ClearLumen II Peripheral Thrombectomy System

K163051 · Walk Vascular, LLC · QEZ · Feb 21, 2017 · Cardiovascular

Device Facts

Record IDK163051
Device NameClearLumen II Peripheral Thrombectomy System
ApplicantWalk Vascular, LLC
Product CodeQEZ · Cardiovascular
Decision DateFeb 21, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ClearLumen II Peripheral Thrombectomy System is intended to remove/aspirate fluid and thrombus from the peripheral vasculature.

Device Story

ClearLumen II Peripheral Thrombectomy System is a multi-lumen catheter system for peripheral vascular thrombus removal. It uses a hydro-mechanical approach: a peripheral Saline Drive Unit (SDU) powers a piston pump in a disposable pump set to deliver high-pressure saline to the catheter's distal tip, macerating thrombus. Simultaneously, a user-provided 60 ml VacLok syringe creates suction to aspirate the macerated material through a primary lumen. The system is used in clinical settings by physicians. The SDU is IV-pole mounted, microprocessor-controlled, and provides status via LED indicators. The device benefits patients by enabling minimally invasive removal of peripheral thrombi. No saline is injected into the patient during operation.

Clinical Evidence

No clinical data were generated. Evidence consists of bench testing (pull, leak, flow, kink, torque, guidewire compatibility, deliverability, simulated clot removal) and pre-clinical histological evaluation in peripheral arteries. All testing met pre-determined acceptance criteria.

Technological Characteristics

Multi-lumen catheter system; dual-lumen design. Components: aspiration catheter, pump set, and reusable SDU. Energy: 24V external power supply. SDU features microprocessor-controlled circuit and firmware for motor/vacuum monitoring. Sterilization: validated per AAMI TIR 28. Biocompatibility: ISO 10993-1 compliant.

Indications for Use

Indicated for the removal/aspiration of fluid and the break-up of soft emboli and thrombus from the peripheral vasculature.

Regulatory Classification

Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 21, 2017 Walk Vascular, LLC % Mr. Paul Gasser Medical Device RA/QA Consultant 13612 Rushmore Lane Santa Ana, CA 92705 Re: K163051 Trade/Device Name: ClearLumen II Peripheral Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: January 10, 2017 Received: January 11, 2017 Dear Mr. Gasser: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, M.A. Willemsen for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163051 #### Device Name ClearLumen II Peripheral Thrombectomy System Indications for Use (Describe) The ClearLumen II Peripheral Thrombectomy System is intended to remove/aspirate fluid and thrombus from the peripheral vasculature. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <span style="text-decoration: underline;"><b> </b></span> <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Traditional 510(k) Summary | Submitter: | Walk Vascular, LLC<br>17171 Daimler Street<br>Irvine, CA 92614 | |------------------------|------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Brad Culbert<br>VP, Engineering<br>Walk Vascular LLC<br>Telephone: 949-752-9642<br>Fax: 949-752-9658<br>Email: bsculbert@yahoo.com | | Date Summary Prepared: | October 28, 2016 | | Device Trade Name: | ClearLumen II Peripheral Thrombectomy System | | Common Name: | Embolectomy/Thrombectomy Catheter | | Classification Name: | Embolectomy Catheter (21 CFR 870.5150) | | Product Code: | DXE | | Predicate Device: | Walk Vascular ClearLumen II Thrombectomy System (K161786) | # Device Description: The ClearLumen II Peripheral Thrombectomy System is a multi-lumen device that allows for simultaneous hydro-mechanical thrombus disruption and thrombus aspiration. It is designed to simultaneously deliver a stream of high pressure saline via a displacement pump to the distal tip of the catheter, while aspirating thrombotic material macerated by the saline stream. The ClearLumen II Peripheral Thrombectomy System consists of one aspiration catheter, which connects to the ClearLumen II Pump Set and the ClearLumen Peripheral Saline Drive Unit (SDU). In use, thrombus enters the distal catheter tip via the suction force provided by a 60 ml VacLok syringe, which is connected to the vacuum sensor. The peripheral SDU and pump set deliver a stream of sterile saline through the secondary lumen to break up and dilute the thrombus within the catheter. The diluted thrombus and saline is drawn back through the primary lumen and deposited into the syringe. No saline is injected into the patient during normal operation. The proximal end of the pump set consists of a spike and an in-line drip chamber that is used to pierce the saline bag and connect the pump set to the saline source. The cassette, which is centered in the pump set, contains a piston pump and is mounted onto the peripheral SDU. The cassette is powered by the motor contained in the peripheral SDU. The distal end of the pump {4}------------------------------------------------ set has a connector, which mounts to the proximal end of the Y-connector of the peripheral catheter and delivers the saline to the peripheral catheter. The peripheral SDU is a reusable, IV pole mounted device. The fork drive of the peripheral SDU is designed to run the piston pump contained in the pump set to deliver the stream of saline to the peripheral catheter, when activated by the vacuum sensor. Vacuum is achieved by a vacuum syringe (not provided), which is connected to the vacuum sensor. The vacuum sensor is connected to the aspiration lumen on the Y-connector of the peripheral catheter. The peripheral SDU contains a microprocessor controlled circuit and firmware that monitors various functions of the motor and vacuum to assure that the device is functioning as expected. Various colored LED lights on the front panel indicate to the user the current status of the SDU. Energy is provided by a 24 volt external power supply, which is connected to mains power. # Indications for Use: The ClearLumen II Peripheral Thrombectomy System is intended to remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature. # Statement of Equivalence: The subject device and the predicate share a similar intended use and technological characteristics. | Design Features | Predicate ClearLumen II<br>Catheter | Subject ClearLumen II<br>Peripheral Catheter | |---------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------| | Intended use | Coronary and peripheral | Peripheral | | Catheter overall length<br>(cm) | 135 | 115 | | Minimum guide catheter<br>ID (inches) | 0.070 | 0.113 | | Number of lumens | Tri-lumen | Dual-lumen | | Proximal design | Saline lumen is bonded to the<br>straight-away main port of the<br>Y-site adaptor | Saline lumen is bonded to the<br>side-branch port of the Y-site<br>adaptor | | Tip design | Beveled | Blunt, soft | | Guidewire | Rapid exchange | Over the wire | | SDU to be used with | Coronary SDU | Peripheral SDU | Key differences between the two are reflected in the table below. The ClearLumen II Peripheral Thrombectomy System is substantially equivalent to the predicate device with regard to its similar intended use, design, function, materials and sterilization method. {5}------------------------------------------------ # Summary of Non-Clinical Performance Data: Device evaluation consisted of in vitro testing performed pursuant to Walk Vascular's risk analysis. All data met the acceptance criteria and fell within pre-determined product specifications and external standard requirements. The following testing was performed: # Design Verification Testing: - Pull test of welded and glued joints Leak Pressure Flow Kink resistance Torque strength Guidewire compatibility Deliverability in torturous model Performance (simulated clot removal) # Biocompatibility Testing: Biocompatibility testing was conducted on the peripheral catheter and pump set combination, as the two devices are used concurrently during the surgical procedure. Biocompatibility testing was conducted in accordance with ISO 10993-1. # Sterilization Testing: A sterilization adoption assessment was conducted in accordance with AAMI TIR 28 in order to support the ability of the current sterilization cycle to adequately sterilize the subject device. # Transportation and Shelf Life Testing: Shipping and distribution testing was conducted in accordance with ISTA 2A. Shelf life testing was performed. The data from the in vitro testing above supports the substantial equivalence of the subject device to the predicate device. # Summary of Pre-Clinical and Clinical Data: A pre-clinical study was conducted to evaluate the safety of the peripheral catheter. Histological evaluation established the safety of the peripheral catheter when used in peripheral arteries. No clinical data were generated to establish substantial equivalence. Bench data and pre-clinical data are considered adequate to support a determination of substantial equivalence. # Summary: Based on the intended use, in vitro performance and biocompatibility information provided in this premarket notification, the ClearLumen II Peripheral Thrombectomy System is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...