K162559 · Biomet, Inc. · LLZ · Mar 31, 2017 · Radiology
Device Facts
Record ID
K162559
Device Name
Move Forward 3D Motion Simulation Service
Applicant
Biomet, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Move Forward 3D Motion Service is intended as pre-operative or post-operative software for simulating/evaluating hip preservation surgical treatment options and historical case review, respectively.
Device Story
Move Forward 3D Motion Simulation Service is an online image analysis service for hip joint evaluation. Input consists of patient CT or MR image data uploaded to the CGOnline module. Zimmer Biomet operators use internal software modules (Arbiter and Articulis) to perform image segmentation, creating 3D bone models of the pelvis and femur. The system performs 3D motion simulations to visualize bone contact and range of motion, calculates intersection zones, and computes morphological shape parameters. Output is an interactive PDF report delivered to clinicians (radiologists and orthopedic surgeons) to assist in pre-operative or post-operative planning for hip preservation surgery, specifically regarding femoroacetabular impingement (FAI). Clinicians do not interact with the software directly; the service is managed by trained company operators.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical software verification and validation activities demonstrating safety and effectiveness comparable to the predicate device.
Technological Characteristics
Stand-alone software service; utilizes DICOM files from CT or MRI scanners. Features include image segmentation, 3D bone model rendering, motion simulation algorithms, and morphological parameter calculation. Delivered via online portal (CGOnline).
Indications for Use
Indicated for skeletally mature individuals to obtain 3D motion simulation reports based on CT or MR image data of hip joints for clinical evaluation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Dyonics Plan Hip Impingement Planning Software (K132636)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2017
Biomet, Inc. % Mr. Paul Hardy Regulatory Affairs Senior Specialist 56 East Bell Drive PO Box 587 WARSAW IN 46581
Re: K162559
Trade/Device Name: Move Forward 3D Motion Simulation Service Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 23, 2017 Received: February 24, 2017
Dear Mr. Hardy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162559
Device Name
Move Forward 3D Motion Simulation Service
### Indications for Use (Describe)
Zimmer Biomet's Move Forward 3D Motion Simulation Service is an online image analysis service indicated for skeletally mature individuals that enables clinicians to obtain 3D motion simulation reports based on CT or MR image data.
The service can be used by uploading either CT or MR image data of hip joints. The image data is processed by Zimmer Biomet to create 3D anatomy models. These 3D models are then used to perform 3D motion simulations of the hip joint. The report also provides several calculated intersection zones that may improve the simulated range of motion. Move Forward also calculates several morphological shape parameters for each of the 3D anatomy models.
The software generates an interactive report as output. Clinicians do not interact with the image analysis software directly. The image analysis software is only operated by Zimmer Biomet operators who have been specifically trained for this purpose. End users of the generated Move Forward reports are trained medical professionals, including radiologists and orthopedic surgeons.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are dark gray, and the image has a slightly blurred effect. The word appears to be a logo or brand name.
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Move Forward 3D Motion Simulation Service 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Biomet, Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number: 1825034 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Paul Hardy<br>Regulatory Affairs Senior Specialist<br>574-372-6799 |
| Date: | February 22, 2017 |
| Subject Device: | Trade Name: Move Forward 3D Motion Simulation Service<br>Common Name: Hip impingement image analysis system<br>Classification Name:<br>• LLZ- Picture Archive and Communications System (PACS) (21 CFR<br>892.2050) |
# Legally marketed devices to which substantial equivalence is claimed:
- Dyonics Plan Hip Impingement Planning Software manufacturered by Smith & Nephew (K132636)
### Device Description
The Move Forward 3D Motion Simulation Service is an online image analysis service that enables clinicians to obtain 3D motion simulation reports based on CT or MR image data. This service consists of three modules: CGOnline, Arbiter, and Articulis.
During the segmentation process, 3D bone models of the pelvis and femur bones are created from the image data. Any required editing of the segmentation images are performed by Zimmer Biomet operators. These 3D bone models are then used to perform 3D motion simulations of the hip joint. The 3D motion simulations can be used to visualize rigid shapes such as bones that come into contact with one another, thus potentially limiting range of motion. The system includes several calculated intersection zones that each individually improve the simulated range of motion. The system also calculates several morphological shape parameters for each of the 3D bone models.
{5}------------------------------------------------
The software generates an interactive PDF report as output, which Zimmer Biomet delivers to clinicians on a per case basis through the CGOnline module. In particular, the report that is generated provides information to the clinicians associated with femoroacetabular impingement (FAI). Clinicians do not interact with the image analysis software directly. The Articulis and Arbiter modules are only operated by Zimmer Biomet operators who have been specifically trained for this purpose. End users of the generated Move Forward reports are trained medical professionals, including radiologists and orthopedic surgeons.
# Indications for Use
Zimmer Biomet's Move Forward 3D Motion Simulation Service is an online image analysis service indicated for skeletally mature individuals that enables clinicians to obtain 3D motion simulation reports based on CT or MR image data.
The service can be used by uploading either CT or MR image data of hip joints. The image data is processed by Zimmer Biomet to create 3D anatomy models. These 3D models are then used to perform 3D motion simulations of the hip joint. The report also provides several calculated intersection zones that may improve the simulated range of motion. Move Forward also calculates several morphological shape parameters for each of the 3D anatomy models.
The software generates an interactive report as an output. Clinicians do not interact with the image analysis software directly. The image analysis software is only operated by Zimmer Biomet operators who have been specifically trained for this purpose. End users of the generated Move Forward reports are trained medical professionals, including radiologists and orthopedic surgeons.
# Intended Use
The Move Forward 3D Motion Service is intended as pre-operative or post-operative software for simulating/evaluating hip preservation surgical treatment options and historical case review, respectively.
# Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Indications for Use: The Dyonics Plan Hip Impingement Planning System software is intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner such as a CT scanner to an output file. It is also intended as pre-operative or post-operative software for simulating/evaluating hip preservation surgical treatment options and historical case review, respectively.
- . Intended Use: The Dyonics Plan is intended as a pre-operative or post-operative software for simulating/evaluating hip preservation surgical treatment options and historical case review respectively.
- Design Features: Stand-alone software packages that have the following features:
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are dark gray, and the word is slightly elongated horizontally. The overall impression is of a clean and modern design.
- Input files consist of DICOM files from CTs for Dyonics and CTs and MRIs for 3D O Move Forward Simulation Service
- o The proposed and predicate device both provide image processing tools including image segmentation and 3D rendering tools
- The proposed and predicate device both have tools for surgical simulation and O planning
- The proposed device allows for the display of anatomical parameters which the o predicate device also provides for anatomical parameters.
- The proposed and predicate device provide an output report which can be O referenced pre-operatively, post-operatively, and inter-operatively
- The proposed and predicate device provide range of motion simulations o
## Summary of Performance Data (Nonclinical and/or Clinical)
- . Non-Clinical Tests
- o Software Verification and Validation activities demonstrate that the Move Forward 3D Motion Simulation Service does not raise any new issues of safety and effectiveness as compared to the predicate device.
## Substantial Equivalence Conclusion
The differences between the proposed and predicate device do not introduce new types of safety and effectiveness questions. The Move Forward 3D Motion Simulation Service is substantially equivalent to the Dyonics Plan software and the intended uses are identical.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.