1012WCC, 1012WGC

K162518 · Rayence Co., Ltd. · MQB · Oct 6, 2016 · Radiology

Device Facts

Record IDK162518
Device Name1012WCC, 1012WGC
ApplicantRayence Co., Ltd.
Product CodeMQB · Radiology
Decision DateOct 6, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1680
Device ClassClass 2

Intended Use

1012WCC and 1012WGC Digital Flat Panel X-Ray Detector are indicated for digital imaging solution designed for human anatomy including head, neck, cervical spine, arm, leg and peripheral (foot, hand, wrist, fingers, etc.). They are intended to replace film based radio diagnostic systems and provide a case diagnosis and treatment planning for physicians and other health care professionals. Not to be used for mammography.

Device Story

1012WCC/1012WGC are wired/wireless digital flat-panel X-ray detectors using amorphous silicon (a-Si) TFT sensors. 1012WCC uses a Cesium Iodide (CsI:Tl) scintillator; 1012WGC uses a Gadolinium Oxysulfide (Gd2O2S:Tb) scintillator. Devices capture X-ray signals, convert them to digital RAW files, and transmit data via wireless LAN (IEEE 802.11a/g/n/ac) or wired connection to a console system. Used in clinical settings by physicians and healthcare professionals to replace film-based radiography. Output is a digital image for diagnostic analysis and treatment planning. Integration with a separate console software (e.g., Xmaru View V1/PACS) is required to convert RAW files to DICOM 3.0 format. Benefits include high-speed processing, improved operability, and digital image clarity for extremity and skeletal imaging.

Clinical Evidence

Clinical evaluation involved a comparative review of radiographs taken with subject devices versus predicates (1012WCA/1210SGA) by a licensed US radiologist. Images were assessed for spatial and soft tissue contrast resolution. Results indicated subject devices provided superior clarity, particularly in extremity films, with no difficulty in evaluating anatomical structures for diagnostic conclusions.

Technological Characteristics

Amorphous Silicon (a-Si) TFT flat-panel detector. Scintillators: CsI:Tl (1012WCC) or Gd2O2S:Tb (1012WGC). Dimensions: 328x268x15mm. Pixel pitch: 127 μm. Resolution: 3.9 lp/mm. Connectivity: Wireless LAN (IEEE 802.11a/g/n/ac) and wired. Power: Battery-operated. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62220-1 (image quality).

Indications for Use

Indicated for digital X-ray imaging of human anatomy including head, neck, cervical spine, extremities (arm, leg, foot, hand, wrist, fingers). Intended to replace film-based systems for diagnostic and treatment planning. Contraindicated for mammography.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, overlaid on a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 6, 2016 Rayence Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025 Re: K162518 Trade/Device Name: 1012WCC / 1012WGC Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: September 1, 2016 Received: September 9, 2016 Dear Mr. Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert Oals Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) ## K162518 Device Name 1012WCC 1012WGC Indications for Use (Describe) 1012WCC and 1012WGC Digital Flat Panel X-Ray Detector are indicated for digital imaging solution designed for human anatomy including head, neck, cervical spine, arm, leg and peripheral (foot, hand, wrist, fingers, etc.). They are intended to replace film based radio diagnostic systems and provide a case diagnosis and treatment planning for physicians and other health care professionals. Not to be used for mammography. Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |--------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 1. 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date 510k summary prepared: October 6, 2016 Submitter's Name, address, telephone number, a contact person: | Submitter's Name : | Rayence Co., Ltd. | |-------------------------|---------------------------------------------------------| | Submitter's Address: | 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea | | Submitter's Telephone: | +82-31-8015-6459 | | Contact person: | Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459 | | Official Correspondent: | Dave Kim (davekim@mtech-inc.net) | | (U.S. Designated agent) | | | Address: | 8310 Buffalo Speedway, Houston, TX 77025 | | Telephone: | +713-467-2607 | | Fax: | +713-583-8988 | Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known: | Trade/proprietary name: | 1012WCC / 1012WGC | |-------------------------|-----------------------------------------| | Common Name: | Digital Flat Panel X-ray Detector | | Classification Name : | 21CFR892.1680 / Stationary x-ray system | | Product Code: | MQB | | Predicate Device : | | | Manufacturer : | Rayence Co., Ltd. | | Device : | 1012WCA | | 510(k) Number : | K123345 | | Common Name: | Digital Flat Panel X-ray Detector | | Classification Name : | 21CFR892.1680 / Stationary x-ray system | | Product Code: | MQB | {4}------------------------------------------------ | Reference Device : | | |-----------------------|-----------------------------------------| | Manufacturer : | Rayence Co., Ltd. | | Device : | 1210SGA | | 510(k) Number : | K113630 | | Common Name: | Digital Flat Panel X-ray Detector | | Classification Name : | 21CFR892.1680 / Stationary x-ray system | | Product Code: | MQB | ## 2. Device Description 1012WCC / 1012WGC is a wired/wireless digital solid state X-ray detector that is based on flat-panel technology. The wireless LAN(IEEE 802.11a/g/n/ac) communication signals images captured to the system and improves the user operability through high-speed processing. This radiographic image detector and processing unit consists of a scintillator coupled to an a-Si TFT sensor. This device needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis. The RAW files can be further processed as a DICOM compatible image file by a separate console SW program (K160579 / Xmaru View V1 and Xmaru PACS/ Rayence Co.,Ltd.) for a diagnostic analysis. ### 3. Indication for use 1012WCC and 1012WGC Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for human anatomy including head, neck, cervical spine, arm, leg and peripheral (foot, hand, wrist, fingers, etc.). It is intended to replace film based radiographic diagnostic systems and provide a case diagnosis and treatment planning for physicians and other health care professionals. Not to be used for mammography. ### 4. Summarv of Design Control Risk management The 1012WCC / 1012WGC digital X-ray detector is a modification of 1012WCA(K123345), the predicate device which was developed for the purpose of portable imaging. 1012WCC / 1012WGC is slightly thinner, smaller and lighter than 1012WCA. The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted. {5}------------------------------------------------ # 5. Summary of the technological characteristics of the device compared to the predicate device: 1012WCC and 1012WGC detector described in this 510(k) has the same indications for use and similar technical characteristics as its predicate device, 1012WCA. | 5.1<br>Comparison table | | | |-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristic | Proposed<br>Rayence Co.,Ltd.<br>1012WCC / 1012WGC | Predicate<br>Rayence Co.,Ltd.<br>1012WCA | | Feature | Image: Rayence Co.,Ltd. 1012WCC / 1012WGC | Image: Rayence Co.,Ltd. 1012WCA | | 510(k) number | | K123345 | | Intended Use | 1012WCC and 1012WGC Digital Flat<br>Panel X-Ray Detector are indicated for<br>digital imaging solution designed for<br>human anatomy including head, neck,<br>cervical spine, arm, leg and peripheral<br>(foot, hand, wrist, fingers, etc.). They are<br>intended to replace film based radiographic<br>diagnostic systems and provide a case<br>diagnosis and treatment planning for<br>physicians and other health care<br>professionals. Not to be used for<br>mammography. | 1012WCA Digital Flat Panel X-Ray<br>Detector is indicated for digital imaging<br>solution designed for human anatomy<br>including head, neck, cervical spine,<br>arm, leg and peripheral (foot, hand,<br>wrist, fingers, etc.). It is intended to<br>replace film based radiographic<br>diagnostic systems and provide a case<br>diagnosis and treatment planning for<br>physicians and other health care<br>professionals. Not to be used for<br>mammography. | | Detector Type | Amorphous Silicon, TFT | Amorphous Silicon, TFT | | Scintillator | 1012WCC CsI:Tl<br>1012WGC Gd2O2S:Tb | CsI:Tl | | Imaging Area | 10 x 12 inches | 11 x 13 inches | | Pixel matrix | 2304 X 1792 | 2560 X 2080 | | Pixel pitch | 127 μm | 127 μm | | Resolution | 3.9 lp/mm | 3.9 lp/mm | | A/D conversion | 14 / 16 bit | 14 / 16 bit | | Preview time | ≤2 | ≤3 | | Data output | RAW<br>*The RAW files are convertible into DICOM<br>3.0 by console S/W | RAW<br>*The RAW files are convertible into DICOM<br>3.0 by console S/W | | Dimensions | 328 X 268 X 15 mm | 395 x 337 x 18 mm | | Weight | 3.5 kg (incl. battery) | 3.15 kg | | Application | Wireless/Wired portable system | Wireless portable system | | | stand. | stand | | <b>Added</b> | Battery & Battery Charger | - Battery & Battery Charger | | <b>Optional</b> | Interface Box | Interface Box | | <b>Components</b> | IrDA module | | #### Comparison table 5.1 {6}------------------------------------------------ ### 5.2 Scintillator layer 1012WCC and 1012WGC have the same hardware, software and components. The type of a scintillator layer is different. ( * scintillator : a phosphor that produces scintillations) | | Proposed | Predicate | Reference | |--------------------------------------|----------|-----------|-----------| | CsI<br>(Cesium Iodide) | 1012WCC | 1012WCA | | | Gd2O2S:Tb<br>(Gadolinium Oxysulfide) | 1012WGC | | 1210SGA | ## 5.3 Added Optional Components (Comparison with Predicate device) | Components | Description | |-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Battery & Battery Charger<br>Image: Battery and battery charger | Sources of electricity. | | Mobile Battery Charger<br>Image: Mobile battery charger | | | Interface Box<br>Image: Interface box | 1) Connector to synchronize the detector and the generator.<br>2) Data transfer and battery charge while the detector is in use<br>(Connect between the detector and Interface Box), Up to three<br>detectors can be connected.<br>3) Transmitting an image/command between the detector and PC.<br>4) Wireless AP. | | IrDA module<br>Image: IrDA module | Sharing function for PC and the detector. | {7}------------------------------------------------ ### Recommended Generator Specification | Model | Manufacture | | 32kW | 40kW | 50kW | |----------------|--------------------------------------|-----|--------------------|--------|--------| | CMP 200 | Communications & Power<br>Industries | kVp | 40-125 | | 40-150 | | | | mA | 10-400 | 10-500 | 10-630 | | EDITOR HFe 501 | Rontgenwerk Bochum | kVp | 40-150 | | | | | | mA | 10-630 | | | | UD150L-40E/40F | Shimadzu | kVp | 40-150 | | | | | | mA | @100 kVp- 500(320) | | | | | | mA | @80 kVp- 630(400) | | | | PXR-321B | Poskom Co.,Ltd. | kVp | 125/150 | | | | | | mA | 500 | | | Image /page/7/Picture/3 description: The image shows a caution sign. The sign is a yellow triangle with a black border and a black exclamation point in the center. Below the triangle, the word "CAUTION" is written in black, bold letters. To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators which are not listed above, please contact vour Ravence representative. # 6. Summary of Performance Testing 1012WCC and 1012WGC Digital Flat Panel X-Ray Detectors have the same indications for use, material, form factor, performance, and safety characteristics compared to the predicate devices, 1012WCA. The non-clinical test and clinical consideration test for each subject device were performed to demonstrate the substantial equivalency of the subject devices compared to the predicate device. The nonclinical test report contains the MTF, DQE and NPS test results of 1012WCC by using the identical test equipment and same analysis method described by IEC 62220-1. The comparative result of the DQE test for 1012WCC detector with respect to the predicate device demonstrated that the DQE of the both subject devcies performed with the predicate device. The DQE represents the ability to visualize object details of a certain size and contrast. 1012WCC has higher DQE performance at high spartial frequencies, especially from 2.5 lp/mm to 4 lp/mm, compared with 1012WCA. The comparison of the MTF and DQE for 1012WGC detector {8}------------------------------------------------ demonstrated that the MTF and DQE performance results for the subject device were almost same as those of 1210SGA (K113630), the reference device. | Spatial<br>Frequency | MTF Value | | |----------------------|-----------|-------| | 1.0 lp/mm | 0.560 | 0.575 | | 2.0 lp/mm | 0.259 | 0.278 | | 3.0 lp/mm | 0.143 | 0.143 | | 3.5 lp/mm | 0.108 | 0.101 | | 3.93 lp/mm | 0.086 | 0.074 | | Spatial<br>Frequency | MTF Value | | | 1.0 lp/mm | 0.588 | 0.585 | | 2.0 lp/mm | 0.275 | 0.264 | | 3.0 lp/mm | 0.131 | 0.119 | | 3.5 lp/mm | 0.087 | 0.076 | | 3.93 lp/mm | 0.062 | 0.053 | | | 1012WCC | 1012WCA | |--------|---------|---------| | DQE(0) | 0.778 | 0.753 | | | 1012WGC | 1210SGA | | DQE(0) | 0.437 | 0.470 | To further demonstrate the substantial equivalency of two devices, clinical images are taken from both subject devices and reviewed by a licensed US radiologist to render an expert opinion. Both the test subjects.1012WCC and 1012WGC, have been evaluated and compared to 1012WCA and 1210SGA, respectively, by taking sample radiographs of similar age groups and anatomical structures in accordance with the test protocol of diagnostic radiography evaluation procedure. After a broad review of plain radiographic images taken with the 1012WCC/1012WGC and the 1012WCA/1210SGA, the images obtained with the 1012WCC/1012WGC were superior to the same view obtained from a similar patient with the 1012WCA/1210SGA, respectively. In general, both the spatial and soft tissue contrast resolution are superior using the 1012WCC/1012WGC. Specifically, the soft tissues on extremity films were seen with better clarity. There is little difficulty in evaluating a wide range of anatomic structures necessary to provide a correct conclusion. Based on the non-clinical and clinical consideration test and the outcome of a comparative review by an expert for both devices, the sponsor claims the substantial equivalency between the subject devices and their predicate device in terms of diagnostic image quality. {9}------------------------------------------------ The manufacturing facility is in conformance with the design control procedure requirements and the relevant EPRC standards as specified in 21 CFR 802.30 and the records are available for review. # 7. Summary for any testing in the submission: Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:2005 (3rd Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (Medical electrical equipment Part 1:General requirements for basic safety and essential performance) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2: 2007. Non-clinical & Clinical considerations according to FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory. # 8. Summary for risk management activities in the submission For 1012WCC vs 1012WCA | | Added function | Risk factor | Discussion | |---|------------------|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | Cassette<br>size | - Essential<br>performance<br>(Image size) | Implemented the cassette size to change total pixel matrix<br>(pixels): 2304x1792, total pixel area: 292.6x227.6, and<br>effective pixel matrix (pixels): 264x1752. The image<br>quality was tested through SSXI Nonclinical/ Clinical test<br>and to be equivalent compared to the predicate device. | | | | | The revised RM table (Att3-1) include the following risk: for<br>1012WCC - Operational hazard (Incorrect image data)<br>Highlighted in yellow (HI-601-1-4.3) | ### For 1012WGC vs 1012SGA | | Added function | Risk factor | Discussion | |---|----------------|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | Cassette size | -. Essential performance (Image size) | Implemented the cassette size to change total pixel matrix (pixels): 2304x1792, total pixel area: 292.6x227.6, and effective pixel matrix (pixels): 264x1752. The image quality was tested through SSXI Nonclinical/ Clinical test and to be equivalent compared to the predicate device.<br>The revised RM table (Att3-1) include the following risk: for 1012WGC - Operational hazard (Incorrect image data) Highlighted in yellow (HI-601-1-4.3) | | 2 | Wireless | -. Battery | The risk related to a wireless product that uses a battery has been identified and mitigated according to IEC 60601-1, IEC62366 test, and SW validation.<br>The revised RM table; Att3-1 and Att 3-2 include the following risk: Operation hazard (Use error)-Incorrect replacement. Hazard ID (HI-601-1-7.3.3) and Operation | {10}------------------------------------------------ | | hazard-Batter connection, Hazard ID (HI-601-1-15.4.3.2)<br>Att3-3 SW validation also includes HZ 0017-power<br>instability caused by low battery power which is highlighted<br>in yellow. | |--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Wireless<br>signal | Att 3-3. SW validation also identified the following risk<br>and risk management activities regarding the wireless<br>signal; (HZ 0017-power instability caused by low battery<br>power.<br>(HZ 0019)-Misdiagnosis | | | Not possible to check the data validity of all data packets, | | | including CRC value, based on the IrDA communication | | | transmission in the PC and the detector<br>/HZ 0020)-Misdiagnosis | | | Image not transmitted to the User PC due to the network | | | transmission failure. | | | Check the connection between the detector and the User-PC | # 9. Conclusions: In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Rayence Co., Ltd. concludes that 1012WCC and 1012WGC are substantially equivalent in comparison with 1012WCA, the predicate device as described herein.