Invacare Tracer SX5 Manual Wheelchair, Invacare Tracer IV Heavy Duty/Extra Wide Manual Wheelchair
Applicant
Invacare Corporation
Product Code
IOR · Physical Medicine
Decision Date
Apr 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Invacare® Tracer Manual Wheelchair, Invacare® Tracer® SX5 Manual Wheelchair, Invacare Tracer SX5 Recliner Manual Wheelchair, and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair are intended to provide mobility to persons limited to a sitting position.
Device Story
Manual wheelchairs providing mobility for individuals restricted to sitting position; intended for daily indoor/outdoor use on firm surfaces. Models include standard, recliner (90°-180° backrest), and heavy-duty/extra-wide configurations. Features include X-design cross-brace, adjustable armrests, anti-tippers, and wheel locks (push-to-lock/pull-to-lock). Operated by user or caregiver; provides customized fit to improve comfort, propulsion, and transfer. Recliner models allow backrest angle adjustment. Heavy-duty models support higher weight capacities (up to 450 lbs). Benefits include improved mobility, stability, and postural support for patients with limited ambulatory function.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ANSI/RESNA WC/Volume 1 2009 standards for static stability, brake effectiveness, dimensions, maneuvering space, seating/wheel dimensions, and static/impact/fatigue strength. Biocompatibility testing per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (skin irritation). Flammability testing per CAL117:2013 and ISO 8191.
Indicated for persons limited to a sitting position requiring mobility assistance.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with a design element above them that resembles a stylized wing or feather.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 17, 2017
Invacare Corporation Elijah Wreh Regulatory Affairs Manager (pre-market) One Invacare Way Elyria, Ohio 44035
Re: K162621
Trade/Device Name: Invacare Tracer SX5 Manual Wheelchair, Invacare Tracer SX5 Recline Manual Wheelchair, Invacare Tracer IV Heavy Duty/Extra Wide Manual Wheelchair
Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: October 5, 2016 Received: October 6, 2016
Dear Elijah Wreh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements
{2}------------------------------------------------
as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Michael J. Hoffmann -S" in a large, sans-serif font. The text is black and appears to be centered on a white background. The letters are evenly spaced, and the overall impression is clean and professional.
for
Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162621
#### Device Name
Invacare Tracer SX5 Manual Wheelchair, Invacare Tracer SX5 Recliner Manual Wheelchair, Invacare Tracer IV Heavy Duty/extra Wide Manual Wheelchair
#### Indications for Use (Describe)
The Invacare® Tracer Manual Wheelchair, Invacare® Tracer® SX5 Manual Wheelchair, Invacare Tracer SX5 Recliner Manual Wheelchair, and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair are intended to provide mobility to persons limited to a sitting position.
Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| SUBMITTER: | Invacare Corporation<br>One Invacare Way<br>Elyria, OH 44035 |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Phone: (440) 329-6840<br>Email: ewreh@invacare.com |
| CONTACT PERSON: | Elijah N. Wreh<br>Regulatory Affairs Manager (Pre-Market) |
| MANUFACTURER: | INVAMEX S.DE R.L. DE C.V.<br>Avenida El Puente S/N<br>Parque Industrial El Puente Manimex<br>Reynosa Tamaulipas, Mexico C.P. 88783 |
| Date Prepared: | April 17, 2017 |
| DEVICE<br>Name of Devices: | Invacare Tracer SX5 Manual Wheelchair<br>Invacare Tracer SX5 Recline Manual Wheelchair<br>Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair |
| Common or Usual Name:<br>Classification Name: | Wheelchair, Mechanical<br>Mechanical Wheelchair 21 CFR §890.3850 |
| Regulatory Class: | I |
| Product Code: | IOR: Wheelchair, Mechanical |
| Regulation Description: | Mechanical Wheelchair |
| PREDICATE DEVICE: | Tracer Series Mechanical Wheelchairs (K935398) |
| REFERENCE DEVICE: | Ride Lite 9000 Series Wheelchair (K881951) |
{5}------------------------------------------------
### DEVICE DESCRIPTION
The Invacare® Tracer® SX5 Manual Wheelchair is intended to provide mobility to persons who may be restricted to a sitting position. The manual wheelchair is intended for ongoing daily use. The models offer a wide range of customization which allows for a better fit for the end user resulting in comfort, lightweight, easier propelling, transferring, and smooth ride. The manual wheelchair is intended for indoor and outdoor use on firm surfaces free of climbing obstacles.
The Invacare® Tracer® SX5 Recliner Manual Wheelchair is a Tracer SX5 model with the addition of a recline feature. This recliner model incorporates a mechanism that allows the backrest to recline in angles from 90° to 180°. Both models of the chair have a weight capacity of 250 to 300 lbs. depending on the seat width. Both models also come with an anti-tipper device to ensure the chair remains stable while in use.
The Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair has the same intended use as the Invacare® Tracer® SX5 Manual Wheelchair. However, the Tracer IV does not come with a recliner model and has a higher weight capacity of 350 lbs. depending on the configuration. Both models of Tracer IV chairs come with a push-to-lock or pull-to-lock feature to engage and disengage the wheel locks. Both models also come with anti-tippers that allow clearance of 1.5 to 2 inches. There is no prior submission for the subject devices. The associated model and accessories include:
- . Invacare Tracer SX5 Manual Wheelchair
- o Seat Width 14" to 18" 250 lbs. (114 Kg)
- o Seat width 20" to 22" 300 lbs. (136 Kg)
- Invacare Tracer SX5 Recline Manual Wheelchair .
- o Seat Width 14" to 18" 250 lbs. (114 Kg)
- o Seat width 20" to 22" 300 lbs. (136 Kg)
- Invacare Tracer IV & Heavy Duty Manual Wheelchair . o Weight capacity at 350 lbs. and 450 lbs.
- Tipper aid ●
- Cane holder
- Anti-theft device
- Non-folding device
- O2 holder ●
- Wheel lock extension ●
- IV rod telescopic fixed back .
{6}------------------------------------------------
### INDICATIONS FOR USE
The Invacare® Tracer Manual Wheelchair, Invacare® Tracer® SX5 Manual Wheelchair, Invacare Tracer SX5 Recliner Manual Wheelchair, and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair are intended to provide mobility to persons limited to a sitting position.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE and REFERENCE DEVICES
The device comparison showed that the subject devices are substantially equivalent in intended use, design, materials, and operational principles to the previously cleared Tracer Series Mechanical Wheelchairs (K935398) to provide mobility to persons limited to a sitting position.
# BASIS OF SUBSTANTIAL EQUIVALENCE
The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" and the technological characteristics which include materials, design, energy source, and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A).
The subject device components are as safe and effective as the predicate and reference devices and do not raise different questions of safety and effectiveness.
The major frame components for the subject Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair and the previously cleared Tracer Series Manual Wheelchairs (K935398) are the same. The data generated from the subject Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair test report supports a finding of substantial equivalence in regards to the device comparison, dimensional analysis, device specifications, design characteristics and to provide mobility to persons limited to a sitting position.
The difference in weight capacity between the subject Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair and the previously cleared Tracer Series Manual Wheelchairs (K935398) do not raise no new issues of safety or effectiveness.
{7}------------------------------------------------
In conclusion, the performance testing, device comparison, and dimensional analysis demonstrates that the subject device components are substantially equivalent to the predicate and reference devices in regards to static stability, effectiveness of brakes, dimensions, mass and maneuvering space, measurement of seating and wheel dimensions, static impact and fatigue strength, coefficient of friction, information disclosures, documentation and labeling.
{8}------------------------------------------------
# Design Characteristics Comparison – Invacare® Tracer® SX5 Manual Wheelchair
| Device | Invacare® Tracer® SX5 Manual Wheelchair | Tracer Series Manual Wheelchairs |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| 510(k) Number | Subject Devices<br>Pending Submission (K162621) | Predicate Device<br>(K935398) |
| Intended Use | The Invacare® Tracer® Manual Wheelchair is intended to provide mobility to persons limited to a sitting position. | To provide mobility to persons who may be restricted to a sitting position. |
| Indications for Use | The Invacare® Tracer® SX5 Manual Wheelchair and Invacare®<br>Tracer® SX5 Recliner Manual Wheelchair are intended to provide<br>mobility to persons limited to a sitting position. | Persons who may be restricted to a sitting position. |
| Design | Tracer SX5<br>(Subject Device) | Tracer Series Manual Wheelchairs<br>(K935398) |
| Weight Limit | 250 lbs.<br>300 lbs. | 250 lbs. |
| Seat Width 14" to 18"<br>Seat Width 20" to 22" | X-design | X-design |
| Cross-brace Configuration | 14", 16", 18", 20" or 22 " | 16.5 ", 18 " |
| Seat Width | 14" to 18" with the weight limitation is 250 lbs.<br><br>20" to 22" with the weight limitation is 300 lbs. | 14" to 18" |
| Seat Depth | 16" to 18" | 13", 14", 15", 16", 17", 18", 19", 20", 21" or 22" |
| Back Style | Fixed | Fixed |
| Arm Types | Fixed or adjustable height; desk or full length; removable | Permanent and/or Removable |
| Anti-Tippers | Optional | Optional |
| Wheel Locks | Push-to-lock and pull-to-lock | Push to lock |
| Rear Wheel Sizes | 20", 22", 24 " | 20", 22", 24 " composite |
| Caster Sizes | 6 " – 8" | 6 " – 8" |
| Front riggings | Swing away | Swing away |
{9}------------------------------------------------
| Device | Invacare Tracer SX5 Recliner Manual Wheelchair | Ride Lite 9000 Series Wheelchair |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Subject Device<br>(K162621) | Reference Device<br>(K881951) |
| Intended Use | The Invacare® Tracer Manual Wheelchair is intended to provide<br>mobility to persons limited to a sitting position. | The Ride Lite 9000 Series Wheelchair is intended to provide mobility to persons limited<br>to a sitting position. |
| Indications for Use | The Invacare® Tracer® SX5 Manual Wheelchair and Invacare®<br>Tracer® SX5 Recliner Manual Wheelchair are intended to<br>provide mobility to persons limited to a sitting position | The Ride Lite 9000 Series Wheelchair is indicated to provide mobility to person ages 12<br>and up (adolescents and adults) with a weight capacity of 350 lb. depending on the seat<br>width. |
| Design | | |
| Weight Limit | 250 lbs.<br>300 lbs. | 350 lb. |
| Seat Width | 16" and 18" | 14" and 18" |
| Recline Feature | Reclining | Reclining |
| Cross-brace Configuration | X-design | X-design |
| Seat Depth | 16" and 18" | 16", 17" and 18" |
| Back Style | Reclining (5° – 78° vertical measurement) | Reclining (5° – 78° vertical measurement) |
| Arm Types | Fixed or adjustable height; desk or full length; removable | Fixed or adjustable height; desk or full length; removable |
| Anti-Tippers | Required | Required |
| Wheel Locks | Push-to-lock and pull-to-lock | Push-to-lock and pull-to-lock |
| Rear Wheel Sizes | 20", 22", 24" | 20", 22", 24" |
| Caster Sizes | 6" – 8" | 6" – 8" |
| Front riggings | Swing away | Swing away |
# Design Characteristics Comparison – Invacare Tracer SX5 Recliner Manual Wheelchair
Note: The reference device is used for the recline feature and technical characteristics comparison.
{10}------------------------------------------------
| Device | Invacare® Tracer IV® Heavy Duty/Extra Wide Manual<br>Wheelchair | Tracer Series Manual Wheelchairs |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| 510(k) Number | Subject Device<br>Pending Submission (K162621) | Primary Predicate Device<br>(K935398) |
| Intended Use | The Invacare® Tracer® Manual Wheelchair is intended to provide<br>mobility to persons limited to a sitting position. | To provide mobility to persons who may be restricted to a sitting position. |
| Indications for Use | The Invacare® Tracer® IV Manual Wheelchair and Invacare®<br>Tracer® IV Heavy Duty/Extra Wide Manual Wheelchair are intended<br>to provide mobility to persons limited to a sitting position. | Persons who may be restricted to a sitting position. |
| Design | | |
| Weight Limit | 350lbs.<br>450 lbs. | 250 lbs. |
| Cross-brace Configuration | X-design | X-design |
| Seat Width | 18", 20", 22" or 24" | 16.5", 18" |
| Seat Depth | 18" and 20" | 16", 16" |
| Back Style | Fixed | Fixed |
| Arm Types | Fixed or adjustable height; desk or full length; removable | Permanent and/or Removable |
| Anti-Tippers | Optional | Optional |
| Wheel Locks | Push to Lock (Adult), Pull to Lock (Hemi) | Push to lock |
| Rear Wheel Sizes | 24" Composite; 24" pneumatic flat free insert | 20", 22", 24" composite |
| Caster Sizes | 8 x 1 $1/4$ "Urethane<br>8 x 1 $3/4$ "Semi Pneumatic | 6"- 8" * |
| Front riggings | Swing away | Swing away |
# Design Characteristics Comparison – Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair
{11}------------------------------------------------
## PERFORMANCE DATA
## Non-Clinical Test
Non-clinical laboratory testing was performed on the subject Invacare® Tracer Manual Wheelchairs to determine substantial equivalence. The following testing was performed:
- ANSI / RESNA WC/Volume 1 2009, Section 1: Determination of Static Stability .
- ANSI / RESNA WC/Volume 1 2009, Section 3: Determination of Effectiveness of . Brakes
- ANSI / RESNA WC/Volume 1 2009, Section 5: Determination of Dimensions, Mass . and Maneuvering Space
- . ANSI / RESNA WC/Volume 1 - 2009, Section 7: Measurement of Seating and Wheel Dimensions
- ANSI / RESNA WC/Volume 1 2009, Section 8: Static, Impact and Fatigue Strengths . Tests
- RESNA WC-1:2009 Section 11: Test Dummies .
- RESNA WC-1:2009 Section 13: Determination of Coefficient of Friction of Test . Surfaces
- ANSI / RESNA WC/Volume 1 2009, Section 15: Requirements for Information . Disclosure, Documentation and Labeling
- CAL117:2013, Section 1: Flammability Testing .
- ISO 8191-1:1987 & 8191-2:1988: Flammability Testing .
Verification Testing demonstrated that the subject Invacare® Tracer® SX5 Manual and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual are substantially equivalent to the marketed predicate and reference devices.
## Biocompatibility Testing
The biocompatibility evaluation for the subject Invacare® Tracer® SX5 Manual and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual were conducted in accordance with the FDA Blue Book Memorandum #G95 – 1 "Use of International Standard ISO – 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"' May 1, 1995, and International Standard ISO 10993 - 1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- AAMI / ANSI / ISO 10993-5:2009, Biological Evaluation of Medical Devices Part 5: Tests for in vitro Cytotoxicity
{12}------------------------------------------------
- AAMI / ANSI / ISO 10993-10:2010, Biological Evaluation of Medical Devices Part 10: Tests for Skin Irritation
### Animal Study
Animal testing was not required for this submission.
## Clinical Testing
Clinical testing was not required for this submission.
## CONCLUSIONS
The subject devices have the same intended use and similar technological characteristics as the predicate and reference devices. Design Verification Testing support the safety and performance of the subject devices and demonstrate that the subject devices should perform as intended in the specified use conditions. Therefore, the subject Invacare Tracer SX5 Manual Wheelchair, Invacare Tracer SX5 Recline Manual Wheelchair, and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair are substantially equivalent to the predicate and reference devices identified throughout this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.