Portable Slit Lamp Microscope Model KJ5S is intended for use in the examination of the anterior and posterior segment of the eye.
Device Story
Portable Slit Lamp Microscope Model KJ5S is a handheld, battery-powered converging stereomicroscope. Used in healthcare facilities/hospitals by clinicians to examine the anterior and posterior segments of the eye. Device features an illumination system and fixation targets activated by a trigger on the handle; light intensity is adjusted via a rheostat. Clinicians view the eye through the microscope to perform visual examinations; findings inform clinical decision-making regarding ocular health. Benefits include portability and ease of use for point-of-care ophthalmic assessment.
Clinical Evidence
No clinical data. Bench testing only, including IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 15004-1/2 (ophthalmic instrument requirements and light hazard protection), ISO 10939 (slit-lamp microscopes), and IEC 62133 (battery safety). Biocompatibility testing conducted per ISO 10993-1, -5, and -10.
Technological Characteristics
Handheld converging stereomicroscope; white LED light source; 7.4V 680mAh rechargeable Li-ion battery. Magnification 10X/16X; working distance 80mm. Materials evaluated for biocompatibility per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2, ISO 15004-1/2, ISO 10939, IEC 62133. No software-based algorithms.
Indications for Use
Indicated for examination of the anterior and posterior segment of the eye in patients requiring ophthalmic evaluation.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2017
Suzhou Kangjie Medical Inc. % Di Wu. Ph.D. Regulatory Affairs Consultant DIWU Regulatory 3440 Indian Queen Lane, Rear Philadelphia, PA 19129
Re: K162778
Trade/Device Name: Portable Slit Lamp Microscope Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: HJO Dated: May 25, 2017 Received: May 30, 2017
Dear Dr. Di Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K162778
Device Name Portable Slit Lamp Microscope Model KJ5S
Indications for Use (Describe)
Portable Slit Lamp Microscope Model KJ5S is intended for use in the anterior and posterior segment of the eye.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
### Date prepared July 13, 2017
### Submission Sponsor (Manufacturer):
Suzhou Kanqjie Medical Inc. No.129, Weixin Rd., Weiting Town Suzhou, Jiangsu, P.R. China 215000
### Submission Correspondent (Agent):
Mrs. Di Wu, PhD 3440 Indian Queen Lane, Rear Philadelphia, PA 19129 USA Phone: (386) 487-8384 Email: diwuregulatory@gmail.com
#### Trade/Device Name:
Portable Slit Lamp Microscope Model KJ5S
Common or Usual Name: Slit Lamp Microscope Device Class: II Classification Name: AC-Powered Slitlamp Biomicroscope Requlation Number: 21 CFR 886.1850 Product Code: HJO Review Panel: Ophthalmic
#### Predicate Device:
- K131711, SUZHOU 66 VISION TECH CO., LTD ● YZ3 PORTABLE SLIT LAMP
#### Device Description:
Portable Slit Lamp Microscope Model KJ5S
The Portable Slit Lamp Microscope is a hand held converging stereomicroscope system powered by rechargeable batteries.
{5}------------------------------------------------
The illumination system and fixation targets are activated using a single/double click trigger located on the rear of the grip/handle. A rheostat located below the eyepieces on the front of the grip /handle is used to increase or reduce the light intensity.
Environment of Use: healthcare facility/hospital Duration and type of contact: less than 24 hours, surface device
#### Intended Use:
Portable Slit Lamp Microscope Model KJ5S is intended for use in the examination of the anterior and posterior seqment of the eye.
#### Comparison to Predicate Device:
Refer to the Comparison Table on Pages 4 and 5.
#### Discussion of Non-Clinical Tests Performed:
The performance tests were conducted:
IEC 60601-1, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
ISO 15004-1, Ophthalmic instruments - Fundamental requirements and test methods - Part 1: General requirements applicable to all ophthalmic instruments
ISO 15004-2, Ophthalmic instruments - Fundamental requirements and test methods - Part 2: Light hazard protection
ISO 10939, Ophthalmic instruments - Slit-lamp microscopes
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International
{6}------------------------------------------------
Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
ISO 10993-1, Biological Evaluation of Medical Devices -- Part 1: Evaluation And Testing Within A Risk Management Process. (Biocompatibility)
ISO 10993-5, Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
ISO 10993-10, Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Biocompatibility)
IEC 62133, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
### Discussion of Clinical Tests Performed:
None
## Conclusion:
The subject device has the same intended uses and indications, similar technological characteristics, and principles of operation as its predicate device. The minor differences between the subject device and its predicate devices raise no new issues of safety or effectiveness. Thus, the subject devices are substantially equivalent to its predicate devices.
{7}------------------------------------------------
## KJ5S Slit Lamp Microscope vs. K131711
| Descriptive Information | Proposed Device | Predicate Device |
|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| 510(k) Number | K162778 | K131711 |
| Manufacturer | Suzhou Kangjie Medical Inc. | 66 Vision Tech Co., Ltd. |
| Proprietary or Model Name | KJ5S portable slit lamp microscope | YZ3 Portable Slit Lamp microscope |
| Indications for Use | Same | Same |
| Flammability of materials near the light source | NONE | NONE |
| Maximum temperature of parts of the device held by the operator or accessible to the patient | Maximum temperature of parts of the device held by the operator:<br>- Eyepiece: 35°<br>- Grip: 35 °<br>- Slit width control ring: 35°<br>Maximum temperature of parts of the device accessible to the patient:<br>Forehead rest 35° | Same |
| Brightness controls | Maximal Illumination ≥12000Lx | Maximal Illumination ≥30000Lx |
| Operating temperature | 10° to 35° | Same |
| Slit Width | 0 to 10mm continuously adjustable | 0 to 12mm continuously adjustable |
| Slit Length | 1mm, 3mm, 5mm, 10mm | 0.2mm, 1mm, 2mm, 12mm |
| Illumination field diameter | 1mm, 3mm, 5mm, 10mm | 0.2mm, 1mm, 2mm, 12mm |
| Radial movement of the slit light illumination relative to the microscope axis | Horizontal ± 30° | Same |
| Stereo angle | 13° | Same |
| Light sources | White LED Single Light Source | Same |
| Pupil-distance | 48 to 72mm | 50 to 75mm |
| Eye piece | 10X, 16X | Same |
| Objective | 1X | Same |
| Total magnifications | 10X, 16X(Optional) | Same |
| Filter | Heat-absorption, Cobalt blue, Red-free, Gray | Cobalt blue, Red-free, Color Temperature Compensation |
| Descriptive Information | Proposed Device | Predicate Device |
| 510(k) Number | K162778 | K131711 |
| Illumination rotation angle | Horizental±30° | Same |
| Working distance | 80mm | 60mm |
| Power | 7.4V 680mAh Li Battery, Rechargeable | 7.4V 2200mA Li Battery, Rechargeable |
| Working time | 5-6 hours | 2.5 hours |
| Net Weight | 750g | 900g |
{8}------------------------------------------------
# Comparison Table
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.