The Bunegin-Albin Air Aspiration Set is intended to aspirate venous air emboli.
Device Story
The Bunegin-Albin Air Aspiration Set is a manual aspiration system used to remove venous air emboli. The set includes a radiopaque extruded polyethylene aspiration catheter (5.8 French, open distal end with 6 side ports), wire guide, entry access needle, dilator, syringe, stopcock, and scalpel. The device is used by a physician in a clinical setting, typically via the Seldinger technique under fluoroscopic guidance. The physician manually operates the syringe to create suction, aspirating air emboli through the catheter. This process removes air from the venous system, potentially preventing or mitigating complications associated with venous air embolism. The device is for single-use and sterilized by ethylene oxide.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing and an acute animal study. Bench testing included tensile strength, air aspiration, liquid leakage, kink resistance, and wire guide integrity (corrosion, flex, fracture) per BS EN ISO 10555-1, BS EN 13868, and BS EN ISO 11070. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility) was performed per ISO 10993-1. Animal study confirmed the catheter's ability to remove air from target sites with 'adequate' or 'good' performance ratings.
Technological Characteristics
Radiopaque extruded polyethylene catheter (5.8F); stainless steel wire guide and needle; polyethylene dilator. Manual aspiration principle via syringe. Sterilized by ethylene oxide (EtO). Testing standards: BS EN ISO 10555-1 (catheter), BS EN ISO 11070 (wire guide/dilator), BS EN 13868 (kink), ISO 594-2 (luer connections), USP 788 (particulates).
Indications for Use
Indicated for the aspiration of venous air emboli in patients requiring such intervention.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized human profiles stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 27, 2017
Cook Incorporated Johnathan Liu Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404
Re: K163031
Trade/Device Name: Bunegin-Albin Air Aspiration Set Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: June 24, 2017 Received: June 27, 2017
Dear Johnathan Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
2
510(k) Number (if known)
K163031
Device Name Bunegin-Albin Air Aspiration Set
Indications for Use (Describe)
The Bunegin-Albin Air Aspiration Set is intended to aspirate venous air emboli.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font. The logo is simple and clean, and the red and white color scheme is eye-catching.
COOK INCORPORATED 750 DANIELS WAY, P.O. BOX OMINGTON, IN 474
# 2.0 510(k) SUMMARY
#### K163031 Bunegin-Albin Air Aspiration Set 21 CFR §870.5150 Date Prepared: July 25, 2017
#### Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|-----------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact: | Johnathan Liu |
| Email: | regsubmissions@cookmedical.com |
| Contact Phone Number: | (812) 335-3575 x104509 |
| Contact Fax Number: | (812) 332-0281 |
### Device Information:
| Trade Name: | Bunegin-Albin Air Aspiration Set |
|-----------------------|----------------------------------|
| Common Name: | Catheter, Embolectomy |
| Classification Name: | Embolectomy Catheter |
| Regulation: | 21 CFR §870.5150 |
| Product Code: | DXE |
| Device Class: | II |
| Classification Panel: | Cardiovascular |
### Predicate Device:
The Bunegin-Albin Air Aspiration Set is substantially equivalent to the following device: the Export Advance" Aspiration Catheter (K130536, Medtronic Inc.) cleared on July 16, 2013.
### Device Description:
The Bunegin-Albin Air Aspiration Set is comprised of an air aspiration catheter, wire guide, entry access needle, dilator, syringe, stopcock, and scalpel. The air aspiration catheter is manufactured from radiopaque extruded polyethylene tubing and is designed
{4}------------------------------------------------
with a pre-molded winged fitting. The catheter is 5.8 French in size with an endhole diameter of 0.035 inches. The wire guide is manufactured from stainless steel coils, a safety wire, and a mandril wire. The access needle is manufactured from stainless steel. The dilator is manufactured from polyethylene and has a pre-molded proximal hub. The set also includes a 12 cc syringe, a low pressure three way stopcock, and a scalpel. The Bunegin-Albin Air Aspiration Set is sterilized by ethylene oxide and intended for onetime use.
#### Indications for Use:
The Bunegin-Albin Air Aspiration Set is intended to aspirate venous air emboli.
#### Comparison to Predicate Device:
The Bunegin-Albin Air Aspiration Set and the predicate device, the Export Advance™ Aspiration Catheter (K130536), are substantially equivalent in that these devices are similar in intended use and principle of operation. Additionally, the subject device has similar indications for use, materials, and dimensions as the predicate device. The differences between the subject device and the predicate device, including the materials, dimensions, and indications for use do not raise any new issues of safety and effectiveness. The substantial equivalence comparison of the subject device to the predicate is provided in the table below.
{5}------------------------------------------------
| | PREDICATE DEVICE | SUBJECT DEVICE |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| | Export AdvanceTM Aspiration Catheter | Bunegin-Albin Air Aspiration Set |
| Manufacturer | Medtronic Incorporated | Cook Incorporated |
| 510(k) | K130536 | Subject of this submission |
| Regulation | 21 CFR §870.5150 | Identical |
| Product Code | DXE | Identical |
| Regulation<br>Description | Catheter, Embolectomy | Identical |
| Classification | II | Identical |
| Indications for<br>Use | The Export AdvanceTM Aspiration Catheter<br>is indicated for:<br>- Removal/aspiration of embolic<br>material (thrombus/debris) from<br>vessels of the arterial system, and<br>- To sub selectively infuse/deliver<br>diagnostics or therapeutics agents<br>with or without vessel occlusion. | To aspirate venous air emboli. |
| Principle of<br>Operation | Manually aspirate emboli with syringe | Identical |
| Visualization<br>Guidance | Fluoroscopy | Identical |
| Method of Use | Single use | Identical |
| Catheter<br>Placement | Percutaneously using Seldinger technique<br>under fluoroscopy | Identical |
| Catheter<br>Material | Polytetrafluoroethylene (PTFE) | Radiopaque extruded polyethylene |
| Catheter Outer<br>Diameter | 6 French | 5.8 French |
| Catheter<br>Length | 140 cm | 24, 60 cm |
| Catheter Distal<br>End | Open | Open with 6 side ports |
| Catheter<br>Lumens | Dual lumen | Single lumen |
| Wire Guide<br>Compatibility | 0.014 inch, 0.035 inch | 0.035 inch |
| Accessory Set<br>Components | 2 Syringes<br>1 Wire Guide<br>1 Filter Cup | 1 Syringe<br>1 Wire Guide<br>1 Needle<br>1 Dilator<br>1 Stopcock<br>1 Scalpel |
| Sterilization<br>Method | EtO | Identical |
| Sterility<br>Assurance<br>Level | 10-6 | Identical |
| Packaging | Tyvek pouch | Tyvek Peel Pouch |
{6}------------------------------------------------
# Technological Characteristics:
The subject Bunegin-Albin Air Aspiration Set was subjected to applicable testing to assure reliable design and performance under the testing parameters. The tests are listed below:
# Performance
- Catheter Shaft Tensile (Zero time) The peak load of catheter shaft section shall be . greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
- . Catheter Sideport Tensile (Zero time) – The peak load of catheter sideport section shall be greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
- Catheter Hub-to-Shaft Tensile (Zero time) The peak load of catheter hub-to-shaft . section shall be greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
- Catheter Air Aspiration (Zero time) No air shall enter the hub when tested in . accordance with Annex D of BS EN ISO 10555-1. The acceptance criterion was met.
- Catheter Liquid Leakage (Zero time) No part of the catheter shall leak liquid when . tested in accordance with Annex C of BS EN ISO 10555-1. The acceptance criterion was met.
- . Catheter Kink Length (Zero time) - Catheter shall not kink when tested in accordance with Annex B of BS EN 13868. The acceptance criterion was met.
- . Wire Guide Corrosion (Zero time) – Wire guide shall show no evidence of corrosion that could affect the functional performance when tested in accordance with Annex B of BS EN ISO 11070. The acceptance criterion was met.
- . Wire Guide Flex (Zero time) – Wire guide shall show no signs of defects or damage, including flaking or material loss when tested in accordance with Annex G of BS EN ISO 11070. The acceptance criterion was met.
- Wire Guide Fracture (Zero time) Wire guide shall not fracture when wound around . an approximate former in accordance with Annex F of BS EN ISO 11070. The acceptance criterion was met.
- . Wire Guide Tensile (Zero time) – The peak load to failure shall be greater than or equal to 10 N in accordance with Annex H of BS EN ISO 11070. The acceptance criterion was met.
- Wire Guide Rotations to Failure Testing (Zero time) The number of rotations to . failure shall be characterized for the wire guide.
{7}------------------------------------------------
- Wire Guide Tip Deflection (Zero time) The tip deflection side loads will be . characterized for the wire guide.
- . Dilator Shaft Tensile (Zero time) – the peak load of dilator shaft section shall be greater than or equal to 15 N in accordance with BS EN ISO 11070. The acceptance criterion was met.
- Dilator Hub-to-Shaft Tensile (Zero time) The peak load of dilator hub-to-shaft . section shall be greater than or equal to 15 N in accordance with BS EN ISO 11070. The acceptance criterion was met.
- . Dimensional, Compatibility, and Surface Analysis (Zero-time) - All measurements and dimensional requirements shall be within the listed tolerances. The set components shall be compatible for each test specimen. The external surface of the effective length of the catheter, dilator, and wire guide shall appear free of extraneous matter for each test specimen. The acceptance criterion was met.
- Particulate Testing (Zero-time) Each device must meet the USP 788 thresholds for . small volume injections. The acceptance criterion was met.
- . Resistance to Overriding Testing (Zero-time) – Testing was conducted in accordance with ISO 594-2. The acceptance criterion was met.
- . Separation Force Testing (Zero-time) - Testing was conducted in accordance with ISO 594-2. The acceptance criterion was met.
- Unscrewing Torque Testing (Zero-time) Testing was conducted in accordance with . ISO 594-2. The acceptance criterion was met.
- Catheter Shaft Tensile (3-year accelerated aging) The peak load of catheter shaft . section shall be greater than or equal to 15 N in accordance with BS EN ISO 1055-1. The acceptance criterion was met.
- Catheter Sideport Tensile (3-year accelerated aging) The peak load of catheter . sideport section shall be greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
- . Catheter Hub-to-Shaft Tensile (3-year accelerated aging) - The peak load of catheter hub-to-shaft section shall be greater than or equal to 15 N in accordance with BS EN ISO 10555-1. The acceptance criterion was met.
- Catheter Air Aspiration (3-year accelerated aging) No air shall enter the hub when . tested in accordance with Annex D of BS EN ISO 10555-1. The acceptance criterion was met.
{8}------------------------------------------------
- . Catheter Liquid Leakage (3-year accelerated aging) - No part of the catheter shall leak liquid when tested in accordance with Annex C of BS EN ISO 10555-1. The acceptance criterion was met.
- . Catheter Kink Length (3-year accelerated aging) - Catheter shall not kink (based on a 20 mm kink radius) when tested in accordance with Annex B of BS EN 13868. The acceptance criterion was met.
- Dilator Shaft Tensile (3-year accelerated aging) The peak load of dilator shaft . section shall be greater than or equal to 15 N in accordance with BS EN ISO 11070. The acceptance criterion was met.
- Dilator Hub-to-Shaft Tensile (3-year accelerated aging) The peak load of dilator . hub-to-shaft section shall be greater than or equal to 15 N in accordance with BS EN ISO 11070. The acceptance criterion was met.
# Animal Testing
- Acute performance of aspiration catheter in animal model Catheter shall receive a grading of "adequate" or "good" in each of the performance parameters, and it shall be able to remove air from the target sites. The acceptance criterion was met.
# Biocompatibility
- Per ISO 10993-1 and FDA guidance, testing for cytotoxicity, sensitization, . intracutaneous irritation, acute systemic toxicity, material-mediated pyrogenicity, and hemocompatibility (hemolysis, complement activation, and in vivo thrombogenicity) were performed to ensure the biocompatibility of the subject device set. The acceptance criterion was met.
### Conclusion:
The results of these tests support a conclusion that the Bunegin-Albin Air Aspiration Set met the design input requirements based on the intended use and support the conclusion that this device does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate device, the Export Advance™ Aspiration Catheter (Medtronic Inc., K130536).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.