← Product Code [OLP](/productcode/OLP) · K162371

# Pulsaderm ACE-All Blue (K162371)

_Nutra Luxe MD, LLC · OLP · Dec 15, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162371

## Device Facts

- **Applicant:** Nutra Luxe MD, LLC
- **Product Code:** [OLP](/productcode/OLP.md)
- **Decision Date:** Dec 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Pulsaderm ACE-All Blue is indicated to treat mild to moderate acne on the face.

## Device Story

Pulsaderm ACE-All Blue is an OTC light therapy device using blue light-emitting diodes (LEDs) to treat mild to moderate facial acne. The device emits blue light at a wavelength of 415 +/- 5 nm with a dose rate of 25 mW/cm2 and a total dose of 12 J/cm2. It is intended for patient self-use at home. Users apply the device to the skin based on instructions provided in the user manual. The device is a smaller version of the predicate Nutra Light Blue and shares similar technological characteristics and treatment regimens with the Tanda Mini Skincare System. The device aims to provide therapeutic light exposure to the skin to manage acne conditions.

## Clinical Evidence

No clinical efficacy trials were performed. Evidence consists of bench testing for electrical safety (IEC 60601-1, IEC 60601-1-2), photobiological safety (IEC 62471), and biocompatibility (ISO 10993). A lay-user study (n=unspecified) demonstrated that 98% of participants could correctly interpret labeling and operate the device without assistance.

## Technological Characteristics

LED-based light therapy system; wavelength 415 +/- 5 nm; dose rate 25 mW/cm2; total dose 12 J/cm2. Conforms to IEC 60601-1:2012, IEC 60601-1-2:2014, IEC 62471:2006, and ISO 10993. Developed under ISO 14971:2007 risk management.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Nutra Light Blue ([K132838](/device/K132838.md))
- Tanda Mini Skincare System ([K124042](/device/K124042.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines above them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 15, 2016

Nutra Luxe MD, LLC Ms. Gloria Avendano Regulatory Affairs Manager 12801 Commonwealth Dr., Unit 5 Fort Myers, FL 33913

Re: K162371

Trade/Device Name: Pulsader ACE-All Blue Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery And in dermatology Regulatory Class: Class II Product Code: OLP Dated: October 31, 2016 Received: November 16, 2016

Dear Ms. Avendano

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known): pending K162371

Device Name: Pulsaderm ACE-All Blue

Indications for Use:

Pulsaderm ACE-All Blue is indicated to treat mild to moderate acne on the face.

Prescription Use (21 CFR Part 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR Part 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

# 510(k) Summary Pulsaderm ACE - All Blue

# 1. General Information

Submitter: NutraLuxe MD, LLC 12801 Commonwealth Dr. Unit 5 Fort Myers, FL 33913

# Contact Person:

Gloria Avendano Regulatory Affairs Manager Nutra Luxe MD 12801 Commonwealth Dr. Unit 5 Fort Myers, FL 33913 Gloria@nutraluxemd.com

Summary Preparation Date: August 20th, 2016

# 2. Device

| Device Name:               | Pulsaderm ACE - All Blue                            |
|----------------------------|-----------------------------------------------------|
| Common/Usual Name:         | Acne Light Therapy System                           |
| Classification Name:       | Over-the-counter powered light based laser for acne |
| Classification Regulation: | 21CFR 878.4810                                      |
| Classification Panel:      | General and Plastic Surgery                         |
| Product Code:              | OLP                                                 |

# 3. Predicate Devices:

Pulsaderm ACE-All Blue is substantially equivalent to the following predicate devices;

| Device                     | 510(k) Number | Manufacturer        |
|----------------------------|---------------|---------------------|
| Nutra Light Blue           | K132838       | Nutra Luxe MD, LLC  |
| Tanda Mini Skincare System | K124042       | Syneron Beauty Inc. |

{4}------------------------------------------------

#### 4. Device Description:

The Pulsaderm ACE-All Blue is a device that uses Blue Light Emitting Diodes to provide LED Light therapy to the skin. Blue light is emitted for the treatment of mild to moderate acne.

#### 5. Indications for use:

Pulsaderm ACE-A11 Blue is indicated to treat mild to moderate acne on the face.

## 6. Technological Characteristics and substantial equivalence:

Pulsaderm ACE-All Blue is substantially equivalent to Nutra Light Blue (k132838) and Tanda Mini Skin Care System (k124042). Pulsaderm ACE-All Blue is a smaller version of the predicate device Nutra Light Blue (k132838) and shares the same features and characteristics as the predicate. Pulsaderm ACE-All Blue uses the same technology to deliver blue light and same indications for use as the predicates. Pulsaderm ACE-All Blue uses the same dose rate and treatment regimen as Tanda Mini Skin Care System (k124042). Any minor technological differences between the Pulsaderm ACE-All Blue and the predicate devices raise no new issues of safety or effectiveness.

## Comparison table:

| Device name         | Pulsaderm ACE-All Blue                                     | Nutra Light Blue                                                                                                                                            | Tanda mini skin care<br>system                                                         |
|---------------------|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| 510(K)              | Pending                                                    | K132838                                                                                                                                                     | K124042                                                                                |
| Intended use        | Indicated to treat<br>mild to moderate<br>acne on the face | Indicated to treat<br>dermatological<br>conditions. Specifically,<br>blue light modules are<br>indicated to treat mild to<br>moderate inflammatory<br>acne. | Indicated for over the<br>counter use for the<br>treatment of mild to<br>moderate acne |
| Energy type         | Light Emitting diodes<br>(LED's)                           | Light Emitting diodes<br>(LED's)                                                                                                                            | Light Emitting diodes<br>(LED's)                                                       |
| OTC                 | Yes                                                        | Yes                                                                                                                                                         | Yes                                                                                    |
| Wavelength<br>(nm)  | Blue light 415 +/-5 nm                                     | Blue light 415 +/- 5 nm                                                                                                                                     | Blue light 415+/- 5 nm                                                                 |
| Dose Rate<br>mW cm2 | Blue light: 25 mW cm2                                      | Blue light: 75 mW cm2                                                                                                                                       | Blue light 22.4 mW cm2                                                                 |
| Dose<br>J/cm2       | 12 J/cm2                                                   | 14 J/cm2                                                                                                                                                    | 12 J/cm2                                                                               |

{5}------------------------------------------------

# 7. Performance Data:

Performance testing was conducted and confirm compliance to design specifications; similar wavelength, output power and energy type as FDA cleared predicate devices. Pulsaderm ACE-All Blue has been tested and found in conformance with test standards: IEC 60601-1:2012, 60601-1-2:2014, and IEC 62471:2006. Biocompatibility testing to ISO 10993. Pulsaderm ACE-All Blue was designed and developed under a Risk Management System conforming to ISO 14971:2007.

## 8. Non clinical-testing

A lay-user study and self-selection study was conducted with Institutional Review Board (IRB) approval and oversight to determine if lay users could read the product labeling and then selfassess if the Pulsaderm ACE-ALL Blue would be beneficial for them to use. Study data was collected and demonstrated that the intended users of the device could successfully follow the instructions and use the device as intended. The performance data supplied in this 510(k) demonstrated that 98% of lay users were able to properly use the device by reading instructions in the user manual without any assistance.

In summary, the conclusion was that an average lay user can read and comprehend correctly the user manual and package labeling.

## 9. Conclusion

Based upon the analysis of the performance characteristics, safety characteristics and intended use for the Pulsaderm ACE-All Blue, Nutra Luxe MD, believes that no significant differences exist between this device and the predicates. Therefore, Substantial equivalence is requested.

---

**Source:** [https://fda.innolitics.com/device/K162371](https://fda.innolitics.com/device/K162371)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
