← Product Code [DXN](/productcode/DXN) · K163045

# Omron HEM-6400T-Z Wrist Blood Pressure Monitor (K163045)

_Omron Healthcare, Inc. · DXN · Feb 21, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K163045

## Device Facts

- **Applicant:** Omron Healthcare, Inc.
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Feb 21, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 6 inches to 8 inches (15.0 cm to 20.5 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

## Device Story

Battery-powered, automatic, non-invasive wrist blood pressure monitor for home use. Device uses piezoelectric pump to inflate wrist cuff and electric valve for deflation. Oscillometric method extracts pulse waveform data during inflation; microprocessor analyzes data to determine systolic/diastolic blood pressure and pulse rate. Includes Advanced Positioning Sensor (APS) using accelerometer to ensure cuff is at heart level. Features irregular heartbeat detection. Optional smartphone app ("Omron connect") displays trend graphs but performs no diagnostic/measurement functions. Operated by patient; results shown on device display. Benefits include convenient, self-administered blood pressure monitoring for home management.

## Clinical Evidence

Clinical investigation validated accuracy of oscillometric measurements against auscultation method using calibrated sphygmomanometer by trained staff. Results demonstrated equivalent performance. Nonclinical testing included comparative bench testing, cleaning verification, usability testing in simulated home environment, biocompatibility (ISO 10993-1), electrical safety, EMC, and software V&V.

## Technological Characteristics

Cuff oscillometric measurement; semiconductor pressure sensor; piezoelectric pump; active electronic control valve. Rechargeable lithium-polymer battery. Organic electroluminescent display. Wrist cuff (elastomer strap, air bag, metal buckle). APS via accelerometer. Biocompatibility per ISO 10993. Non-sterile. Home use environment.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Omron BP652N (HEM-6300-Z) with APS ([K142917](/device/K142917.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 21, 2017

Omron Healthcare, Inc. % Ronald S. Warren Senior Director, Regulatory Affairs Experien Group LLC 224 Airport Parkway Suite 250 San Jose, CA 95110 US

Re: K163045

Trade/Device Name: Omron HEM-6400T-Z Wrist Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN Dated: January 20, 2017 Received: January 23, 2017

Dear Ronald S. Warren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Form Approved: OMB No.   | 0910-0120        |
|--------------------------|------------------|
| Expiration Date:         | January 31, 2017 |
| See PRA Statement below. |                  |

**DEPARTMENT OF HEALTH AND HUMAN SERVICES**  

 Food and Drug Administration

**Indications for Use**

| 510(k) Number (if known)       | K163045                                                                                                                                                                                                                                                                                                                                        |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | Omron HEM-6400T-Z Wrist Blood Pressure Monitor                                                                                                                                                                                                                                                                                                 |
| Indications for Use (Describe) | The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population             with wrist circumference ranging from 6 inches to 8 inches (15.0 cm to 20.5 cm).<br>The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. |

| Type of Use (Select one or both, as applicable) | <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label><br><label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

 This section applies only to requirements of the Paperwork Reduction Act of 1995.  

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***  

 The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete
 and review the collection of information. Send comments regarding this burden estimate or any other aspect
 of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

 "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
 information unless it displays a currently valid OMB number."

 FORM FDA 3881 (1/14)
 Page 1 of 1
 PSC Publishing Services (301) 443-6740 EF

a marka mana mana mana mana marka masa mara marka masa mara mana mana mara mara mara mara mara mara mara mara mara mara mara mara mara mara mara mara mara mara mara mara mar

{3}------------------------------------------------

## 510(k) Notification K_

#### GENERAL INFORMATION [807.92(a)(1)]

# Applicant:

Omron Healthcare, Inc. 1925 West Field Court Lake Forest, IL 60045 USA Phone: 847-247-5626 FAX: 847-680-6269

### Contact Person:

Ronald S. Warren Senior Director, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA Phone: 408-505-3926 FAX: 408-400-0856

# Date Prepared: October 27, 2016

#### DEVICE INFORMATION

Trade Name:

Omron HEM-6400T-Z Wrist Blood Pressure Monitor

#### Generic/Common Name:

Noninvasive blood pressure measurement system

#### Classification:

Class II per 21CFR§870.1130

#### Product Code:

DXN

{4}------------------------------------------------

## PREDICATE DEVICE(S)

The Omron HEM-6400T-Z Wrist Blood Pressure Monitor is substantially equivalent to the Omron BP652N (HEM-6300-Z) with APS (K142917).

## DEVICE DESCRIPTION

The Omron HEM-6400T-Z Wrist Blood Pressure Monitor is a battery- powered automatic non-invasive blood pressure system intended for home use. The device is powered by a rechargeable lithium-polymer battery. There is no connection to external power. The device inflates a wrist cuff with an integral pump, then deflates the cuff via an electric valve. During inflation, the wrist cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic blood pressure. The systolic and diastolic blood pressures are measured using the oscillometric method. The cuff pressure range is 0 to 299mmHg and the pulse rate range is 30 to 199 beats/min.

HEM-6400T-Z is intended for use in adult patient population with wrist circumference ranging from 6 inches to 8 inches (15.0 cm to 20.5 cm). The device also detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. The device displays the latest blood pressure reading, while up to 100 readings can be stored in memory. In addition, the device includes an Advanced Positioning Sensor (APS) as an aid to the user to determine if the wrist cuff is at the correct height in relation to the heart. It makes this determination based on the reading of an accelerometer (to measure the angle of the arm in relation to the table) integral to the device. The APS feature is similar to the predicate device. The operation of the device is intended for home use. Functions and other features that are controlled by the end user include applying the wrist cuff to the wrist, powering on/off the system, starting or stopping the BP and pulse measurement cycle, and charging the batteries as needed. As an optional feature, the user can also pair the HEM-6400T-Z to a smartphone when employing the "Omron connect" app. This app is an optional feature and is only intended to display trend graphs of measured systolic and diastolic blood pressure, and pulse rate. This app does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user. Aside from this optional app for smartphones, HEM-6400T-Z does not connect with other collateral devices.

### INDICATIONS FOR USE

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 6 inches to 8 inches (15.0 cm to 20.5 cm).

The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

{5}------------------------------------------------

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

The proposed HEM-6400T-Z has similar technological characteristics as compared to the predicate BP652N device. Both devices are intended for home use and employ the cuff oscillometric method for measuring blood pressure and pulse rate from the wrist. Both devices have the same cuff pressure range of 0 to 299mmHg. The pulse rate range is similar between the two devices with a range of 30 to 199 beats/min for HEM-6400T-Z and range of 40 to 180 beats/min for the predicate BP652N. The devices accommodate similar sized wrists. HEM-6400T-Z is intended for a wrist circumference of 6 to 8 inches (15.0 cm to 20.5 cm) while the BP652N is intended for a wrist circumference of 5.25 to 8.5 inches (13.5 cm to 21.5 cm). The accuracy of pressure reading is ±3mmHg for both devices, and accuracy of pulse rate is ±5% in both devices. Both devices include detection of irregular heartbeats and give a warning signal with readings. In addition, both devices include an APS as an aid to the user to determine if the wrist cuff is at the correct height in relation to the heart. Both devices utilize a piezoelectric pump for wrist cuff inflation. Both the proposed device and the predicate device employ a semiconductor pressure sensor, and utilize an active electronic control valve that performs cuff air bleeding and release.

There are minor differences between HEM-6400T-Z and BP652N but these do not raise different questions of safety or effectiveness. The HEM-6400T-Z uses a rechargeable lithium-polymer battery as the power source, whereas BP652N uses 2 "AAA" batteries. On the device display, HEM-6400T-Z adds time and date as compared to BP652N display, and the HEM-6400T-Z control buttons have been updated for user convenience. Specifically, the HEM-6400T-Z has a secondary "mini display" on the bottom of the device which also serves as the start/stop button. During the BP measurement, this mini display cycles through a display of systolic pressure, diastolic pressure, and then pulse rate. At the conclusion of the BP measurement, the mini display goes blank and the final results are shown on the main display screen on the top of the device. The BP652N only has a single display screen and start/stop button located on the top of the device. The HEM-6400T-Z wrist cuff main components include an elastomer strap, air bag and metal buckle, while the BP652N wrist cuff main components are a cloth bag, air bag, and hookand-loop strap (aka Velcro). See Table 5.1 for a summary of the comparison between the proposed device and the predicate device.

{6}------------------------------------------------

| Feature                                         | Proposed Device<br>Omron HEM-<br>6400T-Z<br>K                                                                                                                                                                                                                                                                                                                                      | Predicate Device<br>Omron BP652N<br>K142917                                                                                                                                                                                                                                                                                                                                                | Analysis of Technological<br>Differences                                                                                                                                                                                                                                                           |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification-<br>Regulation                   | 21 CFR§870.1130,<br>Noninvasive blood<br>pressure<br>measurement<br>system.                                                                                                                                                                                                                                                                                                        | 21 CFR§870.1130,<br>Noninvasive blood<br>pressure<br>measurement system.                                                                                                                                                                                                                                                                                                                   | No difference. Proposed device and<br>predicate device have the same<br>medical device classification<br>number                                                                                                                                                                                    |
| Classification -<br>Product Code                | DXN - Noninvasive<br>blood pressure<br>measurement                                                                                                                                                                                                                                                                                                                                 | DXN - Noninvasive<br>blood pressure<br>measurement                                                                                                                                                                                                                                                                                                                                         | No difference. Proposed device and<br>predicate device have the same<br>medical device product code                                                                                                                                                                                                |
| Indications for<br>Use                          | The device is a<br>digital monitor<br>intended for use in<br>measuring blood<br>pressure and pulse<br>rate in adult patient<br>population with<br>wrist circumference<br>ranging from 6<br>inches to 8 inches<br>(15.0 cm to 20.5<br>cm).<br>The device detects<br>the appearance of<br>irregular heartbeats<br>during measurement<br>and gives a warning<br>signal with readings. | The device is a<br>digital monitor<br>intended for use in<br>measuring blood<br>pressure and pulse<br>rate in adult patient<br>population with wrist<br>circumference<br>ranging from 5 1/4<br>inches to 8 1/2<br>inches (13.5cut to<br>21.5 cm). The device<br>detects the<br>appearance of<br>irregular heartbeats<br>during measurement<br>and gives a warning<br>signal with readings. | No difference of intended purpose.<br>The proposed indication for use is<br>the same as the cleared indication<br>for use of the predicate device. The<br>wrist circumference is slightly<br>different between the two devices,<br>but this difference does not affect<br>safety or effectiveness. |
| Environment of<br>Use                           | Home Use                                                                                                                                                                                                                                                                                                                                                                           | Home Use                                                                                                                                                                                                                                                                                                                                                                                   | No difference.  The proposed<br>device and predicate device have<br>the same environment of use.                                                                                                                                                                                                   |
| Patient Population                              | Adults                                                                                                                                                                                                                                                                                                                                                                             | Adults                                                                                                                                                                                                                                                                                                                                                                                     | No difference. Intended for same<br>population.                                                                                                                                                                                                                                                    |
| Contraindications/<br>Warnings/<br>Precautions  | There are no known<br>contraindications.                                                                                                                                                                                                                                                                                                                                           | There are no known<br>contraindications.                                                                                                                                                                                                                                                                                                                                                   | No difference. Proposed device<br>and predicate device have no<br>known contraindications                                                                                                                                                                                                          |
| Single Use                                      | No                                                                                                                                                                                                                                                                                                                                                                                 | No                                                                                                                                                                                                                                                                                                                                                                                         | No difference.  Proposed and<br>predicate devices are durable<br>medical equipment intended for<br>multiple uses                                                                                                                                                                                   |
| Sterility                                       | External contacting<br>device, nonsterile                                                                                                                                                                                                                                                                                                                                          | External contacting<br>device, nonsterile                                                                                                                                                                                                                                                                                                                                                  | No difference. Proposed and<br>predicate device are provided<br>nonsterile                                                                                                                                                                                                                         |
| Label Information                               | Labeled for OTC<br>(Home Use)                                                                                                                                                                                                                                                                                                                                                      | Labeled for OTC<br>(Home Use)                                                                                                                                                                                                                                                                                                                                                              | No difference, Label information<br>provided as required by US medical<br>device regulations for home use.                                                                                                                                                                                         |
| Feature                                         | Proposed Device<br>Omron HEM-<br>6400T-Z<br>K                                                                                                                                                                                                                                                                                                                                      | Predicate Device<br>Omron BP652N<br>K142917                                                                                                                                                                                                                                                                                                                                                | Analysis of Technological<br>Differences                                                                                                                                                                                                                                                           |
| Specifications/Features                         |                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                    |
| Measurement<br>method/Principle<br>of operation | Cuff oscillometric<br>method                                                                                                                                                                                                                                                                                                                                                       | Cuff oscillometric<br>method                                                                                                                                                                                                                                                                                                                                                               | No difference                                                                                                                                                                                                                                                                                      |
| Measurement<br>range                            | Pressure: 0 to<br>299mmHg<br>Pulse Rate: 30 to 199<br>beats/min.                                                                                                                                                                                                                                                                                                                   | Pressure: 0 to<br>299mmHg<br>Pulse Rate: 40 to 180<br>beats/min.                                                                                                                                                                                                                                                                                                                           | This minor difference in pulse rate<br>range does not impact safety and<br>effectiveness of the device.                                                                                                                                                                                            |
| Pressure senor                                  | Semiconductor<br>pressure sensor                                                                                                                                                                                                                                                                                                                                                   | Semiconductor<br>pressure sensor                                                                                                                                                                                                                                                                                                                                                           | No difference                                                                                                                                                                                                                                                                                      |
| Applicable cuff<br>(Wrist<br>Circumference)     | 15.0 to 20.5cm                                                                                                                                                                                                                                                                                                                                                                     | 13.5 to 21.5cm                                                                                                                                                                                                                                                                                                                                                                             | This minor difference does not<br>impact safety and effectiveness of<br>the device.                                                                                                                                                                                                                |
| Accuracy of<br>pressure<br>indicator            | Within ±3mmHg                                                                                                                                                                                                                                                                                                                                                                      | Within ±3mmHg or<br>2% of reading                                                                                                                                                                                                                                                                                                                                                          | There is no difference in the<br>accuracy range based on mmHg<br>scale                                                                                                                                                                                                                             |
| Accuracy of<br>pulse rate                       | Within ±5 % of<br>reading                                                                                                                                                                                                                                                                                                                                                          | Within ±5 % of<br>reading                                                                                                                                                                                                                                                                                                                                                                  | No difference                                                                                                                                                                                                                                                                                      |
| Inflation method                                | Automatic inflation<br>with piezoelectric<br>pump                                                                                                                                                                                                                                                                                                                                  | Automatic inflation<br>by piezoelectric pump                                                                                                                                                                                                                                                                                                                                               | No difference                                                                                                                                                                                                                                                                                      |
| Deflation<br>method                             | Automatic rapid<br>deflation valve                                                                                                                                                                                                                                                                                                                                                 | Automatic rapid<br>deflation valve                                                                                                                                                                                                                                                                                                                                                         | No difference                                                                                                                                                                                                                                                                                      |
| Display                                         | Organic<br>electroluminescent<br>display                                                                                                                                                                                                                                                                                                                                           | LCD digital display                                                                                                                                                                                                                                                                                                                                                                        | Similar. Both devices provide<br>digital display of BP and pulse.                                                                                                                                                                                                                                  |
| Power source                                    | Rechargeable<br>lithium-polymer<br>battery                                                                                                                                                                                                                                                                                                                                         | Two (2) "AAA"<br>batteries                                                                                                                                                                                                                                                                                                                                                                 | HEM-6400T-Z includes<br>rechargeable battery for user<br>convenience and reduced<br>environmental impact                                                                                                                                                                                           |
| Operating<br>conditions                         | 5 to 40 °C (41 to 104<br>°F)<br>15 to 85 %RH (non-<br>condensing)<br>800 to 1060 hPa                                                                                                                                                                                                                                                                                               | 10 to 40 °C (50 to<br>104 °F)<br>15 to 85 %RH                                                                                                                                                                                                                                                                                                                                              | Similar. HEM-6400T-Z operating<br>conditions specified to comply with<br>IEC60601-1 requirements.                                                                                                                                                                                                  |
| Charging/Data<br>transmission<br>conditions     | 10 to 35 °C (+50 to<br>95 °F)<br>15 to 85 %RH(non-<br>condensing)                                                                                                                                                                                                                                                                                                                  | NA                                                                                                                                                                                                                                                                                                                                                                                         | HEM-6400T-Z has specific battery<br>charging conditions for recharging<br>lithium-ion battery.                                                                                                                                                                                                     |
| Storage<br>conditions                           | -20 to 40 °C (-4 to<br>104 °F)<br>10 to 90 %RH(non-<br>condensing)                                                                                                                                                                                                                                                                                                                 | -20 to 60 °C (-4 to<br>140 °F)<br>10 to 95 %RH                                                                                                                                                                                                                                                                                                                                             | Similar. HEM-6400T-Z has specific<br>storage conditions since lithium-ion<br>battery requires specific storage<br>conditions.                                                                                                                                                                      |
| Transporting<br>conditions                      | -20 to 60 °C (-4 to<br>140 °F)<br>10 to 90 %RH(non-<br>condensing)                                                                                                                                                                                                                                                                                                                 | -20 to 60 °C (-4 to<br>140 °F)<br>10 to 95 %RH                                                                                                                                                                                                                                                                                                                                             | Similar. HEM-6400T-Z has specific<br>transporting conditions since<br>lithium-ion battery requires specific<br>transporting conditions.                                                                                                                                                            |
| Feature                                         | Proposed Device<br>Omron HEM-<br>6400T-Z<br>K                                                                                                                                                                                                                                                                                                                                      | Predicate Device<br>Omron BP652N<br>K142917                                                                                                                                                                                                                                                                                                                                                | Analysis of Technological<br>Differences                                                                                                                                                                                                                                                           |
| Dimensions<br>(mm)                              | 54 (W) × 63 (D) × 16<br>(H) mm                                                                                                                                                                                                                                                                                                                                                     | 89 (W) × 61 (D) × 13<br>(H) mm                                                                                                                                                                                                                                                                                                                                                             | Similar. HEM-6400T-Z has smaller<br>dimensions for added convenience<br>in home use.                                                                                                                                                                                                               |
| Weight                                          | Approx. 110g (3 7/8<br>oz.)<br>(including batteries)                                                                                                                                                                                                                                                                                                                               | Approx. 80g (2 7/8<br>oz.)<br>(not including<br>batteries)                                                                                                                                                                                                                                                                                                                                 | This minor difference does not<br>impact safety and effectiveness of<br>the device.                                                                                                                                                                                                                |
| Irregular Heart<br>Beat Feature                 | Yes                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                        | No difference                                                                                                                                                                                                                                                                                      |
| Body movement<br>detection                      | Yes                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                        | No difference                                                                                                                                                                                                                                                                                      |
| Hypertension<br>indicator                       | No                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                        | Hypertension Indicator not included<br>in HEM-6400T-Z. Device not<br>intended to make diagnosis of<br>hypertension. Instruction Manual<br>includes statement that only a<br>physician is qualified to diagnose<br>and treat high blood pressure.                                                   |
| Advanced<br>positioning<br>sensor (APS)         | Yes                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                        | No difference                                                                                                                                                                                                                                                                                      |
| Technology/Features Comparison                  |                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                    |
| Power supply                                    | Regulates power<br>voltage regardless of<br>battery voltage.                                                                                                                                                                                                                                                                                                                       | Regulates power<br>voltage regardless of<br>battery voltage.                                                                                                                                                                                                                                                                                                                               | No difference                                                                                                                                                                                                                                                                                      |
| Microprocessor                                  | • determines blood<br>pressure and pulse<br>rate<br>• controls the pump,<br>the valve, and the<br>display<br>• detects switch<br>operations<br>• stores<br>measurement results<br>• manages date and<br>time                                                                                                                                                                       | • determines blood<br>pressure and pulse<br>rate<br>• controls the pump,<br>the valve, and the<br>display<br>• detects switch<br>operations<br>• stores measurement<br>results<br>• manages date and<br>time                                                                                                                                                                               | No difference                                                                                                                                                                                                                                                                                      |
| Pressure sensor                                 | Semiconductor<br>pressure sensor.                                                                                                                                                                                                                                                                                                                                                  | Semiconductor<br>pressure sensor                                                                                                                                                                                                                                                                                                                                                           | No difference                                                                                                                                                                                                                                                                                      |
| Rapid exhaust<br>Valve                          | Active electronic<br>control valve that<br>performs cuff air<br>bleeding and release                                                                                                                                                                                                                                                                                               | Active electronic<br>control valve that<br>performs cuff air<br>bleeding and release                                                                                                                                                                                                                                                                                                       | No difference                                                                                                                                                                                                                                                                                      |
| Inflation source                                | Piezoelectric pump                                                                                                                                                                                                                                                                                                                                                                 | Piezoelectric pump                                                                                                                                                                                                                                                                                                                                                                         | No difference                                                                                                                                                                                                                                                                                      |
| Feature                                         | Proposed Device<br>Omron HEM-<br>6400T-Z<br>K                                                                                                                                                                                                                                                                                                                                      | Predicate Device<br>Omron BP652N<br>K142917                                                                                                                                                                                                                                                                                                                                                | Analysis of Technological<br>Differences                                                                                                                                                                                                                                                           |
| Display                                         | Organic<br>electroluminescent<br>display<br>• current cuff<br>pressure<br>• systolic blood<br>pressure<br>• diastolic blood<br>pressure<br>• pulse rate<br>• error messages<br>• time and date<br>• latest results in<br>the memory                                                                                                                                                | LCD (Liquid Crystal<br>Display) displays;<br>• current cuff<br>pressure<br>• systolic blood<br>pressure<br>• diastolic blood<br>pressure<br>• pulse rate<br>• error messages<br>• measurement<br>results in the<br>memory                                                                                                                                                                  | Different technology for digital<br>display. Similar information<br>displayed.                                                                                                                                                                                                                     |
| Controls                                        | • START/STOP<br>Button<br>• Connection Button<br>• Memory<br>• Home Button                                                                                                                                                                                                                                                                                                         | • START/STOP<br>Button<br>• Date/Time setting<br>Button<br>• Up/Down Button<br>• User ID Selections<br>Button                                                                                                                                                                                                                                                                              | Similar. Minor differences relate to<br>user convenience.                                                                                                                                                                                                                                          |
| Cuff                                            | Wrist Cuff,<br>Composed of<br>elastomer belt, air<br>bag and metal buckle                                                                                                                                                                                                                                                                                                          | Wrist Cuff,<br>Composed of cloth<br>bag, air bag and<br>hook-and-loop<br>fastener                                                                                                                                                                                                                                                                                                          | Similar. Difference in method to<br>apply device to wrist was evaluated<br>in usability study and found<br>acceptable.                                                                                                                                                                             |
| Materials                                       | Patient contact<br>materials of the cuff<br>have been tested in<br>accordance with ISO<br>10993 and FDA<br>guidance                                                                                                                                                                                                                                                                | Patient contact<br>materials of the cuff<br>have been tested in<br>accordance with ISO<br>10993 and FDA<br>guidance                                                                                                                                                                                                                                                                        | No difference                                                                                                                                                                                                                                                                                      |

Table 5.1: Substantial Equivalence Table

{7}------------------------------------------------

## Table 5.1: Substantial Equivalence Table (cont.)

{8}------------------------------------------------

Table 5.1: Substantial Equivalence Table (cont.)

{9}------------------------------------------------

Tablo 5 1: Substantial Equivolonco Tablo (cont )

### SUBSTANTIAL EQUIVALENCE

The indications for use statement for HEM-6400T-Z is identical to that of the predicate BP652N. Comparison testing demonstrated that the proposed device is equivalent to the predicate device with regard to measurement of blood pressure in a pulse wave generator test. Minor differences in technological features relate to convenience considerations for home use but do not impact safety or performance of blood pressure or pulse rate measurements. These minor differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness. Thus, HEM-6400T-Z is substantially equivalent to the predicate BP652N device.

{10}------------------------------------------------

## PERFORMANCE DATA [807.92(b)]

All necessary bench and clinical testing was conducted on HEM-6400T-Z to support a determination of substantial equivalence to the predicate device.

## Nonclinical Testing Summary:

The nonclinical, bench testing included:

- Comparative blood pressure and pulse rate testing to the predicate device
- Performance verification testing of HEM-6400T-Z to confirm acceptable performance of device features and functions
- Cleaning verification testing to confirm device retains its performance when cuff is exposed to surfactants as may be experienced in home use environment
- Usability testing with a representative population of study participants in a ● simulated home use environment

Other nonclinical safety testing included:

- . Biocompatibility of patient-contacting materials per ISO 10993-1 requirements
- . Electrical safety, electromagnetic compatibility, and electrostatic discharge testing
- Software verification and validation

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of HEM-6400T-Z meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that HEM-6400T-Z does not raise different questions of safety or effectiveness for measurement of blood pressure and pulse in a home use environment when compared to the predicate device.

# Clinical Testing Summary:

A clinical investigation was conducted with the objective of validating the accuracy of blood pressure measurements by HEM-6400T-Z based on an oscillometric method as compared to an auscultation method using a calibrated sphygmomanometer by trained medical staff. The results demonstrated that HEM-6400T-Z performed equivalently to the auscultation method.

### Conclusions

Based on the results from the nonclinical and clinical tests performed in support of HEM-6400T-Z, it is concluded that that the proposed device is safe, is effective, and performs at least as safely and effectively as the legally marketed predicate device.

# Summarv

HEM-6400T-Z and BP652N are designed for the measurement of blood pressure, pulse rate and detection of irregular pulses in adult population for home use. These devices have the same indications for use. Performance characteristics related to BP measurement and pulse rate are comparable between the two devices. The minor differences in labeling and technological characteristics between the proposed device and the predicate devices have been evaluated and determined to not raise any different issues of safety or effectiveness. As such, the proposed HEM-6400T-Z is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K163045](https://fda.innolitics.com/device/K163045)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
