The ArtPix Mobile EZ2GO, is intended for use in general radiographic examinations by qualified/trained doctors or technicians, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ArtPix Mobile EZ2GO allows imaging of the skull, chest, shoulders, spine, abdomen, and extremities. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
Device Story
ArtPix Mobile EZ2GO is a digital radiography (DR) system for general radiographic examinations. It captures X-ray images using portable flat-panel detectors (FPDs) and processes them via a tablet computer. The system is used by qualified doctors or technicians in clinical settings. Inputs are X-ray exposures; the system transforms these into digital images displayed on the tablet. Output features include window/level adjustment, image rotation, inversion, zoom, and contrast enhancement. Images are transferred via DICOM to PACS or printers. The device improves workflow efficiency and portability compared to traditional systems. It benefits patients by enabling rapid, high-resolution imaging in various positions (sitting, standing, lying).
Clinical Evidence
Bench testing only. Validation included software unit testing, system integration testing, and image quality performance measurements. All system requirements and risk management files were verified with no open critical safety or efficacy issues.
Technological Characteristics
Digital radiography system using Pixium Portable 3543EZ/2430EZ flat-panel detectors (CsI or Gadox scintillators). Tablet-based workstation (i5-4310U, 8GB RAM, Windows 7). Connectivity via WiFi AP and 1Gig LAN (DICOM 3.0). Complies with IEC 60601-1, IEC 60601-1-2, IEC 62304, and ISO 14971. Power: 100-240V AC.
Indications for Use
Indicated for general radiographic examinations of the skull, chest, shoulders, spine, abdomen, and extremities in patients of any age or gender, excluding fluoroscopy, angiography, and mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, wave-like appearance. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2016
CMT Medical Technologies, Ltd. % Ms. Lilia Schlosberg Director of Quality and Regulatory Affairs Hacarmel St. Bld 7/2, POB 111, Industrial Park Yoqneam Ilit, 20692 ISRAEL
Re: K162224
Trade/Device Name: ArtPix Mobile EZ2GO Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR, MQB Dated: August 21, 2016 Received: August 24, 2016
Dear Ms. Schlosberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K162224
Device Name ArtPix Mobile EZ2GO
Indications for Use (Describe)
The ArtPix Mobile EZ2GO, is intended for use in general radiographic examinations by qualified/trained doctors or technicians, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography.
ArtPix Mobile EZ2GO allows imaging of the skull, chest, shoulders, spine, abdomen, and extremities. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1. 510(k) Summary K162224
#### 1.1. Proprietary Device Name:
ArtPix Mobile EZ2GO
#### 1.2. Establishment Name and Registration Number of Submitter
Manufacture Name: CMT Medical Technologies Ltd. Hacarmel St, Building 7/2 Industrial Park Yokneam Illit 20692 Israel
Registration: 8030112
Submission contact: Lilia Schlosberg Regulatory Affairs Director Tel: +972- 4- 8566233 Mobile: +972- 53 -6247071 Fax: E-mail: lilia.schlosberg@cmt-med.com
#### 1.3. Device Classification
| Classification Name: | Stationary X-Ray System |
|------------------------------|--------------------------------------------------------------|
| Classification Regulation: | 21CFR 892.1680 |
| Classification Panel: | Radiology |
| Device Class: | Class II |
| Classification Product Code: | KPR (System X-Ray Stationary) |
| Subsequent Product Code: | MQB - solid-state X-Ray imager (flat panel / digital imager) |
#### 1.4. Reason for 510K Submission
Special 510K Submission
## 1.5. Identification of Legally Marketed equivalent Devices
PrestoDR 4143 K110849
| Classification Name: | Stationary X-Ray System |
|------------------------------|--------------------------------------------------------------|
| Classification Regulation: | 21CFR 892.1680 |
| Classification Panel: | Radiology |
| Device Class: | Class II |
| Classification Product Code: | KPR (System X-Ray Stationary) |
| Subsequent Product Code: | MQB - solid-state X-Ray imager (flat panel / digital imager) |
{4}------------------------------------------------
# 1.6. Device Description
The modified ArtPix Mobile EZ2GO is a flexible high-resolution Digital Radiography (DR) system that is designed to provide fast and smooth radiography examinations of sitting, standing or lying patient Imaging. The modified ArtPix Mobile EZ2GO consists of the following components: Tablet Computer, Flat Panel Detectors, Detector Removable batteries & Charger, and other optional components such as External Monitor for Docking Station, USB HUB, Bluetooth DAP device and Fix Room Access Point Support. Images may be transferred via a network using DICOM protocol for printing and storage in PACS.
The modified ArtPix Mobile EZ2GO can be used in combination with any conventional analog radiography system currently using film screen or CR, fixed or mobile, or other system capable of producing general radiography exposures.
The modified Artpix Mobile EZ2GO detector can be shared between multiple Artpix Mobile EZ2GO systems.
# 1.7. The Device Intended and Indications for Use
## Indications for use
The ArtPix Mobile EZ2GO, is intended for use in general radiographic examinations by qualified/trained doctors or technicians, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography.
ArtPix Mobile EZ2GO allows imaging of the skull, chest, shoulders, spine, abdomen, and extremities. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
Statement: CMT Medical Technologies Ltd hereby states that in terms of fundamental scientific technology, indication for use, safety and effectiveness the submitted modified The ArtPix Mobile EZ2GO device is substantially equivalent to the currently marketed predicate device PrestoDR 4143 K110849.
# 1.8. Standards & Guidance
The Modified ArtPix Mobile EZ2GO complies with the following International and FDA recognized consensus and FDA Guidance:
- > EN/IEC/UL IEC 60601-1 31d edition "Medical Electrical equipment. Part 1: General requirements for safety".
- > IEC 60601-1-2 3te edition Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- > IEC 60601-1-6:2013-Ed.3.1 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- > IEC 62304:2006 Medical device software Software life cycle processes
{5}------------------------------------------------
- > IEC 60601-2-54:2009-Ed.1.0 Medical electrical equipment Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy IEC 60601-2-54:2009-Ed.1.0/Cor.1:2010
- > IEC 60601-1-3:2008-Ed.2.0 Medical electrical equipment Part 1-3: General requirements for basic safety and essential performance - Collateral standard: Radiation protection in diagnostic X-ray equipment
- > IEC TR 60878, "Graphical symbols for electrical equipment in medical practice"
- > NEMA PS 3.1 1 3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) Set.
- > EN ISO 14971:2012 Application of risk management to medical devices
- > ISO 15223-1:2012, Medical devices -- Symbols to be used with medical device labels, labeling and information to be supplied
- > CFR 1020.30 Diagnostic x-ray systems and their major components.
- > CFR 1020.31 Radiographic equipment.
- Device specific guidance document, titled "Guidance for the Submission of 510(k)'s for A Solid State X-ray Imaging Devices August 6, 1999,"
- > ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process ISO 10993-1:2009/Cor.1:2010
{6}------------------------------------------------
# 1.9. Summary of the Basis for Substantial Equivalence
This submission of the ArtPix Mobile EZ2GO modification presents changes of the legally marketed unmodified device PRESTODR 4143 K110849 .
The ArtPix Mobile EZ2GO is a Digital Radiography system, featuring the Pixium Portable 3543EZ / 2430EZ family Portable Flat Panel Detectors manufactured by Trixell (Pixium Portable 3543EZ, Pixium Portable 3543EZ-C,Pixium Portable 3543EZh-C, Pixium Portable 3543EZ-G, Pixium Portable 2430EZ-C, Pixium Portable 2430EZ-G, Pixium Portable 3543EZh Siemens branded "MAX wi-D", Pixium Portable 2430EZ Siemens branded "MAX mini") and CMT's proprietary technology, which incorporates a tablet computer and state of the art objectoriented software.
The legally marketed PRESTODR 4143 K110849 has been modified: To integrate the Pixium 3543EZ / 2430EZ family Portable Flat Panel Detectors and improve cost effectiveness. Aging technologies and components (hardware and software) have been redesigned. The devices major functions, intended use and principle of operation were not changed.
The modified device ArtPix Mobile EZ2GO consists of almost the same architecture, technologies and algorithms of the predicate PRESTODR 4143 K110849. The topics of the design modifications are presented below:
## Hardware modifications:
The predicate device PRESTODR 4143 K110849 was minorly modified:
- 1. Predicate device PRESTODR 4143 K110849 supports Pixium 4143 FPD (Flat Panel Detector) and Pixium Portable 3543 FPD. The modified device ArtPix Mobile EZ2GO is using the Pixium 3543EZ / 2430EZ family Portable Flat Panel Detectors. The Pixium Flat Panel Detectors supported in the modified device ArtPix Mobile EZ2GO have the same architecture and functionality as in the PRESTODR 4143 K110849, and have already been cleared by FDA in the currently marketed (K141440) dicomPACS DX-R -Oehm und Rehbein GmbH.
- 2. The PRESTODR 4143 K110849 PC computer was replaced by ArtPix Mobile EZ2GO tablet. In terms of technical performance (Processing Capabilities, Display and Touch), the tablet is no worse than the stationary PC used in PrestoDR 4143, except for hard drive capacity, which is a marginal feature with no impact on the fundamental scientific technology, indication for use, safety and effectiveness.
## Software modifications:
The software basic functions were not changed. User interface was changed in order to be upto-date with the modern applications. Additional code was added to support the Pixium 3543EZ / 2430EZ family Portable Flat Panel Detectors.
The submitted device description includes enhanced connectivity, processing time, convenience of use, production ability, and serviceability.
Summary: This submission presents design changes. The modified device ArtPix Mobile EZ2GO is compared to the predicate PRESTODR 4143 K110849. It has the same intended use
{7}------------------------------------------------
(see the comparison table, section 4.2below). The changed device has almost the same technological and performance characteristics of the predicate devices (see the comparison table, section 0) and, in CMT's opinion, these modifications do not raise new types of safety or effectiveness concerns. This submission includes detailed device descriptive and performance information (see sections 1-6 and appendices A-H) that demonstrate that the device is substantially equivalent to the predicate device.
#### 1.9.1. Comparison of indications for use
| | Comparison<br>content | PrestoDR 4143 | ArtPix Mobile EZ2GO |
|---|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | K number | K110849 | |
| 2 | Owner | CMT Medical Technologies Ltd | CMT Medical Technologies Ltd |
| 3 | Indication for<br>use | The PrestoDR 4143 , is intended for<br>use in general radiographic<br>examinations, wherever conventional<br>screen-film systems may be used,<br>excluding fluoroscopy, angiography<br>and mammography. PrestoDR 4143<br>allows imaging of the skull, chest,<br>shoulders, spine, abdomen, pelvis<br>and extremities. | The ArtPix Mobile EZ2GO, is<br>intended for use in general<br>radiographic examinations by<br>qualified/trained doctors or<br>technicians, wherever conventional<br>screen-film systems may be used,<br>excluding fluoroscopy, angiography<br>and mammography.<br>ArtPix Mobile EZ2GO allows<br>imaging of the skull, chest,<br>shoulders, spine, abdomen, and<br>extremities. Applications can be<br>performed with the patient sitting,<br>standing, or lying in the prone or<br>supine position. |
The indications of use of the predicate PrestoDR 4143 K110849 and the modified ArtPix Mobile EZ2GO are identical, except for slight language modifications.
{8}------------------------------------------------
Performance & Technological Characteristics Compariso 1.9.2.
Detectors 1.9.2.1.
| Divinne D | Discium Dorbolo | Divitume | Divinne 41490 | |
|-----------|-----------------|----------|---------------------------------|---|
| | | | PrestoDR 414 | |
| | | | Predicate K1108 | |
| | | | Detectors with CsI scintillator | ﺗ |
| | | | | |
| System | Model | Pixel Pitch | Active Image Area | Dynamic Range | DQE (typical values) | Weight |
|----------------------------------------------------|------------------------------------|-------------|------------------------------------------------------------------------|---------------|----------------------------------------------------------------------|---------------------|
| Predicate K110849<br>PrestoDR 4143<br>Pixium 4143C | Pixium 4143C | 148 μm | 41 x 42.5 cm<br>(16.2" x 16.7")<br>2775 x 2874 usable pixels | 16 bits | 64% at 0 lp/mm<br>52% at 1 lp/mm<br>42% at 2 lp/mm<br>25% at 3 lp/mm | 11.5 Kg<br>25.3 lbs |
| ArtPix EZ2GO | Pixium Portable 3543 | 144 μm | 34.1x 43.2 cm2<br>(13.4" x17" /<br>2372 x 3000 usable pixels | 16 bit | 66% at 0 lp/mm<br>51% at 1 lp/mm<br>39% at 2 lp/mm<br>23% at 3 lp/mm | 4.7 Kg<br>8.5 lbs |
| | Pixium Portable 3543EZ | | 33.7 x 41.4 cm<br>(13.26" x 16.30") / 2280 x 2800 usable pixels | | | 2.8 Kg<br>6.17 lbs |
| | Pixium Portable 3543EZ-C | 148 μm | 34.4 x 42.1 cm<br>(13.54" x 16.57" /<br>2330 x 2846 usable pixels) | 16 bit | 66% at 0 lp/mm<br>50% at 1 lp/mm<br>40% at 2 lp/mm<br>24% at 3 lp/mm | 3.0 Kg<br>6.61 lbs |
| | Pixium Portable 3543EZh-C | | 34.4 x 42.1 cm<br>(13.54" x 16.57" /<br>2330 x 2846 usable pixels) | | | 2.8 Kg<br>6.17 lbs |
| | Pixium Portable 2430 EZ-C | | 22.2 x 28.4 cm<br>(8.74" x 11.18" /<br>1500 x 1920 usable pixels) | | | 1.6 Kg<br>3.52 lbs |
| | Pixium Portable 3543EZh (MAX wi-D) | | 34.4 x 42.1 cm2<br>(13.54" x<br>16.57" / 2330 x<br>2846 usable pixels) | | | 3.2Kg<br>7.05 lbs |
| | Pixium Portable 2430EZ (MAX mini) | | 22.2 x 28.4 cm2<br>(8.74" x 11.18" /<br>1500 x 1920 usable pixels) | | | 1.6 Kg<br>3.52 lbs |
The Pixium Portable 3543EZ / 2430EZ FPD family of detectors that are used in the modified ArtPix Mobile EZ2GO ha
already been cleared by FDA in in the currently marketed (K14
{9}------------------------------------------------
# Detectors with Gadox scintillat ് എ
| System | | ArtPix EZ2GO | |
|-----------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | Predicate K110849<br>PrestoDR 4143<br>Pixium 4143G | Pixium Portable<br>3543EZ-G | Pixium Portable<br>2430 EZ-G |
| Pixel Pitch | 148 μm | 148 μm | 148 μm |
| Active Image Area | 41 x 42.5 cm (16.2" x 16.7")<br>2775 x 2874 usable pixels | 33.7 x 41.4 cm<br>(13.26" x 16.30" / 2280 x 2800 usable<br>pixels) | 22.2 x 28.4 cm<br>(8.74" x 11.18" / 1500 x 1920 usable<br>pixels) |
| Dynamic Range | 16 bits | 16 bit | 16 bit |
| MTF (typical values) | 37% at 0 lp/mm<br>25% at 1 lp/mm<br>17% at 2 lp/mm<br>7% at 3 lp/mm | 37% at 0 lp/mm<br>25% at 1 lp/mm<br>17% at 2 lp/mm<br>7% at 3 lp/mm | 37% at 0 lp/mm<br>25% at 1 lp/mm<br>17% at 2 lp/mm<br>7% at 3 lp/mm |
| Weight | 11.5 Kg<br>25.3 lbs | 2.8 Kg<br>6.17 lbs | 1.6 Kg<br>3.52 lbs |
| System | | Predicate K110849<br>PrestoDR 4143 | ArtPix EZ2GO |
| Workstation<br>Computer<br>* min requirements | Processor* | i3-540 | i5-4310U |
| | Memory* | 2GB, DDR3, 1333 MHz RAM | 8GB |
| | Operating<br>System | Windows XP | Windows 7 |
| | Interfaces | One 1Gig LAN for FPD (WiFi access point)<br>and One 1Gig LAN for DICOM network<br>data connection | One WiFi AP for FPD and One 1Gig LAN<br>for DICOM network data connection |
| | Display | Color 17"/19" LCD Touch Screen,<br>1280x1024 pixels | 10.1" LCD 10-Point Multi Touch Screen<br>Optional 24" (with docking station)<br>1920x1200 pixels |
| | Acquisition -<br>Patient data<br>entry | Manually or automatically from the<br>DICOM Worklist | Manually or automatically from the<br>DICOM Worklist |
| | Acquisition -<br>Patient data<br>entry | Urgent patient registration (including<br>update) | Urgent patient registration (including<br>update) |
| | | Multiple Examinations Handling | Multiple Examinations Handling |
| | Reviewing &<br>Processing | User-selectable images | User-selectable images |
| | Reviewing &<br>Processing | Window/level | Window/level |
| | | Window polarity inversion | Window polarity inversion |
| | | Image inversion: Rotation of the image by<br>90°, 180° and 270°. Horizontal and vertical<br>reverse. | Image inversion: Rotation of the image by<br>90°, 180° and 270°. Horizontal and vertical<br>reverse. |
| | | Electronic zoom with pan & scroll<br>capabilities | Electronic zoom with pan & scroll<br>capabilities |
| | | Segmentation based image processing<br>and contrast enhancement | Segmentation based image processing<br>and contrast enhancement |
| | | Operator selectable anatomical<br>programs | Operator selectable anatomical<br>programs Automatic background filming |
| | | Preview image typically within 6-7 seconds<br>after exposure @ 0.5sec integration time | Preview image typically within 4 seconds<br>after exposure @ 0.5sec integration time |
| | | Selectable automatic, semi-automatic or<br>manual storage and printing | Selectable automatic, semi-automatic or<br>manual storage and printing |
| | DICOM 3.0<br>Connectivity | Print: Interface to up to 10 DICOM-<br>compliant printers | Print: Interface to up to 10 DICOM-<br>compliant printers |
| | | Store: Interface to up to 10 DICOM<br>Storage servers | Store: Interface to up to 10 DICOM<br>Storage servers |
| | | MWL: Interface to Hospital / Radiology<br>information system's Modality Worklist<br>including patient registration and study<br>information (optional) | MWL: Interface to Hospital / Radiology<br>information system's Modality Worklist<br>including patient registration and study<br>information (optional) |
| | Disk Storage* | Hard-disk: 500 GB (or higher capacity)<br>enabling at least 15,000 image storage. | Hard-disk: 250 GB SSD (or higher<br>capacity) enabling at least 7,500 image<br>storage. Drive is used for temporary<br>storage only, and not for archive<br>purpose. From usability stand-point 7,500<br>images is much more than needed. |
| | Dimensions | 36 (H) x 18.3 (W) x 42 (L) cm<br>14.2" (H) x 7.2" (W) x 16.5" (L) | 27 (H) x 18.8 (W) x 2 (L)cm<br>10.6" (H) x 7.4" (W) x 0.8" (L) |
| | Weight | 10.1 Kg | 1.09 Kg |
| System | | Predicate<br>PrestoDR 4143 | ArtPix EZ2GO |
| Bucky | | Optional item, including grid and AEC | Provided by OEM. In stationary rooms only<br>(including grid and AEC). No electrical<br>connection to Artpix EZ2GO. |
| Power Distribution<br>Unit | | Provides power to FPD | N/A (FPD is powered by a battery) |
| Power Requirements | AC Voltage | 100-240V | 100-240V |
| | Max Power | 700VA | 92VA (27VA (Monitor), 65VA (Tablet)) |
| | Line<br>Frequency | 50-60 Hz | 50-60 Hz |
| Digital System<br>Storage and<br>Transport | Temperature | -20 - 60°C | -20 - 60°C |
| | Relative<br>Humidity | 10 - 90% | 10 - 90% |
| | Atmospheric<br>Pressure | 70 - 106 kPa | 70 - 106 kPa |
| Regulatory Compliance | | IEC 60601-1-1, IEC 60601-1-2 (only when all<br>system's hardware supplied by Thales)<br>ISO 15223, IEC 60878, ISO 14971<br>Compliant with HIPAA requirements. | IEC 60601-1-1, IEC 60601-1-2 (only when all<br>system's hardware supplied by Thales)<br>ISO 15223, IEC 60878, ISO 14971<br>Compliant with HIPAA requirements. |
The Pixium Portable 3543EZ / 2430EZ FPD family of detectors that are used in the modified ArtPix Mobile EZ2GO I
already been cleared by FDA in in the currently marketed (K141
{10}------------------------------------------------
#### 1.9.2.2. System
{11}------------------------------------------------
## 1.10. Verification, Validation (V&V)
The V&V processes of the modified PrestoDR Portable have been preformed in several steps as follows:
- a. The programmers performed software unit tests during the coding phase. Detected bugs were corrected on line.
- b. The Software was integrated into a software version, which was installed into the target ArtPix Mobile EZ2GO modified system. A software test document was prepared prior to the tests. That document presents the case scenario (test protocol) and includes the pass/ fails ("expected results") criteria. The software was integrated and tested with the target system. The (positive /negative) results were documented in the test document.
- c. System bench tests were performed by CMT Medical Technologies Ltd. Residual software anomalies that were detected during this stage were recorded to and corrected.
- d. The system performance of the modified ArtPix Mobile EZ2GO was validated by measuring the image quality.
- e. RMF and System Requirements were tested. No critical, Safety, efficacy or blocking remained open.
## Conclusion:
The modified ArtPix Mobile EZ2GO system passed its acceptance criteria and is recommended for release.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.