← Product Code [MCW](/productcode/MCW) · K162326

# Pantheris Catheter, Lightbix HS Imaging Console, Lightbox Sled (K162326)

_Avinger, Inc. · MCW · Oct 18, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162326

## Device Facts

- **Applicant:** Avinger, Inc.
- **Product Code:** [MCW](/productcode/MCW.md)
- **Decision Date:** Oct 18, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4875
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0mm, using OCT-assisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures and vessel morphologies. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

## Device Story

Pantheris System is a peripheral atherectomy catheter system utilizing Optical Coherence Tomography (OCT) for real-time image-guided plaque removal. The system comprises a sterile, single-use catheter (7Fr or 8Fr), a Lightbox Sled, and a Lightbox HS Imaging Console. During procedures, the catheter is inserted into the peripheral vasculature; the Lightbox transmits light through an optical fiber in the catheter to the intraluminal environment. The system receives and interprets backscattered light signals to generate real-time images of vessel lumen, wall structures, and morphologies. These images serve as an adjunct to fluoroscopy, allowing physicians to orient the atherectomy device precisely. The system is operated by physicians in a clinical setting to remove plaque from partially occluded vessels, potentially improving blood flow and patient outcomes. The PC-based console processes the OCT signals to provide the visualization platform.

## Clinical Evidence

No new clinical testing was performed for this submission. Substantial equivalence is supported by clinical data provided in the initial Pantheris System (K152275) submission and non-clinical bench testing, including design verification, validation, biocompatibility, electrical safety, and sterilization studies.

## Technological Characteristics

System utilizes Optical Coherence Tomography (OCT) for imaging. Catheter dimensions: 7Fr/8Fr, 110 cm length, compatible with 0.014" guidewires. Components: Catheter, Lightbox Sled, Lightbox HS Imaging Console (PC-based). Connectivity: Optical fiber interface. Sterilization: Single-use, sterile. Software: PC-based processing system for OCT signal interpretation.

## Regulatory Identification

An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)

## Predicate Devices

- Avinger Pantheris System ([K160827](/device/K160827.md))
- Avinger Pantheris System ([K152275](/device/K152275.md))
- Avinger Ocelot System ([K140185](/device/K140185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 18, 2016

Avinger, Inc. Ms. Patty Hevey Vice President, Clinical and Regulatory Affairs 400 Chesapeake Drive Redwood City, CA 94063

Re: K162326

Trade/Device Name: Pantheris System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW, NQQ Dated: August 17, 2016 Received: August 19, 2016

Dear Ms. Hevey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162326

Device Name Pantheris System

#### Indications for Use (Describe)

The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm, using OCT-assisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures and vessel morphologies.

The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

| Type of Use (Select one or both, as applicable)                                                |                                             |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 1.0 510(K) SUMMARY

## 510(k) Notification K162326

#### GENERAL INFORMATION

Applicant:

Avinger, Inc. 400 Chesapeake Drive Redwood City, CA 94063 U.S.A. Phone: 650-241-7900 Fax: 650-241-7901

Contact Person:

Patty Hevey Vice President, Clinical and Regulatory Affairs Avinger, Inc. Phone: 650-222- 3666 Fax: 650-241-7901

Date Prepared:

October 17th 2016

# DEVICE INFORMATION

Trade Name:

Pantheris System

Generic/Common Name:

Peripheral Atherectomy Catheter

Classification:

21 CFR§870.4875, Intraluminal Artery Stripper, Class II 21 CFR\$892.1560, Ultrasonic Pulsed Echo Imaging System, Class II

Product Codes:

MCW, NQQ

{4}------------------------------------------------

# PREDICATE DEVICES

- Primary Predicate Device Avinger Pantheris System (K160827), hereafter referred to ● as the current Pantheris System
- Secondary Predicate Device Avinger Pantheris System (K152275), hereafter referred ● to as the initial Pantheris System
- Secondary Predicate Device - Avinger Ocelot System (K140185), hereafter referred to as the Ocelot System

# DEVICE DESCRIPTION

The current Pantheris System (K160827), listed as the primary predicate device, consists of the Pantheris Catheter, Lightbox Sled with integrated Umbilical (referred to as "Sled") the Lightbox HS Imaging Console (referred to as "Lightbox") and the accessories that are packaged with the Pantheris Catheter: flush fixture, tweezers, syringes and stopcocks, and a Sterile Drape accessory that is packaged separately. The subject Pantheris System submitted as part of this 510(k) submission does not introduce any changes to the device specifications, materials or manufacturing processes and is identical to the current Pantheris System.

The Pantheris System combines the use of Avinger's Optical Coherence Tomography (OCT) technology (identical to the OCT Technology used in the Ocelot System (K140185)) with peripheral vascular atherectomy capabilities.

The Pantheris Catheter comes in both a 7Fr and 8Fr size. Both sizes have a working length of 110 cm and are sterile, single-use devices that are compatible with 7Fr and 8Fr sheaths (respectively) and 0.014" guidewires. The Pantheris Catheter incorporates an Optical Fiber that allows for real-time OCT guided directional atherectomy during the procedure.

The Pantheris Catheter is connected to the Lightbox via the Sled. The Lightbox is an optical transceiver, transmitting light to the intraluminal environment through the optical fiber on the Pantheris Catheter and receiving and interpreting the signal from the tissue using a PCbased processing system. The Lightbox provides a visualization platform for the real-time OCT-assisted directional atherectomy. The Lightbox consists of a cart with two monitors; a PC based processing system, an isolation transformer and an OCT system.

### PROPOSED INDICATIONS FOR USE

The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0mm, using OCT-assisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures and vessel morphologies. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

{5}------------------------------------------------

#### SUBSTANTIAL EQUIVALENCE

The subject Pantheris System and the primary predicate device, the current Pantheris System (cleared under K160827) are identical. They have the same device specifications, materials, manufacturing processes, device packaging, principles of operation for vessel visualization and OCT technological characteristics. Post the clearance of K160827 on March 1, 2016, the current Pantheris System's Lightbox software was updated to version 4.0.1. The proposed Indications for Use for the subject Pantheris System is very similar to the current Pantheris System (K160827) Indications for Use.

The subject Pantheris System and the secondary predicate device, the initial Pantheris System (cleared under K152275) are very similar. They have almost identical device specifications, materials, manufacturing processes, device packaging, principles of operation for vessel visualization and OCT technological characteristics.

The subject Pantheris System and the secondary predicate device, the Ocelot System (K140185) have very similar intended use, principles of operation for vessel visualization and OCT technological characteristics.

Any differences between the subject Pantheris System and the predicate devices do not alter the intended use or impact the safety and effectiveness of the Pantheris System.

### NONCLINICAL TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

The subject Pantheris System is identical to the primary predicate device, the Pantheris System cleared under K160827. The changes proposed in this submission do not alter the Pantheris System specifications, design, functionality or manufacturing process. The changes only pertain to updates to the labeling to accurately reflect the Pantheris System's current imaging functionality.

The non-clinical, bench testing conducted on the Pantheris System and provided in the prior K160827 and K152275 submissions, are applicable to the subject Pantheris System, including the following:

- Design verification and bench validation studies ●
- Packaging and shelf-life ●
- Software verification and validation
- o Electrical safety, electromagnetic compatibility, and laser safety testing
- o Biocompatibility
- In-vivo animal validation study
- Sterilization o

The collective results of the non-clinical testing demonstrate that the subject Pantheris System meets the established specifications necessary for consistent performance for its intended use.

{6}------------------------------------------------

Use of the same OCT technology between the subject and the predicate devices, along with successful OCT image comparisons between the subject Pantheris System, the Ocelot System (K140185) and images from the submitted compendium, confirm a determination of substantial equivalence between the subject device and the predicate devices regarding the updates to the subject Pantheris System labeling.

### CLINICAL TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

Clinical testing of the Pantheris System was included for the secondary predicate, the initial Pantheris System original 510(k) submission (K152275). Due to the equivalence between the Systems no new clinical testing is included in this 510(k) submission.

# CONCLUSION

The subject Pantheris System discussed in this submission is equivalent to the primary predicate device (current Pantheris System, K160827) in terms of specifications, design, functionality and manufacturing process. No new issues of safety or effectiveness are raised in this submission. The use of identical OCT technology in the predicate devices (the current Pantheris System (K160827), the intial Pantheris System (K152275) and the Ocelot System (K140185) and the successful OCT image comparisons, confirm that the Pantheris System is substantially equivalent to the predicate devices listed above.

---

**Source:** [https://fda.innolitics.com/device/K162326](https://fda.innolitics.com/device/K162326)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
