K162378 · Nanoentek USA, Inc. · LTJ · May 17, 2017 · Immunology
Device Facts
Record ID
K162378
Device Name
FREND PSA PLUS (reagent cartridge)
Applicant
Nanoentek USA, Inc.
Product Code
LTJ · Immunology
Decision Date
May 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.6010
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162378 · May 17, 2017
FREND PSA PLUS (reagent cartridge)
Nanoentek USA, Inc.
Longitudinal serial patient samples; Medical records/clinical status documentation (laboratory tests, patient interviews, physical exams, imaging)
Retrospective analysis of serial PSA measurements from patients with prostate cancer to demonstrate the device's ability to mirror clinical status (NED, Responding, Stable, Progression) compared to a predicate device.
Prostate cancer; Serial monitoring; Retrospective clinical cohort; Clinical status correlation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Serial monitoring study; Retrospective longitudinal cohort study
Patients previously diagnosed with prostate cancer and treated via various methods (prostatectomy, cryoablation, lymph node resection, TURP, chemotherapy, hormone therapy); Sample Size: 194 serial determinations
Predicate device (FREND™ PSA Plus, K124056)
Concordance of PSA concentration changes with clinical status (NED, Responding, Stable, Progression)
Indications for Use
The NanoEnTek FREND™ PSA Plus is designed for in vitro DIAGNOSTIC USE ONLY for the quantitative measurement of total Prostate Specific Antigen (PSA) in human serum, Li-heparinized plasma, and K3-EDTA plasma using the FREND™ System. This device is indicated for the serial measurement of total PSA to be used as an aid in the management of patients with prostate cancer.
Device Story
FREND™ PSA Plus is a single-use fluorescence immunoassay cartridge for quantitative measurement of total PSA in human serum, Li-heparinized plasma, and K3-EDTA plasma. Operator applies 35 µL sample to cartridge inlet; capillary action moves sample through micro-fluidics pathway to form immune complexes. Cartridge is inserted into portable, automated FREND™ System reader. System measures laser-induced fluorescence; concentration calculated via ratio of test zone to reference zone intensity. Results displayed on touchscreen; used by clinicians to monitor PSA levels longitudinally in prostate cancer patients. Aids in assessing disease status (NED, responding, stable, progression). Calibration via lot-specific code-chip.
Clinical Evidence
Clinical study evaluated 257 serial serum samples from 63 prostate cancer patients. Performance compared to clinical status (progression vs. no-progression). Sensitivity 63.0% (95% CI: 51.5%–73.2%), specificity 81.0% (95% CI: 73.1%–87.0%). Concordance with predicate device was 74.2%. Cut-off for significant change defined as >0.2 ng/mL for PSA ≤1 ng/mL or >20% increase for PSA >1 ng/mL.
Technological Characteristics
Fluorescence immunoassay using microfluidic cartridges. Sensing principle: sandwich immune-complex binding detected via fluorescence intensity. Energy source: FREND™ System. Connectivity: Code chip for lot-specific calibration. Standards: CLSI EP05-A3, EP6-A, EP07-A2, EP09-A2, EP14-A3, EP17-A2, EP25-A, C28-A3.
Indications for Use
Indicated for serial measurement of total PSA as an aid in the management of patients with prostate cancer. For prescription use only.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K162378
B. Purpose for Submission:
Modification of Previously Cleared Device
C. Measurand:
Total Prostate Specific Antigen
D. Type of Test:
Fluorescence Immunoassay, Quantitative
E. Applicant:
NanoEnTek, Inc.
F. Proprietary and Established Names:
FREND™ PSA Plus
G. Regulatory Information:
1. Regulation section:
21 CFR §866.6010 – Tumor-Associated Antigen Immunological Test System
2. Classification:
Class II
3. Product code:
LTJ – Prostate Specific Antigen (PSA) for Management of Prostate Cancers
4. Panel:
Immunology (82)
H. Intended Use:
{2}
1. Intended use:
The NanoEnTek FREND™ PSA Plus is designed for *in vitro* DIAGNOSTIC USE ONLY for the quantitative measurement of total Prostate Specific Antigen (PSA) in human serum, Li-heparinized plasma, and K3-EDTA plasma using the FREND™ System. This device is indicated for the serial measurement of total PSA to be used as an aid in the management of patients with prostate cancer.
2. Indication for use:
Same as Intended Use above
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
NanoEnTek FREND™ System (K124056)
I. Device Description:
FREND™ PSA Plus is comprised of the following materials (to perform 25 tests):
- FREND™ PSA Plus cartridges (25): each cartridge contains monoclonal anti-PSA1, monoclonal anti-PSA2, and fluorescent particle.
- Disposable pipette tips (30)
- FREND™ PSA Plus Code Chip
- FREND™ PSA Plus Package Insert
J. Substantial Equivalence Information:
1. Predicate device name:
NanoEnTek FREND PSA Plus,
2. 510(k) number:
K124056
3. Comparison with predicate:
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device Modified FREND™ PSA PLUS | Predicate FREND™ PSA Plus |
| Intended Use | The NanoEnTek FREND™ PSA Plus is designed for in vitro DIAGNOSTIC USE ONLY for the quantitative measurement of total Prostate Specific Antigen (PSA) in human serum, heparinized plasma, and EDTA plasma using the FREND™ System. This device is indicated for the serial measurement of total PSA to be used as an aid in the management of patients with prostate cancer. | Same |
| Analyte | Total PSA | Same |
| Type of Test | Fluorescent immunoassay | Same |
| Instrument | FREND system | Same |
| Sample type | Serum, heparinized plasma, EDTA plasma | Same |
| Interpretation of results | Interpolation from a lot-specific calibration curve | Same |
| Calibration standard | WHO International Standard Prostate Specific Antigen (90:10) NIBSC code: 96/670 | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device Modified FREND™ PSA PLUS | Predicate FREND™ PSA PLUS |
| Sample size | 35 μL | 30 μL |
| LoD/LoQ | LoD: 0.03 ng/mL
LoQ: 0.08 ng/mL | LoD: 0.1 ng/mL
LoQ: not determined |
| Assay measuring range (AMR) | 0.08–25 ng/mL | 0.10–25 ng/mL |
# K. Standard/Guidance Document Referenced:
- CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition
- CLSI EP6-A, Evaluation of Linearity of Quantitative Measurement, Approved Guideline.
- CLSI EP07-A2, Interference Testing in Clinical Chemistry, Approved Guideline, Second Edition—Second Edition
- CLSI EP09-A2, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline—Second Edition
{4}
- CLSI EP14-A3, Evaluation of Commutability of Processed Samples; Approved Guideline
- CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline—Second Edition
- CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guidelines
- CLSI C28-A3, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline
## L. Test Principle:
The FREND™ PSA Plus is a fluorescence immunoassay. The FREND™ Test Cartridge utilizes microfluidic technology and detects sandwich immune-complexes bound to PSA. Each FREND™ PSA Plus Test Cartridge contains a test zone and as well as a reference zone. The specimen is added by the operator to the sample inlet with a transfer pipet, allowing the appropriate volume of sample to be delivered into the FREND™ PSA Plus Test Cartridge. The Cartridge containing the sample is placed into the FREND™ System, which is programmed to begin analysis once the sample has reacted with the reagents. The PSA quantification is based on the amount of fluorescence detected by the FREND™ System at the FREND™ PSA Plus Test Cartridge window. The magnitude of the fluorescent signal is directly proportional to the amount of total PSA in the sample.
## M. Performance Characteristics:
1. Analytical performance: The test results met the sponsor's pre-determined acceptance criteria for all analytical performance studies.
### a. Precision/Reproducibility:
The precision and lot-to-lot reproducibility of the assay was evaluated per CLSI EP05-A3 by testing four serum sample pools containing various concentrations of PSA. Each sample was run in duplicate, in two runs per day, for 20 days with three lots of cartridge on one FREND Plus system by one operator (n=240 per sample). The results are summarized in the table below.
| Sample | Mean (ng/mL) | Within-Run | | Between-Run | | Between-Day | | Between-Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 0.08 | 0.01 | 14.1 | 0.00 | 2.6 | 0.00 | 2.4 | 0.00 | 0.0 | 0.01 | 14.6 |
| 2 | 0.10 | 0.01 | 12.3 | 0.00 | 0.0 | 0.00 | 2.0 | 0.00 | 0.0 | 0.01 | 12.5 |
| 3 | 4.00 | 0.29 | 7.2 | 0.06 | 1.6 | 0.00 | 0.0 | 0.00 | 0.0 | 0.30 | 7.4 |
| 4 | 21.40 | 1.20 | 5.6 | 0.51 | 2.4 | 0.00 | 0.0 | 0.09 | 0.4 | 1.30 | 6.1 |
The site-to-site reproducibility of the assay was evaluated by testing three serum samples at three sites. Each sample was tested in replicates of five per run, one run
{5}
per day, for five days using one lot of reagent (n=75 per sample). The results are summarized in the table.
| Sample | Mean (ng/mL) | Within-Run | | Between-Day | | Within-Site | | Between-Site | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 0.26 | 0.03 | 10.1 | 0.01 | 4.4 | 0.03 | 11.0 | 0.00 | 0.0 | 0.03 | 11.0 |
| 2 | 2.87 | 0.22 | 7.6 | 0.03 | 1.2 | 0.22 | 7.7 | 0.05 | 1.8 | 0.23 | 7.9 |
| 3 | 11.49 | 1.14 | 9.9 | 0.49 | 4.2 | 1.24 | 10.8 | 0.22 | 1.9 | 1.26 | 10.9 |
b. Linearity/assay reportable range:
Linearity:
Linearity was evaluated according to the CLSI guideline EP6-A by preparing two dilution series. The first dilution series was made by diluting a serum sample containing a PSA concentration above the upper limit of measuring range with a serum sample containing a low concentration of PSA. Each dilution was tested in replicates of five using three lots of cartridges. The second dilution series was made by diluting a serum sample with a PSA concentration at 1.0 ng/mL with a serum sample with a PSA concentration below the lower limit of measuring range. Each dilution was tested in replicates of four using three lots of cartridge to determine the observed value of PSA. Percent recovery and linear regression analysis was determined by comparing the observed PSA result with the expected value. The results are summarized as follows:
| Dilution Range (ng/mL) | Slope (95% CI) | Y-intercept (95% CI) | R² | % Recovery |
| --- | --- | --- | --- | --- |
| 0.10–25.53 | 1.00 (0.98–1.02) | 0.04 (-0.26–0.35) | 0.99 | 96.0–102.0 |
| 0.05–1.00 | 0.99 (0.96–1.03) | -0.00 (-0.02–0.02) | 0.98 | 97.2–100.0 |
The results support that the assay is linear for the claimed measuring range, 0.08-25 ng/mL.
Hook effect:
The presence of high dose hook effect was tested using a sample with PSA concentration of 1200 ng/mL. No high dose hook effect was observed in the sample with a PSA concentration as high as 1200 ng/mL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
{6}
The FREND™ PSA Plus assay is traceable to the WHO International Standard Prostate Specific Antigen (90:10) NIBSC code: 96/670.
**Value assignment**: The internal standard is manufactured by gravimetric methods based on the PSA Ag ACT complex (Scripps P0624-90010) or WHO standard material (NIBSC 96/670). The concentrations of calibrators are confirmed based on the measurement on ARCHITECT $i$ total PSA assay (P910007) or TOSOH AIA ST PA Test (P910065). The calibration information is then coded in the individual cartridge.
**Stability**:
The stability of the cartridges for FREND™ PSA Plus was established previously per K124056. The cartridges are stable for one year from the date of manufacture if stored refrigerated as directed.
**d. Detection limit**:
**Limit of Blank (LoB)** was determined by assaying five PSA free serum samples. Each sample was tested in replicates of two per day for six days with three lots of cartridges to generate 60 data points. The LoB was calculated based on the $95^{\text{th}}$ percentile of 60 measurements and determined to be $0.02\,\mathrm{ng/mL}$.
**The Limit of Detection (LoD)** was determined by assaying five samples with PSA concentrations close to the lower limit of measuring range. Each sample was tested in replicates of two per day for six days with three lots of cartridges to generate 60 data points. The LoD value was calculated as the $\mathrm{LoB} + 1.645 \times \mathrm{SD}$ of the replicates for the low level samples and was determined to be $0.03\,\mathrm{ng/mL}$.
**The Limit of Quantification (LoQ)** was evaluated using the precision profile approach and the data from the precision study following CLSI EP05-A3. Within-laboratory precision was determined using four samples with PSA concentration at 0.025, 0.05, 0.075 and $0.1\,\mathrm{ng/mL}$. The LoQ was calculated as the sample concentration at which the pre-determined goal of within-lab imprecision was met. The LoQ was determined to be $0.08\,\mathrm{ng/mL}$.
**e. Analytical specificity**:
**Endogenous Interference**: Two serum base samples (1 ng/mL and 4 ng/mL) were spiked separately with endogenous interferents. The control samples were prepared for each interfering substance by spiking with the same volume of diluents. Each sample was tested in three replicates and the percent recovery was calculated by comparing to a corresponding control sample. No interference was detected in the samples using interferents up to the concentrations listed in the table below.
{7}
7
| Potential Interfering Substances | Concentration Tested | Range of % Recovery |
| --- | --- | --- |
| Hemoglobin | 500 mg/dL | 86.79–106.68 |
| Bilirubin (unconjugated) | 200 μg/mL | 89.52–108.14 |
| Triglycerides | 3 g/dL | 89.10–116.30 |
| Total Protein | 50 mg/mL | 88.89–112.83 |
| Rheumatoid Factor | 1075 IU/mL | 102.99–118.48 |
| HAMA | 70 ng/mL | 87.18–96.46 |
Exogenous Interference: The same protocol used for evaluation of endogenous interference was used to evaluate the potential interference of common drugs. No interference was observed for each drug at the concentrations listed below.
| Name of Agent | Concentration Tested | Range of % Recovery |
| --- | --- | --- |
| Acetaminophen | 250 ng/mL | 87.47–114.99 |
| Aspirin | 600 μg/mL | 94.23–109.28 |
| Ibuprofen | 500 μg/mL | 88.62–115.63 |
| Flutamide | 10 μg/mL | 88.68–99.47 |
| Diethylstilbestrol (DES) | 5 μg/mL | 89.42–110.00 |
| Goserelin | 40 ng/mL | 94.90–110.20 |
| Leuprolide | 275 ng/mL | 89.15–106.38 |
| Finasteride | 250 ng/mL | 89.52–108.14 |
| Tamsulosin | 100 ng/mL | 89.71–120.05 |
| Docetaxel | 10 μg/mL | 90.83–105.50 |
| Prostatic acid phosphatase | 10 ng/mL | 87.71–110.42 |
f. Assay cut-off:
See clinical cut-off
2. Comparison studies:
a. Method comparison with predicate device:
A total of 64 serum samples were tested with the modified FREND™ PSA Plus using a sample volume of 35 μL and the predicate device using a sample volume of 30 μL. The concentrations of PSA for these samples covered the measuring range for both assays. Passing-Bablok regression analysis was performed by comparing the results obtained from the modified FREND™ PSA Plus (y) and the predicate (x). The results are summarized below:
| N | Range (ng/mL) | Slope (95% CI) | Y-Intercept (95% CI) | Correlation Coefficient |
| --- | --- | --- | --- | --- |
| 64 | 0.11–20.23 | 0.98 (0.96–1.01) | -0.09 (-0.09–0.07) | 0.99 |
{8}
# b. Matrix comparison:
A study was performed using the modified FREND™ PSA Plus to compare PSA values obtained for 40 matched samples each with serum, lithium heparin plasma and K3-EDTA plasma. The concentrations of PSA for these matched serum/plasma samples covered the measuring range of the assay. Passing-Bablok regression analysis was performed by comparing the results of samples from different plasma samples (y) to the results of corresponding serum samples (x). The results are summarized below:
| N=40 | Range (ng/mL) | Slope (95% CI) | Y-Intercept (95% CI) | Correlation Coefficient |
| --- | --- | --- | --- | --- |
| Li-heparin plasma vs. Serum | 0.08–23.59 | 0.96 (0.89–1.03) | 0.03 (-0.15–0.24) | 0.99 |
| K3-EDTA plasma vs. Serum | 0.08–23.59 | 1.03 (0.99–1.09) | -0.08 (-0.37–0.08) | 0.99 |
# c. Other supporting comparison studies:
A total of 207 samples were tested with the modified FREND™ PSA Plus on FREND™ System and Abbott ARCHITECT total PSA (P910007) assay on the Abbott ARCHITECT i System. The PSA concentrations of these samples covered the measuring range for both assays. The results generated from the modified FREND™ PSA Plus (y) were compared to the results generated from ATCHITECT total PSA assay (x). Results from the Passing-Bablok regression analysis are summarized in the following table:
| N | Range (ng/mL) | Slope (95% CI) | Y-Intercept (95% CI) | Correlation Coefficient |
| --- | --- | --- | --- | --- |
| 207 | 0.08–23.56 | 0.98 (0.94–1.01) | -0.05 (-0.101–0.00) | 0.98 |
# 3. Clinical studies:
# a. Clinical Sensitivity and Specificity:
The effectiveness of the modified FREND PSA Plus assays as an aid in monitoring of disease status in prostate cancer subjects was determined by assessing changes in PSA levels in serial serum samples from 63 subjects diagnosed with prostate cancer compared to changes in their disease status. The sample inclusion and exclusion criteria are as follows:
# Inclusion criteria:
- Sample must be serum and is entered into the study only once.
- Sufficient volume for the sample exists to allow testing on both the predicate and the experimental device.
{9}
- Serial sets must include a minimum of three draws. Samples are to be from blood draws performed at or after definitive diagnosis throughout the clinical course of the disease.
- Clinical information that details the status of the subject’s disease must be available for each sample as must the treatment regimen being used at the time of the blood collection.
- Sample may be fresh or stored for one week at 2–8 °C. Samples stored for more than one week must be held at −20 °C or colder.
## Exclusion criteria:
- Samples that appear to be contaminated, turbid, lipemic, icteric or hemolyzed will be excluded.
- Samples failing to meet all the inclusion criteria will be excluded.
Among 63 subjects, the majority of subjects (n = 49) were Caucasian, the remaining subjects included nine African American, one Asian, and four Hispanics. 62 out of 63 subjects had staging information of initial diagnosis based on the recommendations by the American Joint Committee on Cancer (AJCC) TNM system: three were stage I; 25 were stage II; 22 were stage III; and 12 were stage IV. Of 63 subjects, 40 subjects were treated, including prostatectomy, radical perirectal prostatectomy, bilateral scrotal orchiectomy, lymphadenectomy, bilateral pelvic lymphadenectomy, prostate curettage, cranial resection mets, cryoblation, seminal vesiculectomy, and transurethral resection of the prostate (TURP). A total of 257 values consisting of 63 baseline values and 194 monitoring observations were obtained with the mean number of sample draws of 4.08 per subject.
## Results:
In this study, all 257 serum samples (including 63 baseline and 194 follow-up measurements) from 63 subjects were measured using both the modified FREND PSA Plus on the FREND system and the predicate FREND PSA Plus which was previously cleared under K124056. At each follow-up visit, the percentage change of the PSA was calculated by comparing the test result to the results obtained from the previous visit. The performance of the FREND PSA Plus as an aid in the management of patients with prostate cancer subjects was evaluated by assessing the changes in PSA levels correlated to the clinical assessment at the time of assay.
For both modified and predicate FREND™ PSA Plus assays, a positive change in the PSA concentration was defined as measurable increase in the value that is 20% greater than the previous value of the test for samples with the PSA concentration > 1 ng/mL, and increase >0.2 ng/mL for sample with the PSA concentration ≤ 1 ng/mL.
For each pair of samples in a serial set, the change in the PSA concentration was compared to the change in the clinical status of the patients as determined by the physician based on other laboratory tests, patient interviews, physical examinations, and medical imaging. The following definition was used to categorize the patient’s
{10}
disease status:
1. No Evidence of Disease (NED) – All detectable tumors had disappeared.
2. Stable – The tumors stayed the same size. To account for measurement errors on scans and to discount “insignificant” changes, stable disease includes either a small amount of growth (typically less than 20%–25%) or a small amount of shrinkage (typically less than 25%).
3. Responding – Significant (≥50%) amount of decrease in the total tumor volume responding to the treatment, with evidence of some residual disease still remaining.
4. Progression – Clinical evidence of growth in the primary tumor or new tumors have appeared (*see reference below).
* Costelloe CM, Chuang HH, Madewell JE, Ueno NT. (2010). Cancer response criteria and bone metastases: RECIST 1.1, MDA and PERCIST. J. Cancer. 1:80-92.
The following table represents the summary of all follow-ups (n=194) for subjects at which a clinical evaluation occurred. A positive percentage change of the subjects at these clinical evaluations was defined by the cut-off of >20% change for samples with PSA concentration ≥1 ng/mL or >0.2 ng/mL for samples with PSA concentrations <1 ng/mL
| | Clinical Status | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | No-Progression | | | Progression | Total |
| | | NED | Stable | Responding | | |
| Change in PSA value | Positive | 16 | 6 | 1 | 46 | 69 |
| | Negative | 80 | 9 | 9 | 27 | 125 |
| | Total | 96 | 15 | 10 | 73 | 194 |
Based on clinical disease status using progression or no-progression (NED, stable, responding), the performances of the modified FREND PSA Plus are summarized in the following table:
| Performance Measurement | Value | 95% CI |
| --- | --- | --- |
| Sensitivity | 63.0% | 51.5%–73.2% |
| Specificity | 81.0% | 73.1%–87.0% |
| Positive Predictive Value (PPV) | 66.7% | 57.1%–75.0% |
| Negative Predictive Value (NPV) | 78.4% | 72.7%–83.2% |
| Positive Likelihood Ratio | 3.32 | 2.23–5.01 |
| Negative Likelihood Ratio | 0.46 | 0.33–0.61 |
In this study, 62 out of 63 subjects had staging information from the initial diagnosis. The summary statistics of the change of the FREND PSA Plus results from the preceding sample for all follow-up samples by disease stage are shown in the following tables:
{11}
| Change in FREND PSA Plus | Clinical Status | | | | |
| --- | --- | --- | --- | --- | --- |
| | No-Progression | | | Progression | Total |
| | NED | Stable | Responding | | |
| Stage I (N=8 from 3 subjects) | | | | | |
| Positive | 0 | 0 | 0 | 2 | 2 |
| Negative | 4 | 0 | 1 | 1 | 6 |
| Total | 4 | 0 | 1 | 3 | 8 |
| Stage II (N = 85 from 25 subjects) | | | | | |
| Positive | 5 | 2 | 0 | 22 | 29 |
| Negative | 39 | 2 | 4 | 11 | 56 |
| Total | 44 | 4 | 4 | 33 | 85 |
| Stage III (N=60 from 22 subjects) | | | | | |
| Positive | 7 | 2 | 0 | 15 | 24 |
| Negative | 21 | 3 | 4 | 8 | 36 |
| Total | 28 | 5 | 4 | 23 | 60 |
| Stage IV (N=37 from 12 subjects) | | | | | |
| Positive | 3 | 2 | 1 | 7 | 13 |
| Negative | 13 | 4 | 0 | 7 | 24 |
| Total | 16 | 6 | 1 | 14 | 37 |
| Stage unknown (N = 4 from 1 subject) | | | | | |
| Positive | 1 | 0 | 0 | 0 | 1 |
| Negative | 3 | 0 | 0 | 0 | 3 |
| Total | 4 | 0 | 0 | 0 | 4 |
As 40 out of 63 subjects had surgery, the following tables summarized the change of the FREND PSA Plus results from the preceding sample for all follow-up samples and clinical performance of the FREND PSA Plus values over successive visits by surgery status:
| Change in FREND PSA Plus | Clinical Status | | | | |
| --- | --- | --- | --- | --- | --- |
| | No-Progression | | | Progression | Total |
| | NED | Stable | Responding | | |
| Non-Surgery Group | | | | | |
| Positive | 5 | 5 | 0 | 19 | 29 |
| Negative | 26 | 6 | 4 | 10 | 46 |
| Total | 31 | 11 | 4 | 29 | 75 |
| Surgery Group | | | | | |
| Positive | 11 | 1 | 1 | 27 | 40 |
| Negative | 54 | 3 | 5 | 17 | 79 |
| Total | 65 | 4 | 6 | 44 | 119 |
{12}
| Performance Measurement | Non-Surgery (N=78) | Surgery (N=116) |
| --- | --- | --- |
| Sensitivity (95 % CI) | 66.7 % (47.2–82.7%) | 60.5% (44.4–75.0%) |
| Specificity (95% CI) | 79.2 % (65.0–89.5%) | 82.2% (71.5–90.2%) |
| PPV (95% CI) | 66.7% (47.2–82.7%) | 66.7% (49.8–80.9%) |
| NPV (95% CI) | 79.2% (65.0–89.5%) | 77.9% (67.0–86.8%) |
| Positive likelihood ratio | 3.20 (1.74–5.87) | 3.40 (1.96–5.88) |
| Negative likelihood ratio | 0.42 (0.25–0.71) | 0.48 (0.33–0.71) |
# b. Other clinical supportive data:
The clinical performance of the modified FREND PSA Plus was evaluated against that of the predicate, FREND PSA Plus (K124056). The table below compared the clinical sensitivity and specificities between these two assays using the same sample cohort in the clinical study described above.
| | N=257 (63 subjects) | |
| --- | --- | --- |
| | The Modified FREND PSA Plus | The FREND PSA Plus (K124056) |
| Sensitivity | 63.0% (51.5–73.2%) | 65.8% (54.3–75.6%) |
| Specificity | 81.0% (73.1–87.0%) | 81.0% (73.1–87.0%) |
| Total Concordance | 74.2% (67.4–80.2%) | 75.3% (68.6–81.2%) |
| PPV | 66.7% (57.1–75.0%) | 67.6% (58.2–75.8%) |
| NPV | 78.4% (72.7–83.2%) | 79.7% (73.8–84.5%) |
| Positive Likelihood Ratio | 3.32 (2.23–5.01) | 3.46 (2.34–5.21) |
| Negative Likelihood Ratio | 0.46 (0.33–0.61) | 0.42 (0.30–0.57) |
# 4. Clinical cut-off:
An absolute cutoff is not applicable for monitoring. The significance of PSA increase was defined as a change greater than $0.2\mathrm{ng / mL}$ for samples with PSA concentration $\leq 1$ ng/mL and the percentage increase $>20\%$ for samples with PSA concentration $>1\mathrm{ng / mL}$ .
# 5. Expected values/Reference range:
The reference range for the modified FREND™ PSA Plus was determined by testing serum samples from a total of 121 apparently healthy males aged 50 to 88 years (average age of 63, median age of 63). The results are summarized in the table below:
| N | 121 |
| --- | --- |
| Mean | 1.27 ng/mL |
| Median | 1.01 ng/mL |
| 95th percentile | 2.92 ng/mL |
| 97th percentile | 3.19 ng/mL |
| 99th percentile | 4.59 ng/mL |
{13}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.