CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas(PICS-AF)
K162388 · Curaseal, Inc. · FTM · Feb 3, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162388
Device Name
CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas(PICS-AF)
Applicant
Curaseal, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Feb 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162388 · Feb 3, 2017
CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas(PICS-AF)
Curaseal, Inc.
Historical clinical site registry/records
Historical control data from patients treated with a predicate device (GORE BIO-A) at the same clinical site were used to compare safety and effectiveness outcomes against the PICS-AF device.
PICS-AF Clinical Study; Prospective, non-randomized clinical study with historical control; Follow-up/Duration: 6 months
Subjects with anorectal fistulas; Sample Size: 19 historical control subjects; Number of Sites: 1
GORE BIO-A Fistula Plug
Fistula closure success at 6 months; serious adverse events
Indications for Use
The CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas (PICS-AFTM) is for implantation to reinforce soft tissue for the repair of anal and rectal fistulas.
Device Story
PICS-AF is a sterile, single-use, sphincter-sparing device for anorectal fistula repair. Components include a silicone Sealing Disk, Sheath, Dilator, bovine-derived cross-linked collagen matrices, and resorbable sutures. During minimally invasive surgery, the device is implanted into the fistula tract; the Sealing Disk covers the inner ostium to inhibit enteric matter entry, while collagen matrices provide a scaffold for tissue ingrowth. The Sealing Disk is designed to be expelled as waste or removed by a physician if discomfort occurs. The external fistula opening remains open for drainage. The device is used in a clinical setting by a physician. It benefits patients by providing a scaffold for wound healing and fistula closure while sparing the sphincter.
Clinical Evidence
Prospective, non-randomized clinical study (N=30) compared PICS-AF to historical controls (N=19). Primary endpoint: fistula closure at 6 months (MRI-adjudicated). Success rate: 46.67% (ITT) and 66.67% (PP). Safety endpoint: serious adverse events (SAE) through 6 months. SAE rate was acceptable; overall adverse event rate (50%) was not statistically different from historical controls (P=0.3955).
Technological Characteristics
Materials: Bovine dermis-derived cross-linked collagen, silicone. Components: Sealing Disk, Sheath, Dilator, resorbable sutures. Principle: Mechanical occlusion of inner ostium and biological scaffold for tissue ingrowth. Sterilization: Ethylene Oxide. Form factor: Two models (M1/M2) with varying disk/dowel sizes.
Indications for Use
Indicated for implantation to reinforce soft tissue for the repair of anal and rectal fistulas in patients requiring surgical intervention.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 3, 2017
CuraSeal, Inc. % Mr. Kit Cariquitan Experien Group, LLC 755 North Mathilda Ave Sunnyvale, CA 94085
Re: K162388
Trade/Device Name: CuraSeal PICS Fistulae Closure Device -PICS - AF - M1™ PICS -AF - M2"M Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM Dated: December 16, 2016 Received: December 19, 2016
Dear Mr. Cariquitan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Binita S. Ashar -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K162388
Device Name
CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas (PICS-AFTM)
Indications for Use (Describe)
The CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas (PICS-AFTM) is for implantation to reinforce soft tissue for the repair of anal and rectal fistulas.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# CONSTITUTION OF THE FOR FOR FOR FOR FOR FDA USE ONLY CONTRACTOR CONSULTION OF
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the lime to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
#### 510(k) Notification K
K162388, pg. 1 of 11
#### GENERAL INFORMATION [807.92(a)(1)]
#### Applicant:
CuraSeal, Inc. 2231 Calle de Luna Santa Clara, CA 95054 USA Phone: 408-565-8787 Fax: 408-986-8643
#### Contact Person:
Kit Cariquitan Chief Regulatory Officer Experien Group, LLC 755 N. Mathilda Ave, Suite 100 Sunnyvale, CA 94085 USA Phone: 408-400-0856 Fax: 408-400-0865
# Date Prepared: August 12, 2016
# DEVICE INFORMATION [807.92(a)(2)]
The CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas (PICS-AF") is a medical device consisting of a Sealing Disk, Sheath, Dilator, Collagen Matrices and resorbable sutures. The PICS-AF Closure System is designed as a sphincter-sparing device that inhibits the movement of enteric matter into the anorectal fistula tract and to provide a scaffold for tissue ingrowth to occur to close the anorectal fistula.
#### Trade Name:
CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas (PICS-AF")
Generic/Common Name: Surgical Mesh
Classification: 21 CFR§878.3300, Class II
Product Code: FTM
{4}------------------------------------------------
# PREDICATE DEVICE(S) [807.92(a)(3)]
# K162388, pg. 2 of 11
- W.L. Gore & Associates, GORE BIO-A Fistula Plug (K083266)
- . Cook Biotech, SIS Fistula Plug (K050337)
The Cook Biotech, SURGISIS® RVP™ Recto-Vaginal Fistula Plug (K062729) is added as a reference device due to its same intended use and similar technological characteristics.
The predicate devices have not been subject to a design-related recall.
# DEVICE DESCRIPTION [807.92(a)(4)]
The PICS-AF Closure System is a sterile, single-use medical device that is designed to be implanted, using a minimally invasive surgical technique, to reinforce soft tissue for the repair of anal and rectal fistulas like the predicate devices. As the PICS-AF Closure System is specifically designed to be sphincter sparing, it does not preclude performing a repeat PICS-AF procedure or any other fistula repair procedure, if required. The key features of the CuraSeal PICS-AF Closure System include the Sealing Disk, Sheath, Dilator, Collagen Matrices and resorbable sutures. There are two models of the PICS-AF device (M1 and M2). Model M1 has a smaller Sealing Disk and smaller diameter Collagen Matrices than Model M2. The Collagen Matrices consist of up to six collagen dowels that are implanted in the fistula tract. After placement, any excess collagen dowels are removed or trimmed at the time of surgery. Placement of the PICS-AF device to cover the inner ostium opening of the fistula tract inhibits the movement of enteric matter into the fistula tract and the collagen provides a scaffold for wound healing to occur to close the anorectal fistula. The Collagen Matrices are delivered directly into the fistula tract and are secured to the fistula tract or the surrounding tissue by resorbable sutures. The Collagen Matrices are resorbed by the body during the natural healing process. The silicone Sealing Disk is designed to be expelled from the body as waste once the resorbable sutures have resorbed. Alternatively, the silicone Sealing Disk can be removed by the physician during the first two months following the procedure if the Sealing Disk appears in the patient's anal canal or causes any discomfort. Similar to the predicate devices, the external fistula opening (EFO) is left open during the healing process to allow for drainage to prevent infection or abscess formation.
# INDICATIONS FOR USE [807.92(a)(5)]
The CuraSeal Percutaneous Intraluminal Closure System for Anorectal Fistulas (PICS-AF") is for implantation to reinforce soft tissue for the repair of anal and rectal fistulas.
# TECHNOLOGICAL CHARACTERISTICS [807.92(a)(6)]
The technological characteristics of the CuraSeal PICS-AF Closure System are similar and substantially equivalent to the predicate devices. Table 1 lists the technological characteristics of the PICS-AF device and the predicate devices. Table 1 also provides the rationale to support a determination of substantial equivalence. Any differences between the devices do not raise different questions of safety or effectiveness. Performance data, including clinical testing, were provided to support the determination
{5}------------------------------------------------
of substantial equivalence. Clinical testing was performed by CuraSeal. The results from the clinical testing support the safe and effective use of the PICS-AF Closure System for anal and rectal fistula repair and further establish the substantial equivalence to the predicate devices.
| Feature | CuraSeal<br>PICS-AF | W.L. Gore &<br>Associates,<br>GORE BIO-A<br>Fistula Plug<br>(Predicate) | Cook Biotech,<br>SIS Fistula Plug<br>(Predicate) | Cook Biotech,<br>Modified SIS<br>Fistula Plug<br>(Reference) | Substantial<br>Equivalence<br>Rationale |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | TBD | K083266 | K050337 | K062729 | -- |
| Indications<br>for Use | The CuraSeal<br>Percutaneous<br>Intraluminal<br>Closure System<br>for Anorectal<br>Fistulas<br>(PICS-AF™) is<br>for implantation<br>to reinforce soft<br>tissue for the<br>repair of anal<br>and rectal<br>fistulas. | The GORE<br>BIO-A™<br>Fistula Plug<br>device is<br>intended for<br>use in the<br>reinforcement<br>of soft tissue<br>for the repair of<br>anorectal<br>fistulas. | The SIS Fistula<br>Plug is for<br>implantation to<br>reinforce soft<br>tissue where a<br>rolled<br>configuration is<br>required, for<br>repair of anal,<br>rectal, and<br>enterocutaneous<br>fistulas. The<br>device is<br>supplied sterile<br>and is intended<br>for one-time use. | The Modified<br>SIS Fistula Plug<br>is for<br>implantation to<br>reinforce soft<br>tissue for repair<br>of recto-vaginal<br>fistulas. The<br>device is<br>supplied sterile<br>and is intended<br>for one-time<br>use. | Like the predicate<br>devices, the<br>PICS-AF device<br>reinforces soft<br>tissue for the repair<br>of anal and rectal<br>fistulas. It is also<br>supplied sterile and<br>is intended for<br>one-time use. The<br>minor differences in<br>the indications for<br>use statements do<br>not affect the safety<br>and effectiveness of<br>the PICS-AF<br>device. |
| Classification/<br>Product Code | §878.3300<br>FTM | §878.3300<br>FTL | §878.3300<br>FTM | §878.3300<br>FTM | Same classification.<br>Same or similar<br>Product Code |
| Anatomical<br>Location | Anal and rectal<br>fistulas | Anal and rectal<br>fistulas | Anal, rectal, or<br>enterocutaneous<br>fistulas | Rectal and<br>vaginal fistulas | Same or similar<br>anatomical location |
| Feature | CuraSeal<br>PICS-AF | W.L. Gore &<br>Associates,<br>GORE BIO-A<br>Fistula Plug<br>(Predicate) | Cook Biotech,<br>SIS Fistula Plug<br>(Predicate) | Cook Biotech,<br>Modified SIS<br>Fistula Plug<br>(Reference) | Substantial<br>Equivalence<br>Rationale |
| | Bovine dermis-<br>derived, cross-<br>linked collagen | Synthetic<br>PGA/TMC<br>copolymer | Porcine-derived<br>small intestine<br>submucosa | Porcine-derived<br>small intestine<br>submucosa | Like the predicate<br>devices, the<br>resorbable material<br>acts as a scaffold to<br>specifically<br>reinforce soft tissue<br>and promote<br>healing of the<br>fistula tract. Design<br>verification and<br>clinical testing<br>demonstrate that the<br>PICS-AF device is<br>safe and performs<br>as intended. The<br>PICS-AF is<br>substantially<br>equivalent to the<br>predicate devices. |
| Resorbable<br>Plug Shape | Round dowels,<br>segmented up to<br>6 dowels/device | One piece<br>device consists<br>of Cap and six<br>round hollow<br>tube segments.<br>Tube segments<br>are cut to<br>length and/or<br>cut off to suit<br>fistula<br>diameter. | Tapered rolled<br>sheet, one piece<br>cut to length | Tapered rolled<br>sheet, one piece<br>cut to length | All devices have<br>plugs that occupy<br>the fistula tract and<br>any excess material<br>is trimmed from the<br>device at the time<br>of surgery. The<br>PICS-AF is<br>substantially<br>equivalent to the<br>predicate devices. |
| Attachment<br>Mechanism | Internal disk and<br>resorbable<br>sutures | Internal cap<br>and resorbable<br>sutures | Resorbable<br>sutures | Internal disk<br>and resorbable<br>sutures | All devices have<br>similar attachment<br>mechanism. Design<br>verification and<br>clinical testing<br>demonstrate that the<br>PICS-AF device is<br>safe and performs<br>as intended. The<br>PICS-AF is<br>substantially<br>equivalent to the<br>predicate devices. |
| Feature | CuraSeal<br>PICS-AF | W.L. Gore &<br>Associates,<br>GORE BIO-A<br>Fistula Plug<br>(Predicate) | Cook Biotech,<br>SIS Fistula Plug<br>(Predicate) | Cook Biotech,<br>Modified SIS<br>Fistula Plug<br>(Reference) | Substantial<br>Equivalence<br>Rationale |
| Disk Design | Sealing Disk | Disk | N/A | Button/Flange | Like the GORE<br>BIO-A Fistula Plug,<br>the internal opening<br>of the ostium is<br>closed to mitigate<br>enteric fluids from<br>entering the fistula<br>tract. Design<br>verification and<br>clinical testing<br>demonstrate that the<br>PICS-AF device is<br>safe and performs<br>as intended. The<br>PICS-AF is<br>substantially<br>equivalent to the<br>predicate devices. |
| Delivery<br>Sheath Design | Protects<br>collagen and<br>keeps it clean<br>when pulled into<br>tract | N/A | N/A | N/A | Design verification<br>and clinical testing<br>demonstrate that the<br>addition of the<br>Sheath/Dilator for<br>the PICS-AF device<br>does not increase<br>the risk to the<br>patient and is<br>therefore<br>substantially<br>equivalent to the<br>predicate devices. |
| Sterilization<br>Method | Ethylene Oxide<br>Gas | Gamma<br>Irradiation | Ethylene Oxide<br>Gas | Ethylene Oxide<br>Gas | Same sterility<br>assurance level,<br>same or similar<br>sterilization method |
Table 1: Summary of Technological Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE
The indications for use for the CuraSeal PICS-AF Closure System is substantially equivalent to the indications for use of the predicate devices. Any differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness. Thus, the CuraSeal PICS-AF Closure System is substantially equivalent to the predicate devices.
# PERFORMANCE DATA [807.92(b)]
All necessary bench and clinical testing were conducted on the CuraSeal PICS-AF Closure System to support a determination of substantial equivalence to the predicate devices.
#### TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
All necessary bench and clinical testing were conducted on the CuraSeal PICS-AF Closure System to support a determination of substantial equivalence to the predicate devices.
# Non-clinical Testing Summary [807.92(b)(1)]:
The non-clinical, bench testing included:
- . Design verification and bench validation studies, including dimensional analysis
- Physician simulated use
- Biocompatibility ●
- Sterilization ●
- Packaging and shelf-life
The collective results of the non-clinical testing demonstrate that the materials chosen, the manufacturing processes utilized and the design of the CuraSeal PICS-AF Closure System meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrate that the CuraSeal PICS-AF Closure System does not raise different questions of safety or effectiveness for the repair of anorectal fistulas when compared to the predicate devices.
{9}------------------------------------------------
# Clinical Testing Summary [807.92(b)(2)]:
A prospective, non-randomized clinical study was conducted to evaluate the safety and effectiveness of the PICS-AF Closure System for repair of anorectal fistulas. The safety and effectiveness rates for the PICS-AF devices were compared to historical control data generated from commercially available anorectal fistula repair devices.
The PICS-AF Closure System was used in 30 subjects enrolled in the study from one clinical site in Europe. Eight different investigators treated the 30 PICS-AF subjects. There were also 19 Historical Control subjects that were treated using a commercially available fistula plug (GORE BIO-A Fistula Plug) at this same clinical site and were evaluated as part of the PICS-AF clinical study.
The primary effectiveness endpoint for this study was fistula closure success at 6 months defined as complete healing of the fistula tract and associated external opening without drainage or abscess. Fistula closure success for the PICS-AF device was assessed using MRIs from Month 6. The MRI results were reviewed and adjudicated by three independent U.S. radiologists. The rate of the PICS-AF closure success at 6 months was 46.67% (14/30) based on the Intent-to-Treat (ITT) population results and 66.67% (14/21) based on the Per Protocol (PP) population results. Nine subjects had inclusion/exclusion deviations, based on the independent MRI assessment, and thus should have been disqualified from participating in the PICS-AF study.
The primary safety endpoint for this study was the proportion of subjects experiencing a serious adverse event (i.e., an infection, enlargement of the fistula, an allergic reaction to the PICS-AF device, etc.) through 6 months of post-procedure follow-up. There were three subjects who experienced a Serious Adverse Event during this study (two subjects had a fistula abscess and the third subject had a seton placed). The two subjects with a fistula abscess have recovered without sequelae and the subject with the seton is stable. The PICS-AF safety results have demonstrated that the overall primary safety endpoint was met. The PICS-AF safety data was comparable to the Historical Control group. The overall PICS-AF adverse event data was 50% as compared to 36.84% for the Historical Controls; however, this difference was not statistically significant (P=0.3955).
Table 2 provides the baseline demographics and characteristics for the control (GORE BIO-A Fistula Plug) and PICS-AF treatment groups.
| Characteristic | Control<br>N=19 | PICS-AF<br>N=30 | P-value |
|----------------------|-------------------|-------------------|---------|
| Gender: n/N (% Male) | 17/19 (89.47%) | 23/30 (76.67%) | 0.4511a |
| Age: Mean (SD) N | 50.11 (7.522) 18c | 53.70 (12.086) 30 | 0.2112b |
| Median (Min, Max) | 49.5 (35, 64) | 56.5 (31, 76) | |
| Smoker: n/N (%) | 5/19 (26.32%) | 11/30 (36.67%) | 0.5412a |
| Obese: n/N (%) | 4/19 (21.05%) | 7/30 (23.33%) | 1.0000a |
| Diabetes: n/N (%) | 5/19 (26.32%) | 2/30 (6.67%) | 0.0926a |
Table 2: Baseline Demographics and Characteristics
aTwo-sided Fisher's exact test.
bTwo-sided unequal variance t-test.
One Historical Control subject did not record age.
{10}------------------------------------------------
Table 3 provides a comparison of the fistula variables at baseline for the control (GORE BIO-A Fistula Plug) and PICS-AF treatment groups.
| Characteristic | Control<br>N=19 | PICS-AF<br>N=30 |
|---------------------------------------------------------|---------------------------|---------------------------|
| Onset of Fistula: Mean (Min, Max) | 32.6 mo. (5 mo., 121 mo.) | 57.2 mo. (4 mo., 246 mo.) |
| Type of Fistula: Transsphincteric | 19/19 (100%) | 30/30 (100%) |
| Length of Fistula Tract (cm):<br>Mean (SD) N | 4.1000 (1.8529) 10ª | 5.6733 (2.4057) 30 |
| Median (Min, Max) | 4.00 (2.00, 8.00) | 5.25 (2.50, 12.00) |
| No. of Fistula Openings: n/N (%)<br>Internal 1 opening: | 19/19 (100%) | 30/30 (100%) |
| Internal 2 openings: | -- | -- |
| External 1 opening: | 16/19 (84%) | 26/30 (87%) |
| External 2 openings: | 3/19 (16%) | 4/30 (13%) |
| Inner Fistula Diameter (cm):<br>Mean (SD) | Not Reported | 0.6017 (0.8391) |
| Median (Min, Max) | -- | 0.30 (0.05, 3.80) |
| Outer Fistula Diameter (cm):<br>Mean (SD) | Not Reported | 0.6137 (0.7834) |
| Median (Min, Max) | -- | 0.50 (0.01, 4.00) |
| Recurrent Fistula: n/N (%) | 12/19 (63%) | 20/30 (66.67%) |
| No. of Previous Fistula Treatments | 16 | 93 |
| Incontinence Score: Mean (Min, Max) | Not Performed | -- |
| Solid | -- | 0.23 (0, 4) |
| Liquid | -- | 0.67 (0, 4) |
| Gas | -- | 0.33 (0, 3) |
| Wears Pad | -- | 0.40 (0, 4) |
| Lifestyle Alteration | -- | 0.57 (0, 4) |
| Pain Scale:<br>Mean (Min, Max) | Not Performed | 3.83 (0, 8) |
Table 3: Comparison of Fistula Variables at Baseline
4Nine Historical Control subjects did not have fistula length recorded.
Incontinence Scale: 0=Never, 1=Rarely, 2=Sometimes, 3=Usually, 4=Always
Pain Scale: 0-10 (0=No pain, 5=Moderate pain, 10=Worst pain)
Table 4 provides the Month 6 fistula closure results for the control (GORE BIO-A Fistula Plug) and PICS-AF treatment groups.
#### Table 4: Fistula Closure Results at Month 6
| Treatment Group | Fistula Not Closed<br>n/N (%) | Fistula Closed<br>n/N (%) |
|-----------------|-------------------------------|---------------------------|
| PICS-AF | 16/30 (53.33%) | 14/30a (46.67%) |
| Control | 17/18b (94.44%) | 1/18 (5.56%) |
'Nine subjects were determined to have inclusion deviations that were detected by independent MRI readings. These subjects should not have been enrolled into this study. None of these nine subjects had healed fistulas at Month 6. The fistula closure success rate was 14/21 (66.67%) for subjects who met all specified enrollment criteria. bone Historical Control subject did not have an outcome recorded at six months.
{11}------------------------------------------------
Table 5 provides an analysis of the impact on fistula closure results based on the baseline characteristics for the combined treatment groups.
| Baseline Characteristic | Covariate P-value | P-value (Combined<br>Treatment Groups) |
|----------------------------|-------------------|----------------------------------------|
| Age | 0.6973 | 0.0179 |
| Weight | 0.3005 | 0.0244 |
| Gender | 0.0400 | 0.0174 |
| History of Diabetes | 0.7948 | 0.0176 |
| History of Cardiac Disease | 0.5648 | 0.0133 |
| History of Renal Disease | 0.9801 | 0.0115 |
| History of Smoking | 0.3127 | 0.0111 |
Table 5: Logistic Regression for Baseline Characteristic Impact on Fistula Closure
Table 6 provides a comparison of Month 6 outcomes for the control (GORE BIO-A Fistula Plug) and PICS-AF treatment groups.
| Category | Control<br>N=19 | PICS-AF<br>N=30 |
|-------------------------------------|-----------------|-----------------|
| Fistula Healed: n/N (%) | 1/18ª (5.56%) | 14/30 (46.67%) |
| Incontinence Score: Mean (Min, Max) | Not Performed | |
| Solid | -- | 0.27 (0, 3) |
| Liquid | -- | 0.97 (0, 4) |
| Gas | -- | 1.03 (0, 4) |
| Wears Pad | -- | 1.30 (0, 4) |
| Lifestyle Alteration | -- | 1.00 (0, 4) |
| Pain Scale: | Not Performed | |
| Mean (Min, Max) | -- | 2.37 (0, 10) |
4One Historical Control subject did not have an outcome recorded at six months. Incontinence Scale: 0=Never, 1=Rarely, 2=Sometimes, 3=Usually, 4=Always
Pain Scale: 0-10 (0=No pain, 5=Moderate pain, 10=Worst pain)
{12}------------------------------------------------
Table 7 provides a summary of the adverse events (including serious adverse events) for the control (GORE BIO-A Fistula Plug) and PICS-AF treatment groups.
| Table 7: Adverse Events | |
|-------------------------|--|
|-------------------------|--|
| | Control | | PICS-AF | |
|--------------------------------------------------------------------|----------------------------------|------------------------|----------------------------------|------------------------|
| Adverse Event | Number of<br>Patients n/N<br>(%) | Number<br>of<br>Events | Number of<br>Patients n/N<br>(%) | Number<br>of<br>Events |
| Primary Safety Endpointa | 0/19 (0%) | 0 | 3/30 (10.00%) | 3 |
| Any Adverse Event | 7/19 (36.84%) | 7 | 15/30 (50.00%) | 17d |
| Hypertension | 0/19 (0%) | 0 | 1/30 (3.33%) | 1 |
| Fistula-related Infection | 0/19 (0%) | 0 | 2/30 (6.67%) | 2 |
| Skin Irritation | 0/19 (0%) | 0 | 1/30 (3.33%) | 1 |
| Pain and Discomfort | 0/19 (0%) | 0 | 4/30 (13.33%) | 4 |
| Other Fistula-related<br>Complicationb | 0/19 (0%) | 0 | 2/30 (6.67%) | 2 |
| Premature Pullout of the<br>Anchor Suture and Partial<br>Migration | 1/19 (5.26%) | 1 | 0/30 (0%) | 0 |
| Fistula Abscessc | 2/19 (10.53%) | 2 | 1/30 (3.33%) | 1 |
| New Fistula | 2/19 (10.53%) | 2 | 0/30 (0%) | 0 |
| Recurrence of Treated Fistula<br>Tract | 1/19 (5.26%) | 1 | 0/30 (0%) | 0 |
| Fistula Tract Bleeding | 1/19 (5.26%) | 1 | 0/30 (0%) | 0 |
| Infection – Fistula Tract | 0/19 (0%) | 0 | 1/30 (3.33%) | 1 |
| Other | 0/19 (0%) | 0 | 2/30 (6.67%) | 2 |
4Fistula abscess requiring IV antibiotics and extended hospitalization (SAE).
°One mild external orifice and one anal canal ulcer.
6Excludes the two fistula abscesses that were primary safety endpoint events.
dTwo subjects had two Events each.
Table 8 summarizes the PICS-AF fistula healing outcomes based upon: subjects with adverse events and not healed fistulas; subjects with adverse events and healed fistulas; and subjects with no adverse events and not healed fistulas.
{13}------------------------------------------------
| PICS-AF Subjects with Adverse Events and Not Healed Fistulas (n=11) | | | | | |
|-----------------------------------------------------------------------|--------------------|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------|
| No. Subjects | No. Adverse Events | No. Serious Adverse Events | Type of Intervention n/N (%) | Solid Incontinence n/N (%) | Pain Scale Score Mean (Min, Max) |
| 11 | 11 | 2 | Sealing Disk removed = 3/11 (27%)<br>Setons placed = 1/11 (9%)<br>Drainage performed = 2/11 (18%)<br>Medications prescribed = 3/11 (27%) | 0=Never: 8/11 (73%)<br>1=Rarely: 2/11 (18%)<br>2=Sometimes: 0/11 (0%)<br>3=Usually: 1/11 (9%)<br>4=Always: 0/11 (0%) | 3.27 (0, 8) |
| PICS-AF Subjects with Adverse Events and Healed Fistulas (n=4) | | | | | |
| No. Subjects | No. Adverse Events | No. Serious Adverse Events | Type of Intervention n/N (%) | Solid Incontinence n/N (%) | Pain Scale Score Mean (Min, Max) |
| 4 | 3 | 1 | Sealing Disk removed = 1/4 (25%)<br>Setons placed = 0/4 (0%)<br>Drainage performed = 2/4 (50%)<br>Medications prescribed = 3/4 (75%) | 0=Never: 4/4 (100%)<br>1=Rarely: 0/4 (0%)<br>2=Sometimes: 0/4 (0%)<br>3=Usually: 0/4 (0%)<br>4=Always: 0/4 (0%) | 3.0 (0, 10) |
| PICS-AF Subjects with No Adverse Events and Not Healed Fistulas (n=5) | | | | | |
| No. Subjects | No. Adverse Events | No. Serious Adverse Events | Type of Intervention n/N (%) | Solid Incontinence n/N (%) | Pain Scale Score Mean (Min, Max) |
| 5 | 0 | 0 | Sealing Disk removed = 0/5 (0%)<br>Setons placed = 0/5 (0%)<br>Drainage performed = 0/5 (0%)<br>Medications prescribed = 0/5 (0%) | 0=Never: 5/5 (100%)<br>1=Rarely: 0/5 (0%)<br>2=Sometimes: 0/5 (0%)<br>3=Usually: 0/5 (0%)<br>4=Always: 0/5 (0%) | 1.6 (0, 5) |
| Table 8: Summary of PICS-AF Fistula Healing Outcomes at Month 6 | | | |
|-----------------------------------------------------------------|--|--|--|
| | | | |
In conclusion, the collective data from the PICS-AF Clinical Study support the overall safety and effectiveness of the PICS-AF Closure System for the repair of anorectal fistulas. The PICS-AF Closure System has been demonstrated to have favorable effectiveness results for fistula closure and to have an acceptable safety profile.
# CONCLUSIONS [807.92(b)(3)]
Extensive bench and clinical testing have been performed on the CuraSeal PICS-AF Closure System to evaluate the overall performance of the device. The collective results confirm that the CuraSeal PICS-AF Closure System is safe and effective, functions according to its specifications, is biocompatible and exhibits the appropriate mechanical and functional characteristics for an anorectal fistula repair device.
# SUMMARY
The CuraSeal PICS-AF Closure System is considered to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.