K162790 · Kci USA, Inc. (Kinetic Concepts, Inc.) · OMP · Jan 26, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162790
Device Name
V.A.C.ULTA Negative Pressure Wound Therapy System
Applicant
Kci USA, Inc. (Kinetic Concepts, Inc.)
Product Code
OMP · General, Plastic Surgery
Decision Date
Jan 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The V.A.C.ULTA™ Negative Pressure Wound Therapy System is an integrated wound management system that provides Negative Pressure Wound Therapy with an instillation option. Negative Pressure Wound Therapy in the absence of instillation is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Instillation Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed. The V.A.C.ULTA™ Negative Pressure Wound Therapy System with and without is indicated for patients with chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure and venous insufficiency), flaps and grafts. Negative Pressure Wound Therapy in the absence of instillation may also be used for: The temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary and for open abdominal wounds with exposed viscera including, but not limited to, abdominal compartment syndrome. The intended care setting is a closely monitored area within the acute care hospital, such as the ICU. The abdominal dressing will most often be applied in the operating theater. The management of the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.
Device Story
Integrated wound management system providing negative pressure wound therapy (NPWT) and instillation therapy. System comprises vacuum pump and instillation pump; delivers negative pressure to remove exudate and instills topical solutions/suspensions to wound bed via tubing and foam dressings under occlusive drape. Used in acute care hospitals (e.g., ICU, OR) by clinicians. Software monitors negative/positive pressure, manages therapy settings, and provides alarm/help functions. Recent modifications include updated user pathways for therapy selection (V.A.C., VERAFLO, PREVENA, ABTHERA), on-screen parameter selection, and ancillary tools (Wound Imaging, History, Information). Clinicians use output to manage wound environment, promote healing, and reduce edema. Benefits include wound bed preparation, granulation tissue formation, and perfusion.
Clinical Evidence
No clinical tests were necessary. Human factors engineering assessment conducted with 30 nurses and doctors demonstrated that new software features could be safely and effectively used.
Technological Characteristics
Powered suction pump system. Components: vacuum pump, instillation pump, tubing, foam dressings, occlusive drape. Software-controlled. Standards: AAMI/ANSI ES60601-1:2005/(R)2012, IEC 60601-1-6:2013, IEC 60601-1-8:2012-11, IEC 60601-1-2:2007-03. Connectivity: Standalone therapy unit. Software: Embedded, provides therapy-specific alerts, help features, and user pathways.
Indications for Use
Indicated for patients with chronic, acute, traumatic, sub-acute, and dehisced wounds, partial-thickness burns, ulcers (diabetic, pressure, venous insufficiency), flaps, and grafts. Also indicated for temporary bridging of abdominal wall openings/open abdominal wounds with exposed viscera and management of draining surgical incisions.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
V.A.C.ULTA™ Negative Pressure Wound Therapy System (K100657)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three overlapping heads, representing the department's mission to protect the health of all Americans. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 26, 2017
KCI USA, Inc. (Kinetic Concepts, Inc.) Margaret Marsh Regulatory Affairs Technical Director 6203 Farinon Drive San Antonio, Texas 78249
Re: K162790
Trade/Device Name: V.A.C. Ulta Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: December 29, 2016 Received: December 30, 2016
Dear Margaret Marsh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162790
#### Device Name
V.A.C.ULTA Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
The V.A.C.ULTA™ Negative Pressure Wound Therapy System is an integrated wound management system that provides Negative Pressure Wound Therapy with an instillation option.
· Negative Pressure Wound Therapy in the absence of instillation is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material.
· Instillation Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.
The V.A.C.ULTA™ Negative Pressure Wound Therapy System with and without is indicated for patients with chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure and venous insufficiency), flaps and grafts.
· Negative Pressure Wound Therapy in the absence of instillation may also be used for:
o The temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary and for open abdominal wounds with exposed viscera including, but not limited to, abdominal compartment syndrome. The intended care setting is a closely monitored area within the acute care hospital, such as the ICU. The abdominal dressing will most often be applied in the operating theater.
o The management of the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.
Type of Use (Select one or both, as applicable)
| <div><span style="font-size:10px"></span>Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|---------------------------------------------------------------------------------------------|
| <div><span style="font-size:10px"></span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
# V.A.C.ULTA™ Negative Pressure Wound Therapy System
| Date prepared | January 24, 2017 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Submitter information [21 CFR 807.929(a)(1)] | |
| Name | KCI USA, Inc. (Kinetic Concepts, Inc.) |
| Address | 6203 Farinon Drive; San Antonio, Texas 78249 |
| Fax number | 210 255-6727 |
| Establishment<br>Registration<br>Number | 3009897021 |
| Name of contact<br>person | Margaret Marsh, Technical Director, Regulatory Affairs |
| Name of the device [21 CFR 807.92(a)(2)] | |
| Trade or<br>proprietary name | V.A.C.ULTA™ Negative Pressure Wound Therapy System<br>(V.A.C.ULTA™ Therapy System) |
| Common or<br>usual name | Negative pressure wound therapy system with an instillation option |
| Classification<br>name | Negative Pressure Wound Therapy Powered Suction Pump<br>(and components) |
| Legally marketed device(s) to which equivalence is claimed<br>[21 CFR 807.92(a)(3)] | |
| V.A.C.ULTA™ Negative Pressure Wound Therapy System (K100657) | |
| Device description [21 CFR 807.92(a)(4)] | |
| The V.A.C.ULTA™ Negative Pressure Wound Therapy System is a negative pressure wound therapy<br>system with an instillation feature which allows controlled delivery and drainage of topical wound treatment<br>solutions and suspensions | |
| The unit is comprised of a vacuum pump and an instillation pump. The vacuum pump delivers negative<br>pressure therapy for the removal of wound exudate and when applicable instilled solutions. The<br>instillation pump provides controlled delivery of topical wound solutions and suspensions. Both pumps are<br>software controlled. Instillation solutions and negative pressure are delivered through tubing to foam<br>dressings in the wound covered by an occlusive drape. Software monitors both negative pressure during<br>negative pressure wound therapy as well as positive pressure during instillation of fluids to the wound bed.<br>Software also provides controls for help and alarm functions | |
| Indications for Use [21 CFR 807.92(a)(5)] | |
| The V.A.C.ULTA™ Negative Pressure Wound Therapy System is an integrated wound management<br>system that provides Negative Pressure Wound Therapy with an instillation option. | |
| Negative Pressure Wound Therapy in the absence of instillation is intended to create an<br>environment that promotes wound healing by secondary or tertiary (delayed primary) intention by<br>preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and<br>perfusion, and by removing exudate and infectious material.Instillation Therapy is indicated for patients who would benefit from vacuum assisted drainage and<br>controlled delivery of topical wound treatment solutions and suspensions over the wound bed.The V.A.C.ULTA™ Negative Pressure Wound Therapy System with and without instillation is indicated for<br>patients with chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers<br>(such as diabetic, pressure and venous insufficiency), flaps and grafts. | |
- The temporary bridging of abdominal wall openings where primary closure is not possible O and/or repeat abdominal entries are necessary and for open abdominal wounds with exposed viscera including, but not limited to, abdominal compartment syndrome. The intended care setting is a closely monitored area within the acute care hospital, such as the ICU. The
{4}------------------------------------------------
# 510(k) SUMMARY V.A.C.ULTA™ Negative Pressure Wound Therapy System
abdominal dressing will most often be applied in the operating theater.
- The management of the environment of surgical incisions that continue to drain following o sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.
### Comparison of the technological characteristics with the predicate device [21 CFR 807.92(a){6)]
At a high level, the subject devices and predicate devices are based on the following same technological elements:
- Both device systems provide the same previously indicated therapies (V.A.C. Therapy, ● V.A.C.VERAFLO Therapy, PREVENA Incision Management Therapy and ABTHERA Open Abdomen Therapy.
- They both use the same dressing components and therapy unit disposables. ●
- . They both have the same technology and specifications or delivery of negative pressure and instillation therapies.
# The following technological differences exist between the subject and predicate devices:
The V.A.C.ULTA Therapy Unit software has been revised to provide the following user enhancements:
- The therapy unit now provides on screen options for selection of labeled parameters for PREVENA and ABTHERA Therapies; the options also provide therapy specific alert notifications and help features.
- . There are now updated user pathways for the V.A.C. and VERAFLO Therapy settings to allow for immediate selection of the default settings with an advanced settings pathway option for experienced users.
- Optional ancillary features have been added to facilitate use, such as a Wound Imaging Tool. History . Tool to access therapy, patient and alarm histories, and an Information Tool to access current therapy settinas.
### Performance Data [21 CFR 807.92(b)]
# Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]
- Software has been assessed in accordance with FDA Guidance, Guidance for the Content of ● Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005.
- . Therapy unit has been certified as conforming with the following electrical safety and electromagnetic standards:
- AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012: o Edition 1.)
- IEC 60601-1-6:2013, Medical electrical equipment Part 1-6: General Requirements for Basic O Safety and Essential Performance - Collateral Standard: Usability: 2013-10, Edition 3.1
- IEC 60601-1-8: 2012-11, Medical Electrical equipment Part 1-8: Basic Safety and Essential O Performance - Collateral Standard: General Requirements, Tests and Guidance for Alarm Systems In Medical Electrical Equipment and Medical Electrical System; Edition 2.1
- IEC 60601-1-2: 2007-03, Medical Electrical Equipment Part 1-2: General Requirements for Basic O Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests. (General II (ES/EMC); Edition 3.
- Neqative pressure performance testing was conducted wound exudate, maximum air ● leak rate, worst case dressing configuration and for the maximum use life of the dressings. The results documented that selection of PREVENA and ABTHERA Therapy settings on the Therapy Unit resulted in delivery of neqative pressure within established parameters for PREVENA Therapy (-125 mmHg, continuous mode) and for ABTHERA Therapy (-100, -125 and -150 mmHg, continuous mode).
#### Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]
No clinical tests were necessary. However human factors engineering assessment with 30 subject nurses
{5}------------------------------------------------
# 510(k) SUMMARY V.A.C.ULTA™ Negative Pressure Wound Therapy System
and doctors indicated that the new features could be safely and effectively used by all test subjects.
### Conclusions drawn [21 CFR 807.92(b)(3)]
The subject device with software modifications is equivalent to the predicate device:
- There has been no change to the intended use of the unit; the proposed modifications to the ● indications for use statement reflect a consolidation of uses for which the therapy unit has been previously cleared.
- There has been no change to the technology delivering negative pressure and instillation therapy. . Performance specifications remain unchanged.
- Human factors engineering assessment documents that the changes to the therapy unit software and ● its associated labeling to provide ease of use features are safe and effective for their intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.