The Medcomp® Pro-Lock CT Safety Infusion Set is intended for use in the administration of fluids and drugs as well as blood sampling through implanted vascular access ports. The Medcomp Pro-Lock CT Safety Infusion Set is also indicated for power injection of contrast media into the central venous system with implanted vascular access ports indicated for power injection. The maximum recommended infusion rate at 11.8 cPs is 5ml/sec for 20 gauge non-coring Huber style needles.
Device Story
Pro-Lock CT Safety Infusion Set is a non-coring Huber style needle for accessing implanted vascular access port septums. Device includes a safety mechanism to prevent accidental needle sticks during manual removal. Components include stainless steel needle cannula, plastic star needle guard, extension tubing, female Luer connector, and non-removable pinch clamp. Used in clinical settings for fluid/drug administration, blood sampling, and power injection of contrast media. Healthcare providers operate the device. Safety feature requires manual activation (lift) to shield needle after use. Benefits include reduced risk of sharps injury to clinicians and reliable vascular access for therapy or imaging.
Clinical Evidence
Bench testing only. Testing included priming volume, gravity flow, needle insertion/extraction force, air/liquid leakage, Luer lock fittings, occlusion, tensile strength, power injection simulation, coring evaluation, pull tests, corrosion resistance, shelf life, shipping, sharps injury prevention, fatigue, and MRI compatibility (ASTM F2052, F2119, F2182, F2213, F2503). Biocompatibility testing per ISO 10993-1 included sensitization, irritation, acute systemic toxicity, cytotoxicity, pyrogenicity, hemocompatibility, and material characterization.
Technological Characteristics
Huber style stainless steel needle; extension set with PVC female Luer and Pellethane tubing; polypropylene clamp/base; PVC sleeve. Sterilized via EO. Connectivity: None (mechanical). Software: None.
Indications for Use
Indicated for patients requiring administration of fluids, drugs, or blood sampling via implanted vascular access ports, and for power injection of contrast media into the central venous system via ports specifically indicated for power injection.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Medcomp® Power Injectable Safety Huber Needle (K132880)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three figures in profile facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 26, 2017
Medcomp® Ms. Courtney Nix Regulatory Affairs Manager, North America and EU 1499 Delp Drive Harleysville, Pennsylvania 19438
Re: K162271
Trade/Device Name: Pro-lock™ CT Safety Infusion Set Regulation Number: 21 CFR 880.5570 Regulation Name: Single Lumen Hypodermic Needle Regulatory Class: Class II Product Code: PTI Dated: March 20, 2017 Received: March 21, 2017
Dear Ms. Nix:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162271
Device Name Pro-lock™ CT Safety Infusion Set
## Indications for Use (Describe)
The Medcomp® Pro-Lock CT Safety Infusion Set is intended for use in the administration of fluids and drugs as well as blood sampling through implanted vascular access ports. The Medcomp Pro-Lock CT Safety Infusion Set is also indicated for power injection of contrast media into the central venous system with implanted vascular access ports indicated for power injection. The maximum recommended infusion rate at 11.8 cPs is 5ml/sec for 20 gauge non-coring Huber style needles.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| | | 510(k) SUMMARY | Traditional 510K |
|----|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|------------------|
| | K162271 | | |
| 1. | Submitter Information: | | |
| | Submitter: | Medcomp®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Tel: (215) 256-4201, x 2285<br>Fax: (215) 256-9191 | |
| | Registration Number: | 2518902 | |
| | Contact: | Courtney Nix<br>Cnix@Medcompnet.com<br>Regulatory Affairs Manager: North America and | |
| | EU | | |
| | Date Prepared: | 04/26/2017 | |
| 2. | Proposed or Subject Device Information: | | |
| | Trade Name: | Pro-Lock™ CT Safety Infusion Set | |
| | Device: | Non-Coring (Huber) Needle | |
| | Product Code: | PTI | |
| | Regulation Name: | Single Lumen Hypodermic Needle | |
| | C.F.R. Section: | 21 CFR 880.5570 | |
| | Class: | II | |
| | Regulation Medical<br>Specialty and Review<br>Panel: | General Hospital | |
| 3. | Predicate Device Information: | | |
| | 510(k) Number: | K132880 | |
| | Trade Name:<br>Needle | Medcomp® Power Injectable Safety Huber | |
| | Device: | Non-Coring (Huber) Needle | |
| | Product Code: | PTI | |
| | C.F.R. Section: | 21 CFR 880.5570 | |
| | | Medcomp®: Pro-Lock™ CT Safety Infusion Set | Page 1 of 6 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the MedComp logo. The logo features a stylized, geometric shape in red, resembling an abstract letter or symbol. Below the shape, the word "medCOMP" is written, with "med" in red and "COMP" in black. A registered trademark symbol is present next to the word.
4.
1499 Delo Drive
Harlevsville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
### Device Description:
The 20G x ¾" (19mm) Pro-Lock™ CT Safety Infusion Set is composed of a Huber style needle for port septum access having a safety feature designed to aid in the prevention accidental needle sticks when manually activated during needle removal. The needle is connected to a conventional style extension set. The proximal end of the extensions tubing attaches to a female Luer connector with removable cap creating a fluid path to the port. A nonremovable pinch clamp is located between the female Luer and needle cannula which is designed to restrict fluid flow through the extension tubing when engaged.
The needle cannula is constructed with a Huber style needle. The cannula is stainless steel and is shielded by a removable star needle guard of plastic construction.
#### 5. Indications for Use:
The Medcomp® Pro-Lock™ CT Safety Infusion Set is intended for use in the administration of fluids and drugs as well as blood sampling through implanted vascular access ports. The Medcomp® Pro-Lock™ CT Safety Infusion Set is also indicated for power iniection of contrast media into the central venous system with implanted vascular access ports indicated for power injection. The maximum recommended infusion rate at 11.8 cPs is 5 ml/sec for 20 gauge non-corina Huber style needles.
#### Comparison to Predicate Devices: 6.
The 20G x ¾" (19mm) Pro-Lock™ CT Safety Infusion Set is substantially equivalent to the predicate device, Medcomp® Power Injectable Safety Huber Needle (K132880), in terms of indications for use, intended use, fluid path materials, biocompatibility, basic design, performance, labeling, and manufacturing process.
The difference between the predicate device, Medcomp® Power Injectable Safety Huber Needle (K132880), and the 20G x 34" (19mm) Pro-Lock™ CT Safety Infusion Set are the changes to the clamp material from acetal to Polypropylene, and the ergonomics of the safety feature to allow better grip during safety feature activation.
The indications for use differ in regards to the name of the device. There are no other differences in the indications for use statements.
### Table 6.1: 510(K) Summary: Design Comparison Matrix
| Device | Proposed Device: | Predicate Device: |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 20G x 3/4" (19mm) Pro-<br>LockTM CT Safety Infusion<br>Set | Medcomp® Power Injectable<br>Safety Huber Needle<br>(K132880) |
| Design | Anti-coring Huber style needle for port septum access having a safety feature which that will prevent accidental needle sticks. The needle is connected to a conventional style extension set for attachment to standard IV/Drug infusion line sets. | Anti-coring Huber style needle for port septum access having a safety feature which that will prevent accidental needle sticks. The needle is connected to a conventional style extension set for attachment to standard IV/Drug infusion line sets. |
| Dimensions or Lengths | $3/4$ " | $1/2$ ", $3/4$ " 1", 1- $1/2$ " |
| Indications for Use | The Medcomp® Pro-Lock™ CT Safety Infusion Set is intended for use in the administration of fluids and drugs as well as blood sampling through implanted vascular access ports. The Medcomp® Pro-Lock™ CT Safety Infusion Set is also indicated for power injection of contrast media into the central venous system with implanted vascular access ports indicated for power injection. The maximum recommended infusion rate at 11.8 cPs is 5 ml/sec for 20 gauge non-coring Huber style needles. | The Medcomp® Power Injectable Safety Huber Needle is intended for use in the administration of fluids and drugs as well as blood sampling through implanted vascular access ports. The Medcomp® Power Injectable Safety Huber Needle is also indicated for power injection of contrast media into the central venous system with implanted vascular access ports indicated for power injection. The maximum recommended infusion rate at 11.8 cPs is 5 ml/sec for 20 gauge non-coring Huber style needles. |
| Gauge Sizes | 20 GA | 19, 20, 22 GA |
| Sterilization Method | EO | EO |
| Fluid Path Materials | Cannula: Stainless Steel<br>Female Luer: PVC<br>Extension: Pellethane | Cannula: Stainless Steel<br>Female Luer: PVC<br>Extension: Pellethane |
| Skin Contacting Materials | Clamp: Polypropylene<br>Base: Polypropylene<br>Sleeve: PVC | Clamp: Acetal<br>Base: Polypropylene<br>Sleeve: PVC |
| Performance Testing | Maximum Infusion Rate:<br>11.8 cPs is 5 ml/sec for 20 gauge<br><br>Priming Volume: 0.20cc<br><br>Maximum Flow Rate:<br>5ml/second 325 psi max | Maximum Infusion Rate:<br>11.8 cPs is 5 ml/sec for 20 gauge<br><br>Priming Volume: 0.18cc<br><br>Maximum Flow Rate:<br>5ml/second 325 psi max |
| Patient<br>Ergonomics | Lift required to engage<br>safety lock, when locked<br>needle is a blind hole, and<br>catch for extension tubing | Lift and squeeze required to<br>engage safety lock, when<br>locked needle is in an open<br>hole, and no catch for<br>extension tube |
Medcomp®: Pro-Lock™ CT Safety Infusion Set
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image features the logo for medCOMP. The logo consists of a stylized, geometric shape in red, resembling a house or an abstract letter. Below the shape, the word "medCOMP" is written, with "med" in red and "COMP" in black.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape above the word "medCOMP". The word "med" is in red, and the word "COMP" is in black.
9.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
### Bench / Performance Data / Non-Clinical Testing:
The results of performance testing, in conjunction with the substantial equivalence claims, effectively demonstrate the proposed device, 20G x ¾" (19mm) Pro-Lock™ CT Safety Infusion Set, is equivalent to the predicate device, Medcomp® Power Injectable Safety Huber Needle (K132880).
| Test Performed | Test Method |
|-------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Priming Volume | Per Internal Test Method |
| Gravity Flow | ISO 10555-1: 2013 – Incorporating corrigendum April 2014<br>– Intravascular catheters – Sterile and single-use catheters<br>Part 1: General Requirements |
| Needle<br>Insertion/Extraction<br>Force | ISO 10555-6: 2015 – Intravascular catheters – Sterile and<br>single-use catheters – Part 6: Subcutaneous implanted |
| Air Leakage | ISO 8536-8: 2004 – Infusion equipment for medical use –<br>Part 8: Infusion equipment for use with pressure infusion<br>apparatus |
| Test Performed | Test Method |
| Liquid Leakage | ISO 10555-1: 2013 – Incorporating corrigendum April 2014<br>– Intravascular catheters – Sterile and single-use catheters<br>Part 1: General Requirements, Annex C |
| Luer Lock Fittings | ISO 594-2: 1998 – Conical fittings with 6 % (Luer) taper for<br>syringes, needles and certain other medical equipment |
| Occlusion with<br>Clamp | Per Internal Test Method |
| Extension Tensile<br>and % Elongation | ISO 10555-1: 2013 – Incorporating corrigendum April 2014<br>– Intravascular catheters – Sterile and single-use catheters<br>Part 1: General Requirements |
| Power Injection<br>Simulation | ISO 10555-1: 2013 – Incorporating corrigendum April 2014<br>– Intravascular catheters – Sterile and single-use catheters<br>Part 1: General Requirements |
| Port Septum/Needle<br>Evaluation for<br>Coring | Per Internal Test Method |
| Break Pull Test/<br>Static Load Pull<br>Test | ISO 10555-1: 2013 – Incorporating corrigendum April 2014<br>– Intravascular catheters – Sterile and single-use catheters<br>Part 1: General Requirements |
| Needle to Extension<br>Joint Pull | ISO 10555-1: 2013 – Incorporating corrigendum April 2014<br>– Intravascular catheters – Sterile and single-use catheters<br>Part 1: General Requirements |
| Corrosion<br>Resistance | ISO 11070: 2014 – Sterile single-use intravascular<br>introducers, dilators and guidewires |
| Shelf Life (3 years) | ISO 11607-1: 2009+A1:2014- Packaging for terminally<br>sterilized medical devices - Part 1: Requirements for<br>materials sterile barrier systems and packaging<br>ASTM F1980-16, Standard guide for accelerated aging of<br>sterile barrier systems for medical<br>ISO 10555-1: 2013 – Incorporating corrigendum April 2014<br>– Intravascular catheters – Sterile and single-use catheters-<br>Part 1: General Requirements<br>ASTM F-1929-15-, Standard Test Method for Detecting<br>Seal Leaks in Porous Medical Packaging by Dye<br>Penetration<br>ASTM F-1140- Standard Test Methods for Internal<br>Pressurization Failure Resistance of Unrestrained<br>Packages |
| Shipping Test | ISO 11607-1: 2009+A1:2014- Packaging for terminally<br>sterilized medical devices - Part 1: Requirements for<br>materials sterile barrier systems and packaging<br>ISTA-2A-2011 - Performance Tests for Packaged-Product<br>Packaged-Products 150lb (68 kg) or Less<br>ASTM F-1929-15-, Standard Test Method for Detecting<br>Seal Leaks in Porous Medical Packaging by Dye<br>Penetration<br>ASTM F-1140- Standard Test Methods for Internal<br>Pressurization Failure Resistance of Unrestrained<br>Packages |
| Simulated Use<br>Study: Sharps Injury<br>Prevention | FDA's Guidance: Medical Devices with Sharps Injury<br>Prevention Features |
| Living Hinge<br>Fatigue | Per Internal Test Method |
| Force at Break | Per Internal Test Method |
| Evaluation of<br>Magnetic Field<br>Interactions,<br>Heating, and<br>Artifacts | ASTM F2052-15 – Standard Test Method for Measurement<br>of Magnetically Induced Displacement Force on Medical<br>Devices in the Magnetic Resonance Environment |
| | ASTM F2119-07 – Standard Test Method for Evaluation of<br>MR Image Artifacts from Passive Implants<br>F2182-11a- Standard Test Method for Measurement of<br>Radio Frequency Induced Heating On or Near Passive<br>Implants During Magnetic Resonance Imaging<br>F2213-06– Standard test Method for Measurement of<br>Magnetically Induced Torque on Medical Devices in the<br>Magnetic Resonance Environment<br>ASTM F 2503-13 - Standard practice for Marking Medical<br>Devices and Other Items for Safety in the Magnetic |
Medcomp®: Pro-Lock™ CT Safety Infusion Set
Page 4 of 6
{7}------------------------------------------------
Medcomp®: Pro-Lock™ CT Safety Infusion Set
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the MedComp logo. The logo consists of a red geometric shape that resembles a stylized letter "M" or a house outline. Below the shape, the word "medCOMP" is written, with "med" in red and "COMP" in black.
1499 Delo Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
## Biocompatibility:
10.
Biocompatibility was performed on the final, finished 20G x ¾" (19mm) Pro-Lock™ CT Safety Infusion Set per ISO 10993-1 for an externally communicating device, with indirect blood path exposure for a prolonged duration with the body. The biological endpoints include:
- Sensitization/Irritation: ISO 10993-10: 2010 Biological Evaluation of o Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
- Acute Systemic Toxicity: ISO 10993-11: 2006 Biological Evaluation of o Medical Devices - Part 11: Tess for Systemic Toxicity
- Cytotoxicity: ISO 10993-5: 2009 Biological Evaluation of Medical o Devices - Part 10: Tests for in vitro Cytotoxicity
- Pvrogenicity: ISO 10993-11: 2006 Biological Evaluation of Medical o Devices - Part 11: Test for Systemic Toxicity
- Hemocompatibility: ISO 10993-4: 2002 Amended 2006 Biological o Evaluation of Medical Devices - Part 4: Selection for tests for interactions with blood
- Material Characterization: ISO 10993-18: 2005 Biological Evaluation of . Medical Devices - Part 18: Chemical Characterization of Materials
#### 11. Summary of Substantial Equivalence:
In conclusion, the proposed device, 20G x ¾" (19mm) Pro-Lock™ CT Safety Infusion Set, is considered substantially equivalent to the predicate device, Medcomp® Power Injectable Safety Huber Needle (K132880) as demonstrated through non-clinical testing performed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.