← Product Code [DRF](/productcode/DRF) · K162892

# Achieve Advance Mapping Catheter (K162892)

_Medtronic, Inc. · DRF · Nov 15, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162892

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [DRF](/productcode/DRF.md)
- **Decision Date:** Nov 15, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The intended use of the Achieve Advance Mapping Catheter is to record intracardiac signals and provide cardiac stimulation during electrophysiology studies.

## Device Story

Intracardiac electrophysiology diagnostic catheter; circular loop distal section with electrodes; used for recording electrical signals and cardiac stimulation in atrial regions. Operated by physicians in clinical electrophysiology settings. Device inserted via delivery catheter; electrodes capture intracardiac electrograms; signals transmitted to external recording/stimulation equipment. Provides electrophysiological data to assist clinicians in mapping cardiac structures and diagnosing arrhythmias. Benefits include precise atrial mapping and stimulation capabilities.

## Clinical Evidence

Bench testing only. No clinical data. Testing included thermal conditioning, packaging distribution (ASTM 4332/D4169), radial loop compliance, distal stiffness, axial load, torque response, turns to failure, electrical continuity, device functionality, kink resistance, joint strength (ISO 10555-1), stiffness, radiopacity (ASTM F640), and life testing.

## Technological Characteristics

Intracardiac mapping catheter; 3.3F diameter; 146cm length. Distal circular loop (15mm/20mm) with 8 electrodes. Materials: Nitinol, Pebax, Stainless Steel, polyimide, PET. Standards: ASTM F640 (radiopacity), ISO 10555-1 (joint strength), ASTM D4169/4332 (packaging/thermal). Connectivity: Wired to external recording/stimulation system. Non-powered/passive sensing.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Achieve ST Mapping Catheter ([K153139](/device/K153139.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures in profile.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 15, 2016

Medtronic, Inc. Heather Taylor Principal Regulatory Affairs Specialist 8200 Coral Sea Street NE Mail Stop MVS46 Mounds View, Minnesota 55112

Re: K162892

Trade/Device Name: Achieve Advance Mapping Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: DRF Dated: October 14, 2016 Received: October 17, 2016

Dear Heather Taylor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely

Minda Jellison
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162892

Device Name Achieve Advance Mapping Catheter

Indications for Use (Describe)

The Achieve Advance mapping catheter is indicated for multiple electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The Achieve Advance mapping catheter is designed to obtain electrograms in the atrial regions of the heart.

| Type of Use (Select one or both, as applicable)             |                                                            |
|-------------------------------------------------------------|------------------------------------------------------------|
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

| 510(k) Summary per 21 CFR §807.92 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |  |  |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Date Summary Prepared:            | 26 Sep 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |  |  |
| Applicant:                        | Medtronic Inc.<br>710 Medtronic Parkway<br>Minneapolis, MN 55432<br>Establishment Registration No.: 3001504994                                                                                                                                                                                                                                                                                                                                                                           |  |  |  |
|                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |  |  |
| Contact Person:                   | Heather Taylor<br>Principal Regulatory Affairs Specialist<br>Medtronic Inc.<br>Coral Sea Street<br>Mounds View, MN  55112<br>Telephone: 763.526.9066<br>Fax: 763.367.9903<br>Email: heather.m.taylor@medtronic.com                                                                                                                                                                                                                                                                       |  |  |  |
| Trade Name:                       | Achieve Advance™ Mapping Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |  |  |
| Common Name:                      | Catheter, electrode recording, or probe, electrode<br>recording                                                                                                                                                                                                                                                                                                                                                                                                                          |  |  |  |
| Classification Name:              | Electrode Recording Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  |  |  |
| Classification & Panel:           | Class II, 21 CFR 870.1220                                                                                                                                                                                                                                                                                                                                                                                                                                                                |  |  |  |
| Product Code:                     | DRF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |  |  |
| Predicate Device(s):              | Achieve™ ST Mapping Catheter (K153139)                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |  |  |
| Device Description:               | The Achieve Advance Mapping diagnostic catheter is an intra-<br>cardiac electrophysiology recording catheter and can be used for<br>cardiac stimulation during electrophysiology studies. The distal<br>mapping section of the Achieve Advance Mapping Catheter is a<br>circular loop with evenly spaced electrodes to map electrical<br>conduction within the atrium. The Achieve Advance Mapping<br>Catheter is available in a 15mm, 20mm, or 25mm pre-shaped distal<br>loop diameter. |  |  |  |

{4}------------------------------------------------

| Intended Use of Device: | The intended use of the Achieve Advance Mapping Catheter is to<br>record intracardiac signals and provide cardiac stimulation during<br>electrophysiology studies. |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|

The Achieve Advance mapping catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The Achieve Advance mapping catheter is designed to obtain electrograms in the atrial regions of the heart.

Comparison of Technological The Achieve Advance Mapping Catheter uses the same technology, Characteristics and has the same intended use, functions, materials and method of operation as the following predicated device:

|                         | Achieve Advance<br>Catheter - Line<br>Extension (Subject<br>Device)                                                                                                                                                                                                                                 | Achieve ST Catheter<br>(predicate device, now<br>named Achieve<br>Advance Catheter)<br>K153139                                                                                                                                                                                   |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use            | Map intracardiac<br>structures of the<br>heart                                                                                                                                                                                                                                                      | Map intracardiac<br>structures of the heart                                                                                                                                                                                                                                      |
| Indications for<br>Use  | The Achieve<br>Advance is<br>indicated for<br>multiple electrode<br>electrophysiological<br>mapping of cardiac<br>structures (i.e.,<br>recording and<br>stimulation only).<br>The Achieve<br>Advance Catheter<br>is designed to<br>obtain<br>electrograms in the<br>atrial regions of the<br>heart. | The Achieve ST is<br>indicated for multiple<br>electrode<br>electrophysiological<br>mapping of cardiac<br>structures (i.e.,<br>recording and<br>stimulation only). The<br>Achieve ST Catheter is<br>designed to obtain<br>electrograms in the<br>atrial regions of the<br>heart. |
| Catheter Body<br>Tubing | Nitinol/Pebax<br>(distal body)<br>Stainless Steel/<br>polyimide<br>(proximal)                                                                                                                                                                                                                       | Nitinol/Pebax (distal<br>body)<br>Stainless Steel/<br>polyimide (proximal)                                                                                                                                                                                                       |
| Diameter                | 3.3F                                                                                                                                                                                                                                                                                                | 3.3F                                                                                                                                                                                                                                                                             |

{5}------------------------------------------------

| Effective Length                      | 146cm                                      | 146cm                                      |
|---------------------------------------|--------------------------------------------|--------------------------------------------|
| Number of<br>Electrodes               | 8 (15mm & 20mm)                            | 10 (25mm)                                  |
| Electrode Spacing                     | 5.8mm (20mm)<br>4.1mm (15mm)               | 5.8mm (25mm)                               |
| Distal End Shape                      | Circular Loop                              | Circular Loop                              |
| Loop Diameter                         | 15mm and 20mm                              | 25mm                                       |
| Loop Material                         | Nitinol insulated with PET (Pebax covered) | Nitinol insulated with PET (Pebax covered) |
| Delivered through a delivery catheter | Yes                                        | Yes                                        |

Performance Data In vitro bench testing was completed to support the line extension of the Achieve Advance Mapping Catheter. Performance testing verified that the 15mm and 20mm Achieve Advance Mapping Catheter performs as designed and is suitable for its intended use.

> Performance testing included the following: Thermal Conditioning & Packaging Distribution TT0224 (per ASTM 4332 and D4169) Radial Loop Compliance Distal Stiffness Axial Load Torque Response Turns to Failure Electrical Continuity Device Functionality Kink Resistance Joint Strength (per 10555-1) Stiffness Radiopacity (per ASTM F640) Life Testing

## Conclusion

The data presented in this submission demonstrate that the Achieve Advance Mapping Catheter is substantially equivalent to the predicate device identified in regards to device design, materials, and intended use.

---

**Source:** [https://fda.innolitics.com/device/K162892](https://fda.innolitics.com/device/K162892)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
