K162570 · Miromatrix Medical, Inc. · FTM · Feb 16, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162570
Device Name
Miromatrix Biological Mesh TW
Applicant
Miromatrix Medical, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Feb 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Miromatrix Biological Mesh TW is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Device Story
Implantable, acellular surgical mesh derived from porcine liver tissue via perfusion decellularization; composed primarily of collagen type I. Designed as interface between tendon and tendon sheath or surrounding tissues; placed under, around, or over injured tendon to provide protective scaffold during healing. Used in surgical settings by clinicians. Device is hydrated, flexible, and supplied in various sizes (1x2 cm to 10x20 cm); trimmable and suturable. Terminal sterilization via electron beam irradiation. Benefits patient by providing protective environment for tendon repair.
Clinical Evidence
No human clinical data. Bench testing evaluated tensile strength and suture pullout strength, showing higher performance than predicate. Animal testing (4 and 10-week partial Achilles tendon defect model) demonstrated comparable tissue attachment and healing response to predicate.
Technological Characteristics
Acellular porcine liver-derived collagen type I mesh. Thickness 0.1-1.5 mm. Dimensions 1x2 cm to 10x20 cm. Sterilization: electron beam irradiation. Standalone implantable device.
Indications for Use
Indicated for management and protection of tendon injuries without substantial loss of tendon tissue.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 16, 2017
Miromatrix Medical Inc. % Miriam Provost, Ph.D. Biologics Consulting Group, Inc. 400 N. Washington St., Suite 100 Alexandria, Virginia 22314
Re: K162570
Trade/Device Name: Miromatrix Biological Mesh TW Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM, OWW Dated: January 16, 2017 Received: January 18, 2017
Dear Dr. Provost:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162570
Device Name Miromatrix Biological Mesh TW
Indications for Use (Describe)
The Miromatrix Biological Mesh TW is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.97(h) and 21 CFR 807.92, the 510(k) Summary for the Miromatrix Biological Mesh TW is provided below.
## 1. SUBMITTER
Miromatrix Medical Inc. 10399 West 70th Street Eden Prairie, MN 55334
Contact Person: Jeff Ross, Ph.D. Phone: 763-458-8801 Email: jross@miromatrix.com Date Prepared: September 13, 2016
# DEVICE 2.
510(k) Number: K162570 Name of Device: Miromatrix Biological Mesh TW Common Name: Surgical Mesh Classification Regulation: 21 CFR 878.3300 Regulatory Class: II Product Code: FTM, OWW Panel: General and Plastic Surgery
# PREDICATE DEVICE 3.
Predicate Device: Integra Tendon Wrap™ Tendon Protector (K053655)
Reference Device: Miromatrix Biological Mesh (K134033)
# DEVICE DESCRIPTION 4.
The Miromatrix Biological Mesh TW is an implantable, animal-sourced, acellular surgical mesh that is derived from porcine liver tissue. The liver tissue undergoes perfusion decellularization and the resulting mesh is comprised primarily of collagen type I. The Miromatrix Biological Mesh TW provides a scaffold during tissue repair and is designed to be an interface between the tendon and tendon sheath or the surrounding tissues. The Miromatrix Biological Mesh TW is designed for placement under, around or over the injured tendon. The device is terminally sterilized by electron beam irradiation in its packaging and is hydrated, moist and flexible when its packaging is opened. The mesh is available in sizes ranging from 1 cm x 2 cm to 10 cm x 20 cm, and may be trimmed or cut as required before being sutured to the surgical site.
# 5. INDICATION FOR USE
The Miromatrix Biological Mesh TW is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
{4}------------------------------------------------
This indication for use is identical to that of the predicate device, the Integra Tendon Wrap™ Tendon Protector (K053655).
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6.
The similarities and differences in technological characteristics between the subject Miromatrix Biological Mesh TW and the predicate mesh are summarized below.
## Similarities 6.1.
The Miromatrix Biological Mesh TW is similar in design to the predicate Integra Tendon Wrap™ Tendon Protector (K053655). Both are designed to provide a protective environment while the tendon is healing for the management and protection of the tendon injuries in which there has been no substantial loss of tendon tissue.
In addition, the Miromatrix Biological Mesh TW is identical in materials (i.e., porcine liver tissue) to the reference device, the Miromatrix Biological Mesh (K134033).
## 6.2. Differences
Although similar in design, there are minor differences in dimensions between the subject Miromatrix Biological Mesh TW and the predicate Integra Tendon Wrap™ Tendon Protector (K053655). The Miromatrix Biological Mesh TW has a thickness of 0.1-1.5 mm and sizes from 1x2 cm to 10x20 cm. Based on Miromatrix's own measurements, the predicate Integra Tendon Wrap™ Tendon Protector has a thickness range that slightly exceeds that of the subject device. Based on publically available information, the sizes of the predicate are similar, i.e., 5x5cm and 10x12.5cm.
When compared to the reference device (i.e., the Miromatrix Biological Mesh of K134033), the only difference in design are a slight change in dimensions, i.e., a reduction in the mesh thickness (i.e., from 0.5-3 mm to 0.1-1.5 mm) and a reduction in the maximum mesh size (i.e., from 20x30 cm to 10x20 cm).
In terms of materials, the predicate Integra Tendon Wrap™ Tendon Protector is comprised of a porous matrix of cross-linked bovine Type I collagen and glycosanminoglycan (GAG). While the predicate material is different from the porcine liver tissue that comprises the subject device, the difference does not raise a new question of safety and effectiveness.
The subject device has the same material as the reference device (K134033). Thus, the biocompatibility of the subject device material (i.e., porcine liver) and, therefore, its safety for the proposed intended use has already been established as part of K134033.
# 7. PERFORMANCE DATA
Because the materials of the subject Miromatrix Biological Mesh TW are identical to that of the reference device, the Miromatrix Biological Mesh (K134033), no new biocompatibility testing is required to demonstrate substantial equivalence.
Shelf life testing was performed to establish the proposed shelf life of the device.
{5}------------------------------------------------
Bench and animal data show that the subject Miromatrix Biological Mesh TW is substantially equivalent to the predicate Integra Tendon Wrap™ Tendon Protector with regard to performance. Specifically:
- . The total tensile strength and suture pullout strength were evaluated and measured for both the Miromatrix Biological Mesh TW and the predicate device. The Miromatrix Biological Mesh TW was shown to have higher tensile and suture pullout strength than the predicate device. These differences do not have any impact on safety and effectiveness, and therefore, on substantial equivalence.
- . Animal testing comparing the subject and predicate devices in a 4 and 10-week partial Achilles tendon defect model was performed. After 4 and 10 weeks, the average tissue attachment was comparable. The healing response at 4 and 10 weeks was also comparable.
## 8. CONCLUSIONS
The Miromatrix Biological Mesh TW is identical in indications to the predicate Integra Tendon Wrap (K053655).
The Miromatrix Biological Mesh TW is similar in design to the predicate Integra Tendon Wrap (K053655). In terms of materials, the Integra Tendon Wrap is comprised of a porous matrix of cross-linked bovine Type I collagen and glycosanminoglycan (GAG). While the Integra Tendon Wrap material is different from the porcine liver tissue that compromises the subject device, the difference does not raise a new question of safety and effectiveness. The biocompatibility of the subject device material (i.e., porcine liver) has already been established as part of K134033.
Performance testing (bench and animal testing) of the subject device versus the predicate device was performed and the results demonstrate that the subject device is substantially equivalent to the predicate device for the proposed indication for use.
In conclusion, based on the exact same indication for use and comparable design and performance to the predicate Integra Tendon Wrap (K053655) and based on the same materials and comparable design to the reference device Miromatrix Biological Mesh (K134033), as well as favorable preclinical testing, the subject Miromatrix Biological Mesh TW may be found substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.