The Proture Double Electric Breast Pump to be used by lactating women in a home setting to express and collect milk from their breasts. This device is intended for a single user.
Device Story
Proture Double Electric Breast Pump is a powered, single-user device for home use by lactating women to express and collect breast milk. It consists of a pump unit and an expression collection kit (tubing, breast shield, silicone cushion). The device uses a single-chip microcomputer to control vacuum levels (15-247 mmHg) and cycle rates (up to 107 cycles/min), mimicking a baby's natural sucking rhythm. It features a 2-phase expression cycle: a Stimulation phase (fast cycles, low vacuum) to initiate milk flow and an Expression phase (slower cycles, higher vacuum) for milk collection. The user operates the device via a touch-button control panel with an LCD screen to adjust settings. It supports single or double pumping and operates via AC adapter or batteries. The system is closed with anti-backflow protection. The device benefits the patient by providing an automated, adjustable, and hygienic method for milk expression.
Clinical Evidence
No clinical data or animal studies were required to support substantial equivalence. Evidence consists of bench testing, including biocompatibility (cytotoxicity, sensitization, skin irritation per ISO 10993-5/10), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing of technical parameters (vacuum stability, cycle rates, noise, and use-life).
Technological Characteristics
Materials: polypropylene, liquid/solid silicone, thermoplastic elastomer. Sensing/Actuation: reciprocating pump driven by a single-chip microcomputer. Energy: AC adapter or battery. Form factor: pump unit with LCD interface. Connectivity: none. Sterilization: manual cleaning/boiling. Software: minor level of concern, microprocessor-based control.
Indications for Use
Indicated for lactating women in a home setting to express and collect milk from their breasts. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract design of an eagle with three human profiles incorporated into its body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 8, 2016
Guangdong Horigen Mother & Baby Products Co., Ltd. % Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd. 8-9th Floor. R&D Building. No. 26 Oinglan Street Panyu District, Guangzhou, 510006 China
Re: K162747 Trade/Device Name: Proture Double Electric Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: Class II Product Code: HGX Dated: October 9, 2016 Received: October 11, 2016
Dear Mike Gu.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
For Division
Douglas Silverstein -S 2016.12.08 10:57:16 -05'00'
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162747
Device Name Proture Double Electric Breast Pump
Indications for Use (Describe)
The Proture Double Electric Breast Pump to be used by lactating women in a home setting to express and collect milk from their breasts. This device is intended for a single user.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image features a logo for "HORIGEN". The logo consists of a red square with rounded corners. Inside the square, there is a white graphic that resembles a crescent moon with a profile of a face within it. Below the square, the word "HORIGEN" is written in a bold, serif font, also in red.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
- 1. SUBMITTER
Guangdong Horigen Mother & Baby Products Co., Ltd. NO.8, Yi Road, Pingbei Industrial Zone, Chaoyang District, Shantou, 515100, Guangdong, China Phone: +86-754-83613668 Fax: +86-754-83843338
| Primary Contact Person: | Jun Deng |
|---------------------------|--------------------------------------------------------------|
| | General Manager |
| | Guangdong Horigen Mother & Baby Products Co., Ltd. |
| | Tel: (+86)-754-83613668-866 |
| | Fax: (+86)-754-83843338 |
| Secondary Contact Person: | Mike Gu |
| | Regulatory Affairs Manager |
| | Guangzhou Osmunda Medical Device Technical Service Co., Ltd. |
| | Tel: (+86)-20-6231 6262 |
| | Fax: (+86) -20-8633 0253 |
| Date prepared | Sep 27, 2016 |
# 2. DEVICE
| Device Name: | Proture Double Electric Breast Pump |
|--------------------|-------------------------------------|
| Common/Usual Name: | Powered breast pump |
| Regulation number | 21 CFR 884.5160 Powered breast pump |
| Regulation Class: | II |
| Product Code: | HGX Pump, Breast, Powered |
# 3. PREDICATE DEVICE
K160511, Naya Breast Pump System This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo featuring a red square with rounded corners. Inside the square, there's a white crescent moon shape with a profile of a face looking towards the right. Below the square, there is a word that appears to be a brand name, but the text is cut off at the bottom of the image.
### DEVICE DESCRIPTION
The Proture Double Electric Breast Pump is designed, manufactured by the Guangdong Horigen Mother & Baby Products Co., Ltd. The device is intended for lactating women in a home setting and is for a single user. It comprises a pump unit and the expression collection kit including tubing, and its raw materials include polypropylene, liquid and solid silicone, and thermoplastic elastomer.
This electric breast pump imitates baby's sucking rhythm with help of a single-chip microcomputer. It has multiple stimulation levels for breast massage, and multiple milk suction speed intensities to imitate rhythm of a baby's suction. The keyboard of the control panel is soft. The screen is an LCD, allowing for process viewing. The pump's electronic memory takes over, mimicking the rhythm. The electric breast pump is capable of providing vacuum levels from 15 to 247 mmHg with cycle rates up to 107 cycles per minute.
The Proture Double Electric Breast Pump provides the following user features:
- . Closed system with anti-backflow
- . LCD screen touch button operation
- 2-phase expression: Stimulation and Expression phases
- Stimulation Phase: Suction pattern with fast cycles (53-107 cycles/min) and low vacuum (15-120mmHg) to start milk flowing, 5 levels
- Expression Phase: Suction pattern with slower cycles (19~47T/min) and higher vacuum (82-247mmHg) to express milk, 9 levels
- Double-pumping ability: single or double pumping
- Dual power source: it can be operated by A/C adapter or batteries
All parts of the device that directly contact with breast and milk are to be sanitized by boiling in water for 5 minutes, followed by drying with a clean towel. After use, all parts that directly contact the breast and milk are to be cleaned in warm, soapy water.
#### 5. INDICATIONS FOR USE
The Proture Double Electric Breast Pump is a powered breast pump to be used by lactating women in a home setting to express and collect milk from their breasts. This device is intended for a single user.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for HORIGEN. The logo features a red square with rounded corners. Inside the square is a white graphic of a crescent moon shape with a silhouette of a face looking towards the right. Below the square, the word "HORIGEN" is written in red, with a serif font.
# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Specification | Predicate Device | Proposed Device | Discussion of Differences |
|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Device name | Naya Breast Pump<br>System | Proture Double Electric<br>Breast Pump | |
| K number | K160511 | K164727 | |
| Product code | HGX | HGX | Identical |
| Indication for Use | The Naya Breast Pump<br>System is a powered<br>breast pump to be used<br>by lactating women in<br>the hospital or home<br>setting to express and<br>collect milk from their<br>breasts. | The Proture Double Electric<br>Breast Pump is a powered<br>breast pump to be used by<br>lactating women in a home<br>setting to express and<br>collect milk from their<br>breasts. This device is<br>intended for a single user. | Different.<br>The proposed device is only for<br>single users in a home<br>environment.<br>Same Intended Use |
| Patient<br>Population | Lactating women | Lactating women | Identical |
| Environment of<br>Use | Home and Hospital | Home | Different.<br>The proposed device is only<br>intended for a home<br>environment.<br>Same Intended Use |
| Pump Style | Reciprocating Pump | Reciprocating Pump | Identical |
| Single/double<br>pump | both | both | Identical |
| Adjustable<br>suction levels | 10 levels | 9 levels | Similar.<br>Less suction levels.<br>This difference does not raise<br>different questions of safety<br>and effectiveness. |
| Cycle speed<br>(cycles/min) | 34–120 | 19-107 | Similar.<br>Lower cycle speed.<br>This difference does not raise<br>difference questions of safety<br>and effectiveness. |
| Visual indicator | a mobile App | LCD Screen | Different.<br>Different user interface.<br>This difference does not raise<br>different questions of safety<br>and effectiveness. |
| HORIGEN | | | |
| 510(k) Premarket Notification Submission | | | |
| Specification | Predicate Device | Proposed Device | Discussion of Differences |
| Vacuum range<br>(mmHg) | 50-260 | 15-247 | Similar.<br>Lower vacuum levels.<br>This difference does not raise<br>difference questions of safety<br>and effectiveness. |
| Cycling/Suction<br>Control<br>Mechanism | Microprocessor | Microprocessor | Identical |
| Power source | Input: 100-240 VAC,<br>50/60Hz,2.1A | Input: 100-240 V, 50/60Hz,<br>1.6A | Similar.<br>Different batteries.<br>Complied with IEC 60601-1 |
| | Rechargeable Li-ion<br>Battery | Battery or A/C Adapter | and IEC 60601-1-2.<br>This difference does not raise<br>different questions of safety<br>and effectiveness. |
| Software | Yes | Yes | Identical |
| Anatomical Sites | breast | breast | Identical |
| Cleaning method<br>for Accessories | Soap and warm water | Soap and warm water | Identical |
| Materials in<br>contact with user<br>and expressed<br>milk tested per<br>ISO 10993-1 | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Identical |
| Electrical Safety | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC60601-1-2 | Identical |
# Table 6.1 Comparison with Predicate Device
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo with a red square background. Inside the square is a white crescent moon shape. Within the crescent moon shape are two faces facing each other in profile. The faces are also white and appear to be in a silhouette style.
According to the above comparison, the proposed device is comparable to the predicate device in pump style, single/double pump, cycling/suction control mechanism, anatomical sites, cleaning method for accessories, biocompatibility and electrical safety.
Compared with the predicate device, the proposed device is indicated for single users in a home environment, while the predicate device is indicated for use by single or multiple users in a home or hospital environment. These differences in physical location of use do not impact the intended use (expressing and collecting milk from breasts), which is the same for the proposed and predicate devices.
In terms of technology compared with the predicate device, the proposed device has less suction levels, different cycle speeds, different vacuum range, a different device interface, and type of batteries. However, as noted in the table above, these differences in technological characteristics do not raise different questions of safety and effectiveness as compared to the predicate device.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for a company or organization. The logo features a red square with rounded corners. Inside the square, there is a white crescent shape with a profile of a face looking towards the right. Below the square, there is text that appears to be the name of the company or organization, but the text is blurry and difficult to read.
### PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
The biocompatibility evaluation for the Double Electric Breast Pump was conducted in accordance with the International Standard ISO 10993-1:2009/(R)2013, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:
- Cytotoxicity
- Sensitization
- Skin Irritation
The breast shield body and silicone cushion of the Proture Double Electric Breast Pump are considered to contact directly with the human body and breast milk for a duration of less than 24 hours. Testing for cytotoxicity, sensitization and skin irritation complied with ISO 10993-5:2009/(R)2014, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity and ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. Testing demonstrates that the materials are non-cytotoxic, non-sensitizing, and non-irritating.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Proture Double Electric Breast Pump. The device complies with the IEC 60601-1:2012, standard for electrical safety and the IEC 60601-1-2:2007 standard for EMC. It demonstrates substantial equivalences to the predicate device.
#### Performance testing
Performance testing was conducted on the Proture Double Electric Breast Pump. Technical parameters of stimulation mode and suction mode, including working current, pressure, cycle rate and noise were evaluated in the performance testing, as well as the stability of vacuum level and use-life. All of the tested parameters meet the predefined acceptance criteria. The performance parameters were compared with the predicate device in Table 6.1, and it is concluded that the proposed device is substantially equivalent to the predicate device.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image features the logo for HORIGEN. The logo consists of a rounded square with a pink background. Inside the square is a white graphic that appears to be a stylized representation of a human face in profile, possibly a mother and child. Below the square, the word "HORIGEN" is written in a serif font, also in pink.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "minor" level of concern. Software validation demonstrated that the software functions as specified in the software requirement specifications. It demonstrates substantial equivalences to the predicate device.
# Animal Study
The subject of this premarket submission, Proture Double Electric Breast Pump, does not require animal studies to support substantial equivalence.
# Clinical Study
The subject of this premarket submission, Proture Double Electric Breast Pump, did not require clinical studies to support substantial equivalence.
# 8. CONCLUSION
The differences between the Proture Double Electric Breast Pump and its predicate device do not raise new issues of safety and effectiveness. The non-clinical data support the safety of the device and the performance testing report demonstrate that the Proture Double Electric Breast Pump should perform as intended in the specified use conditions.
From the results of non-clinical data including the performance testing described, Guangdong Horigen concludes that the Proture Double Electric Breast Pump is as safe and as effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.