Elecsys Toxo IgM, Elecsys Toxo IgM PreciControl

K162678 · Roche Diagnostics · LGD · Jun 23, 2017 · Microbiology

Device Facts

Record IDK162678
Device NameElecsys Toxo IgM, Elecsys Toxo IgM PreciControl
ApplicantRoche Diagnostics
Product CodeLGD · Microbiology
Decision DateJun 23, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3780
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K162678 · Jun 23, 2017Elecsys Toxo IgM, Elecsys Toxo IgM PreciControlRoche DiagnosticsClinical laboratory routine testing samples; Retrospective commercial sample vendor collectionThe sponsor used clinical routine testing samples (prospective and retrospective) to evaluate the performance of the Elecsys Toxo IgM assay compared to an FDA-cleared predicate device.Retrospective clinical samples; Routine clinical care; Clinical performance evaluation

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
US retrospectively collected study; Retrospective analysis60 patients (4 males, 56 females, including 50 pregnant women) who had anti-Toxo testing ordered per clinical routine.; Sample Size: 60FDA-cleared predicate devicePercent agreement with predicate device

Indications for Use

The Elecsys Toxo IgM immunoassay is for the in vitro qualitative detection of IgM antibodies to Toxoplasma gondii in human serum and plasma. This assay may be used as an aid in the diagnosis of an acute or recent Toxoplasma gondii infection in suspected patients and pregnant women. Patient testing must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immuno-assay analyzers. This assay has not been cleared by the FDA for blood/plasma donor screening. PreciControl Toxo IgM is used for quality control of the Elecsys Toxo IgM immunoassay on the Elecsys and cobas e immunoassay analyzers.

Device Story

Two-step sandwich immunoassay; uses streptavidin microparticles and electrochemiluminescence (ECLIA) detection. Input: 10 µL human serum or plasma sample. Process: Sample prediluted 1:20; reacts with labeled T. gondii-specific recombinant antigen; biotinylated monoclonal IgM-specific antibodies and streptavidin-coated microparticles added; complex binds to solid phase; magnetically captured on electrode surface; unbound substances removed via ProCell; voltage applied to induce chemiluminescent emission measured by photomultiplier. Output: Qualitative result (Reactive/Non-reactive/Equivocal) based on cutoff index (COI). Used in clinical laboratories on Elecsys 2010 or cobas e 411 analyzers. Results aid clinicians in diagnosing acute/recent Toxoplasmosis infection when used alongside IgG testing.

Clinical Evidence

Clinical performance evaluated via three studies (two prospective, one retrospective) and a CDC panel (total 601 samples + 97 panel members). Studies compared Elecsys Toxo IgM to FDA-cleared predicate. CDC panel results: 91% positive agreement (29/32) and 98% negative agreement (64/65). Prospective EU study (n=101) showed 94.2% positive agreement and 50% negative agreement. US prospective study (n=440) showed 66.7% positive agreement and 98.6% negative agreement. Equivocal results were treated as discrepant.

Technological Characteristics

Two-step sandwich immunoassay; electrochemiluminescence (ECLIA) detection. Reagents: streptavidin-coated microparticles, ruthenium-labeled T. gondii recombinant antigen, biotinylated anti-human IgM. Platform: Elecsys/cobas e automated analyzers. Sample types: serum, lithium heparin, K2-EDTA, sodium citrate plasma. Software-based signal processing and cutoff determination. No specific material standards cited.

Indications for Use

Indicated for in vitro qualitative detection of IgM antibodies to Toxoplasma gondii in human serum and plasma. Used as an aid in diagnosis of acute or recent infection in suspected patients and pregnant women. Must be used with anti-Toxoplasma gondii IgG antibody assay. Not for blood/plasma donor screening.

Regulatory Classification

Identification

Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.

Predicate Devices

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K162678 B. Purpose for Submission: To obtain substantial equivalence for the Elecsys Toxo IgM assay and Elecsys Toxo IgM PreciControl set C. Measurand: Toxoplasma IgM antibody D. Type of Test: Immunoglobulin Class-Capture Chemiluminescence Immunoassay E. Applicant: Roche Diagnostics F. Proprietary and Established Names: Elecsys Toxo IgM, Elecsys Toxo IgM PreciControl G. Regulatory Information: 1. Regulation section: 21 CFR 866.3780 - Toxoplasma gondii serological reagents 2. Classification: Class II 3. Product code: LGD; enzyme linked immunoabsorbent assay, Toxoplasma gondii JJX; single (specified) analyte controls (assayed and unassayed) 4. Panel: {2} 83 Microbiology H. Intended Use: 1. Intended use(s): Elecsys Toxo IgM Immunoassay The Elecsys Toxo IgM immunoassay is for the in vitro qualitative detection of IgM antibodies to *Toxoplasma gondii* in human serum and plasma. This assay may be used as an aid in the diagnosis of an acute or recent Toxoplasmosis infection in suspected patients and pregnant women. Patient testing must be performed in conjunction with an anti-*Toxoplasma gondii* IgG antibody assay. The electrochemiluminescence immunoassay “ECLIA” is intended for use on Elecsys and cobas e immuno-assay analyzers. This assay has not been cleared by the FDA for blood/plasma donor screening. PreciControl Toxo IgM PreciControl Toxo IgM is used for quality control of the Elecsys Toxo IgM immunoassay on the Elecsys and cobas e immunoassay analyzers. 2. Indication(s) for use: Same as intended use 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Elecsys 2010 and cobas e 411 analyzers I. Device Description: The Elecsys Toxo IgM is a two-step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection. In the first incubation, 10 µL of sample is automatically prediluted 1:20 with Elecsys Diluent Universal, with *T. gondii*-specific recombinant antigen labeled complex then added. Anti-Toxo IgM antibodies present in the sample react with the labeled *T. gondii*-specific recombinant antigen. Biotinylated monoclonal IgM specific antibodies and streptavidin-coated microparticles are subsequently added. The complex becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then {3} induces chemiluminescent emission which is measured by a photomultiplier. Results are determined automatically by the Elecsys software by comparing the electrochemiluminescence signal obtained from the reaction. The assay can be run on either the Elecsys 2010 of the cobas e 411 analyzers platforms. # J. Substantial Equivalence Information: 1. Predicate device name(s): VIDAS Toxoplasma Gondii IgM Assay (K923166) 2. Predicate 510(k) number(s): K923166 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate (K923166) | | Intended Use | Elecsys Toxo IgM Immunoassay: The Elecsys Toxo IgM immunoassay is for the in vitro qualitative detection of IgM antibodies to Toxoplasma gondii in human serum and plasma. This assay may be used as an aid in the diagnosis of an acute or recent Toxoplasmosis infection in suspected patients and pregnant women. Patient testing must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay. The electrochemiluminescence immunoassay “ECLIA” is intended for use on Elecsys and cobas e immuno-assay analyzers. This assay has not been cleared by the FDA for blood/plasma donor screening. PreciControl Toxo IgM: PreciControl Toxo IgM is used for quality control of the Elecsys | The VIDAS TOXO IgM (TXM) assay is intended for use with a VJDAS (Vitek ImmunoDiagnostic Assay System) instrument as an automated enzyme-linked fluorescent immunoassay (ELFA) for the presumptive qualitative detection of anti-Toxoplasma gondii IgM antibodies in human serum, as an aid in the diagnosis of acute, recent, or reactivated Toxoplasma gondii infection. This assay must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay. VIDAS TXM assay performance has not been established for prenatal screening or newborn testing. This assay has not been cleared by the FDA for blood/plasma donor screening. | {4} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate (K923166) | | | Toxo IgM immunoassay on the Elecsys and cobas e immunoassay analyzers. | | | Calibrator | Included with the kit | Same | | Assay type | Sandwich ELISA | Same | | Instrument Platform | Fully automated system | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate (K923166) | | Assay Protocol | Electrochemiluminescence Immunoassay (ECLIA) | Enzyme-linked fluorescent immunoassay (ELFA) | | Sample type | Serum, lithium heparin plasma, potassium EDTA plasma and sodium citrate plasma | Serum | | Calibrators | Two | One | ## K. Standard/Guidance Document Referenced (if applicable): - EP05-A2, Evaluation of Precision Performance of Quality Measurement Methods; Approved Guideline - Second Edition 2004 - EP17-A, Protocols For the Determination of the Limits of Detection and Limits of Quantitation ## L. Test Principle: Electrochemiluminescence immunoassay ## M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision was determined at one sites using Elecsys reagents, three human sera pools and two controls. Testing was performed in accordance with the Clinical and Laboratory Standards Institute Document EP5-A with the following modification: Samples were tested six times daily for 10 days (n = 60 data points per panel member). The following results were obtained and summarized below in Table 1 and were determined to be acceptable. {5} Table 1. Precision | | | | Repeatibility | | Between-run | | Intermediate precision | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | N | Mean [s/co] | SD [s/co] | CV (%) | SD [s/co] | CV (%) | SD [s/co] | CV (%) | | PC Toxo IgM 1 | 60 | 0.146 | 0.004 | 2.5 | 0.003 | 1.9 | 0.005 | 3.2 | | PC Toxo IgM 2 | 60 | 1.61 | 0.037 | 2.3 | 0.045 | 2.8 | 0.059 | 3.6 | | Serum Pool 1 | 60 | 0.147 | 0.003 | 2.3 | 0.005 | 3.1 | 0.006 | 3.9 | | Serum Pool 2 | 60 | 2.06 | 0.075 | 3.7 | 0.031 | 1.5 | 0.082 | 4.0 | | Serum Pool 3 | 60 | 3.49 | 0.172 | 4.9 | 0.060 | 1.7 | 0.183 | 5.2 | Reproducibility Representative performance data on the analyzers are given below. Reproducibility was determined at two sites using Elecsys reagents using three human sera panels. Testing was performed in accordance with the Clinical and Laboratory Standards Institute Document EP5-A with the following modification: Samples were tested six 6 times daily for 10 days at each site (n = 60 data points per site). The following results were obtained and summarized below in Table 2. Table 2. Reproducibility | | | | Repeatibility | | Between-run | | Intermediate precision | | Between-site | | Total precision | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Serum Pool | N | Mean [s/co] | SD [s/co] | CV (%) | SD [s/co] | CV (%) | SD [s/co] | CV (%) | SD [s/co] | CV (%) | SD [s/co] | CV (%) | | Pool 1 | 120 | 0.161 | 0.003 | 1.9 | 0.006 | 3.5 | 0.006 | 4.0 | 0.021 | 13.0 | 0.022 | 13.6 | | Pool 2 | 120 | 2.01 | 0.065 | 3.3 | 0.03 | 1.5 | 0.072 | 3.6 | 0.067 | 3.3 | 0.098 | 4.9 | | Pool 3 | 120 | 3.37 | 0.135 | 4.0 | 0.073 | 2.2 | 0.153 | 4.6 | 0.174 | 5.2 | 0.232 | 6.9 | b. Linearity/assay reportable range: Not applicable - This assay is qualitative c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: {6} There is no international standard available for measuring anti-Toxoplasma IgM in serum; therefore, calibrators are traceable to an in-house reference preparation. ## Calibrators Calibration of an Elecsys Toxo IgM reagent lot is recommended every 28 days (1 month). Elecsys Toxo IgM reagent kits can be stored on-board the analyzers for up to 2 weeks. A new calibration of a kit kept on-board is recommended every 7 days ## Stability: ### Reagents Reagent stability was determined by testing five human samples and two control samples. The results support the claim of reagent stability for 12 weeks of storage after first opening at 2-8°C (not on board). Reagents are only stable on-board the analyzer up to 84 hours total time. The real-time data currently available supports a shelf-life claim of 14 months at 2-8°C for unopened reagents. ### Calibrators Calibrator stability was determined by testing the calibrators in duplicate once after first opening and once again after 8 weeks without loss of signal. The results support the claim of calibrator stability for 8 weeks of storage after first opening at 2-8°C (not on board). Calibrators are also stable beyond 5 hrs. when tested every hour on board the analyzer at 20-25°C. The data provided supports a 5 hr. on-board stability claim. The real-time data currently available supports a shelf-life claim of 14 months at 2-8°C without opening. ### Controls The PreciControl Toxo IgM is stable at 2-8°C after first opening for up to 8 weeks. The PreciControl Toxo IgM is stable for 5 h under on-board conditions. The real-time data currently available supports a shelf-life claim of 14 months. ### Samples Samples are stable for 3 days at 15-25°C, 3 weeks at 2-8°C and 3 months at -20°C. ## Expected values: - Calibrator 1 is manufactured to have a target Index value range of 400 - 2500. - Calibrator 2 is manufactured to have a target Index value range of 4500 - 35000. - The negative control is manufactured to have a target Index value of 0.18 COI*. - The positive control is manufactured to have a target Index value of 2.0 COI*. *cutoff index (COI) is calculated as a ratio of sample signal over cutoff. The target Index value range for each control is reported in the rackpack or PerciControl kit to make sure that the correct target values are used. {7} d. Detection limit: Not applicable - This assay is qualitative e. Analytical specificity: Cross reactivity: The cross-reactivity study for the Elecsys Toxo IgM assay was designed to evaluate potential interference from the presence of potentially cross-reactive antibodies or substances that may cause symptoms similar to or that may mimic Toxoplasmosis infection. Only samples that were sero-positive for the cross reactant and sero-negative for Toxoplasma gondii IgM by a commercially available Toxoplasma IgM assay were used to test for potentially cross-reactive organisms. The results are summarized in table 3. below. Table 3. Cross Reactivity Data | Cross- reactant | # of Toxo IgM Elecsys reactants / Total Tested | | --- | --- | | AMA* | 1/15 | | ANA | 0/26 | | Chlamydia | 0/5 | | CMV | 0/13 | | EBV | 1/10 | | Gonorrhea | 0/5 | | HAV | 0/10 | | HBV | 1/21 | | HCV | 0/12 | | HIV | 0/13 | | HSV | 0/9 | | Influenza | 0/16 | | Malaria | 2/25 | | Parvo B19 | 0/11 | | Rubella** | 1/10 | | Syphilis | 0/5 | | TPHA | 0/4 | | VZV | 0/8 | *Contains a false positive result ** Equivocal results were counted as positive AMA Anti-mitochondria antibodies. ANA Antinuclear antibodies TPHA Treponema pallidum human antibodies {8} 8 High Dose Hook Effect (false negative): High-dose hook effect was evaluated by testing three samples with Toxoplasma gondii IgM levels out-of-range &gt; 64 COI (i.e., 64 x minimum positive value). In this study, three samples with very high Toxo IgM levels were serially diluted with negative sample to cover the entire assay range. The neat sample and dilutions were tested using one lot on one instrument. The signal obtained was plotted versus the dilution. The samples resulted in calculated concentration values above the measuring range with no sample misclassification, indicating no hook effect was observed. Interference Testing was performed to determine whether the presence of endogenous or exogenous substances (compounds) may interfere with assay results. Matched sample pools (3 negative and 3 positive) were tested neat and spiked with the respective interferent. Acceptance criteria were defined as the % change in signal must not be more than +20% and no change the qualitative result. The assay is unaffected at the concentration for each substance listed in Table 4 below on the Elecsys Toxo IgM assay. Table 4. Endogenous Interference Results | Compound | Concentration | | --- | --- | | Bilirubin | < 684 μmol/L or < 40 mg/dL | | Hemoglobin | < 1.24 mmol/L or < 2 g/dL | | Intralipid | < 2000 mg/dL | | Biotin | < 246 nmol/L or < 60 ng/mL | f. Assay cut-off: The cut-off for the Elecsys Toxo IgM assay was established by the use of 900 clinical samples from which a preliminary cutoff was set by ROC analysis. The results of the internal and external testing to evaluate the performance of the Elecsys Toxo IgM assay demonstrate that the cutoff of 1.0 COI is appropriate to obtain good sensitivity and specificity. 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: {9} Studies were conducted to evaluate the suitability of the following 5 types of blood collection tubes; serum/gel separation tubes, lithium heparin plasma, K₂-EDTA plasma, K₃-EDTA plasma and sodium citrate plasma. A minimum of 49 sample pairs were drawn into serum and plasma collection tubes. The recovery of the analyte in the presence of the anti-coagulant was compared to serum. The results indicate there was no difference between any of the plasma types when compared to serum. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Comparative testing: Clinical performance of the Elecsys Toxo IgM assay was evaluated in three studies (two prospective and one retrospective) and with a CDC panel. These studies were conducted to compare the performance of the subject assay to an FDA-cleared predicate device. Overall, there were 601 samples and 97 panel members tested in these studies which comprised of samples from individuals who were sent to the laboratory for *T gondii IgM* testing, or were from a retrospective collection. The demographics of the patients vary widely and will be described with each pooled data set. Study 1 A total of 440 prospective specimens were obtained from US reference laboratory. All specimens represented samples for whom anti-Toxo testing had been ordered for clinical routine testing. The patient population consisted of 162 males and 278 females (158 were from pregnant women). The patient's age ranged from 2 to 86 years. The performance of the Elecsys Toxo IgM immunoassay, relative to an FDA cleared method is presented in the Table 5 below. 9 {10} Table 5. US prospectively collected (overall patient population) | Comparator Toxo IgM Method | | | | | | | --- | --- | --- | --- | --- | --- | | Elecsys Toxo IgM * | | Positive | Equivocal | Negative | Total | | | Reactive | 2 | 0 | 2 | 4 | | | Equivocal | 0 | 0 | 4 | 4 | | | Non-reactive | 0 | 1 | 431 | 432 | | | Total | 2 | 1 | 437 | 440 | *The Elecsys IgM assay calls positive results “reactive” and negative results “non-reactive” | Agreement Classification | Numerator/ Denominator | Percent Agreement | 95 % Confidence Interval | | --- | --- | --- | --- | | Positive agreement* | 2/3 | 66.7 | 9.43 - 99.2 | | Negative agreement* | 431/437 | 98.6 | 97.0- 99.4 | *For agreement calculation, equivocal results are counted as discrepant against the Elecsys Toxo IgM. Study 2 A total of 60 retrospective samples were obtained from a US commercial sample vendor. The patient population consisted of 4 males and 56 females (50 were from pregnant women). All specimens represented samples from whom anti-Toxo testing had been ordered per clinical routine. The performance of the Elecsys Toxo IgM immunoassay, relative to an FDA cleared method is presented in Table 6 below. {11} Table 6. US retrospectively collected (overall patient population) | Comparator Toxo IgM Method | | | | | | | --- | --- | --- | --- | --- | --- | | Elecsys Toxo IgM | | Positive | Equivocal | Negative | Total | | | Reactive | 0 | 0 | 0 | 0 | | | Equivocal | 0 | 0 | 1 | 1 | | | Non-reactive | 0 | 0 | 59 | 59 | | | Total | 0 | 0 | 60 | 60 | | Agreement Classification | Numerator/ Denominator | Percent Agreement | 95 % Confidence Interval | | --- | --- | --- | --- | | Positive agreement | 0/0 | NA | NA | | Negative agreement* | 59/60 | 98.3 | 91.1 - 100 | *For agreement calculation, the equivocal result was counted as discrepant against the Elecsys Toxo IgM # Study 3 One-hundred one (101) samples collected prospectively were obtained from the general population in Europe under suspect of Toxoplasmosis infection consistin of 5 males and 96 females with an age range of 22 to 69 years. The performance of the Elecsys Toxo IgM immunoassay, relative to an FDA cleared method is presented in Table 7 below. {12} Table 7. EU Prospectively collected (overall patient population) | Comparator Toxo IgM Method | | | | | | | --- | --- | --- | --- | --- | --- | | Elecsys Toxo IgM | | Positive | Equivocal | Negative | Total | | | Reactive | 81 | 1 | 5 | 87 | | | Equivocal | 0 | 1 | 1 | 2 | | | Non-reactive | 4 | 1 | 7 | 12 | | | Total | 85 | 3 | 13 | 101 | | Agreement Classification | Numerator/ Denominator | Percent Agreement | 95 % Confidence Interval | | --- | --- | --- | --- | | Positive agreement* | 81/86 | 94.2 | 87.0 - 98.1 | | Negative agreement* | 7/14 | 50.0 | 23.0 - 77.0 | *For agreement calculation, equivocal results are counted as discrepant results against the Elecsys Toxo IgM. In case of unanimously equivocal samples, samples were not used for agreement calculation. Of the 101 prospectively collected European specimens described above, 87 were from pregnant women. The agreement results are shown in Table 8 below. {13} Table 8. EU prospectively collected (pregnant women) | Reference Toxo IgM Method | | | | | | | --- | --- | --- | --- | --- | --- | | Elecsys Toxo IgM | | Positive | Equivocal | Negative | Total | | | Reactive | 68 | 0 | 5 | 73 | | | Equivocal | 0 | 1 | 1 | 2 | | | Non-reactive | 4 | 1 | 7 | 12 | | | Total | 72 | 2 | 13 | 87 | | Agreement Classification | Numerator/ Denominator | Percent Agreement | 95 % Confidence Interval | | --- | --- | --- | --- | | Positive agreement* | 68/73 | 93.2 | 84.7 - 97.7 | | Negative agreement* | 7/13 | 53.9 | 25.1 - 80.8 | *For agreement calculation, equivocal results are counted as discrepant results against the Elecsys Toxo IgM. In case of unanimously equivocal samples, samples were not used for agreement calculation. CDC Panel study: A panel of 97 samples (32 known positive and 65 known negative) was obtained through the US Centers for Disease Control and Prevention (CDC) and was tested for Toxo IgM on the Elecsys 2010 analyzer. The results are presented as agreement with the known results as additional information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC. The results of the CDC panel testing are summarized below in Table 9. Table 9. CDC Panel results | | ElecsysToxo IgM Percent Agreement (n/N) | 95 % Confidence interval | | --- | --- | --- | | CDC positives | 91 % (29/32) | 75 - 98 % | | CDC negatives | 98 % (64/65) | 92 - 100 % | 4. Clinical cut-off: Non-reactive: $&lt; 0.8$ COI {14} Equivocal: ≥ 0.8 - &lt; 1.0 COI Reactive: ≥ 1.0 COI Samples with a cutoff index &lt; 0.8 are non-reactive in the Elecsys Toxo IgM assay. Samples with a cutoff index between ≥ 0.8 and &lt; 1.0 are considered equivocal. The sample should be retested. In case the result is still equivocal, a second sample should be tested e.g. within 2-3 weeks. Samples with a cutoff index ≥ 1.0 are reactive in the Elecsys Toxo IgM assay. 5. Expected values/Reference range: Expected value range will be provided N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 14
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...