K162461 · Aroa Biosurgery (Formerly Mesynthes Limited) · FTM · Dec 20, 2016 · General, Plastic Surgery
Device Facts
Record ID
K162461
Device Name
Endoform Plastics and Reconstructive Matrix
Applicant
Aroa Biosurgery (Formerly Mesynthes Limited)
Product Code
FTM · General, Plastic Surgery
Decision Date
Dec 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Endoform® Plastics and Reconstructive Matrix is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one-time use.
Device Story
Endoform Plastics and Reconstructive Matrix is an advanced collagen matrix composed of >70% natural, non-reconstituted ovine-derived collagen. The device is supplied sterile, dry, and in various sizes/thicknesses. Surgeons trim the matrix to meet patient needs and fix it to surrounding tissue using sutures, staples, or tacks. It is used to reinforce damaged or ruptured soft tissue membranes and muscle flaps. The device is fabricated using a proprietary 'lug lamination' process, which physically bonds individual collagen sheets together without embroidery. By providing a scaffold for soft tissue reinforcement, the device supports surgical repair, potentially improving clinical outcomes in plastic and reconstructive procedures.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench testing, including tensile strength, suture retention, biochemical composition, endotoxin levels, dimensional verification, ball burst, modulus of elasticity, and delamination evaluation. Biocompatibility was assessed per ISO 10993-1.
Technological Characteristics
Material: >70% natural, non-reconstituted ovine-derived collagen. Form factor: Lyophilized, multi-laminate lugged sheets (0.15–1.5 mm thick). Sterilization: Validated to SAL 10^-6. Connectivity: None. Energy source: None (mechanical implant).
Indications for Use
Indicated for patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery where soft tissue weakness exists.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
SIS Plastic Surgery Matrix (Cook Biotech) (K034039)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract design of an eagle with three human profiles incorporated into its body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2016
Aroa Biosurgery c/o Gordon MacFarlane, Ph.D., RAC ICON PLC 62 Forest Street, Suite 300 Marlborough, MA 01752
Re: K162461
Trade/Device Name: Endoform Plastics and Reconstructive Matrix Regulation Number: 21 CFR §878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM Dated: November 17, 2015 Received: November 21, 2016
Dear Dr. MacFarlane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director
> Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162461
Device Name
Endoform® Plastics and Reconstructive Matrix
Indications for Use (Describe)
Endoform® Plastics and Reconstructive Matrix is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one-time use.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
Image /page/3/Picture/2 description: The image contains a logo with an abstract design on the left and the text "AROA" on the right. The abstract design consists of a cluster of orange ovals arranged in a circular pattern, resembling a stylized sun or flower. The text "AROA" is in a bold, sans-serif font, with each letter evenly spaced and aligned horizontally.
# 510(k) Summary for the Aroa Biosurgery Endoform® Plastics and Reconstructive Matrix (per 21CFR 807.92)
#### 1. SUBMITTER/510(K) HOLDER
Aroa Biosurgery, Limited 2 Kingsford Smith Place Auckland 2022 New Zealand
| Contact Person: | Tina O'Brien |
|-----------------|------------------------------|
| | Director, Regulatory Affairs |
| | tina.obrien@aroabio.com |
| Telephone: | +64 9 8693035 |
| Date Prepared: | August 19, 2016 |
#### 2. DEVICE NAME
| Proprietary Name: | Endoform® Plastics and Reconstructive Matrix |
|--------------------|----------------------------------------------|
| Common/Usual Name: | EndoGraft |
| Device Class: | Surgical Mesh |
| Product Code: | FTM |
#### 3. PREDICATE / REFERENCE DEVICES
Predicate Device:
- · SIS Plastic Surgery Matrix (Cook Biotech) (K034039)
- · Biodesign Sling, Biodesign Plastic Surgery Matrix, Biodesign Anal Fistula Plug (K161221)
Reference Devices:
- · Endoform® Reconstructive Template (Aroa Biosurgery) (K130547/K156366)
#### 4. DEVICE DESCRIPTION
Endoform® Plastics and Reconstructive Matrix is an advanced collagen matrix comprised of natural >70%, non-reconstituted collagen. Endoform® Plastics and Reconstructive Matrix is designed to be fixed, via sutures, staples or tacks to the surrounding tissue, at the discretion of the attending physician for applications in plastic and reconstructive surgery.
{5}------------------------------------------------
For example, the device can be surgically implanted to reinforce damaged or ruptured soft tissue membranes and reinforce muscle flaps. The device is supplied sterile and dry in a variety of sizes and thicknesses, which can be trimmed by surgeons to meet the individual patient's needs.
Endoform® Plastics and Reconstructive Matrix is laminated via a proprietary process termed 'lug lamination', which is a physical means of fabricating the devices, such that the individual sheets of the device are physically bonded to one another as opposed to using embroidery to sew the laminate of collagen matrix sheets.
#### 5. INDICATIONS FOR USE
Endoform® Plastics and Reconstructive Matrix is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one-time use.
# 6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICES
The predicate and reference devices are listed in the table below. The predicate and reference devices serve to support substantial equivalence for the technological characteristics. The technological characteristics of Endoform® Plastics and Reconstructive Matrix and SIS Plastic Surgery Matrix are equivalent in that they are both primarily composed of mammalian collagens and are indicated for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery.
| Device Status | Proposed Device | Predicate | Predicate | Reference Predicate |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Aroa | Cook Biotech | Cook Biotech | Aroa |
| 510K Number | Proposed | K034039 | K161221 | K130547/K153633 |
| Device name | Endoform® Plastics & Reconstruction Matrix | SIS Plastic Surgery Matrix | Biodesign Plastic Surgery Matrix | Endoform® Reconstructive Template |
| Classification Name | Surgical mesh (FTM)<br>21 CFR 878.3300<br>Class II | Surgical mesh (FTM)<br>21 CFR 878.3300<br>Class II | Surgical mesh (FTM)<br>21 CFR 878.3300<br>Class II | Surgical mesh (FTL/FTM)<br>21 CFR 878.3300<br>Class II |
| Intended use | Endoform® Plastics and Reconstructive Matrix is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in | The SIS Plastic Surgery Matrix is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. | For implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic or reconstructive surgery. | Endoform® Reconstructive Template is intended for use as a surgical mesh to reinforce and/or repair soft tissue where weakness exists. Indications for use include, but are not limited to the |
### Side-by-Side Comparison of the Proposed Device and the Predicate and Reference Devices
{6}------------------------------------------------
| Device Status | Proposed Device | Predicate | Predicate | Reference Predicate |
|---------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Aroa | Cook Biotech | Cook Biotech | Aroa |
| | plastic and<br>reconstructive surgery.<br>The device is supplied<br>sterile and is intended<br>for one-time use. | The device is supplied<br>sterile and is intended<br>for one-time use. | | following procedures;<br>hernioplasty and repair<br>of body wall defects.<br>The device allows<br>reinforcement or<br>bridging of a deficit to<br>obtain the desired<br>surgical outcome.<br>Endoform®<br>Reconstructive<br>Template is intended<br>for single use only. |
| Animal origin | Ovine | Porcine | Porcine | Ovine |
| Tissue type | Forestomach | Small intestinal<br>submucosa | Small intestinal<br>submucosa | Forestomach |
| Nominal sizes | Sizes ranging from 1<br>cm² to 400 cm²,<br>lugged | 2 to 70 mm width<br>up to 200 mm length | 2 to 70 mm width<br>up to 200 mm length | Sizes ranging up to<br>400 cm² |
| Thickness | 0.15 – 1.5 mm | 0.1 – 1.5 mm | 100 to 1500 µm thick | Approximately 0.15 –<br>1.20 mm |
| Presentation | Sterile, lyophilized,<br>multi-laminate lugged<br>sheets in a peel pouch | The device is supplied<br>sterile in double pouch<br>system. | The device is supplied<br>sterile in double pouch<br>system. | Sterile, lyophilized<br>sheets in peel pouch |
| Components | Ovine derived<br>collagen | Porcine derived<br>collagen | Porcine derived<br>collagen an | Ovine derived<br>collagen and<br>polyglycolic acid<br>(PGA) suture material |
The technological characteristics of Endoform® Plastics and Reconstructive Matrix and the reference device Endoform® Reconstructive Template are equivalent in that they are both primarily composed on ovine ('sheep') collagen and indicated for implantation for soft tissue repair or reinforcement in plastic and reconstructive surgery. Both surgical mesh devices are manufactured from the same ovine source material using an identical manufacturing process.
{7}------------------------------------------------
## 7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Bench testing was conducted on the Endoform® Plastics and Reconstructive Matrix to support the substantial equivalence of the proposed device to the predicate device, as detailed below:
- tensile strength ●
- suture retention ●
- biochemical composition ●
- endotoxin
- dimensional verification ●
- ball burst
- modulus of elasticity ●
- delamination evaluation
In addition to bench testing, validation of the manufacturing process, labeling, packaging transportation, and sterilization to achieve a sterility assurance level (SAL) of 10° were performed. Results of the testing demonstrate that the proposed device meets all product specifications for the intended use.
Biocompatibility of the proposed device has been assessed via ISO 10993-1 and has been concluded that the device is biocompatible based on the established safety of the reference device.
As the proposed predicate and reference devices are technologically equivalent, safety and performance of the proposed device is based on the in vivo performance of the predicate and reference devices.
# 8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
There was no clinical testing required to support the indications for use as they are equivalent to the predicate device.
# 9. SUMMARY OF OTHER INFORMATION
No additional information was provided.
# 10. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
Based on the non-clinical and animal testing completed, and the comparisons with the predicate and reference devices, Endoform® Plastics and Reconstructive Matrix does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.