K162313 · Nu Vasive, Incorporated · PDQ · Mar 16, 2017 · Ear, Nose, Throat
Device Facts
Record ID
K162313
Device Name
NuVasive Next Generation NVM5 System
Applicant
Nu Vasive, Incorporated
Product Code
PDQ · Ear, Nose, Throat
Decision Date
Mar 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The Next Generation NVM5® System is a medical device that is intended for intraoperative neurophysiologic monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremities. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. NVM5 provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), transcranial or lumbar motor evoked potential (MEP) or somatosensory evoked potential (SSEP) responses of nerves. The System also integrates Bendini™ software used to locate spinal implant instrumentation for the placement of spinal rods. XLIF (Detection) — The XLIF (Detection) function allows the surgeon to locate and evaluate spinal nerves, and is used as a nerve avoidance tool. Basic & Dynamic Screw Test - The Screw Test functions allow the surgeon to locate and evaluate spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws. Free Run EMG - The Free Run EMG function identifies spontaneous EMG activity of spinal nerves by continually displaying a live stream waveform of any mechanically induced myotome contractions. Twitch Test (Train of Four) — The Twitch Test function allows the surgeon to assess moderate degrees of neuromuscular block in effect by evaluating muscle contraction following a train of four stimulation pulses. MEP - Transcranial or lumbar (i.e., conus in region of L1-L2) stimulation techniques for motor evoked potentials are used to assess for acute dysfunction in axonal conduction of the corticospinal tract and peripheral nerves. The MEP function provides an adjunctive method to allow the surgeon to monitor spinal cord and motor pathway integrity during procedures with a risk of surgically induced motor injury. SSEP - The SSEP function allows the surgeon to assess sensory spinal cord function in surgical procedures during which the spinal cord is at risk. Remote Access - The remote monitoring and local wireless control provides real-time capabilities to the NG-NVM5 System for additional physicians. Bendini - The Bendini Spinal Rod Bending function is used to locate spinal implant system instrumentation (screws, hooks) to determine their relative location to one another to generate bend instructions to shape a spinal rod. A surgeon is able to use those instructions and bend a rod using the Bendini Bender, a mechanical rod bender.
Device Story
NG-NVM5 System provides intraoperative neurophysiologic monitoring; inputs include electrical nerve stimulation via surgical accessory electrodes and EMG, MEP, or SSEP signal monitoring. System processes signals to assess nerve proximity, neuromuscular block, and spinal cord/motor pathway integrity. Outputs include live EMG waveforms, proximity alerts, and rod-bending instructions. Used in OR by surgeons; supports remote monitoring and local wireless control by additional physicians. Bendini software uses IR tracking to locate spinal instrumentation and generate rod-bending instructions. Device aids surgeons in nerve avoidance and monitoring during procedures with risk of surgically induced injury; benefits include real-time assessment of neurophysiologic status and improved spinal rod shaping.
Clinical Evidence
Bench testing only. Verification and validation testing performed per software requirements. Laboratory bench testing verified accuracy of angle and offset measurements and validated GUI and system component functionality. Current and power densities calculated for cutaneous electrodes.
Technological Characteristics
Class II surgical nerve stimulator/locator. Hardware includes Patient Module (PM), computer, and disposable conductive probes/electrodes. Sensing/actuation via electrical stimulation and EMG/MEP/SSEP monitoring. Bendini function uses optical (IR) tracking technology. Connectivity includes local wireless control and remote access. Compliant with IEC 60601-1. Software includes EMG, MEP, SSEP, Remote Access, and Bendini modalities.
Indications for Use
Indicated for intraoperative neurophysiologic monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremities in patients requiring assessment of neurophysiologic status.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 16, 2017
NuVasive Incorporated Mary Adams Regulatory Affairs Consultant 7475 Lusk Blvd. San Diego, California 92121
Re: K162313
Trade/Device Name: NuVasive Next Generation NVM5 System Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDQ, ETN, GWF, HAW, IKN, OLO Dated: February 15, 2017 Received: February 16, 2017
Dear Ms. Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162313
## Device Name
NuVasive® Next Generation NVM5 System
## Indications for Use (Describe)
The Next Generation NVM5® System is a medical device that is intraoperative neurophysiologic monitoring during spinal surgery, neck dissections, thoracic surger and lower extremities. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. NVM5 provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMC), transcranial or lumbar motor evoked potential (MEP) or somatosensory evoked potential (SSEP) responses of nerves. The System also integrates BendiniTM software used to locate spinal implant instrumentation for the placement of spinal rods. · XLIF (Detection) — The XLIF (Detection) function allows the surgeon to locate and evaltiate spinal nerves, and is used as a nerve avoidance tool.
· Basic & Dynamic Screw Test - The Screw Test functions allow the surgeon to locate and evaluate spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws.
· Free Run EMG - The Free Run EMG function identifies spontaneous EMG activity of spinal nerves by continually displaying alive stream waveform of any mechanically induced myotome contractions.
· Twitch Test (Train of Four) — The Twitch Test function allows the surgeon to assess moderate degrees of neuromuscular block in effect by evaluating muscle contraction following a train of four stimulation pulses.
· MEP - Transcranial or lumbar (i.e., conus in region ofLI-L2) stimulation techniques for motor evoked potentials are used to assess for acute dysfunction in axonal conduction of the corticospinal tract and peripheral nerves. The ME? function provides an adjunctive method to allow the surgeon to monitor spinal cord and motor pathway integrity during procedures with a risk of surgically induced motor injury.
· SSEP - The SSEP function allows the surgeon to assess sensory spinal cord function in surgical procedures during which the spinal cord is at risk.
• Remote Access - The remote monitoring and local wireless control provides real-time capabilities to the NG-NVM5 System for additional physicians.
· Bendini - The Bendini Spinal Rod Bending function is used to locate spinal implant system instrumentation (screws, hooks) to determine their relative location to one another to generate bend instructions to shape a spinal rod. A surgeon is able to use those instructions and bend a rod using the Bendini Bender, a mechanical rod bender.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
|-----------------------------------------------------------------------------------|--|
|-----------------------------------------------------------------------------------|--|
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font size.
## 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR \$807.92, the following summary of information is provided:
#### A. Submitted by:
Mary Adams Regulatory Affairs Consultant NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (909) 991-5235 Date Prepared: March 13, 2017
### B. Device Name
| | Trade or Proprietary Name: | NuVasive® Next Generation NVM5® System |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Common or Usual Name: | Neurological surgical monitor;<br>Stereotaxic Instrument | |
| Classification Name: | Surgical Nerve Stimulator/Locator;<br>Evoked response electrical stimulator;<br>Neurological stereotaxic instrument;<br>Electromyography (EMG) monitor/stimulator | |
| Device Class: | Class II | |
| Classification: | §874.1820, §882.1870, §882.4560, §890.1375 | |
| Product Code: | PDQ, ETN, GWF, HAW, IKN, OLO | |
#### C. Predicate Devices
The subject NuVasive Next Generation (NG) NVM5 System is substantially equivalent to the following predicate devices:
NuVasive NVM5 System K152942 (Primary Predicate) Medtronic NIM 3.0 Nerve Integrity Monitor K083124 (Additional Predicate) NuVasive NVM5 System K132694 (Reference Predicate for current density of the cutaneous electrodes)
#### D. Device Description
The NG-NVM5 System is a medical device that is intended for intraoperative neurophysiologic monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremities. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. NG-NVM5 provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), motor evoked potential (MEP) or somatosensory evoked potential (SSEP) responses of nerves. Moreover. a Twitch Test function is utilized to test the ability of the nerve to respond, or contract. following four stimulation pulses to determine the presence of neuromuscular block.
The System also integrates Bendini® software used to locate spinal implant instrumentation for the placement of spinal rods as well as providing guidance to support the delivery of pedicle
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Nuvasive. The logo consists of a purple and gray abstract leaf-like shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font size and a gradient of gray to purple.
screws during EMG monitoring. Lastly, the system offers real-time control capabilities (local wireless control) to the NG-NVM5 System by additional monitoring physicians, in addition to remote monitoring capabilities.
In summary, the NG-NVM5 System includes the following five (5) software functionalities / modalities:
- 1. Electromyography (EMG)
- Motor Evoked Potential (MEP) 2.
- 3. Somatosensory Evoked Potential (SSEP)
- 4. Remote Access
- 5. Bendini
The NG-NVM5 System hardware consists of a Patient Module (PM) and computer, as well as accompanying accessory components which consist of an assortment of disposable conductive probes, electrodes, and electrode leads.
#### Intended Use E.
The Next Generation NVM5® System is a medical device that is intended for intraoperative neurophysiologic monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremities. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. NVM5 provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), transcranial or lumbar motor evoked potential (MEP) or somatosensory evoked potential (SSEP) responses of nerves. The System also integrates Bendini™ software used to locate spinal implant instrumentation for the placement of spinal rods.
- XLIF (Detection) The XLIF (Detection) function allows the surgeon to locate and ● evaluate spinal nerves, and is used as a nerve avoidance tool.
- Basic & Dynamic Screw Test - The Screw Test functions allow the surgeon to locate and evaluate spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws.
- . Free Run EMG – The Free Run EMG function identifies spontaneous EMG activity of spinal nerves by continually displaying a live stream waveform of any mechanically induced myotome contractions.
- Twitch Test (Train of Four) - The Twitch Test Function allows the surgeon to assess moderate degrees of neuromuscular block in effect by evaluating muscle contraction following a train of four stimulation pulses.
- MEP - Transcranial or lumbar (i.e., conus in region of L1-L2) stimulation techniques for motor evoked potentials are used to assess for acute dysfunction in axonal conduction of the corticospinal tract and peripheral nerves. The MEP function provides an adjunctive method to allow the surgeon to monitor spinal cord and motor pathway integrity during procedures with a risk of surgically induced motor injury.
- SSEP The SSEP function allows the surgeon to assess sensory spinal cord function in ● surgical procedures during which the spinal cord is at risk.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a purple and gray leaf-like shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font size.
- Remote Access - The remote monitoring and local wireless control provides real-time capabilities to the NG-NVM5 System for additional physicians.
- Bendini - The Bendini Spinal Rod Bending function is used to locate spinal implant system instrumentation (screws, hooks) to determine their relative location to one another to generate bend instructions to shape a spinal rod. A surgeon is able to use those instructions and bend a rod using the Bendini Bender, a mechanical rod bender.
### F. Technological Characteristics
As was established in this submission, the subject NG-NVM5 System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, intended use, material composition, and functions. The technological differences within this 510(k) that were shown to be substantially equivalent to the predicates include additional angular and offset spinal parameter assessment.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font and purple color.
| Table 1 – Comparison of Technical Characteristics | | | |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Specification/<br>Property | Predicate Device<br>Medtronic NIM 3.0 Nerve<br>Monitoring System (K083124) | Predicate Device<br>NuVasive NVM5 System (K152942 and K132694) | Subject Device<br>NuVasive NG-NVM5 System |
| Intended Use /<br>Indications for<br>Use | Intended Use: The NIM 3.0 is intended for<br>locating and identifying cranial and<br>peripheral motor and mixed motor-sensory<br>nerves during surgery, including spinal cord<br>and spinal cord nerve roots. The APS<br>electrode is an accessory intended for<br>providing automatic periodic stimulation to<br>nerves when used with the Medtronic Nerve<br>Monitoring Systems.<br><br>Indications for NIM 3.0 EMG Monitoring<br>Procedures include:<br><br>Intracranial, Extracranial, Intratemporal,<br>Extratemporal, Neck Dissections, Thoracic<br>Surgeries, and Upper and Lower<br>Extremities<br><br>Indications for Spinal procedures which<br>may use NIM 3.0 EMG monitoring include:<br><br>Degenerative Treatments, Pedicle Screw<br>Procedures, Fusion Cages, Rhizotomy,<br>Orthopedic Surgery, Open and<br>Percutaneous Lumbar and Cervical Surgical<br>Procedures, and Thoracic Surgical<br>Procedures. | The NVM5® System is a medical device that is intended for intraoperative<br>neurophysiologic monitoring during spinal surgery. The device provides<br>information directly to the surgeon, to help assess a patient's<br>neurophysiologic status. NVM5 provides this information by electrically<br>stimulating nerves via electrodes located on surgical accessories and<br>monitoring electromyography (EMG), transcranial or lumbar motor evoked<br>potential (MEP), or somatosensory evoked potential (SSEP) responses of<br>nerves. The System also integrates Bendini® software used to locate spinal<br>implant instrumentation for the placement of spinal rods.<br>• XLIF (Detection) – The XLIF (Detection) function allows the surgeon to<br>locate and evaluate spinal nerves, and is used as a nerve avoidance tool.<br>• Basic & Dynamic Screw Test - The Screw Test functions allow the<br>surgeon to locate and evaluate spinal nerves by providing proximity<br>information before, during or after bone preparation and placement of<br>bone screws.<br>• Free Run EMG - The Free Run EMG function identifies spontaneous<br>EMG activity of spinal nerves by continually displaying a live stream<br>waveform of any mechanically induced myotome contractions.<br>• Twitch Test (Train of Four) - The Twitch Test Function allows the<br>surgeon to assess moderate degrees of neuromuscular block in effect by<br>evaluating muscle contraction following a train of four stimulation pulses.<br>• MEP - Transcranial or lumbar (i.e., conus in region of L1-L2) stimulation<br>techniques for motor evoked potentials are used to assess for acute<br>dysfunction in axonal conduction of the corticospinal tract and peripheral<br>nerves. The MEP function provides an adjunctive method to allow the<br>surgeon to monitor spinal cord and motor pathway integrity during<br>procedures with a risk of surgically induced motor injury.<br>• SSEP - The SSEP function allows the surgeon to assess sensory spinal<br>cord function in surgical procedures during which the spinal cord is at<br>risk.<br>• Remote Reader - The Remote reader function provides real time remote<br>access to the NVM5 System for a monitoring physician outside of the<br>operating room<br>• Guidance - The Guidance function is intended as an aid for use in either<br>open or percutaneous pedicle cannulation procedures in the lumbar and<br>sacral spine (L1-S1) of adult patients, and when used in conjunction with<br>radiographic imaging and EMG, allows the surgeon to assess the<br>angulation of system accessories relative to patient spinal anatomy for the<br>creation of a cannulation trajectory for bone screw placement.<br>• Bendini - The Bendini Spinal Rod Bending function is used to locate<br>spinal implant system instrumentation (screws, hooks) to determine their<br>relative location to one another to generate bend instructions to shape a<br>spinal rod. A surgeon is able to use those instructions and bend a rod<br>using the Bendini Bender, a mechanical rod bender. | The Next Generation NVM5® System is a medical device that is<br>intended for intraoperative neurophysiologic monitoring during spinal<br>surgery, neck dissections, thoracic surgeries, and upper and lower<br>extremities. The device provides information directly to the surgeon,<br>to help assess a patient's neurophysiologic status. NG-NVM5 provides<br>this information by electrically stimulating nerves via electrodes<br>located on surgical accessories and monitoring electromyography<br>(EMG), transcranial or lumbar motor evoked potential (MEP) or<br>somatosensory evoked potential (SSEP) responses of nerves. The<br>System also integrates Bendini™ software used to locate spinal implant<br>instrumentation for the placement of spinal rods.<br>• XLIF (Detection) - The XLIF (Detection) function allows the<br>surgeon to locate and evaluate spinal nerves, and is used as a nerve<br>avoidance tool.<br>• Basic & Dynamic Screw Test - The Screw Test functions allow the<br>surgeon to locate and evaluate spinal nerves by providing proximity<br>information before, during or after bone preparation and placement<br>of bone screws.<br>• Free Run EMG - The Free Run EMG function identifies<br>spontaneous EMG activity of spinal nerves by continually<br>displaying a live stream waveform of any mechanically induced<br>myotome contractions.<br>• Twitch Test (Train of Four) – The Twitch Test Function allows the<br>surgeon to assess moderate degrees of neuromuscular block in<br>effect by evaluating muscle contraction following a train of four<br>stimulation pulses.<br>• MEP - Transcranial or lumbar (i.e., conus in region of L1-L2)<br>stimulation techniques for motor evoked potentials are used to<br>assess for acute dysfunction in axonal conduction of the<br>corticospinal tract and peripheral nerves. The MEP function<br>provides an adjunctive method to allow the surgeon to monitor<br>spinal cord and motor pathway integrity during procedures with a<br>risk of surgically induced motor injury.<br>• SSEP - The SSEP function allows the surgeon to assess sensory<br>spinal cord function in surgical procedures during which the spinal<br>cord is at risk.<br>• Remote Access - The remote monitoring and local wireless control<br>provides real-time capabilities to the NG-NVM5 System for<br>additional physicians.<br>• Bendini - The Bendini Spinal Rod Bending function is used to<br>locate spinal implant system instrumentation (screws, hooks) to<br>determine their relative location to one another to generate bend<br>instructions to shape a spinal rod. A surgeon is able to use those<br>instructions and bend a rod using the Bendini Bender, a mechanical<br>rod bender. |
| Specification/<br>Property | Predicate Device<br>Medtronic NIM 3.0 Nerve<br>Monitoring System (K083124) | Predicate Device<br>NuVasive NVM5 System (K152942 and K132694) | Subject Device<br>NuVasive NG-NVM5 System |
| Software<br>Modalities /<br>Functionalities | • Stimulating EMG (manual<br>and APS)<br>• Free run EMG | • XLIF Detection<br>• Basic & Dynamic Screw Test<br>• Free Run EMG<br>• Twitch Test (Train of Four)<br>• MEP<br>• SSEP<br>• Remote Reader<br>• Guidance<br>• Bendini<br>o NuvaMap O.R.<br>• NuvaLine spinal parameter assessment tools (optional) | • XLIF (Detection)<br>• Basic & Dynamic Screw Test<br>• Free Run EMG<br>• Twitch Test<br>• MEP<br>• SSEP<br>• Remote Access (Remote Monitoring and Local<br>Wireless Control)<br>• Bendini<br>o NuvaMap O.R. |
| Algorithms | • APS | • XLIF Detection<br>• Basic & Dynamic Screw Test<br>• Free Run EMG<br>• Twitch Test (Train of Four)<br>• MEP<br>• SSEP<br>• Guidance<br>• Bendini - Rod bending and spinal parameter assessment | • XLIF Detection<br>• Basic & Dynamic Screw Test<br>• Free Run EMG<br>• Twitch Test (Train of Four)<br>• MEP<br>• SSEP<br>• Remote Access<br>• Bendini – Rod bending and spinal parameter<br>assessment |
| Total Available<br>Channels | Up to 8 | 32 | 32 |
| Headbox/<br>Patient Module | Unknown | Yes | Yes |
| IEC 60601-1<br>Compliant | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.