K163019 · Careray Digital Medical System Co., Ltd. · MQB · Dec 1, 2016 · Radiology
Device Facts
Record ID
K163019
Device Name
CareView 750Cw/750C X-ray Flat Panel Detectors
Applicant
Careray Digital Medical System Co., Ltd.
Product Code
MQB · Radiology
Decision Date
Dec 1, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The CareView 750Cw/750C detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This product is not intended for mammography applications.
Device Story
CareView 750Cw/750C is a portable digital X-ray flat panel detector; captures X-ray photons via CsI scintillator; converts to visible light then electrical signals via a-Si thin film transistor array; digitizes signals for display on monitor. Device integrates with operating PC and X-ray generator; supports wired (Gigabit Ethernet) and wireless (IEEE 802.11a/b/g/n) communication. Used in clinical settings by healthcare professionals for general radiography. Detector does not provide feedback to generator for exposure termination. Output facilitates diagnostic imaging, aiding clinical decision-making by replacing traditional film/screen systems.
Clinical Evidence
Bench testing only. Performance evaluated per IEC 61223-3-1, IEC 6220-1, and FDA's Solid State X-ray Imaging (SSXI) guidance. Metrics included Offset Stability, Dark Noise, Spatial Resolution, Low-contrast Resolution, Dynamic Range, Sensitivity, Lag, Ghost, MTF, and DQE. Electrical safety and EMC testing conducted per IEC/ES 60601-1 and IEC/EN 60601-1-2. All results were satisfactory.
Technological Characteristics
CsI scintillator; a-Si TFT readout; 120μm pixel pitch; 2048 x 2560 image matrix; 16-bit grayscale; carbon fiber plate; air cooling. Connectivity: Gigabit Ethernet, IEEE 802.11a/b/g/n. Power: DC +24V. Dimensions: 282mm x 333mm x 15mm. Software: API V4.5.0 (NDT V3.5.5). Standards: IEC 60601-1, IEC 60601-1-2, ANSI C63.4-2003, FCC Part 15 Subpart C.
Indications for Use
Indicated for general radiographic diagnosis of human anatomy in patients requiring general-purpose diagnostic procedures. Not intended for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2016S
CareRay Digital Medical System Co., Ltd. % Mr. Xu Wei Manager B3-501, Biobay, 218 Xinghu Street SuZhou Industrial Park SuZhou, Jiangsu 215123 P. R. CHINA
Re: K163019
Trade/Device Name: CareView 750Cw/750C X-ray Flat Panel Detectors Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: July 1, 2016 Received: November 3, 2016
Dear Mr. Wei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K163019
#### Device Name
CareView 750Cw/750C X-ray Flat Panel Detectors
Indications for Use (Describe)
The CareView 750Cw/750C detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This product is not intended for mammography applications.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <svg height="16" width="16"> <polygon points="0,0 0,16 16,16 16,0" style="fill:none;stroke:black;stroke-width:1"></polygon> <polygon points="2,2 2,14 14,14 14,2" style="fill:none;stroke:black;stroke-width:1"></polygon> <polygon points="4,4 4,12 12,12 12,4" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg height="16" width="16"> <polygon points="0,0 0,16 16,16 16,0" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> </span> Over The Counter Use (21 CFR 801 Sub |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
|_ | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
{4}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR807.92 (a) (1)]
Aug.01, 2016
### 2. Submitter's Information [21 CFR807.92 (a) (1)]
| Company Name: | CareRay Digital Medical System Co., Ltd. |
|------------------|-------------------------------------------------------------------------------------------------|
| Company Address: | A2-201/B3-501, Biobay, 218 Xinghu Street, SuZhou<br>Industrial Park, SuZhou 215123, P. R. China |
| Contact Person: | Mr. Xu |
| Phone Number: | (86) 512-86860288 |
| Fax Number: | (86) 512-86860388 |
| E-mail: | wei.xu@careray.com |
#### 3. Trade Name, Common Name, Classification [21 CFR807.92(a)(2)]
| Trade Name: | CareView750Cw/750C X-ray Flat Panel Detectors |
|----------------------|-----------------------------------------------------------|
| | (hereinafter: CareView750Cw/750C or |
| | CareView750Cw/750C detector , apply to the all submission |
| | documents ) |
| Model Name: | CareView 750Cw/750C |
| Classification Name: | Stationary X-ray system |
| Regulation Number: | 21 CFR 892.1680 |
| Regulatory Class: | Class II |
| Product Code: | MQB |
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicates within this submission are as follows: Manufacturer: CareRay Digital Medical System Co., Ltd. Trade Name: X-ray Flat Panel Detectors
{5}------------------------------------------------
| Model Name: | CareView 1500Cw |
|----------------------|-------------------------|
| Classification Name: | Stationary X-ray system |
| Regulation Number: | 21 CFR 892.1680 |
| Regulatory Class: | Class II |
| Product Code: | MQB |
| FDA 510(k) #: | K150929 |
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
CareView 750Cw/750C is a kind of wireless(750C does not have wireless, it has only wired fuction.) portable digital X-ray flat panel detectors which have 244 mm x 307 mm imaging area. The device communicates by not only the wireless communication but also wired communication feature (Giga-bit Ethernet communication mode by connecting the power box) optionally.
The device intercepts X-ray photons and then the scintillator emits visible spectrum photons that illuminate an array of photo detectors (a-Si) that create electrical signals. After the electrical signals are generated, it is converted to a digital value and an image will be displayed on the monitor.
The detector should be integrated with an operating PC and an X-ray generator to utilize as digitalizing X-ray images and transfer for radiography diagnostic.
The detector can't provide feedback to the generator to terminate the x-ray exposure.
System Software Version is API V4.5.0 (NDT V3.5.5) that based on V4.2.0. But V4.5.0 is more powerful than old version. Please see the comparison of section 7 Technological Characteristics.
### 6. Intended Use [21 CFR 807.92(a)(5)]
The CareView 750Cw/750C detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographicfilm/screen systems in all general-purpose diagnostic procedures.This product is not intended for mammography applications.
### 7. Technological Characteristics [21 CFR 807.92(a)(6)]
| Item | Proposed Device:<br>CareView750Cw/750C X-ray Flat<br>Panel Detectors | Predicate Device:<br>CareView 1500Cw X-ray Flat<br>Panel Detectors |
|---------------|----------------------------------------------------------------------|--------------------------------------------------------------------|
| 510(K) Number | K163019 | K150929 |
{6}------------------------------------------------
| Model | CareView 750Cw/750C | CareView 1500Cw | | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name | Stationary X-ray system | Stationary X-ray system | | |
| Product Code | MQB | MQB | | |
| Regulation Number | 892.1680 | 892.1680 | | |
| Panel | Radiology | Radiology | | |
| Class | II | II | | |
| X-ray Absorber | Csl Scintillator | Csl Scintillator | | |
| InstallationType | Wireless (except 750C) ,Wired, Portable | Wireless, Wired, Portable | | |
| ReadoutMechanism | Thin Film Transistor | Thin Film Transistor | | |
| Image Matrix Size | 2048 x 2560 pixels | 2304 x 2816 pixels | | |
| Pixel Pitch | 120μm | 154μm | | |
| Effective Imaging Area | 244 mm x 307 mm | 355 mm x 434 mm | | |
| Grayscale | 16 bit, 65536 grayscale | 16 bit, 65536 grayscale | | |
| Spatial Resolution | Min. 4.2 line pair/mm | Min. 3.3 line pair/mm | | |
| Rated Power Supply Wired Wireless | DC +24 V, Max. 1.3A<br>Powered by the switching power supply using interface cable<br>Powered by the battery pack(except 750C) | DC +24 V, Max.1.5 A<br>Powered by the power box using interface cable<br>Powered by the battery pack | | |
| Power Consumption | Max. 31.2 W | Max. 36 W | | |
| Communications Wired Wireless | Gigabit Ethernet<br>IEEE 802.11a/b/g/n (2.4 / 5 GHz) | Gigabit Ethernet<br>IEEE 802.11a/b/g/n (2.4 / 5 GHz) | | |
| Imaging Plate | Carbon Fiber Plate | Carbon Fiber Plate | | |
| Cooling | Air cooling | Air cooling | | |
| Dimensions | 282mm x 333mm x 15mm | 384 mm x 460 mm x 15 mm | | |
| Operation | Temperature: +5 ~ +35°C<br>Humidity: 30 ~ 75% (Non-Condensing)<br>Atmospheric pressure: 70 ~ 106 kPa<br>Altitude: Max.3000 meters | Temperature: +5 ~ +35°C<br>Humidity: 30 ~ 75% (Non-Condensing)<br>Atmospheric pressure: 70 ~ 106 kPa<br>Altitude: Max.3000 meters | | |
| | Storage and transportation | Temperature: -20 ~ +55°C<br>Humidity: 10 ~ 90% (Non-Condensing)<br>Atmospheric pressure: 70 ~ 106 kPa<br>Altitude: Max.3000 meters | Temperature: -20 ~ +55°C<br>Humidity: 10 ~ 90% (Non-Condensing)<br>Atmospheric pressure: 70 ~ 106 kPa<br>Altitude: Max.3000 meters | |
| | | Intended Use | The CareView 750Cw/750C detector is indicated for digital imaging solution designed for providing general radiographic | The CareView 1500Cw detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of |
| | | | | diagnosis of human anatomy. It is<br>intended to replace radiographic<br>film/screen systems in all<br>general-purpose diagnostic<br>procedures. This product is not<br>intended for mammography<br>applications. |
| MTF | | | ~70%..........(@ 1lp/mm)<br>~40%..........(@ 2lp/mm)<br>~22%..........(@ 3lp/mm) | ~70%..........(@ 1lp/mm)<br>~40%..........(@ 2lp/mm)<br>~22%..........(@ 3lp/mm) |
| DQE | (@RQA5, 30 $\mu$ Gy)<br>~65%..........(@ 0lp/mm)<br>~20%..........(@ 3lp/mm) | | (@RQA5, 30 $\mu$ Gy)<br>~65%..........(@ 0lp/mm)<br>~20%..........(@ 3lp/mm) | |
| Dynamic range | ~82 dB | ~82 dB | | |
| System Software<br>Version | API V4.5.0 (NDT V3.5.5)<br><br>Based on V4.2.0, V4.5.0 includes<br>the following new enhancements:<br>> CarerayTouch handswitch<br>> Resume broken downloads<br>> One-shot calibration<br>> Small-window AED | API V4.2.0 (NDT V3.4.0) | | |
{7}------------------------------------------------
## 8. System requirements to operate with other radiographic system components
- 1) Recommended Generator Specification:
Energy range: 40~150kVp
mA range: 10~1000mA(depending on the generator power)
ms range: 10~6300ms to produce 0.1~1000mAs (depending on the generator power) Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators, please contact your distributor or CareRay.
- 2) Application Program Interface (API) for system integration manufacturer
Peripheral hardware: CareView detector connected via wired communication.
CPU: Intel (R) Core (TM) 2 Duo, 2.93GHz or above
RAM: 2 GB or higher
Hard disk: 160 GB or higher
Monitor: 1280 x 1024 or higher
OS: Windows XP or Windows 7
Development environment: MS Visual Studio 2005
- 3) X-ray exposure mode
{8}------------------------------------------------
The synchronous connection mode is the signal transfer modebetween the X-ray generator which sends the X-ray and the detector which receives the X-ray.
CareView 750Cw/750C supported typical sync mode contains external sync, soft sync, manual sync and auto sync containing FFAED mode.
The detector can't provide feedback to the generator to terminate the x-ray exposure.
## 9. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92(b)(2)]
- A Electrical safety and EMC testing
Electrical, mechanical, environmental safety and performance testing according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC/EN 60601-1-2. All test results are satisfactory.
- A Performance Testing
The performance testing items included Offset Stability, Dark Noise, Spatial Resolution, Low-contrast Resolution, Dynamic Range&Sensitivity, Lag, Ghost, MTF and DQE etc. These testing conducted in accordance with IEC61223-3-1, IEC 6220-1 and FDA's SSXI guidance document (Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices). The RF module that provided by supplier is in compliance with ANSI C63.4-2003 and 47 CFR FCC Part 15 Subpart C and get FCC ID.
- A Nonclinical and clinical considerations
The proposed device (CareView 750Cw/750C) and predicate device (CareView 1500Cw) share most of primary product specifications including intended use, technology, material, and imaging principle, etc. Difference lies in the mechanical structure.
The differences of mechanical structure don't affect the technological parameters and clinical images.
#### 10. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification,theCareView750Cw/750CX-ray flat panel detector is substantially equivalent to predicate deviceCareView1500Cw (K150929). Both propose and predicate devices are same in the intended use, the design principle, the applicable standards and specification. Some characteristics, for example,the mechanical structure are different. However the test reports in this submission documents
{9}------------------------------------------------
provide demonstration that these differences do not raise any new questions of safety an deffectiveness. Therefore, CareRay Digital Medical System Co., Ltd. concludes the CareView750Cw/750C X-ray flat panel detector is substantially equivalent with the predicate device CareView1500Cw (K150929).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.