of motion sensor, three electrodes, a USB-A to micro-USB-B cable for Controller charging, an electrode gel tube, and e-vive mobile app available for
{5}------------------------------------------------
download from the App Store onto a patient's mobile device.
As a TENS device, the e-vive TENS treatment is a safe and effective method of providing a drug-free method of pain relief. Patients with recurring pain, can utilize the e-vive TENS treatment program for immediate and long-term pain relief.
The e-vive™ TENS system consists of a conductive garment with an incorporated Controller. docking receptacle, two electrodes, a USB-A to micro-USB-B cable for Controller charging, an electrode gel tube, and e-vive mobile app available for download from the App Store onto a patient's mobile device.
The e-vive electrogoniometer (battery powered goniometer) is capable of measuring and recording the joint range of motion by using two sensors to detect the knee range of motion. One sensor is located on the Controller board and the second sensor is located within the sensor pod, on a small electronic board. The sensor pod is contained within the conductive garment below the knee area opening. The sensor pod is connected wirelessly to the Controller using a BLE connection. The sensor data from the sensor pod and the Controller sensor is conveyed to the mobile device real time. The sensors can detect the knee extension/flexion angles, and associated range of motion angle through the use of the e-vive mobile app when initiated by the patient or healthcare provider.
The e-vive™ System incorporates a Bluetooth Low Energy, BLE 4.1 connection module to enable wireless communication and can be paired with a Bluetooth enabled mobile device running the e-vive™ App, available from the App stores. The App implements a virtual control panel on the screen of the smart device where on-screen buttons are provided to the user.
The Controller is connected to the conductive garment via a 16-pin interface. The Controller contains the primary safety controls for operation of the device and a push button is available for switching the unit on or off. The power button remains active and can be used in the event of loss of Bluetooth connection to power on/off the device or start/stop stimulation while performing stimulation sessions. The e-vive™ system contains light emitting diodes (LED) which indicate status relating to battery charge, stimulation and Bluetooth™ activity. Power is derived from a 3.7V Li-Po rechargeable battery pack and the unit can be recharged by using the supplied USB cable.
#### 5- INDICATIONS FOR USE
The e-vive™ System is a multifunctional electrotherapy device with two treatment modes that allow for neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS).
Indications for Use:
As an NMES device, indications are for the following conditions:
{6}------------------------------------------------
- 1) Relaxation of muscle spasms
- 2) Retardation or prevention of disuse atrophy
- 3) Increasing local blood circulation
- 4) Re-educating muscles
- 5) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- 6) Maintaining or increasing range of motion
As a TENS device, indications are for the following conditions:
- 1) Symptomatic relief and management of chronic intractable pain
- 2) Adjunctive treatment for post-surgical and post-trauma acute pain
The treatment programs for NMES Post-Op and NMES Strength are shown below.
| Treatment<br>Program | Pulse shape | Duration | Frequency | Pulse width | Duty cycle | Work cycle | Relaxation time | Contraction time | Rest time | Indications numbers |
|----------------------|-------------|----------|-----------|-------------|------------|------------|-----------------|------------------|-----------|---------------------|
| NMES Post-Op | Monophasic | 20 min | 50 pps | 5 ms | 25% | 13 s | 10 s | 3 s<br>2 cycles | 3.4 s | 1, 2, 3, 4, 5, 6 |
| NMES Strength | Monophasic | 20 min | 50 pps | 5 ms | 25% | 12 s | 10 s | 1 s<br>5 cycles | 1.4 s | 1, 2, 3, 4, 5, 6 |
The treatment program for TENS pain management is shown below .
| Treatment<br>Program | Pulse<br>shape | Duration | Frequency | Pulse<br>width | Duty<br>cycle | Work<br>cycle | Interphase<br>interval<br>time | Indications<br>numbers |
|-------------------------|--------------------------|----------|-----------|----------------|---------------|---------------|--------------------------------|------------------------|
| TENS Pain<br>Management | Biphasic,<br>symmetrical | 20 min | 100 pps | 1 ms | 20% | Continuous | 4 ms | 7, 8 |
## 6- COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
CyMedica Orthopedics, Inc. demonstrated that, for the purposes of FDA's regulation of medical devices, the e-vive™ System is substantially equivalent to the predicate device, QB1 System, K150413. The data included in this submission demonstrates substantial equivalence to the predicate device listed above.
The intended use, design, materials and functional characteristics of the e-vive™ NMES and TENS System and the QB1 predicate device are substantially the same. The subject device and predicate device provide neuromuscular electrical stimulation therapy, are portable, hand-held, and for home healthcare environment use. Similar to the predicate device, the power in e-vive™ device is derived from a rechargeable battery that is pre-installed in the User Interface Controller. There are two channels of stimulation in both devices. In both devices, the user needs to select the desired treatment program and adjust the intensity. Both devices employ a ramp-up, work, and rest phases.
{7}------------------------------------------------
CyMedica e-vive™ System K163067
The data included in this submission demonstrates substantial equivalence to the predicate device listed above. The intended use, design, and functional characteristics of the e-vive electrogoniometer are within the limitations of exemption in 21 CFR 888.9, applicable to goniometers which are Class I, 510(k) exempt. The goniometer in the e-vive system measures and records joint range of motion similar to other goniometers under 21 CFR 888.1500. The goniometer is battery operated and uses accelerometers to measure the flexion and extension angles. An embedded software in the e-vive system calculates the range of motion using the measured flexion and extension angles data.
| | Parameter | e-vive | QB1 |
|---|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| | 510(k) Number | K163067 | K150413 |
| | Device Name and Mode | e-vive | QB1 |
| | Manufacturer | CyMedica Orthopedics | CyMedica Orthopedics |
| | Power Source(s)† | Single VD434053:<br>3.7V; 1000mAh;<br>Lithium ion polymer<br>battery | Single H605060: 3.7V;<br>2000mAh; Lithium ion<br>polymer battery |
| - | Method of Line Current Isolation | No line connection | No line connection |
| - | Patient Leakage Current†† | --- | --- |
| - | Normal Condition(µA) | 4.88 | 1.5 |
| - | Single Fault Condition (µA) | 8.00 | 2.6 |
| | Number of Output Modes ††† | 3 | 3 |
| | Number of Output Channels ††††: Synchronous or Alternating? Method of Channel Isolation | Alternating<br>Transistor | Alternating<br>Transistor |
| | Regulated Current or Regulated Voltage? | Regulated Power | Regulated Power |
| | Software/Firmware/Microprocessor Control? | Yes | Yes |
| | Automatic Overload Trip? | No | No |
| | Automatic No-Load Trip? | No | No |
| | Automatic Shut Off? | Yes | Yes |
| | User Override Control? | Yes Stop Buttons | Yes Stop Button |
| | Indicator Display:<br>On/Off Status?<br>Low Battery?<br>Voltage/Current Level? | Yes<br>Yes<br>No | Yes<br>Yes<br>No |
| | Timer Range (minutes) | 20-open | 20-open |
| | Compliance with Voluntary Standards? | Yes.<br>IEC 60601-2-10:2012 Part 2-10;<br><br>IEC 60601-1-11:2015 Part 1-11;<br><br>IEC 60601-1-6:2010;<br><br>ISO 10993-1:2009 Part 1 | Yes.<br>IEC 60601-2-10:2012 Part 2-10;<br><br>IEC 60601-1-11:2010 Part 1-11;<br><br>IEC 60601-1-6:2010;<br><br>ISO 10993-1:2009 Part 1 |
| | Compliance with 21 CFR 898? | Yes | Yes |
| | Weight (lbs., oz.) | 1.76 oz. (50g) | 8.6 oz (243.8g) |
#### Comparison of Technology to Predicate
{8}------------------------------------------------
| Parameter | e-vive | QB1 |
|------------------------------------|-----------------------|-----------------------|
| Dimensions (in.) [W x H x D] | 1.93" x 0.64" x 3.28" | 3.15" x 0.88" x 5.84" |
| Housing Materials and Construction | Molded PC\ABS | Molded PC\ABS |
| | Plastic | Plastic |
| | Bayblend FR3010 | Bayblend FR3010 |
| Parameter | e-vive NMES POST-OP | e-vive NMES STRENGTH | e-vive TENS | QB1 NMES POST-Op | QB1 NMES STRENGTH | QB1 NMES STRENGTH | |
|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------|-------------------------------|------------------------------------------|-----------------------------------------|-----------------------------------------|------------------------------------------|------------------------------|
| 510(k) Number | K163067 | K163067 | K163067 | K150413 | K150413 | K150413 | |
| Mode or Program Name | NMES POST-OP | NMES STRENGTH | TENS | NMES POST-OP | NMES STRENGTH | TENS | |
| Waveform (e.g., pulsed monophasic, biphasic) | Pulsed Monophasic | Pulsed Monophasic | Symmetric Biphasic | Pulsed Monophasic | Pulsed Monophasic | Symmetric Biphasic | |
| Shape (e.g., rectangular, spike, rectified sinusoidal) | Complex | Complex | Complex | Complex | Complex | Complex | |
| Maximum Output Voltage (volts, rms) | 8.5 @500Ω<br>15.5 @ 2 kΩ<br>20.2 @10 kΩ | 8.4 @500Ω<br>@ 2 kΩ<br>@10 kΩ | 10.4 @500Ω<br>17.1 @ 2 kΩ<br>20.9 @10 kΩ | 9.3 @500Ω<br>16.8 @ 2 kΩ<br>21.9 @10 kΩ | 9.3 @500Ω<br>16.6 @ 2 kΩ<br>21.8 @10 kΩ | 10.2 @500Ω<br>16.7 @ 2 kΩ<br>19.8 @10 kΩ | |
| Maximum Output Current (mA, rms) | 17.1 @500Ω<br>7.7 @ 2 kΩ<br>2.0 @10 kΩ | @500Ω<br>@ 2 kΩ<br>@10 kΩ | 20.9 @500Ω<br>8.5 @ 2 kΩ<br>2.1 @10 kΩ | 18.5 @500Ω<br>8.4 @ 2 kΩ<br>2.2 @10 kΩ | 18.6 @500Ω<br>8.3 @ 2 kΩ<br>2.2 @10 kΩ | 20.5 @500Ω<br>8.3 @ 2 kΩ<br>2.0 @10 kΩ | |
| Duration of primary (depolarizing) phase (µsec) | 5000 | 5000 | N/A (Continuous) | 5000 | 5000 | N/A (Continuous) | |
| Pulse Duration (µsec) | 5000 | 5000 | 1000 | 5000 | 5000 | 1000 | |
| Parameter | e-vive NMES<br>POST-OP | e-vive NMES<br>STRENGTH | e-vive TENS | QB1 NMES<br>POST-Op | QB1 NMES<br>STRENGTH | QB1 NMES<br>STRENGTH | |
| Frequency (Hz) [or<br>Rate (pps)] | 50 | 50 | 100 | 50 | 50 | 100 | |
| For interferential<br>modes only: Beat<br>Frequency (Hz) | N/A | N/A | N/A | N/A | N/A | N/A | |
| For<br>multiphasi<br>c<br>waveform<br>s only: | Symmetrical<br>phases? | N/A | N/A | Yes | N/A | N/A | Yes |
| | Phase<br>duration | N/A | N/A | 1 ms | N/A | N/A | 1 ms |
| Net Charge<br>(microcoulombs<br>(µC) per pulse) (If<br>zero, state method<br>of achieving zero<br>net charge.) | 342 @500Ω | 336 @500Ω | 0 @500Ω<br>(Symmetric<br>Biphasic) | 371 @500Ω | 372 @500Ω | 0 @500Ω<br>(Symmetric<br>Biphasic) | |
| Maximum Phase<br>Charge, (µC) | 342 @500Ω | 336 @500Ω | 417 @500Ω | 371 @500Ω | 372 @500Ω | 409 @500Ω | |
| Maximum Current<br>Density (mA/cm², r.m.s.) | 0.66 @500Ω | 0.65 @500Ω | 0.81 @500Ω | 0.72 @500Ω | 0.72 @500Ω | 0.79 @500Ω | |
| Maximum Average<br>Current (average<br>absolute value),<br>mA | 17.1 @500Ω | 16.8 @500Ω | 20.9 @500Ω | 18.5 @500Ω | 18.6 @500Ω | 20.5 @500Ω | |
| Maximum Average<br>Power Density,<br>(W/cm²), (using<br>smallest electrode<br>conductive surface<br>area) | 0.005 @500Ω | 0.006 @500Ω | 0.009<br>@500Ω | 0.007<br>@500Ω | 0.007<br>@500Ω | 0.008<br>@500Ω | |
| | Parameter | e-vive NMES<br>POST-OP | e-vive NMES<br>STRENGTH | e-vive TENS | QB1 NMES<br>POST-Op | QB1 NMES<br>STRENGTH | QB1 NMES<br>STRENGTH |
| Burst<br>Mode | (a) Pulses<br>per burst | 150 | 50 | | 150 | 50 | |
| | (b) Bursts<br>per second | 0.087 | 0.23 | | 0.087 | 0.23 | |
| | (c) Burst<br>duration<br>(seconds) | 3 | 1 | | 3 | 1 | N/A<br>(Continuous<br>Pulse) |
| | (d) Duty<br>Cycle: Line<br>(b) x<br>Line (c) | 0.26 | 0.23 | | 0.26 | 0.23 | |
| | ON Time<br>(seconds) | 312 | 276 | Continuous | 312 | 276 | Continuous |
| | OFF Time<br>(seconds) | 888 | 924 | Continuous | 888 | 924 | Continuous |
| | Additional Features<br>(specify, if<br>applicable) | N/A | N/A | N/A | N/A | N/A | N/A |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
e-vive™ System differs from the QB1 System device in the following area:
| Difference Area | Equivalence Discussion |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology; wireless and<br>mobile app controlled vs.<br>wired:<br>e-vive™ system is a<br>wireless and mobile app<br>controlled device while<br>QB1 system is a wired<br>device. | The mobile app in e-vive system communicates the<br>User Interface Controller within the conductive<br>garment via Bluetooth Low Energy, BLE 4.1<br>connection. The QB1 User Interface is connected to<br>the treatment pod within the conductive garment using<br>a cable. As shown in Section 18, the system has been<br>fully tested per IEC 60601-1, 60601-1-2, 60601-1-6,<br>60601-1-11, 60601-2-10, and IEC 62366 and meets all<br>standard requirements and FDA guidance requirements.<br>It adequately controls the stimulator outputs to the<br>allowable ranges and within the tolerance limits<br>provided in the standards and guidance, evidenced in<br>IEC 60601-2-10. Accordingly, the e-vive wireless and<br>mobile app controlled system poses no new safety risks<br>and is substantially equivalent to the predicate. |
Any differences in the technological characteristics between the subject and predicate device do not raise new issues of safety or effectiveness.
#### 7- PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence.
To demonstrate the safety, the e-vive™ system was tested for electrical safety, electromagnetic compatibility, usability, biocompatibility, and risk management requirements.
To demonstrate the safety, the e-vive™ System was tested per the following standards:
- AAMI/ANSI ES 60601-1:2005/(R)2012 And A1:2012 Medical electrical equipment -. Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: 2014, 4th Edition. Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility requirements
- IEC 60601-2-10: 2012, 2nd Edition, Medical electrical equipment- Part 2-10: Particular . requirements for the basic safety and essential performance of nerve and muscle stimulator
{12}------------------------------------------------
- IEC 60601-1-11: 2015, 2nd Edition, Medical electrical equipment- Part 1-11: General . requirements for basic safety and essential performance- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . IEC 60601-1-6: 2010, 3rd Edition, Medical electrical equipment- Part 1-6: General requirements for basic safety and essential performance- Collateral standard: Usability including IEC 62366: Application of usability engineering to medical devices
- IEC 62366: 2007, 1st Edition, Medical devices -- Application of usability engineering to medical devices
- IEC 62304: 2006, 15T Edition, Medical device software Software life cycle processes ●
- ISO 10993-1: 2009, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process
- ISO 10993-5: 2009, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2010, Biological evaluation of medical devices- Part 10: Tests for ● irritation and skin sensitization
- . ISO 14971: 2007, Application of risk management to medical devices
In addition, the performance of e-vive™ BLE module for wireless co-existence was evaluated and certified by the supplier of the module in an environment with equipment operating in the ISM band i.e. Bluetooth and Wi-Fi devices, cellphones, cordless phones etc. The device met all specified requirements.
BLE module testing was conducted in accordance with the following standards:
- FCC 47CFR PT 15-C, Issued:2007/10/01 Title 47 CFR Part 15 Subpart C: Intentional . Radiators [FCC §15.247]
- FCC 47CFR PT 15 SPT B. Issued: 2013/01/28 Title 47 CFR Part 15 Subpart B: . Unintentional Radiators, FCC Part 15, Subpart B [FCC §15.107 & FCC §15.109]
- . RSS 210, Issue:2010/12/01 Issue:8 Low Power License-Exempt Radio communication Devices (All Frequency Bands) - Category I Equipment
- ICES 003 Issued: 2012/08/01Spectrum Management and Telecommunications . Interference-Causing Equipment Standard
- . ETSI EN 300 328, Issued: 2012/06/08 V. 1.8.1 Electromagnetic Compatibility. & Radio Spectrum Matters; Wideband Trans. Systems; Data Trans. Equip. Operating in the 2.4
Page 10 of 12
{13}------------------------------------------------
GHz ISM Band & Using Wideband Modulation Techniques; Harmon. EN Covering Essen. Req. Under Art. 3.2-R&TTE Directive
- ETSI EN 301 489-1, Issued: 2011/09/22 V1.9.2 Electromagnetic Compatibility & Radio . Spectrum Matters (ERM); Electromagnetic Compatibility (EMC) Radio Equipment & Services; Prt 1: Common Technical Requirements
- ETSI EN 301 489-17, Issue:2009/05/12 Electromagnetic Compatibility And Radio . Spectrum Matters (Erm); Electromagnetic Compatibility (EMC) Standard For Radio Equipment; Part 17: Specific Conditions For Broadband Data Transmission Systems -V2.1.1
Battery testing was conducted in accordance with IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them, for use in portable applications.
To demonstrate the e-vive system effectiveness and performance substantial equivalency of the subject device, e-vive system and the predicate device OB1 System (K150413) were tested according to the following FDA guidance document:
- . FDA Final Guidance Document for Powered Muscle Stimulator 510(k), June 9, 1999
#### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the e-vive system and complies with the IEC 60601-1, IEC 60601-2-10, and 60601-11 standards for safety and the IEC 60601-1-2 standard for EMC.
#### Software Verification & Validation Testing
The device's software has been validated in accordance with the requirements set forth in the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) and IEC 62304: 2006, 180 Edition, Medical device software – Software life cycle processes. The software validation tests demonstrated that the software version meets its design requirements.
#### Human Factors and Usability
The human factors and usability study was conducted to validate the usability of the e-vive system in the home environment. The results of the study support the instructions for successfully using the device as intended. The results of human factors and usability study substantiates the acceptability of the risks identified during the risk assessment activities. The evive system complies with the IEC 60601-1-6: 2010 for usability and IEC 62366: Application of usability engineering to medical devices.
{14}------------------------------------------------
#### 8- CONCLUSION
Based on the performance testing and the supporting documentation, it can be concluded that the e-vive NMES and TENS system is safe, effective, and substantially equivalent to the predicate devices. The e-vive device output pulse parameters provide a safe and effective treatment for the NMES and TENS applications.
Based on the acceptable bench test results and e-vive compliance with the applicable standards, the e-vive NMES and TENS therapies are considered safe and as effective as the predicate device, QB1 System (K150413) for its intended uses and indications for use. The e-vive NMES and TENS pulse parameters and waveform are selected and designed so they would provide a safe and effective treatment for the indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.