The Solero Microwave Tissue Ablation (MTA) System is indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not indicated for cardiac use.
Device Story
Solero Microwave Tissue Ablation (MTA) System is a software-controlled microwave generator with an integrated peristaltic pump; used for soft tissue ablation during open surgical procedures. System inputs include user-selected power/time settings via touchscreen; requires chilled saline source for applicator cooling. Generator delivers 2.45 GHz microwave energy to disposable applicators; energy thermally targets tissue to induce coagulation and ablation. System includes footswitch for alternate activation control. Healthcare providers (surgeons) operate the system in clinical settings; output is visual feedback on touchscreen (ablation time, power, system status). Device benefits include controlled, scalable soft tissue ablation (up to 4cm diameter in ≤ 6 minutes). Safety features include temperature monitoring at applicator tip; system alarms and terminates energy delivery if tip coolant reaches 48°C.
Clinical Evidence
No human clinical data. Evidence includes bench testing (mechanical, electrical, thermal, reliability) per IEC 60601-1 and IEC 60601-1-2, human factors/simulated use testing, and two GLP-compliant acute porcine studies. Animal studies compared Solero to predicate (K122762) and reference (K113237) devices, evaluating safety (ECG, respiration, temperature) and performance at maximum settings (140W, 6 minutes) in liver, kidney, and lung tissue models.
Technological Characteristics
Solid-state microwave generator; 2.45 GHz frequency; 60-140W output range. Applicators feature silicone-coated stainless steel shafts (1.83-2mm OD). Integrated peristaltic pump for saline cooling. Touchscreen interface. Connectivity includes footswitch and power cable. Complies with IEC 60601-1 and IEC 60601-1-2. Software-controlled.
Indications for Use
Indicated for ablation of soft tissue during open surgical procedures. Not for cardiac use.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Acculis Accu2I pMTA Applicator and SulisVpMTA Generator (K122762)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines above them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2017
Ms. Kasey E. Newcomb Specialist I, Regulatory Affairs Angiodynamics, Inc. 26 Forest Street Marlborough, MA 01752
Re: K162449
Trade/Device Name: Solero Microwave Tissue Ablation (MTA) System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NEY Dated: March 31, 2017 Received: April 3, 2017
Dear Ms. Newcomb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and 809]); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS)
{2}------------------------------------------------
regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K162449
Device Name
Solero Microwave Tissue Ablation (MTA) System
Indications for Use (Describe)
The Solero Microwave Tissue Ablation (MTA) System is indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not indicated for cardiac use.
Type of Use (Select one or both, as applicable) Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
# 510(k) Summary – Solero Microwave Tissue Ablation (MTA) System
Date Prepared: April 28, 2017
#### A. Submitter Information
| Submitter | Angiodynamics Inc. |
|-----------|----------------------------------|
| | 26 Forest St |
| | Marlborough, MA 01752 |
| Tel: | (508) 658-7813 |
| Fax: | (508) 263-7976 |
| Contact: | Kasey E. Newcomb, |
| | Regulatory Affairs, Specialist I |
| Email: | knewcomb@angiodynamics.com |
| | |
### Or
Linda J. Varroso AngioDynamics, Inc. 26 Forest Street Marlborough, MA 01752 Tel: (508) 658-7967 Fax: (508) 263-7976 lvarroso@angiodynamics.com
#### B. Trade Name System
| Trade Name: | Solero Microwave Tissue Ablation (MTA)<br>System |
|----------------------|-------------------------------------------------------------------|
| Common Name: | Microwave Tissue Ablation System and<br>Accessories |
| Classification Name: | Electrosurgical Cutting and Coagulation Device<br>and Accessories |
| Regulation Number | 21CFR 878.4400 |
| Product Code: | NEY |
| Class: | II |
{5}------------------------------------------------
#### C. Predicate Devices
| Primary Predicate: | K122762 Acculis Accu2I pMTA Applicator and SulisVpMTA Generator |
|----------------------|-------------------------------------------------------------------|
| Reference Device: | K113237 NeuWave Medical, Inc - Certus 140 System |
| Common Name: | Microwave Ablation System and Accessories |
| Classification Name: | Electrosurgical Cutting and Coagulation Device<br>and Accessories |
| Regulation Number | 21CFR 878.4400 |
| Product Code | NEY |
| Class: | II |
#### D. Device Description
The Solero Microwave Tissue Ablation (MTA) System is a softwarecontrolled, microwave generator with an integrated peristaltic pump that surgically ablates soft tissue through sterile applicators. The system may be used during open procedures for the ablation of soft tissue.
The Solero Generator is distributed with a main power cable and a footswitch, which may be used as an alternate means of controlling microwave activation in place of the microwave button on the front of the generator. Power is delivered through the disposable Solero Applicator, which are provided separately. A chilled saline source is required to maintain the Solero Applicators at an appropriate temperature.
#### E. Indications for Use
The Solero Microwave Tissue Ablation (MTA) System and Accessories is indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not intended for cardiac use.
{6}------------------------------------------------
#### F. Technological Characteristics
| | Subject Device | Predicate Device | Reference Device |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Solero Microwave Tissue | K122762 | K113237 |
| | Ablation (MTA) System | Acculis Accu2i pMTA | NueWave |
| Feature | | Applicator and Sulis VpMTA | Certus 140 2.45 GHz |
| | | Generator | Ablation System and |
| | | | Accessories |
| Indications For Use | The Solero Microwave<br>Tissue Ablation (MTA)<br>System is indicated for the<br>ablation of soft tissue during<br>open procedures. The<br>Solero MTA System is not<br>indicated for cardiac use. | Acculis Accu2i pMTA<br>Applicator with Sulis V<br>pMTA Generator software<br>release 2.1 is indicated for<br>the intraoperative<br>coagulation of soft tissue. | The NeuWave Medical<br>Certus 140'M 2.45 13Hz<br>Ablation System and<br>Accessories are intended for<br>the ablation (coagulation) of<br>soft tissue in percutaneous,<br>open surgical and in<br>conjunction with<br>laparoscopic<br>surgical settings. |
| Applicator Features | Applicator with umbilical<br>cable, IV spike, saline tubing<br>and cartridge connector | Applicator with umbilical<br>cable, IV spike, saline tubing<br>and cartridge connector | Applicator with a handle, a<br>cannula, a radiating section<br>and a faceted tip for<br>insertion |
| Generator Features | Generator touch screen<br>display provides information<br>on ablation/coagulation<br>time, microwave delivered<br>power, and system specific<br>information | Generator touch screen<br>display provides information<br>on ablation/coagulation<br>time, microwave delivered<br>power, temperature probe<br>readings and system specific<br>information | Generator touch screen<br>display provides information<br>on ablation time, power,<br>probe temperature and<br>system specific information |
| Operating Principle | Generator delivers<br>microwave energy to the<br>applicator tip to thermally<br>target tissue, resulting in<br>coagulation and ablation | Generator delivers<br>microwave energy to the<br>applicator tip to thermally<br>target tissue, resulting in<br>coagulation and ablation | Generator delivers<br>microwave energy to the<br>applicator tip to thermally<br>target tissue, resulting in<br>coagulation and ablation |
| Ablated/Coagulated Tissue | Near spherical volumes,<br>scalable to approximately 4<br>cm diameter in ≤ 6 minutes | Near spherical volumes,<br>scalable to approximately 4<br>cm diameter in ≤ 6 minutes | < 4cm ablation in all tissues<br>(PR),<br>< 7 cm ablations in liver,<br>< 6 cm in kidney (LK) |
| Applicator Patient<br>Contacting Materials | Silicone coated shaft | Silicone coated shaft | Stainless steel, insulated<br>shaft |
| Applicators Length | Standard: 14 cm<br>Intermediate: 19 cm<br>Long: 29 cm | Standard: 14 cm<br>Intermediate: 19 cm<br>Long: 29 cm | 15 cm<br>20 cm |
| Applicators Outer Diameter | Stainless steel shaft 1.83mm<br>Stainless steel shaft with<br>silicone coating ≤2mm | 1.83 mm | 1.47mm |
| Thermocouple Location | 2.7cm from the tip | 2.7cm from the tip | Unspecified |
| | Subject Device | Predicate Device | Reference Device |
| | Solero Microwave Tissue<br>Ablation (MTA) System | K122762 | K113237 |
| Feature | | Acculis Accu2i pMTA<br>Applicator and Sulis VpMTA<br>Generator | NueWave<br>Certus 140 2.45 GHz<br>Ablation System and<br>Accessories |
| Max Tip Temperature | Temperatures below 48°C,<br>they alert the user when the<br>coolant begins to become<br>warm (38°C), and an alarm<br>will sound and microwave<br>energy delivery will<br>terminate when the coolant<br>in the tip reaches 48°C | Temperatures below 48°C,<br>they alert the user when the<br>coolant begins to become<br>warm (38°C), and an alarm<br>will sound and microwave<br>energy delivery will<br>terminate when the coolant<br>in the tip reaches 48°C | Unspecified |
| Cable Length | 295.25cm for all<br>applicator lengths | Standard: 200.1cm<br>Intermediate: 192.0cm<br>Long: 176.1cm | Unspecified |
| Generator Frequency | 2.45 GHz | 2.45 GHz | 2.45 GHz |
| Reflected Power Threshold | -5.4 dB | -5.4 dB | Unspecified |
| System Output Power<br>Range | 60 to 140 watts in 20 watt<br>increments | 60 to 140 watts in 20 watt<br>increments | 95W per channel or up to<br>140W in a single channel |
| Microwave Generation | Solid-state microwave<br>generator | Magnetron | Unspecified |
{7}------------------------------------------------
{8}------------------------------------------------
#### G. Safety and Performance
The results of verification and validation activities were performed in accordance with design control requirements per 21 CFR 820.30 and demonstrate that the proposed device Solero Microwave Tissue Ablation (MTA) System meets predetermined performance specifications.
Performance testing that was completed included:
- . Tip Puncture Force
- Flexural Tip Strength ●
- Removal Force
- Flow Rate
- Bag Spike
- Cartridge Connection Force ●
- Cartridge Alignment
- Device Recognition
- Power Output
- Shaft to Applicator Tip Tensile
- Handle to Shaft Tensile
- Umbilical to Handle Tensile ●
- Umbilical to Cartridge ● Tensile
- . Pump Tubing to Cartridge Tensile
- . Pump Tubing to Bag Spike Tensile
- Saline Temperature
- Multiple Ablations
- Track Ablation ●
- Temperature Control ●
- Dielectric Strength ●
- Cartridge Connection
- Pump Compatibility
- Cartridge Attachment Force ●
- Cartridge Removal Force
- . Generator Hardware testing
- Complex programmable logic device
- . Reliability Testing
The performance evaluation plan included testing per the following recognized standards to assess conformance to IEC 60601 (300 Edition).
| IEC 60601-1 | Medical Electrical Equipment – Part 1: General<br>Requirements for Safety |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 | Medical Electrical Equipment – Part 2: General<br>Requirements for Basic Safety and Essential<br>Performance – Collateral Standard Electro<br>Magnetic Compatibility - Requirements and Test |
Human Factors Evaluation:
Simulated Use / Human Factors Testing has been conducted to evaluate the application of the Solero Microwave Tissue Ablation (MTA) System (i.e., Solero System) when used to ablate soft tissue.
{9}------------------------------------------------
Animal Studies:
Performance characterization was performed on three difference exvivo tissue; bovine liver, porcine kidney, and porcine lung, as these are considered to be comparable to human tissue. Testing was performed at varying time settings and power limits. The purpose of this testing was to support the recommended ablation settings, as well as demonstrate substantial equivalence to the predicate device.
Two GLP compliant animal studies were performed. The first study was to evaluate and establish substantial equivalence of the Solero Microwave Tissue Ablation (MTA) System to its predicate device Acculis Accu2I pMTA Applicator and SulisVPMTA Generator (K122762) and reference device Certus 140 System (K113237). The study also looked at the fundamental operational characteristics for the Solero MTA System. Safety was assessed from: animal behavior, ECG's, respiration rates, and temperature.
The second study's objective was to evaluate the safety and performance of the Solero Microwave Tissue Ablation (MTA) System to its predicate device Acculis Accu21 pMTA Applicator and SulisVpMTA Generator (K122762) during open surgical procedure in an acute porcine model. This study specifically evaluated and compared the safety at the highest power (140W) and time (6 minutes) of the two devices.
#### H. Substantial Equivalence Conclusion
The proposed device is equivalent with respect to the basic system design and function to that of the predicate device and reference device. The differences between the predicate, reference device and proposed device do not raise new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.