The intended use of the MIST Systems is for the delivery of low energy ultrasound via a fluid mist for the promotion of wound healing. It is a prescription use only device.
Device Story
UltraMIST System is a low-frequency, low-intensity ultrasound device for wound care. It consists of a generator, treatment wand, and disposable applicator. The system uses a user-provided fluid source (e.g., IV saline) to create an atomized mist. A peristaltic pump on the generator controls fluid flow to the ultrasound radiation surface (titanium alloy horn tip), where fundamental atomization and acoustic streaming generate the mist. The mist is applied to the wound site; the radiation surface does not contact the patient. The device is used in clinical settings by healthcare providers to cleanse wounds and perform maintenance debridement. By removing slough, fibrin, exudates, and bacteria, the device promotes wound healing. The generator includes a microprocessor to manage treatment variables, including duration based on the treatment area.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including IEC 60601-1 3rd Ed. general safety testing and cytotoxicity biocompatibility testing.
Indicated for promoting wound healing through wound cleansing and maintenance debridement by the removal of yellow slough, fibrin, tissue exudates, and bacteria.
Regulatory Classification
Identification
A low energy ultrasound wound cleaner is a device that uses ultrasound energy to vaporize a solution and generate a mist that is used for the cleaning and maintenance debridement of wounds. Low levels of ultrasound energy may be carried to the wound by the saline mist.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Low Energy Ultrasound Wound Cleaner.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" and an emblem in the center. The emblem is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2016
Alliqua Biomedical, Inc. Kathy Simpson Director. OA/RA 11495 Valley View Road Eden Prairie, Minnesota 55344
Re: K162721
Trade/Device Name: Stimucel System, Stimucel Generator, Stimucel Treatment Wand, Stimucel Applicator Regulation Number: 21 CFR 878.4410 Regulation Name: Low Energy Ultrasound Wound Cleaner Regulatory Class: Class II Product Code: NRB Dated: November 28, 2016 Received: December 2, 2016
Dear Kathy Simpson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162721
Device Name UltraMIST
Indications for Use (Describe)
MIST Systems produce a low energy ultrasound generated mist used to promote wound healing through wound cleansing and maintenance debridement by the removal of yellow slough, fibrin, tissue exudates, and bacteria.
Type of Use (Select one or both, as applicable)
| <span> <svg height="10" width="10"> <rect height="10" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="10"></rect> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <svg height="10" width="10"> <rect height="10" style="fill:white;stroke-width:1;stroke:rgb(0,0,0)" width="10"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510k Summary
| 510(k) Summary (As required by 21 CFR §807.92) | |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Alliqua BioMedical, Inc.<br>11495 Valley View Road<br>Eden Prairie, MN 55344 |
| Contact Person: | Kathy Simpson<br>Director QA / RA<br>Telephone: 952-224-8738<br>Fax: 952-224-8797<br>Email: ksimpson@alliqua.com |
| Date Prepared: | August 29, 2016 |
| Trade Name of<br>Device: | UltraMIST™ |
| Common /<br>Usual Name: | UltraMIST™ |
| Classification<br>Name: | Low Energy Ultrasound Wound Cleaner<br>Class: II<br>Regulation: 21 CFR §878.4410 |
| Product Code: | NRB |
| Manufacturer: | Alliqua BioMedical, Inc.<br>11495 Valley View Road<br>Eden Prairie, MN 55344 |
| Establishment<br>Registration: | 3004580659 |
| Predicate<br>Device(s): | K140782, UltraMIST® System<br>Manufacturer: Alliqua BioMedical, Inc. |
| Device<br>Description: | MIST Systems (e.g., UltraMIST System) are low frequency ultrasound<br>devices, which utilize a fluid (e.g., saline) mist to deliver the ultrasound to<br>the treatment site. The atomized fluid allows the treatment to be performed<br>without direct contact to the patient. MIST Systems are made up of a main<br>generator component, treatment wand, and disposable applicator.<br><br>MIST Systems deliver low-frequency ultrasound to the treatment site using<br>an atomized fluid mist. The mist is generated from a small, portable system<br>consisting of a generator, a treatment wand (encasing an ultrasound<br>transducer), and a disposable applicator that has been designed to connect<br>to a user-provided fluid source (e.g., IV saline bag). The system provides a<br>flow of fluid (irrigant) to the tissue surface by the use of low frequency, low<br>intensity ultrasound. Only the mist touches the tissue, not the ultrasound<br>radiation surface (horn tip), thus the term "non-contact" ultrasound. The flow<br>of fluid as an irrigant is controlled by a peristaltic pump on the generator<br>unit. The fluid is dispensed onto the ultrasound radiation surface where the<br>mist is generated. The mist is projected by fundamental atomization or<br>nebulization and acoustic streaming or acoustic radiation force respectively. |
| Intended Use: | The intended use of the MIST Systems is for the delivery of low energy<br>ultrasound via a fluid mist for the promotion of wound healing. It is a<br>prescription use only device. |
| | The overall target market for MIST Systems is the advanced wound care |
{5}------------------------------------------------
| | market. There have been no changes to the intended use of the subject of<br>this submission. | | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Indications for<br>Use | MIST Systems produce a low energy ultrasound generated mist used<br>to promote wound healing through wound cleansing and maintenance<br>debridement by the removal of yellow slough, fibrin, tissue exudates<br>and bacteria. | | |
| Summary of<br>Technical<br>Comparisons: | A summary of the technical components of the subject device and the<br>predicate UltraMIST System are presented in the Table below: | | |
| | Technical<br>Component | UltraMIST System | Subject<br>Device |
| | Control of Fluid Flow | Clamp on tubing line to stop<br>and start fluid flow for priming<br>purposes.<br>Peristaltic pump used to<br>provide consistent flow | Same |
| | Enclosures | Injection molded PC/ABS<br>polymer and compound<br>curvature | Geon HC<br>8210, Rigid<br>Polyvinyl<br>Cyholride,<br>Flammability<br>rated V-0 |
| | Supply Voltage and<br>Current | 100-240VAC Universal input | Same |
| | Input Frequency | 50-60Hz Universal input | Same |
| | Rated Power | 100 watts (fuse limited) | Same |
| | Distal Displacement of<br>Radiation Surface | $65 \pm 10$ microns | Same |
| | Acoustic Frequency | $40 \pm 1$ kHz | Same |
| | Radiation Surface<br>Diameter and Material | 0.390 in. $\pm$ 0.001 in.<br>Titanium alloy (TI-6AL-4V) | Same |
| | Single patient Use<br>Disposable Applicator | Injection molded PC/ABS<br>polymer and compound<br>curvature | Same |
| | Packaging and<br>Labeling | System: Non-sterile, reusable<br>Applicator: Sterile per ISO<br>11607 | Same |
| | Method of Sterilization | Gamma radiation | Same |
| | Shelf Life / Product Life<br>Cycle | System: 5 years<br>Applicator: 3 years | Same |
| | Mode of Exudate and<br>Bacterial Removal | Atomized mist<br>Acoustic and fluid stream<br>delivery / washout | Same |
| | Microprocessor | Yes | Same |
| | Patient contact | None - fluid contact only via<br>atomized mist | Same |
| | Treatment variables | Treatment duration based on<br>area to be treated (algorithm) | Same |
| | - Generator v0.8 | current out of range narrower<br>than original MIST System | modified to<br>be equivalent<br>to that of the<br>original MIST<br>System |
| Performance<br>Data - Non-<br>Clinical Testing: | Testing performed on the subject device to demonstrate Substantial<br>Equivalence:<br>IEC 60601-1 3rd Ed. General Safety Testing -Testing was performed by<br>Intertek to verify that the change in enclosure material for the Generator<br>and Treatment Wand did not have an adverse effect on the safety or<br>performance of the system. The results of the testing demonstrated that<br>the system with the new material enclosures was as safe, as effective,<br>and performed as well as the predicate device with the previous<br>enclosure materials. The complete test report is included in the<br>submission materials and can be found in VOL_16, Appendix C. Biocompatibility testing, Cytotoxicity – Testing was performed by WuXi<br>AppTec to verify that the change in enclosure material for the Generator<br>and the Treatment Wand did not have an adverse effect on the<br>biocompatibility of the system. Since the Generator and Treatment<br>Wand do not come into contact with a breached or compromised<br>surface, or potential indirect contact with a breached or compromised<br>surface like the Applicator, only Cytotoxicity testing was performed to<br>ensure that the new enclosure materials do not present a new or<br>different biocompatibility issue for the user of the device. The results of<br>the testing demonstrated that the new enclosure materials for the<br>Generator and Treatment wand were as safe, as effective, and<br>performed as well as the previous enclosure material as both were<br>determined to be non-cytotoxic. The complete test report is included in<br>the submission materials and can be found in VOL_15, Appendix B. Consensus standards used in the testing and preparation of this<br>premarket notification may be found in VOL_005, Standards Data<br>Report Forms. | | |
| Clinical Studies: | NA for this submission. | | |
| Conclusions: | The modifications made to the subject device are substantially equivalent to<br>its predicate, the UltraMIST System. | | |
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.