← Product Code [DWF](/productcode/DWF) · K162215

# Aortic Arch Cannulae, Optiflow Aortic Arch Cannulae (K162215)

_Sorin Group Italia S.R.L. · DWF · Jun 16, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162215

## Device Facts

- **Applicant:** Sorin Group Italia S.R.L.
- **Product Code:** [DWF](/productcode/DWF.md)
- **Decision Date:** Jun 16, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Aortic Arch Cannulae: The device is intended to be used as perfusion cannulae to return arterial blood from the extracorportal circuit to the patient during cardiopulmonary surgery for periods of up to six hours. Optiflow Aortic Arch Cannulae: The device is intended to be used as perfusion cannulae to return arterial blood from the extracorportal circuit to the patient during cardiopulmonary surgery for periods of up to six hours.

## Device Story

Sterile, single-use perfusion cannulae for cardiopulmonary bypass surgery; used to return arterial blood from extracorporeal circuit to patient. Components: barbed proximal connector for tubing attachment; optional female luer lock for venting; single-lumen polymer tubing (optionally wire-reinforced); distal tip (straight/bent for Aortic Arch Cannulae; 3D basket design for Optiflow). Operated by surgical team in OR. Device facilitates blood flow during bypass; modifications include DINCH plasticizer, polycarbonate basket tip, and specific size offerings (21Fr, 24Fr).

## Clinical Evidence

Bench testing only. No clinical data. In vitro testing performed per ISO 10993-1 and ISO 10555-1, including blood trauma, peak tensile force, structural integrity, kink resistance, pressure-flow, and shipping/packaging validation. All tests met acceptance criteria.

## Technological Characteristics

Materials: Polymer tubing, polycarbonate (basket tip). Plasticizer: DINCH. Design: Single-lumen, optional wire reinforcement, straight/bent or 3D basket tip. Sizes: 21Fr, 24Fr. Energy: Passive fluid flow. Sterilization: Sterile, single-use. Standards: ISO 10993-1, ISO 10555-1.

## Regulatory Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

## Predicate Devices

- Aortic Arch Cannulae ([K870825](/device/K870825.md))

## Reference Devices

- Aortic Arch Cannulae ([K870826](/device/K870826.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 16, 2017

Sorin Group Italia S.r.1. c/o Scott Light Regulatory Affairs Manager Sorin Group USA, Inc. 14401 W. 65th Way Arvada, Colorado 80004

Re: K162215

Trade/Device Name: Aortic Arch Cannulae, Optiflow Aortic Arch Cannulae Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II Product Code: DWF Dated: May 17, 2017 Received: May 18, 2017

Dear Scott Light:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{2}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Nicole G. Ibrahim -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

#### K162215

Device Name Aortic Arch Cannulae Optiflow Aortic Arch Cannulae

#### Indications for Use (Describe)

Aortic Arch Cannulae:

The device is intended to be used as perfusion cannulae to return arterial blood from the extracorportal circuit to the patient during cardiopulmonary surgery for periods of up to six hours.

Optiflow Aortic Arch Cannulae:

The device is intended to be used as perfusion cannulae to return arterial blood from the extracorportal circuit to the patient during cardiopulmonary surgery for periods of up to six hours.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) SUMMARY

| SUBMITTER:                | Sorin Group Italia<br>86, Via Statale 12 Nord<br>41037 Mirandola (MO) Italy |
|---------------------------|-----------------------------------------------------------------------------|
| CONTACT PERSON:           | Luigi Vecchi<br>Phone: 39 0535 29811<br>Fax: 39 0535 25229                  |
| DATE PREPARED:            | May 17, 2017                                                                |
| DEVICE TRADE NAME:        | Aortic Arch Cannulae<br>Optiflow Aortic Arch Cannulae                       |
| COMMON NAME:              | Cardiopulmonary bypass vascular cannulae                                    |
| CLASSIFICATION NAME:      | Cardiopulmonary bypass vascular catheter, cannula, or tubing                |
| CLASSIFICATION CODE:      | DWF                                                                         |
| REGULATION NUMBER:        | 870.4210                                                                    |
| PRIMARY PREDICATE DEVICE: | Aortic Arch Cannulae K870825                                                |
| REFERENCE DEVICE:         | Aortic Arch Cannulae K870826                                                |

### DEVICE DESCRIPTION:

The devices are intended to be used to cannulate the arterial vessels during cardiopulmonary bypass surgery. They are sterile, single use devices.

These cannulae are comprised of three main components:

- 1. A barbed proximal end connector to attach to cardiopulmonary bypass tubing. A female luer lock connector is also optionally available to provide venting and can be oriented up or down.
- Single-lumen polymer tubing with an optional wire reinforcement to prevent kinking. 2.
- 3. A straight or bent tip that is inserted into the patient. The Aortic Arch Cannulae has a monodimensional tip design whereas the Optiflow Aortic Arch Cannulae has a three dimensional basket tip design.

## INDICATIONS FOR USE:

The device is intended to be used as perfusion cannulae to return arterial blood from the extracorporeal circuit to the patient during cardiopulmonary surgery for periods of up to six hours.

{5}------------------------------------------------

# TECHNOLOGICAL CHARACTERISTICS:

The Aortic Arch Cannulae and the Optiflow Aortic Arch Cannulae are modified versions of the cleared Aortic Arch Cannulae (K870825, K870826).

This 510(k) includes the following modifications:

- 1. The plasticizer used in the cannulae body changed from DEHP to DINCH.
- 2. The Optiflow Aortic Arch Cannulae use a basket tip design rather than a mono-dimensional tip design.
- 3. The basket tip design uses polycarbonate rather than ABS as in the mono-dimensional tip design.
- 4. The Optiflow Aortic Arch Cannulae will be offered with two different sizes of outer diameter, 21Fr and 24Fr rather than 9 to 28 Fr for the A221-xx unmodified device, 9 to 28 Fr for the A222-xx and 9 to 28 Fr for the A232-xx unmodified devices.

The modified devices have the same technological characteristics, principles of operation and control mechanisms as the predicate devices. The modifications do not affect the intended use of the devices.

## NON CLINICAL TEST RESULTS:

In order to support the material changes, the applicable tests were conducted in accordance with the requirements of ISO 10993-1 and in conjunction with FDA's guidance titled "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process".

### IN VITRO TEST RESULTS:

Testing was conducted to demonstrate the modified devices are substantially equivalent to the unmodified devices.

The tests were conducted in accordance with internal methods as well as with the relevant requirements of ISO 10555-1 titled "Intravascular catheters - Sterile and single-use catheters Part 1: General requirements".

The testing included the following:

- 1. Blood trauma
- 2. Peak Tensile Force
- 3. Structural Integrity
- 4. Kink resistance
- 5. Pressure-flow test
- 6. Shipping/packaging testing

The devices successfully met all acceptance criteria for each of these tests.

### CONCLUSION:

The testing performed demonstrates the Aortic Arch Cannulae and Optiflow Aortic Arch Cannulae are substantially equivalent to the predicate Aortic Arch Cannulae (K870825).

---

**Source:** [https://fda.innolitics.com/device/K162215](https://fda.innolitics.com/device/K162215)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
