K162478 · Ivwatch, LLC · PMS · Dec 22, 2016 · General Hospital
Device Facts
Record ID
K162478
Device Name
ivWatch
Applicant
Ivwatch, LLC
Product Code
PMS · General Hospital
Decision Date
Dec 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162478 · Dec 22, 2016
ivWatch
Ivwatch, LLC
Clinical study in pediatric patients
The clinical study was used to demonstrate the safety and efficacy of the ivWatch Model 400 in a pediatric population (birth to 17 years old) to support an expansion of the indications for use.
Pediatric clinical study; Clinical validation study
Pediatric patients (birth to 17 years old)
Clinician detection of infiltration
Time difference between ivWatch detection and clinician detection, sensitivity, specificity
Indications for Use
The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for the subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by healthcare practitioners who have been trained in the use of the device.
Device Story
The ivWatch Model 400 provides continuous, non-invasive monitoring of tissue near PIV insertion sites to detect infiltration/extravasation. The system comprises a patient monitor (IPM), a reusable optical sensor cable, and a single-use sensor receptacle. The IPM emits visible and near-infrared light through the cable to the skin and measures reflected light. An internal signal processing algorithm analyzes changes in optical properties to identify infusate pooling in subcutaneous tissue. If an event is detected, the device triggers audible and visual notifications for the clinician. Used in healthcare settings by trained practitioners, the device aids in early detection, allowing for timely intervention to potentially reduce patient harm from infiltration. The subject device adds an RFID tag to the sensor cable for tracking useful life (replacing a USB method) and expands the indicated patient population to include pediatrics.
Clinical Evidence
Clinical study in pediatric patients (2 weeks to 17 years, weight 3.3-196.7 kg) evaluated performance. The device detected 80.0% of infiltrations (95% CI [51.9% to 95.7%]) before clinician detection. In a blinded non-alarming group, sensitivity was 78.3% (95% CI [56.5% to 92.5%]). Average detection time was 29.8 hours prior to clinician diagnosis. No serious adverse events occurred. Results support equivalence to adult clinical data.
Technological Characteristics
System includes IPM, reusable optical sensor cable, and single-use EO-sterilized sensor receptacle (ISO 11135-1). Uses visible/near-infrared light sensing. RFID tag (ISO 15693) for cable life tracking. Compliant with IEC 60601-1-2 (EMC), ANSI/AAMI ES60601-1 (Electrical Safety). Sensor cable is MR Conditional (ASTM F2052-15, F2213-06-11, F2119-07-13, F2182-11a). Biocompatibility per ISO 10993-1, -5, -10.
Indications for Use
Indicated for detection of subcutaneous infiltrations and extravasations (≤10 cc) of optically clear infusates in pediatric and adult patients with peripherally-inserted catheters (PIVs). Adjunctive to clinical assessment; not a substitute for regular clinician site inspection. For use by trained healthcare practitioners.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around a central emblem. The emblem consists of a stylized representation of three human profiles facing to the right, stacked one behind the other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2016
ivWatch, LLC Jaclyn Lautz Director of Regulatory Affairs and Quality Assurance 1100 Exploration Way, Suite 209 Hampton, Virginia 23666
Re: K162478
Trade/Device Name: ivWatch Model 400 Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: PMS Dated: November 22, 2016 Received: November 28, 2016
Dear Jaclyn Lautz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and name. The name "Tina Kiang -S" is written in a simple, sans-serif font. To the left of the name is a complex, looping signature in black ink. The background is a light blue color.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K162478
Device Name ivWatch Model 400
### Indications for Use (Describe)
The iv Watch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for the subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by healthcare practitioners who have been trained in the use of the device.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="true" type="checkbox"/> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <input type="checkbox"/> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized water droplet shape on the left, with the letters "iv" inside the droplet. To the right of the droplet is the word "Watch" in a simple, sans-serif font. The logo is clean and modern, with a focus on the company name.
ivWatch, LLC 1 100 Exploration Way, Suite 209 Hampton, VA 23666
1.855.IVWATCH (489.224) Fax: 757.224.5009 www.ivwatch.com
### TRADITIONAL 510(K) SUMMARY 1
#### 1.1 Administrative
| Submitter Name | ivWatch, LLC |
|--------------------------|------------------------------------------------------------------------------|
| Applicant Address | 1100 Exploration Way, Suite 209<br>Hampton, VA 23666 |
| Phone | 855-489-2824 |
| Fax | 757-224-5009 |
| Primary Contact | Jaclyn Lautz, Ph.D., Director of Regulatory Affairs and<br>Quality Assurance |
| Primary Contact Email | jaclyn.lautz@ivwatch.com |
| Primary Contact<br>Phone | 855-489-2824 x7023 |
| Date Prepared | September 1, 2016 |
### 1.2 Subject Device
| Trade Name | ivWatch Model 400 |
|---------------------|---------------------------------------------------------|
| Manufacturer | ivWatch, LLC |
| 510(k) Number | K162478 |
| Device Class | II |
| Regulation Number | 21 CFR 880.5725 |
| Product Code | PMS (Peripheral Intravenous (PIV) Infiltration Monitor) |
| Classification Name | Infusion Pump |
#### 1.3 Predicate Device
| Trade Name | ivWatch Model 400 |
|-----------------------|-----------------------------------------------------------------------------------------|
| Manufacturer | ivWatch, LLC |
| 510(k) Number | K153605 |
| Device Class | II |
| Regulation Number | 21 CFR 880.5725 |
| Product Code | PMS (Peripheral Intravenous (PIV) Infiltration Monitor) |
| Classification Name | Infusion Pump |
| Clearance Information | Design modification to the sensor cable sheathing of<br>the ivWatch Model 400 (K142374) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "ivWatch." The logo consists of a stylized teardrop shape above the letters "iv" in a sans-serif font, followed by the word "Watch" in a cursive font. The teardrop has a grid-like pattern within it. The logo is in black and white.
ivWatch IIC 1 100 Exploration Way, Suite 209 Hampton. VA 23666
1.855.IVWATCH (489.224) Fax: 757.224.5009 www.ivwatch.com
# Device Description
The ivWatch Model 400 is a medical device that provides continuous, non-invasive monitoring of human tissue adjacent to peripheral intravenous (PIV) insertion sites to aid in the early detection of infiltration and extravasation events. The ivWatch Model 400 consists of the ivWatch Patient Monitor (IPM), a reusable optical sensor cable and a single-use sensor receptacle.
The ivWatch Model 400 uses visible and near-infrared light to measure changes in the optical properties of the tissue near a PIV insertion site. The IPM contains an optical system that generates visible and near-infrared light signals that are sent through the optical sensor cable to the patient's skin. Simultaneously, the IPM measures the light reflected back through the optical sensor cable from the patient's skin. Measured changes between the emitted and reflected signals are processed by the ivWatch signal processing algorithm to determine if an infiltration event may have occurred. If changes in the optical properties of the tissue near the PIV insertion site are consistent with an infusate pooling in the subcutaneous tissue, the IPM emits audible and visual notifications intended to prompt the clinician to inspect the PIV site for a possible infiltration event.
### Indications for Use 1.5
The ivWatch Model 400 is indicated for the detection of subcutaneous infiltrations and extravasations of 10 cc or less of optically clear infusates, as an adjunctive device to the clinical evaluation in the healthcare setting of adults and pediatrics with peripherally-inserted catheters (PIVs). The device is indicated to assess patients for subcutaneous infiltrations and extravasations but should not serve as a substitute for regular clinician assessment of the PIV site. The ivWatch Model 400 is intended for use by healthcare practitioners who have been trained in the use of the device.
# 1.6 Comparison of the Subject Device to the Predicate Device
The subject device (ivWatch Model 400) is a modification to the legally marketed ivWatch Model 400 (K153605). The subject and predicate devices consist of the same components including the ivWatch Patient Monitor, a single-use sensor receptacle and a reusable optical sensor cable. The subject and predicate devices have the same intended use which is to aid in the detection of infiltrations and extravasations during peripheral IV infusion therapy. The subject and predicate devices have the same technological characteristics and principles of operation and there have been no significant changes to the design or materials of the subject device.
This submission includes labeling modifications to the indications for use and to the MR safety information and the addition of an RFID tag in the sensor cable to track sensor cable useful life. The indications for use has been revised to extend the target patient population from patients 18 years old or greater to a patient population including both pediatrics and adults. Clinical performance testing on the pediatric patient population under the age of 18 years demonstrates that the subject device is as safe and effective as the predicate device on patients 18
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for IVWatch. The logo consists of a stylized water droplet shape on the left, with the letters "iv" inside the droplet. To the right of the droplet, the word "Watch" is written in a simple, sans-serif font. The logo is clean and modern, with a focus on the company's name.
1 100 Exploration Way, Suite 209 Hampton. VA 23666
1.855.IVWATCH (489.224) Fax: 757.224.5009 www.ivwatch.com
years old and greater. The MR safety information for the reusable sensor cable has been revised from "MR Unsafe" to "MR Conditional" on the sensor cable label. MR safety testing has been performed and demonstrates that the subject device is as safe and effective as the predicate device with the appropriate MR safety labeling. The RFID replaces a previously used USB method for tracking sensor cable useful life.
# 1.7 Performance Data
# 1.7.1 Sterilization and Shelf Life
The ivWatch sensor receptacle is the only component of the system supplied sterile. The sensor receptacle is sterilized by ethylene oxide (EO) in compliance with ISO 11135-1 and 10993-7. In addition, the sensor receptacle passed package and sterility testing in compliance with ISO 11607-1, ASTM F1140/F1140M-13, ASTM 2096-11 and ASTM D1469-14 supporting a 2-year shelf life.
The ivWatch sensor cable is designed to be cleaned between uses and cleaning and disinfection validation testing results passed all pre-defined acceptance criteria in accordance with AAMI-TIR-2010, AAMI TIR30-2011 and Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling,
# 1.7.2 Biocompatibility
The sensor receptacle and sensor cable sheathing are classified as prolonged duration (greater than 24 hours but less than 30 days) patient-contacting components. The sensor receptacle and sensor cable of the subject device have not changed compared to the predicate device. Therefore, the biocompatibility evaluation of the patient-contacting components conducted in accordance with the FDA Guidance Document "Use of the International Standard ISO 10993-1" and ISO 10993-1, ISO 10993-5 and ISO 10993-10 has not changed.
1.7.3 Electromagnetic Compatibility (EMC) and Electrical Safety
EMC and electrical safety testing were conducted on the subject device and results show that the ivWatch Model 400 is in compliance with the requirements of IEC 60601-1-2: 3rd Ed. 2007 and ANSI/AAMI ES60601-1:2005, 3rd E.
## 1.7.4 Magnetic Resonance (MR) Environment Testing
The ivWatch sensor cable has been assessed in the MR imaging environment in accordance with ASTM F2052-15, ASTM F2213-06-11, ASTM F2119-07-13, and ASTM F2182- 11a. MR compatibility testing results indicate that the sensor cable is MR Conditional.
# 1.7.5 Software Verification and Validation Testing
Software verification and validation testing for the ivWatch Model 400 was conducted and documented in compliance with 21 CFR 820 and in accordance with the FDA's Guidance for Industry and Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and IEC 62304.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "ivWatch". The logo consists of a stylized teardrop shape in black, with a grid-like pattern of small squares in the upper-left portion of the teardrop. To the right of the teardrop is the word "ivWatch" in a stylized font, with the "i" and "v" in lowercase and the "W", "a", "t", "c", and "h" in lowercase. There is a small circle with an "r" inside of it to the right of the word "ivWatch".
ivWatch, LLC 1 100 Exploration Way, Suite 209 Hampton, VA 23666
1.855.IVWATCH (489.224) Fax: 757.224.5009 www.ivwatch.com
# 1.7.6 Performance Testing - Clinical
The predicate device submission included the clinical evaluation of the Model 400 in patients 18 years old or greater. The purpose of this 510(k) submission is to extend the Model 400 indications for use from patients 18 years old or greater to a target patient population including both pediatrics and adults. The pediatric IRB approved clinical study included in this submission demonstrates that the Model 400 is just as safe and effective in the pediatric population under the age of 18 years old as the use of the predicate device in the adult population. Safety and efficacy of the Model 400 has been demonstrated in accepted case studies representing an age distribution from 2 weeks to 17 years, patient weight measurements from 3.3 to 196.7 kg (mean, 34.5 kg), distributed skin pigmentations and similar proportions between male and female subjects.
The clinical validation studies in the pediatric population under the age of 18 years old assessed the performance (e.g. time difference between ivWatch detection and clinician detection of an infiltration, sensitivity and specificity) of the ivWatch Model 400. The ivWatch Model 400 detected 80.0% of infiltrations (95% Cl [51.9% to 95.7%) before detection by a clinician. There were no serious adverse events during the clinical study. Clinical results demonstrate that the ivWatch Model 400 is equivalent in the pediatric population as compared to the adult population.
| Item | Predicate Device<br>K153605 | Subject Device<br>K162478 | Comparison |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The ivWatch Model<br>400 is designed to<br>aid in the detection<br>of infiltrations and<br>extravasations<br>during peripheral IV<br>infusion therapy. | Same | Equivalent |
| Principles of<br>operation | During IV fluid<br>infusion, the ivWatch<br>sensor transmits an<br>optical signal<br>through the tissue;<br>the optical signal is<br>altered if the IV fluid<br>is accumulating in<br>the tissue<br>underneath the<br>sensor, surrounding<br>the intended<br>intravenous<br>administration route | Same | Equivalent |
| | ivWatch, LLC<br>1100 Exploration Way, Suite 209<br>Hampton, VA 23666 | | 1.855.IVWATCH (489.224)<br>Fax: 757.224.5009<br>www.ivwatch.com |
| | which may indicate<br>that fluid is not<br>being delivered to<br>the intended<br>intravenous<br>administration route<br>(i.e., if an infiltration<br>or extravasation has<br>occurred). | | |
| Monitor | ivWatch patient<br>monitor | Same | Equivalent |
| Single-use, sterile<br>patient-<br>contacting<br>component used<br>to secure the<br>sensor-end of<br>the sensor cable<br>onto the patient's<br>skin | ivWatch sensor<br>receptacle | Same | Equivalent |
| Optical cable to<br>transmit visible<br>and near-<br>infrared light<br>signals between<br>the monitor and<br>the patient's skin | ivWatch sensor<br>cable | Same | Equivalent |
| Power source | External power<br>supply and internal<br>lithium backup<br>battery (compliant<br>to IEC 62133:2012) | Same | Equivalent |
| Monitor MR<br>safety | MR Unsafe | Same | Equivalent |
| MR safety<br>labeling of<br>sensor cable | MR Unsafe | MR Conditional<br>Tested in<br>compliance with<br>ASTM F2052-15,<br>ASTM F2213-06-11,<br>ASTM F2119-07-13,<br>and ASTM F2182-<br>11g | MR<br>compatibility<br>testing<br>demonstrated<br>that the sensor<br>cable is MR<br>Conditional. |
| Image: ivWatch logo | ivWatch, LLC<br>1100 Exploration Way, Suite 209<br>Hampton, VA 23666 | | 1.855.IVWATCH (489.224)<br>Fax: 757.224.5009<br>www.ivwatch.com |
| Sterility of the<br>ivWatch sensor<br>receptacle | Sterilized by<br>ethylene oxide (EO)<br>in compliance with<br>ISO 11135-1.<br>Sterilant residuals<br>pass the<br>acceptance criteria<br>in compliance with<br>ISO 10993-7. | Same | Equivalent |
| ivWatch sensor<br>receptacle shelf<br>life | 2 years<br>Tested and passed<br>all acceptance<br>criteria in<br>compliance with ISO<br>11607-1, ASTM 2096-<br>11 and ASTM 2096-<br>11. | Same | Equivalent |
| Biocompatibility<br>of prolonged<br>patient-<br>contacting<br>components<br>(Sensor<br>receptacle and<br>sensor cable) | Tested and passed<br>all biocompatibility<br>acceptance criteria<br>for prolonged intact<br>skin contact in<br>compliance with ISO<br>10993-1, ISO 10993-5<br>and ISO 10993-10. | Same | Equivalent |
| Reprocessing of<br>the ivWatch<br>sensor cable | Cleaning and low-<br>level disinfection<br>testing passed all<br>acceptance criteria<br>in accordance with<br>AAMI-TIR-2010, AAMI<br>TIR30-2011 and<br>Reprocessing<br>Medical Devices in<br>Health Care<br>Settings: Validation<br>Methods and<br>Labeling. | Same | Equivalent |
| Sensor Cable<br>Use limit | 240 hours<br>Sensor cable life is<br>tracked with a USB<br>key tethered to<br>sensor cable. | 240 hours<br>Sensor cable life is<br>tracked with RFID<br>communication<br>between RFID<br>chip embedded<br>in sensor cable | RF wireless<br>testing and<br>results are in<br>compliance<br>with IEC 60601-<br>1-2:3rd Ed. 2007<br>and FCC Part |
| | | strain relief and<br>the optics board<br>in the IPM. | 15.225 and<br>15.702 |
| Electromagnetic<br>compatibility | Tested in<br>compliance with no<br>deviations from IEC<br>60601-1-2: 3rd Ed.<br>2007 | Same | Equivalent |
| Electrical safety | Tested in<br>compliance with no<br>deviations from<br>ANSI/AAMI ES60601-<br>1:2005, 3rd ed.<br>(Medical Electrical<br>Equipment - Part 1:<br>General<br>Requirements for<br>Basic Safety and<br>Essential<br>Performance); IEC<br>60601-1: 2005 +<br>CORR. 1 (2006) +<br>CORR. 2 (2007) | Same | Equivalent |
| Radiofrequency<br>wireless testing | n/a | Tested in<br>compliance with<br>no deviations from<br>IEC 60601-1-2: 3rd<br>Ed. 2007 and FCC<br>Part 15.225 and<br>15.207 | The RF<br>communication<br>frequency<br>between the<br>ISO 15693 tag<br>and antenna<br>inside the<br>patient monitor<br>is not<br>adjustable and<br>does not<br>transmit patient<br>data. Testing<br>demonstrates<br>that the RF<br>communication<br>is as safe as the<br>predicate<br>device. |
| Performance<br>testing - Clinical | Age: Patients 18<br>years old or greater | Age: Pediatrics<br>from birth up to | The patient<br>population of |
| | ivWatch, LLC<br>1100 Exploration Way, Suite 209<br>Hampton, VA 23666 | 1.855.IVWATCH (489.224)<br>Fax: 757.224.5009<br>www.ivwatch.com | |
| target patient<br>population | medium and dark<br>PIV site: Hand and<br>forearm | years old<br>Skin color: Light,<br>medium and ark<br>PIV site: Hand,<br>forearm,<br>antecubital fossa,<br>leg and foot | device<br>compared to<br>the predicate<br>device differs in<br>age to support<br>safety and<br>efficacy in the<br>pediatric<br>population. |
| Performance<br>testing - Clinical<br>results | Sensitivity: 96.4% of<br>infiltrations (95% CI<br>[91.4% to 98.7%])<br>were detected in<br>under 10 cc of fluid<br>False notifications: A<br>false positive<br>notification<br>occurred<br>approximately once<br>every 4 days. | Time difference<br>between ivWatch<br>and clinician<br>detection of<br>infiltration: Red<br>Check IV<br>notification issued<br>on average 29.8<br>hours prior to the<br>clinician<br>detection, 95% CI<br>[14.8, 48.8 hours].<br>Sensitivity: 78.3%<br>of infiltrations (95%<br>CI [56.5 to 92.5%])<br>were detected<br>prior to the<br>clinician<br>diagnosing the<br>infiltration in the<br>blinded non-<br>alarming group.<br>False notifications:<br>A false positive<br>notification<br>occurred<br>approximately<br>once every 4<br>days. A false<br>positive<br>notification<br>occurred<br>approximately<br>once every 10<br>days if<br>notifications | The clinical<br>performance<br>data of the<br>Model 400<br>support the<br>safety and<br>efficacy in the<br>pediatric<br>patient<br>population<br>(from birth up<br>to and<br>including 17<br>years old) as<br>compared to<br>the clinical<br>performance<br>data in<br>patient's 18<br>years or<br>greater. The<br>clinical<br>evaluation<br>demonstrates<br>that the<br>ivWatch Model<br>400 is<br>substantially<br>equivalent to<br>the predicate<br>device<br>submission. |
| Image: iv Watch logo | ivWatch, LLC<br>1100 Exploration Way, Suite 209<br>Hampton, VA 23666 | 1.855.IVWATCH (489.224)<br>Fax: 757.224.5009<br>www.ivwatch.com | |
| Performance<br>testing - Clinical<br>adverse events | There were no<br>serious adverse<br>events during the<br>clinical study. | related to painful<br>IV flushes and<br>device-related<br>issues are<br>excluded. | There are no<br>new safety<br>issues<br>associated with<br>the use of the<br>device in<br>pediatrics. |
## 1.7.7 Summary Table
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "ivWatch". The logo consists of a stylized black teardrop shape with small white squares inside, followed by the text "ivWatch" in a modern, sans-serif font. A vertical line is present on the right side of the logo.
{10}------------------------------------------------
{11}------------------------------------------------
# 1.8 Conclusions
The technological characteristics and principles of operation of the ivWatch Model 400 and the predicate device are the same. The clinical data provided in the submission supports the expansion of the indications for use to now include the pediatric patient population.
The results of the clinical study and performance testing indicate that the ivWatch Model 400 is substantially equivalent to the ivWatch Model 400 (K153605). The ivWatch Model 400 performs as intended and is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.