← Product Code [LOX](/productcode/LOX) · K162434

# NC Gallant PTCA Catheter (K162434)

_Medinol, Ltd. · LOX · Jan 19, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162434

## Device Facts

- **Applicant:** Medinol, Ltd.
- **Product Code:** [LOX](/productcode/LOX.md)
- **Decision Date:** Jan 19, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5100
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The NC Gallant PTCA Catheter is indicated for: - Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion; - Post-deployment stent expansion.

## Device Story

NC Gallant PTCA Catheter is a sterile, single-use, rapid exchange (RX) intravascular device; features double lumen design with expandable non-compliant (NC) coronary balloon. Used by physicians in clinical settings (e.g., cath lab) for percutaneous transluminal coronary angioplasty. Device is inserted into coronary vasculature to dilate stenotic lesions or expand stents; improves myocardial perfusion by restoring blood flow. Performance relies on mechanical inflation of balloon to specific pressures. Benefits include minimally invasive restoration of coronary artery patency.

## Clinical Evidence

Bench testing only. Includes biocompatibility (cytotoxicity, sensitization, toxicity, hemolysis, thromboresistance, mutagenicity, complement activation) and in-vitro performance (handling, crossing profile, burst pressure, fatigue, bond strength, flexibility, torque, radiopacity, coating integrity). Also includes a GLP acute performance study in a porcine model.

## Technological Characteristics

Double lumen rapid exchange (RX) catheter; non-compliant (NC) coronary balloon. Materials and design conform to FDA Class II Special Controls for PTCA Catheters (Sept 2010). Sterilization method not specified. Standalone mechanical device; no software or electronic components.

## Regulatory Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

## Predicate Devices

- NC Trek, Abbott ([K110134](/device/K110134.md))
- NC Quantum, Boston Scientific ([K121667](/device/K121667.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 19, 2017

Medinol Ltd. % Semih Oktay President Cardiomed Device Consultants, LLC 5523 Research Park Drive Suite 205 Baltimore, Maryland 21228

Re: K162434

Trade/Device Name: NC Gallant PTCA Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: December 12, 2016 Received: December 20, 2016

Dear Semih Oktay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Fernando Aguel-S
Fernando Aguel

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K162434

Device Name NC Gallant PTCA Catheter

Indications for Use (Describe)

The NC Gallant PTCA Catheter is indicated for:

- · Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion;
- · Post-deployment stent expansion.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                        |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.3333 2.66663H2.66663C1.93392 2.66663 1.3333 3.26726 1.3333 3.99997V13.3333C1.3333 14.066 1.93392 14.6666 2.66663 14.6666H13.3333C14.066 14.6666 14.6666 14.066 14.6666 13.3333V3.99997C14.6666 3.26726 14.066 2.66663 13.3333 2.66663ZM12.4444 5.33329L7.3333 10.4444L3.55553 6.66663L2.66663 7.55553L7.3333 12.2222L13.3333 6.22219L12.4444 5.33329Z" fill="black"></path> </svg> </span>  Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="12" rx="0.666667" stroke="black" stroke-width="1.33333" width="13.3333" x="1.33325" y="2.66663"></rect> </svg> </span>  Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Medinol. The logo consists of a blue and green abstract shape on the left, followed by the word "Medinol" in green, and the phrase "Ingenuity for life" in a smaller font below it. The logo is clean and modern, with a focus on the company name and tagline.

|                                                   | Medinol LTD's NC Gallant PTCA Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's<br>Name and<br>Address                | Medinol Ltd.<br>Kiryat Atidim, Building 8,<br>Tel Aviv 6158101,<br>Israel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Contact Name<br>and<br>Information                | Michal Hershkovitz<br>VP Clinical and Regulatory Affairs<br>Phone: +972 3 7679000/ext.<br>Fax: +972 3 648 2310<br>michalh@medinol.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date Prepared                                     | August 18 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Proprietary<br>Name(s)                            | NC Gallant PTCA Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Common<br>Name                                    | Catheters, transluminal coronary angioplasty, percutaneous                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                                      | LOX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification                                    | Class II, 21 CFR Part 870.5100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Predicate<br>Devices                              | NC Trek, Abbott (K110134)<br>NC Quantum, Boston Scientific (K121667)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device<br>Description                             | The NC Gallant is a sterile, single-use, intravascular medical device.<br>The NC Gallant is a double lumen catheter comprising an expandable<br>non-compliant (NC) coronary balloon on a rapid exchange (RX)<br>catheter.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Intended Use<br>of Device                         | The NC Gallant PTCA Catheter is indicated for the dilatation of<br>stenosis in coronary arteries or bypass grafts.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Indications for<br>Use                            | The NC Gallant PTCA Catheter is indicated for:<br>• Balloon dilatation of the stenotic portion of a coronary artery or<br>bypass graft stenosis for the purpose of improving myocardial<br>perfusion;<br>• Post-deployment stent expansion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Comparison of<br>Technological<br>Characteristics | The NC Gallant PTCA Catheter incorporates substantially equivalent<br>device design, fundamental technology, manufacturing processes,<br>sterilization and intended use as those featured in the Boston Scientific<br>predicate devices: NC Quantum, Boston Scientific (K121667) and NC<br>Trek, Abbott (K110134)                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Performance<br>Data                               | The NC Gallant PTCA Catheter was subjected to testing according to<br>the requirements of Guidance for Industry and FDA Staff – Class II<br>Special Controls for Certain Percutaneous Transluminal Coronary<br>Angioplasty (PTCA) Catheters, September 8, 2010.<br><br>Bench testing and biocompatibility testing were performed to support a<br>determination of substantial equivalence. The results of these tests<br>provide reasonable assurance that the proposed device has been<br>designed and tested to assure conformance to the requirements for its<br>intended use. No new safety or performance issues were raised during<br>the testing and, therefore, the NC Gallant may be considered<br>substantially equivalent to the predicate devices. |

| Kiryat Atidim, Building 8, POB 58165 Tel Aviv 6158101, Israel | Main: +972-3-767-9000 | Fax: +972-3-648-2310 Medinol Ltd. | Beck Tech Building, 8 Hartom St., POB 45026 Jerusalem 9777508, Israel | Main: +972-2-541-777 | Fax: +972-2-541-7764

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Medinol logo. The logo consists of a blue and green abstract shape on the left, and the word "Medinol" in green on the right. Below "Medinol" is the phrase "Ingenuity for life" in a smaller font, also in green. The logo is clean and modern, with a focus on the company name and tagline.

| Gallant PTCA Catheter:                                                                      |                                                                                                                                                                  |
|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity                                                                                | Sensitization                                                                                                                                                    |
| Intracutaneous Toxicity                                                                     | Material Mediated Pyrogenicity                                                                                                                                   |
| In Vitro Hemolysis                                                                          | Partial Thromboplastin Time                                                                                                                                      |
| In Vivo Thromboresistance                                                                   | Bacterial Reverse Mutation                                                                                                                                       |
| Mouse Lymphoma Assay                                                                        | Mouse Peripheral Blood                                                                                                                                           |
|                                                                                             | Micronucleus Study                                                                                                                                               |
| C3a Complement Activation<br>Assay                                                          | SC5b-9 Complement Activation<br>Assay                                                                                                                            |
| The following in-vitro performance tests were completed on the<br>NC Gallant PTCA Catheter: |                                                                                                                                                                  |
| Visual and Handling Performance                                                             | Dimensional Verification                                                                                                                                         |
| Crossing Profile                                                                            | Simulated Use                                                                                                                                                    |
| Kissing Balloon Technique (KBT)                                                             | Balloon Rated Burst Pressure                                                                                                                                     |
| Balloon Fatigue (Repeat Balloon<br>Inflations)                                              | Balloon Diameter vs. Balloon<br>Pressure (Compliance), Balloon<br>Diameter at NIP, Balloon<br>Diameter at RBP, Markers<br>Positioning, Balloon Working<br>Length |
| Catheter Bond Strength                                                                      | Tip Pull Test                                                                                                                                                    |
| Flexibility and Kink Resistance                                                             | Torque Strength                                                                                                                                                  |
| Catheter Radiopacity                                                                        | Particulate Evaluation                                                                                                                                           |
| Coating Integrity                                                                           | Balloon Rated Burst Pressure (in<br>Stent)                                                                                                                       |
| Balloon Fatigue (Repeat Balloon<br>Inflations; in Stent)                                    | Catheter corrosion resistance                                                                                                                                    |
| Packaging Integrity                                                                         | Environmental & Shipping &<br>Accelerated Shelf-Life                                                                                                             |
| The following in-vitro performance tests were completed on the<br>Gallant PTCA Catheter:    |                                                                                                                                                                  |
| GLP Acute Performance Study of Novel NC PTCA Balloons in<br>Porcine Model                   |                                                                                                                                                                  |

Based on the indications for use, technological characteristics, safety Conclusion and performance testing, the NC Gallant PTCA Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the predicate device.

| Kiryat Atidim, Building 8, POB 58165 Tel Aviv 6158101, Israel | Main: +972-3-767-9000 | Fax: +972-3-648-2310 Medinol Ltd. | Beck Tech Building, 8 Hartom St., POB 45026 Jerusalem 9777508, Israel | Main: +972-2-541-777 | Fax: +972-2-541-7764

---

**Source:** [https://fda.innolitics.com/device/K162434](https://fda.innolitics.com/device/K162434)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
