K162287 · Fujifilm Medical Systems U.S.A, Inc. · LLZ · Apr 6, 2017 · Radiology
Device Facts
Record ID
K162287
Device Name
Synapse 3D Perfusion Analysis
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
LLZ · Radiology
Decision Date
Apr 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Synapse 3D Perfusion Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Perfusion Analysis accepts DICOM compliant medical images acquired from CT and MR. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Addition to Synapse 3D Base Tools, Synapse 3D Perfusion Analysis provides the parameter images by post-processing of dynamic scanned CT arteriography and magnetic resonance (MR) images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time To Peak (TTP).
Device Story
Software module for Synapse 3D Base Tools; processes dynamic contrast-enhanced CT and MR images. Inputs: DICOM-compliant multi-phase 3D images. Operation: post-processing of dynamic scans to calculate perfusion parameters (Blood Volume, Blood Flow, Mean Transit Time, Time to Peak). Outputs: parametric maps and time-density curves. Used in clinical settings by trained medical professionals (radiologists/clinicians) to assess blood flow in brain, heart, and abdomen. Assists in reading, interpreting, and treatment planning. Benefits: provides quantitative perfusion data to support clinical decision-making.
Clinical Evidence
Bench testing only. Performance validated using system-level functionality, component, integration, and usability testing. Accuracy verified by comparing device-calculated perfusion parameters (Pc) against reference values (Pt) derived from perfusion models using arbitrary CT signal inputs. All tests passed with error rates within ±1%. No clinical prospective or retrospective studies were required.
Technological Characteristics
Medical imaging software; runs on Windows-based standalone or server/client configurations. Integrates with Synapse 3D Base Tools via DICOM standard. Processes dynamic CT/MR contrast-enhanced images. Generates parametric maps (BV, BF, MTT, TTP). Software-based analysis; no hardware components.
Indications for Use
Indicated for trained medical professionals to aid in reading, interpreting, reporting, and treatment planning using CT and MR images. Supports assessment of cerebral (CT/MR), myocardial (CT), and abdominal (CT) blood flows. Not for primary diagnostic interpretation of mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
FUJIFILM Medical Systems U.S.A., Inc. % Jyh-Shyan Lin, Ph.D. Senior Manager, Regulatory, Quality and Clinical Affairs 419 West Avenue STAMFORD CT 06902
Re: K162287
Trade/Device Name: Synapse 3D Perfusion Analysis Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ, JAK Dated: March 13, 2017 Received: March 15, 2017
Dear Dr. Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
April 6, 2017
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162287
Device Name Synapse 3D Perfusion Analysis
#### Indications for Use (Describe)
Synapse 3D Perfusion Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals, with tools to aid them in reading, interpreting, and treatment planning. Synapse 3D Perfusion Analysis accepts DICOM compliant medical images acquired from CT and MR. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Addition to Synapse 3D Base Tools, Synapse 3D Perfusion Analysis provides the parameter images by post-processing of dynamic scanned CT arteriography and magnetic resonance (MR) images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time To Peak (TTP),
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# FUJIFILM MEDICAL SYSTEMS U.S.A., Inc.
Synapse 3D Perfusion Analysis
#### 510(k) Summary
#### Date Prepared:
March 11, 2017
#### Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (301) 251-1092 Fax: (203) 602-3785 Contact: Jyh-Shyan Lin
#### Device Trade Name:
Synapse 3D Perfusion Analysis
#### Device Common Names:
Picture Archiving and Communications System (PACS)
#### Device Classification Name:
System, Image Processing, Radiological
#### Product Code:
LLZ
# Regulation Number:
21 CFR 892.2050
#### Device Class:
Class II
#### Panel:
Radiology
#### Predicate Devices:
- Synapse 3D Brain Perfusion (K120637), FUJIFILM Medical Systems U.S.A., Inc. o
- syngo CT Myocardial Perfusion (K150713), Siemens Medical Solutions USA, Inc. ●
- syngo Volume Perfusion-CT Body (K092013), Siemens Medical Solutions USA, Inc. o
{5}------------------------------------------------
# FUJIFILLM MEDICAL SYSTEMS U.S.A., Inc.
Synapse 3D Perfusion Analysis
#### 1. Description of the Device
Synapse 3D Perfusion Analysis is medical device software for Perfusion Analysis that provides the parameter images by post-processing of dynamic scanned CT arteriography and MR images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time to Peak (TTP). The perfusion parameter images can be presented to the trained medical professionals as the time-density curves (TDC) and perfusion characteristics maps (in parametric and summary) to assist them in assessing the blood flows. Synapse 3D Perfusion Analysis includes the following perfusion analysis applications.
- (1) Brain Perfusion (CT and MR) (unchanged from the FDA-cleared version K120637) is an application that analyzes the cerebral blood flow using the contrast-enhanced dynamic scanned CT and MR cerebral arteriography images. The Brain Perfusion post-processes the cerebral arteriography CT and MR images, and generates the parameter images of Cerebral Blood Volume (CBV), Cerebral Blood Flow (CBF), MTT, and TTP.
- (2) 4D Perfusion (CT) is an application that analyzes the changes in the cerebral blood flow over time (in 4D) using the contrast-enhanced multi-phase 3D whole-brain images. The 4D Perfusion post-processes the cerebral arteriography CT and MR images, and generates the 4D (over time) parameter images of CBV, CBF, MTT, and TTP.
- (3) Abdominal Perfusion (CT) is an application that analyzes the blood flow of abdominal organs over time (in 4D) using the contrast-enhanced multi-phase 3D abdomen images. The Abdominal Perfusion post-processes the abdominal CT images, and generates the parameter images of Tissue Blood Volume (TBV), Tissue Blood Flow (TBF), MTT, and TTP.
- (4) Cardiac Perfusion (CT) is an application that analyzes the myocardial blood flow using the multi-phase 3D heart images. The Cardiac Perfusion post-processes the contrast-enhanced Myocardial CT images, and generates the parameter images of Myocardial Blood Volume (MBV), Myocardial Blood Flow (MBF), MTT, and TTP.
Synapse 3D Perfusion Analysis runs on Windows standalone and server/client configuration installed on a commercial general-purpose Windows-compatible computer. Synapse 3D Perfusion Analysis is an optional software module that works with Synapse 3D Base Tools (cleared by CDRH via K120361 on 04/06/2012) which is connected through DICOM standard to medical devices such as CT, MR, CR, US, NM, PT, XA, etc. and to a PACS system storing data generated by these medical devices, and it retrieves image data via network communication based on the DICOM standard. The retrieved image data are stored on the local disk managed by Synapse 3D Base Tools, and the associated image-related information of the image data is registered in the database and is used for display, image processing, analysis, etc. Synapse 3D Perfusion Analysis can handle images of CT and MR. The software can display the images on a display monitor, or printed them on a hardcopy using a DICOM printer or a Windows printer.
{6}------------------------------------------------
# FUJIFILLM MEDICAL SYSTEMS U.S.A., Inc.
Synapse 3D Perfusion Analysis
### 2. Indications for Use
Synapse 3D Perfusion Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Perfusion Analysis accepts DICOM compliant medical images acquired from CT and MR. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Addition to Synapse 3D Base Tools, Synapse 3D Perfusion Analysis provides the parameter images by post-processing of dynamic scanned CT arteriography and magnetic resonance (MR) images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time To Peak (TTP).
### 3. Substantial Equivalence Comparison
Synapse 3D Perfusion Analysis has the same intended use, similar labeling, and clinical application tools as those of the cleared predicate devices. It assesses the blood flows in the same intended body parts of the predicate devices (Table 1).
| Subject body parts | Predicate devices | 510(k) # |
|--------------------|--------------------------------|----------|
| Cerebral region | Synapse 3D Brain Perfusion | K120637 |
| Cardiac region | syngo.CT Myocardial Perfusion | K150713 |
| Abdomen region | syngo Volume Perfusion-CT Body | K092013 |
Table 1. Perfusion analysis in different body parts
Synapse 3D Perfusion Analysis provides the parameter images by post-processing of dynamic scanned CT arteriography and magnetic resonance (MR) images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time To Peak (TTP) (Table 2).
Table 2. Device features and technological characteristics comparison
| Device Parameters | Synapse 3D<br>Perfusion Analysis<br>(This submission) | Synapse 3D<br>Brain Perfusion<br>(K120637) | Syngo.CT<br>Myocardial Perfusion<br>(K150713) | syngo<br>Volume Perfusion-CT Body<br>(K092013) |
|-----------------------------------------------------------------|-------------------------------------------------------|--------------------------------------------|-----------------------------------------------|------------------------------------------------|
| DICOM compliant<br>CT images with<br>contrast image<br>handling | Yes* | Yes | Yes | Yes |
| DICOM compliant<br>MR images with<br>contrast agent<br>handling | Yes | Yes | No | No |
{7}------------------------------------------------
# J | FILM MEDICAL SYSTEMS U.S.A., Inc.
Synapse 3D Perfusion Analysis
| Time-Signal Curve | Yes | Yes | Yes | Yes |
|-----------------------------------------------------------------------------|-----|-----|-----|---------|
| Time to Peak image | Yes | Yes | Yes | Unknown |
| Parameter images:<br>• Blood Volume<br>• Blood Flow<br>• Mean Transit Time | Yes | Yes | Yes | Yes |
*Yes: Equivalent features and technological characteristics
## 4. Safety Information
Synapse 3D Perfusion Analysis introduces no new safety or efficacy issues other than those already indentified with the predicate device. The Risk Management and the results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices." The Synapse 3D Perfusion Analysis labeling contains instructions for use and necessary cautions, warnings and notes to provide the safe and effective use of the device.
## 5. Testing and Performance Information
Synapse 3D Perfusion Analysis is tested successfully with reference to its product requirements, as well as design verification and validation documents and traceability matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Synapse 3D Perfusion Analysis software, which is found to be safe and effective and substantially equivalent to the predicate devices.
Testing involved system level functionality test, component testing, verification testing, integration testing, usability testing, installation/upgrade testing, labeling testing, as well as the testing for risk mitigations associated with the risk management process. In addition, we conducted benchmark performance testing using actual clinical images to help demonstrate that the proposed device achieved the expected accuracy performance.
Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests successfully passed.
The data used for performance testing are defined as follows.
- . Input data (Pi): an image which is created using an arbitrary CT signal value.
- Reference data (Pt): a value derived from perfusion calculation model using the input data (Pi). .
- . Result data (Pc): a value calculated by Synapse 3D Perfusion Analysis using the input data (Pi).
- Error: ((Pt Pc) / Pt)*(100%). o
- Pass criteria: the error is within ± 1%. ●
Based on the above definitions for the performance testing data, all of the test cases were tested and the performance testing results (accuracy test results including uncertainties) were passed.
{8}------------------------------------------------
# FUJIFILLM MEDICAL SYSTEMS U.S.A., Inc.
Synapse 3D Perfusion Analysis
### 6. Conclusion
This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.