Ax Stand-Alone ALIF System

K162236 · Innovasis, Inc. · OVD · Apr 20, 2017 · Orthopedic

Device Facts

Record IDK162236
Device NameAx Stand-Alone ALIF System
ApplicantInnovasis, Inc.
Product CodeOVD · Orthopedic
Decision DateApr 20, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Innovasis Ax Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved levels(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The Ax Stand-Alone interbody implants with a lordotic angle less than 20°, when used with the internal fixation screws, do not require use of supplemental fixation.

Device Story

Intervertebral body fusion device for lumbar spine; anterior (ALIF) approach. Two configurations: HA-enhanced PEEK cage with integrated titanium faceplate or all-titanium version. Includes bone screws and locking mechanism. Interior packed with autograft to facilitate fusion. Used in clinical settings by surgeons. Implants <20° lordosis function as stand-alone devices; implants ≥20° require supplemental fixation (e.g., Excella Spinal System). Supports restoration of sagittal balance. Benefits patient by stabilizing spinal segments and promoting fusion in DDD cases.

Clinical Evidence

Bench testing only. Performance testing conducted per ASTM F2077-14 and F2267-04, including static axial compression, dynamic axial compression, subsidence, expulsion, and side-by-side antibackout testing.

Technological Characteristics

Materials: Invibio PEEK-OPTIMA HA Enhanced, Unalloyed Tantalum (ASTM F560), Ti-6Al-4V ELI, 465/17-4 Stainless Steel (ASTM F899), Silicone, Radel (ASTM D6394), 5052 Aluminum, Polypropylene. Stand-alone interbody fusion cage with integrated titanium faceplate, screws, and locking mechanism. Anterior approach. No software/algorithm.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous lumbar levels (L2-S1) with discogenic back pain, confirmed by history/radiography, and at least 6 months of failed non-operative treatment. May include up to Grade 1 spondylolisthesis or retrolisthesis.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 20, 2017 Innovasis, Inc. Marshall McCarty Director OA/RA 614 East 3900 South Salt Lake City, Utah 84107 Re: K162236 Trade/Device Name: Ax™ Stand-Alone ALIF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD Dated: March 8, 2017 Received: March 21, 2017 Dear Mr. McCarty: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. K162236 Page 1 of 1 510(k) Number (if known) K162236 Device Name AxTM Stand-Alone ALIF System #### Indications for Use (Describe) The Innovasis Ax Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved levels(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The Ax Stand-Alone interbody implants with a lordotic angle less than 20°, when used with the internal fixation screws, do not require use of supplemental fixation. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Innovasis. The logo is black and white and features the word "INNOVASIS" in large, bold letters. Below the word "INNOVASIS" is the phrase "INNOVATE / INVOLVE / INVENT" in smaller letters. The logo is simple and modern. #### 510(k) Summary Report: 4.0 K162236 Page 1 of 3 ## Ax™ Stand-Alone ALIF System Company: Innovasis, Inc. 614 E. 3900 South Salt Lake City, UT 84107 - Marshall C. McCarty Contact: Phone: (801) 261-2236 mmccarty@innovasis.com - Trade Name: Ax™ Stand-Alone ALIF System Common Name: Intervertebral fusion device with bone graft - Classification: Requlation No.: 21CFR 888.3080 Class 2 Product Code: OVD Review Panel: Division of Orthopedic Devices (DOD) - Anterior Spine Devices Branch (ASDB) - Integra Vu a•POD™ Prime System Primary Predicate: K121211 (SeaSpine has separated from Integra and now owns this product line.) This predicate has not been subject to a design-related recall. ## Additional Predicates: Nuvasive® Brigade™ Hyperlordotic Implant System K123045 Innovasis® Px HA™ PEEK IBF System K151785 - Device Description: The Innovasis Ax Stand-Alone ALIF implant will be provided in two configurations. The first is an HA PEEK caqe with an integrated titanium faceplate. The second is an all titanium version of the same implant. Both designs will feature screws, a locking mechanism, and instruments. The implant is a stand-alone anterior intervertebral body fusion device indicated for use in patients with degenerative disk disease (DDD) at one or two contiquous levels from L2 to S1. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). - The interior of the device will be packed with autograft. - Intended to support restoration of the sagittal balance. - It is intended to be used with bone screws and will not require additional supplementary fixation for implants with a lordotic angle less than 20°. - Implants with a lordotic angle ≥20° are indicated for use with supplementary fixation, such as the Excella Spinal System. - It is intended for use with the standard anterior approach {4}------------------------------------------------ Image /page/4/Picture/21 description: The image shows the word "INNOVASIS" in a bold, sans-serif font. The letter "A" in the word is stylized with a diagonal line extending upwards and to the right, creating a dynamic and modern look. The overall design is clean and professional, suggesting a company or brand focused on innovation. April 18, 2017 # K162236 Page 2 of 3 Performance Data: (Non-clinical)—Performance testing per ASTM F2077-14 and F2267-04 for Static Axial Compression, Dynamic Axial Compression, Subsidence, Expulsion and side by side antibackout testing indicates that the Ax Stand-Alone ALIF is substantially equivalent to the predicates. #### Materials: Implants - *Invibio® PEEK-OPTIMA® HA Enhanced . Polyetheretherketone with hydroxyapatite for surgical implant - Unalloyed Tantalum per ASTM F560 ● - . Titanium-6-Aluminum-4-Vanadium ELI (Extra Low Interstitial) Allov Instruments - 465 Stainless Steel per ASTM F899 ● - 17-4 Stainless Steel per ASTM F899 ● - Silicone Handles (Tested for biocompatibility) - o Radel Handle per ASTM D6394 Sterilization Trays - Anodized 5052 Aluminum ● - Polvpropylene ● - Radel *Invibio and PEEK-OPTIMA® are registered trademarks of Invibio Limited. All rights reserved. Indications for Use: The Innovasis Ax Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved levels(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The Ax Stand-Alone interbody implants with a lordotic angle less than 20°, when used with the internal fixation screws, do not require use of supplemental fixation. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Innovasis. The logo is black and features the word "INNOVASIS" in all caps. Below the word is the phrase "INNOVATE / INVOLVE / INVENT" in a smaller font. The logo is simple and modern. April 18, 2017 ## Basis for Substantial Equivalence: K162236 Page 3 of 3 The Ax Stand-Alone ALIF System has been subjected to risk analysis, engineering analysis and testing to recognized standards and has been shown to be substantially equivalent to the predicate device, K121211 Integra Vu a•POD, and reference devices, K151785 Innovasis Px HA PEEK IBF System and K123045 NuVasive Brigade HL. - -Materials (biocompatibility profile and processing) are substantially equivalent to predicate device. - -Technology is substantially equivalent to predicate device. - Design is substantially equivalent to predicate device. - - -Sizes are substantially equivalent to predicate device. - Bone graft window/cavity is substantially equivalent to predicate device. - -Mechanical strength is substantially equivalent to predicate device. - -Indications for use are substantially equivalent to predicate device. Conclusion: The conclusions drawn from the nonclinical tests demonstrate that the subject device is substantially equivalent to the leqally marketed predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...