← Product Code [KWQ](/productcode/KWQ) · K162638

# CETRA Anterior Cervical Plate System (K162638)

_Orthofix, Inc. · KWQ · Dec 16, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162638

## Device Facts

- **Applicant:** Orthofix, Inc.
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Dec 16, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The CETRA Anterior Cervical Plate System is intended for anterior fixation to the cervical spine from C2 to C7. The specific clinical indications include: a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies); b) Spondylolisthesis; c) Fracture; d) Spinal Stenosis; e) Deformities (i.e., scoliosis, kyphosis, and/or lordosis); f) Tumor; g) Pseudoarthrosis; h) Revision of previous surgery

## Device Story

CETRA Anterior Cervical Plate System; temporary anterior cervical implant construct. Components: non-sterile, single-use titanium alloy plates and screws. Surgeon attaches plate to anterior aspect of vertebral body via screws. Used in clinical settings for spinal stabilization. Provides mechanical fixation to support healing/fusion. Benefits: stabilization of cervical spine segments.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1717 (Static Axial Compression, Dynamic Axial Compression, Static Torsion). Additional screw push-out testing conducted to evaluate locking mechanism retention.

## Technological Characteristics

Materials: Titanium alloy (Ti6Al4V ELI per ASTM F136). Components: Plates and screws for anterior cervical fixation. Mechanical testing conducted per ASTM F1717. Non-sterile, single-use.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Hallmark Anterior Cervical Plate System ([K100614](/device/K100614.md))
- Reliant Anterior Cervical Plate System ([K030595](/device/K030595.md))
- SLIM*LOC Anterior Cervical Plate System ([K013877](/device/K013877.md))
- MaxAn Anterior Cervical Plate System ([K133518](/device/K133518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 16, 2016

Orthofix, Inc. Jacki Koch Regulatory Affairs Specialist, II 3451 Plano Parkway Lewisville, Texas 75056

Re: K162638

Trade/Device Name: CETRA Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: December 6, 2016 Received: December 7, 2016

Dear Ms. Koch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| 510(k) Number (if known) | K162638                              |
|--------------------------|--------------------------------------|
| Device Name              | CETRA Anterior Cervical Plate System |

**Indications for Use**

**Indications for Use (Describe)**

The CETRA Anterior Cervical Plate System is intended for anterior fixation to the cervical spine from C2 to C7. The specific clinical indications include:

a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies);

b) Spondylolisthesis;

c) Fracture;

d) Spinal Stenosis;

e) Deformities (i.e., scoliosis, kyphosis, and/or lordosis);

f) Tumor;

g) Pseudoarthrosis;

h) Revision of previous surgery

| Type of Use (Select one or both, as applicable)                                                        |                                                                                     |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  
Food and Drug Administration  
Office of Chief Information Officer  
Paperwork Reduction Act (PRA) Staff  
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14) Page 1 of 1

sinə aid bitki növü. İstinadlar Respublikası Respublikası Say Say Say Alman Say Say Say Say Say Say Say Say Say Say Say Say Say Say

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue, stylized, abstract shape on the left, followed by the word "ORTHOFIX" in black, sans-serif font. The "R" in "ORTHOFIX" is slightly larger than the other letters. There is a trademark symbol after the "X".

K162638 CETRA Anterior Cervical Plate System

# 510(k) SUMMARY

## CETRA Anterior Cervical Plate System

| 510(k) Owner Information<br>Name:<br>Address:       | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056                                                                                                                                                     |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone Number:<br>Fax Number:<br>Email:          | 214.937.2100<br>214-937-3322<br>jackikoch@orthofix.com                                                                                                                                                          |
| Registration Number:                                | 2183449                                                                                                                                                                                                         |
| Contact Person:                                     | Jacki Koch, Regulatory Affairs Specialist, II                                                                                                                                                                   |
| Date Prepared:                                      | September 21, 2016                                                                                                                                                                                              |
| Name of Device<br>Trade Name / Proprietary<br>Name: | CETRA Anterior Cervical Plate System                                                                                                                                                                            |
| Common Name:                                        | Anterior Cervical Plate System                                                                                                                                                                                  |
| Product Code:                                       | KWQ                                                                                                                                                                                                             |
| Regulatory Classification:                          | Class II - 21 CFR § 888.3060                                                                                                                                                                                    |
| Review Panel:                                       | Orthopedic Device Panel                                                                                                                                                                                         |
| Primary Predicate:                                  | Hallmark Anterior Cervical Plate System – K100614 - Orthofix                                                                                                                                                    |
| Additional Predicates:                              | Reliant Anterior Cervical Plate System - K030595 - Orthofix<br>SLIM*LOC Anterior Cervical Plate System - K013877 - Depuy<br>Codman & Shurtleff, Inc.<br>MaxAn Anterior Cervical Plate System - K133518 - Biomet |

Reason for 510(k) Submission: Orthofix is submitting this Traditional 510(k) premarket notification for the introduction of a new Anterior Cervical Plate System called CETRA Anterior Cervical Plate System.

#### Device Description

The CETRA Anterior Cervical Plate System consists of an assortment of non- sterile, single use, titanium alloy (Ti6Al4V ELI per ASTM F136) plates and screws that allow a surgeon to build a temporary anterior cervical implant construct. The plate is attached to the anterior aspect of the vertebral body by means of screws to the cervical spine.

#### Intended Use / Indications for Use

The CETRA Anterior Cervical Plate System is intended for anterior fixation to the cervical spine from C2 to C7. The specific clinical indications include:

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, sans-serif font. The "®" symbol is present to the right of the word "ORTHOFIX".

# K162638 CETRA Anterior Cervical Plate System

a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies);

- b) Spondylolisthesis;
- c) Fracture:
- d) Spinal Stenosis:
- e) Deformities (i.e., scoliosis, kyphosis, and/or lordosis);
- f) Tumor:
- q) Pseudoarthrosis:
- h) Revision of previous surgery

## Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices

The technological characteristics of the CETRA Anterior Cervical Plate System are similar to the predicate device and reference device in terms of design, dimensions, intended use, materials, and performance characteristics. There are no significant differences between the CETRA Anterior Cervical Plate System and the predicate Hallmark Anterior Cervical Plate System (K100614) and the additional predicate devices Reliant Anterior Cervical System (K030595), SLIM*LOC Anterior Cervical Plate System (K013877) and MaxAn Anterior Cervical Plate System (K133518), which would adversely affect the use of the product.

## PERFORMANCE DATA - Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence

Mechanical testing consisting of Static Axial Compression Test, Dynamic Axial Compression Test and Static Torsion Test were conducted in accordance to ASTM F1717. Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model.

Additional mechanical testing included a screw push-out test, which served to mechanically evaluate the CETRA Anterior Cervical Plate System locking mechanism's ability to retain screws after being implanted.

#### Conclusion

The new CETRA Anterior Cervical Plate System has the same intended use, the same indications for use, the same technological characteristics and design, same materials and the same principles of operation as the predicate Hallmark Anterior Cervical Plate System (K100614) and the additional predicate devices Reliant Anterior Cervical System (K030595), SLIM*LOC Anterior Cervical Plate System (K013877) and MaxAn Anterior Cervical Plate System (K133518).

---

**Source:** [https://fda.innolitics.com/device/K162638](https://fda.innolitics.com/device/K162638)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
