Abutments are placed into the dental implants to provide support for prosthetic restoration such as crowns, bridges or overdentures. The Straumann CARES Golden Ti/TiN Abutments are indicated for single tooth replacements and multiple tooth restorations. The prosthetic restoration is cement-retained.
Device Story
Patient-specific dental abutments; designed via traditional waxup/scanning or intraoral scanning; CAD software used for design; design data transmitted via internet portal to manufacturer; manufacturer verifies design against parameters (max 30° angulation, minimum surface area requirements). Manufactured from solid grade 4 titanium; emergence profile and coronal aspect coated with titanium nitride (TiN) for esthetics; implant-to-abutment interface masked during coating. Delivered to dental laboratory for final processing; attached to implant with titanium alloy basal screw; occlusal restoration affixed with dental cement. Used in dental clinics; operated by dental professionals. Benefits include improved esthetics of finished restoration while maintaining mechanical function of titanium abutment.
Clinical Evidence
Bench testing only. No clinical data. Testing included sterilization validation (ISO 17665-1/2), biocompatibility (ISO 10993-1/5/18), dynamic fatigue (ISO 14801), and coating characterization (composition, thickness, porosity, SEM). Adhesion testing performed per ASTM F1044 and F1160. Abrasion testing performed per Tamura et al. methodology.
Technological Characteristics
Substrate: Commercially pure grade 4 titanium (ISO 5832-2). Surface: Titanium nitride (TiN) coating on emergence profile and coronal aspect. Interface: Masked/uncoated. Form factor: One-piece solid abutment. Sterilization: Moist heat (autoclave) to SAL 10^-6. Connectivity: Internet portal for design submission.
Indications for Use
Indicated for patients requiring single tooth replacements or multiple tooth restorations supported by dental implants. Prosthetic restorations must be cement-retained.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or a bird in flight, composed of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2017
Institut Straumann AG % Jennifer Jackson Director of Regulatory Affairs and Quality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K162848
Trade/Device Name: Straumann® CARES® Golden Ti/TiN Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: August 25, 2017 Received: August 28, 2017
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mary S. Runner -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
## K162848
Device Name
Straumann® CARES® Golden Ti/TiN Abutments
Indications for Use (Describe)
Abutments are placed into the dental implants to provide support for prosthetic restoration such as crowns, bridges or overdentures.
The Straumann CARES Golden TiTiN Abutments are indicated for single tooth replacements and multiple tooth restorations. The prosthetic restoration is cement-retained.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 5. | 510(k) Summary | K162848 |
|----|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810 |
| | Contact Person: | Jennifer M. Jackson, MS<br>Director, Regulatory Affairs and Quality<br>+1 (978) 747-2509 |
| | Prepared By &<br>Secondary Contact: | Christopher Klaczyk<br>Head of Regulatory Affairs & Compliance<br>Institut Straumann AG<br>+41 61 965 1260 |
| | Date Prepared: | September 22, 2017 |
| | Product Code(s): | NHA (21 CFR 872.3630) |
| | Device Class: | II (21 CFR 872.3630) |
| | Classification Panel: | Dental |
| | Classification Name: | Endosseous dental implant abutment (21 CFR 872.3630) |
| | Proprietary Name: | Straumann® CARES® Golden Ti/TiN Abutment |
| | Primary Predicate<br>Device: | K082764, Straumann CARES WN Ti Abutments (Institut Straumann) |
| | Reference Device(s): | K052272, Straumann CARES RN Ti Abutments (Institut Straumann)<br>K072151, Straumann CARES RC Ti Abutments (Institut Straumann)<br>K081005, Straumann CARES NC Ti Abutments (Institut Straumann)<br>K992334, 3i TiN Coated Implants and Abutments (Implants Innovations, Incorporated (3i)) |
| | Device Description: | The previously cleared Straumann CARES Titanium<br>Abutments are placed onto dental implants to support prosthetic<br>reconstructions such as crowns and bridges. The abutment<br>allows for individual customization regarding function and<br>esthetics. The design step is typically performed by the dental<br>laboratory to the specifications from a clinician. The abutments<br>are produced by a Straumann Manufacturing milling center. |
{5}------------------------------------------------
The patient-specific abutment is designed by a traditional waxup and subsequent scanning or by scanning of the intraoral setting and designing the element via CAD software. The design information is sent via an Internet portal to Straumann. Straumann verifies the design against the validated design parameters. These design limits include a maximum angulation of 30° and a minimum surface area as follows:
| Tooth Position | Minimum surface area mm² |
|-------------------------|--------------------------|
| 7,10, 24,25 | 37 |
| 4,5,12,13,20,21,28,29 | 47 |
| 26,11,2,27,8,9 | 43 |
| 1-3, 14-16, 17-19,30-32 | 56 |
The abutments are manufactured from solid grade 4 titanium at the Straumann milling center. The abutment is delivered to the dental laboratory for final processing before delivery to the clinician. The abutments are attached to the implant with a titanium alloy basal screw. The occlusal restoration is affixed to the titanium abutment using dental cement.
The proposed devices are a modification of the previously licensed devices. The titanium abutments that serve as the base of the CARES Golden Ti/TiN Abutments (the uncoated abutments) are identical in every respect to the reference devices. The modification consists of the addition of a titanium nitride (TiN) coating on the emergence profile and coronal aspect of the abutment. The apical portion of the abutment consisting of the implant-to-abutment interface is masked during the coating process to assure that the interface remains uncoated.
Indications For Use: Abutments are placed into the dental implants to provide support for prosthetic restoration such as crowns, bridges or overdentures.
> The Straumann CARES Golden Ti/TiN Abutments are indicated for single tooth replacements and multiple tooth restorations. The prosthetic restoration is cement-retained.
- Materials: The substrate material for all Straumann CARES Golden Ti/TiN Abutments is commercially pure Titanium (grade 4) conforming to ISO 5832-2. The surface of the emergence profile and coronal aspect of the abutments are coated with titanium nitride (TiN). The implant-to-abutment interface is masked and remains uncoated.
{6}------------------------------------------------
| Technological Characteristics: | A comparison of the relevant technological characteristics between the subject and primary predicate devices is provided in the table that follows. |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Data: | Test data to support the evaluation of the subject Straumann CARES Golden Ti/TiN abutments is as follows:<br>Sterilization validation in accordance with ISO 17665-1 and ISO/TS 17665-2 Biocompatibility assessment per ISO 10993-1 Cytotoxicity testing per ISO 10993-5 Chemical characterization per ISO 10993-18 Dynamic fatigue testing per ISO 14801 by reference to K052272, K072151, K081005 and K082764 Coating characterization including chemical composition, thickness, porosity, mean volume percent voids, SEM images at 100X Adhesion testing by static shear bonding strength per ASTM F1044 employing minimum shear stress acceptance criteria per ASTM F2068 Adhesion testing by dynamic shear bonding strength per ASTM F1160 Abrasion testing consistent with the method described by Tamura, et al, entitled <i>Mechanical Properties of Surface Nitrided Titanium for Abrasion Resistant Implant Materials</i> . No animal or human clinical studies were conducted. |
| Conclusions: | Based upon our assessment of the design and applicable performance data, the subject devices have been determined to be substantially equivalent to the identified predicate devices. |
{7}------------------------------------------------
| Feature | Primary Predicate Devices<br>CARES Titanium Abutments<br>(K082764) | Subject Devices<br>CARES Titanium Gold Abutments | Equivalence<br>Discussion |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications For Use | Abutments are placed into the dental<br>implants to provide support for prosthetic<br>restoration such as crowns, bridges and<br>overdentures.<br><br>The Straumann Titanium Abutment is<br>indicated for single tooth replacements<br>and multiple tooth restorations. The<br>prosthetic restoration is cement-retained. | Abutments are placed into the dental<br>implants to provide support for prosthetic<br>restoration such as crowns, bridges or<br>overdentures.<br><br>The Straumann CARES Golden Ti/TiN<br>Abutment is indicated for single tooth<br>replacements and multiple tooth<br>restorations. The prosthetic restoration is<br>cement-retained. | Identical |
| Compatible Implants | Straumann Bone Level implants having<br>the NC and RC implant-to-abutment<br>interface geometries.<br><br>Straumann Tissue Level implants having<br>the NN, RN and WN implant-to-<br>abutment interface geometries. | Straumann Bone Level implants having<br>the NC and RC implant-to-abutment<br>interface geometries.<br><br>Straumann Tissue Level implants having<br>the NN, RN and WN implant-to-<br>abutment interface geometries. | Identical |
| Material | Commercially pure grade 4 Titanium<br>per ISO 5832-2 | Commercially pure grade 4 Titanium<br>per ISO 5832-2 | Identical |
| Surface Finish | Uncoated (raw titanium) | Titanium nitride (TiN) coating on<br>emergence profile and coronal surfaces | Equivalent<br>The TiN coating is being added to impart<br>gold color to the emergence profile and<br>coronal surfaces of the abutment in order<br>to improve the esthetics of the finished<br>restoration. The presence of the coating<br>does not change the intended uses or<br>indications of the devices. This use of<br>TiN coating was originally cleared to<br>market per K992334. |
| Construction | One-piece solid devices. | One-piece solid devices. | Identical |
| Feature | Primary Predicate Devices<br>CARES Titanium Abutments<br>(K082764) | Subject Devices<br>CARES Titanium Gold Abutments | Equivalence<br>Discussion |
| Design Limits: | Maximum Angulation: 30°<br>Minimum Surface Area:<br>Pos, 7,10, 24,25 37 mm²<br>Pos, 4,5,12,1320,21,28,29 47 mm²<br>Pos, 26,11,2,27,8,9 43 mm²<br>Pos, 1-3, 14-16, 17-19,30-32 56 mm² | Maximum Angulation: 30°<br>Minimum Surface Area:<br>Pos, 7,10, 24,25 37 mm²<br>Pos, 4,5,12,1320,21,28,29 47 mm²<br>Pos, 26,11,2,27,8,9 43 mm²<br>Pos, 1-3, 14-16, 17-19,30-32 56 mm² | Identical |
| Packaging | The abutment and basal screw in<br>reclosable poly bags, co-packaged in a<br>kraft board box. | The abutment and basal screw inkraft board box. | Identical |
| Sterility | Provided non-sterile.<br>Terminally sterilized via moist heat<br>(autoclave) to a Sterility Assurance Level<br>of 10-6. Validated per ISO 17665-1. | Provided non-sterile.<br>Terminally sterilized via moist heat<br>(autoclave) to a Sterility Assurance Level<br>of 10-6. Validated per ISO 17665-1. | Identical |
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.